The AC3™ Range™ Intra-Aortic Balloon Pump (IABP is clinically indicated for use for the following conditions: - 1. Acute Coronary Syndrome - 2. Cardiac and Non-Cardiac Surgery - 3. Complications of Heart Failure
Device Story
AC3 Range IABP provides counter-pulsation therapy for adult patients with impaired LV function; reduces cardiac work and supports blood pressure via volume displacement. System comprises pump control/display module with touchscreen/keypad and pneumatic drive module on wheels for transport. IAB catheter inserted into femoral artery and positioned in descending thoracic aorta. Device processes real-time physiological patient data to trigger and time IAB inflation. Operates in AutoPilot mode (automated control) or Operator mode (manual settings). Used in clinical settings and during air/ground transport by healthcare professionals. Output provides hemodynamic support; assists clinical decision-making by maintaining precise inflation timing.
Clinical Evidence
Bench testing only. Mechanical and general safety testing performed to demonstrate substantial equivalence to the predicate device.
Technological Characteristics
Compact, transportable intra-aortic balloon pump system. Components: pump control/display module (touchscreen/keypad) and pneumatic drive module. Principle: counter-pulsation via volume displacement. Connectivity: supports air and ground transport. Software-controlled inflation timing and triggering based on real-time physiological data.
Indications for Use
Indicated for adult patients with impaired Left Ventricular (LV) function requiring hemodynamic support for Acute Coronary Syndrome, cardiac/non-cardiac surgery, or complications of heart failure.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
K162820 — AC3 Series IABP System · Arrow International, Inc. · Mar 31, 2017
K192238 — AC3 Series Intra-Aortic Balloon Pump (IABP) · Arrow International, Teleflex · Nov 12, 2019
K070225 — SUPRACOR INTR-AORTIC BALLOON PUMP CONSOLE · Abiomed, Inc. · Dec 13, 2007
K060309 — AUTOCAT INTRA-AORTIC BALLOON PUMP SERIES · Arrow Intl., Inc. · Apr 6, 2006
K232343 — AC3 Series IABP · Arrow International, LLC · Aug 30, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 26, 2025
Arrow International LLC Apurva Gokhale Pr. Regulatory Product Specialist (Subsidiary of Teleflex Incorporated) 3015 Carrington Mill Boulevard Morrisville, North Carolina 27560
Re: K250542
Trade/Device Name: AC3TM RangeTM Intra-Aortic Balloon Pump Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon And Control System Regulatory Class: Class II Product Code: DSP Dated: February 24, 2025 Received: February 24, 2025
Dear Apurva Gokhale:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
TBD
Device Name
AC3 TM Range TM Intra-Aortic Balloon Pump
Indications for Use (Describe)
The AC3™ Range ™ Intra-Aortic Balloon Pump is clinically indicated for use for the following conditions:
- 1. Acute Coronary Syndrome
- 2. Cardiac and Non-Cardiac Surgery
- 3. Complications of Heart Failure
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Teleflex" in a bold, blue font. The word is written in all lowercase letters except for the first letter, which is capitalized. There is a registered trademark symbol to the right of the word. The logo is simple and modern.
510(k) Summary [As required by 21 CFR 807.92] Date Prepared: February 21, 2025 510(k) Number: K250542
# Submitter's Name/Contact Person Manufacturer
Arrow International LLC Subsidiary of Teleflex, Inc. 16 Elizabeth Drive Chelmsford, MA 01824 Establishment Registration #3010532612
Contact Person Apurva N. Gokhale Principal Regulatory Affairs Specialist InterventionalPMRA@teleflex.com Tel. +1 905-808-5313
#### General Information
| Trade Name | AC3™ Range™ Intra-Aortic Balloon Pump (IABP) |
|------------------|------------------------------------------------------------|
| Common Name | Intra-Aortic Balloon and Control System |
| Product Code | DSP |
| Classification | Class II |
| Regulation | 21 C.F.R. §870.3535 |
| Panel | Cardiovascular |
| Predicate Device | K232343 - AC3™ Series IAB Pump (cleared 30 August<br>2023) |
#### Device Description
The AC3 Range IABP system is a professional use device that provides counter-pulsation therapy to adult patients with impaired Left Ventricular (LV) Function. It provides hemodynamic support of blood pressure and reduced cardiac work through volume displacement principles.
The AC3 Range IABP system consists of two (2) components: (1) the pump control/display module which incorporates a display, touch screen, and keypad for system operation and (2) the pneumatic drive module with attached wheels for easy transport.
The IABP is attached to an intra-aortic balloon (IAB) catheter which is inserted into the femoral artery and positioned in the descending thoracic aorta. The pump display incorporates a high-definition touchscreen with color coded icons and a keypad for system operation. The AC3 Range IABP uses software to select and maintain precise IAB catheter inflation timing and triggering based on real time physiological data from the patient. The system offers two modes of operation, (1) the AutoPilot mode, where most functions are automatically selected and controlled by IABP and (2) the Operator mode where the user has control over most settings and selections.
#### Intended Use/Indications for Use
The AC3™ Range™ Intra-Aortic Balloon Pump (IABP is clinically indicated for use for the following conditions:
- 1. Acute Coronary Syndrome
- 2. Cardiac and Non-Cardiac Surgery
- 3. Complications of Heart Failure
#### Technological Characteristics Comparison
The AC3 Range IABP is a compact version of the AC3 Optimus IABP that has a low profile for improved transportability within air and ground transport modes (helicopter, plane, ambulance).
Arrow International LLC - a subsidiary of Teleflex Inc.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "Teleflex" in a stylized, bold, blue font. The word is followed by the registered trademark symbol. The font is sans-serif and the letters are closely spaced together. The overall impression is of a corporate logo.
#### Substantial Equivalence and Summary of Studies
The technological differences between the subject and predicate device have been evaluated through Mechanical and General Safey Testing to provide evidence of substantial equivalence for the AC3 Range IABP.
#### Conclusion
The AC3 Range IABP has identical indications for use, fundamental technologies, principles of operation and performance compared to the predicate device. Substantial equivalence was demonstrated through Mechanical and General Safey Testing. The results of these tests raised no new questions of safety or effectiveness and demonstrated that the subject AC3 Range IABP is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.