GLYCAR PERICARDIAL PATCH

K963967 · Glycar, Inc. · DXZ · Oct 31, 1997 · Cardiovascular

Device Facts

Record IDK963967
Device NameGLYCAR PERICARDIAL PATCH
ApplicantGlycar, Inc.
Product CodeDXZ · Cardiovascular
Decision DateOct 31, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Glycar Pericardial Patch and Glycar Vascular Patch of the Glycar Tissue Repair Patch line of products are intended for pericardial closure, peripheral vascular reconstruction and repair, and cardiac and great vessel reconstruction and repair.

Device Story

Glycar Pericardial Patch is a biological tissue patch derived from bovine pericardium. The tissue undergoes glutaraldehyde fixation, cleaning, additional tanning, and a proprietary anti-inflammation treatment (XX). The device is intended for surgical use by physicians for pericardial closure, vascular reconstruction, and cardiac/great vessel repair. It provides a strong, pliable, biocompatible material for reinforcement of soft tissue defects. The device is supplied sterile, processed in a Class 100 clean room, and stored in a 2% propylene oxide solution. It functions as a structural implant to support natural tissue healing and repair.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on material composition, manufacturing process, and intended use comparisons to the predicate device. FDA noted that claims regarding anticalcification or durability require future human clinical study submission.

Technological Characteristics

Material: Glutaraldehyde-treated bovine pericardium. Sterilization: Formaldehyde and propylene oxide (SAL 10^-6). Manufacturing: Aseptic fill in Class 100 clean room. Proprietary anti-inflammation treatment (XX) applied. Form factor: Tissue patch.

Indications for Use

Indicated for pericardial closure, peripheral vascular reconstruction and repair, and cardiac and great vessel reconstruction and repair in patients requiring soft tissue repair, reinforcement, or reconstruction. Prescription use only.

Regulatory Classification

Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} 510 (K) SUMMARY - GLYCAR PERICARDIAL PATCH K963967 Oct. 31, 1997 GLYCAR # 510(K) SUMMARY ## 1. Official Contact Person Dr. Dirk Frater 4537 Belclaire Ave. Dallas, TX 75205 Phone: (214) 526-5412 ## 2. Regulatory Consultant Bertram Hudson Shotwell & Carr Inc. Suite 100, 3003 LBJ Freeway Dallas, TX 75234 ## 3. Device Name a) Trade Name Glycar Pericardial Patch alternative trade names possible b) Common Name Biological Tissue Patch ## 4. F.D.A. Device Classification Class III Device - we do not believe PMA is needed ## 5. Predicate Device The predicate device is the Tissue Guard™ product line, including Supple Peri-Guard™ product line (K923657), and is a tissue patch made from glutaraldehyde-treated bovine pericardium. The device is produced by Bio-Vascular Inc., 2670 Patton Rd., St. Paul, MN 55113, (800) 255-4018. For copies of Peri-Guard™ labels, see Figures 3 - 4 in the Addenda under Predicate Device Labeling. ## 6. Product Description ### a) Candidate Device Composition The Glycar Pericardial Patches are made of the same material as the predicate device, the Peri-Guard™ tissue patches, which are made from glutaraldehyde-treated bovine pericardium. See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence. Oct. 1, 1996 V {1} 510 (K) SUMMARY - GLYCAR PERICARDIAL PATCH GLYCAR ## b) Material Attributes Glutaraldehyde-treated bovine pericardium is a strong, pliable, biocompatible material which is easy to handle surgically. Glycar’s glutaraldehyde-treated pericardium possesses tissue strength equaling that of the predicate device. It provides for a durable repair, replacement or reinforcement of natural soft tissue. *See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Breaking Strength under Supporting Data* ## c) Pyrogenicity Glycar glutaraldehyde-treated bovine pericardium has been found to have negative pyrogenicity. *See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Pyrogenicity under Supporting Data.* ## d) Manufacturing Process The manufacturing process for the *Glycar Pericardial Patches* involves the harvesting of bovine pericardium, fixation of the pericardium in glutaraldehyde, inspection and cleaning of the pericardium, selection of the pericardium for use in strip production, additional tanning of the bovine pericardium, and then cutting of the glutaraldehyde-treated bovine pericardium for the designated strip size, followed by sterilization in formaldehyde, followed then by treatment with Glycar’s proprietary anti-inflammation treatment, XX. *See in Proprietary Manufacturing Material in Addenda under Confidential Information.* Sterility is maintained by using aseptic fill under Clean Room 100 conditions, during transfer to anti-inflammation treatment and to the final storage medium, 2% propylene oxide in water. *See Outline of Manufacturing Process in Addenda under Manufacturing Information and Sterilization Information.* ## e) Product Sizes Sizes and shape has not been determined at this time. Oct. 1, 1996 ii {2} 510 (K) SUMMARY - GLYCAR PERICARDIAL PATCH GLYCAR ## 7. Intended Use of Product ### a) Candidate Device Intended Use The Glycar Inc., *Pericardial Patch* has the same intended use, pericardial closure, as the predicate device, the Bio-Vascular Inc., Peri-Guard™ patch, which is a tissue patch made from the same glutaraldehyde-treated bovine pericardium. Already approved intended uses for the Glycar Tissue Repair Patch family of glutaraldehyde-tanned bovine pericardial tissue patches are: The *Glycar Tissue Repair Patch* (K942911) was deemed substantially equivalent to the Peri-Guard™ bovine pericardial tissue patch and received approval for U.S. marketing for the intended use of hernia and other intra-abdominal soft tissue defect or deficiency repair in December, 1994. An additional member of the *Glycar Tissue Repair Patch* family of products, the Glycar Staple Strip (K954665) was deemed substantially equivalent to the Peri-Strips™ bovine pericardial surgical staple bolsters and received approval for U.S. marketing for the intended uses of surgical stapling of lung tissue, gastric staplings, rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias or defects of the diaphragm, thoracic and abdominal wall in October, 1995. The Glycar Staple Strip uses the identical material as the Glycar Tissue Repair Patch (with the addition of a synthetic suture material to assist placement of the device on the surgical stapling arm). *See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence.* ### b) Predicate Device Intended Use The predicate device, the Peri-Guard™ family of glutaraldehyde-treated bovine pericardium tissue patch has been used for pericardial closures for over ten years without reported material failure; this same material has in addition been used to repair abdominal and thoracic wall defects, inguinal, paracolostomy, ventral, paraumbilical, scrotal, femoral and other hernias, urethral slings, reconstruction of pelvic floor, repair of rectal and vaginal prolapse, suture and staple bolsters, pledgets and Oct. 1, 1996 iii {3} 510 (K) SUMMARY - GLYCAR PERICARDIAL PATCH GLYCAR slings, and other soft tissue repair, reinforcement, and reconstruction without any reported patient morbidity or mortality attributed to material failures. ## 8. Sterilization a) Method The device is sterilized using exposure to formaldehyde and propylene oxide under environmental control in a class 100 Clean Room environment. See Outline of Sterilization Method in Addenda under Sterilization Information. b) Sterility Assurance Level The *Glycar Pericardial Patches* are labeled as sterile. They are sterilized to an SAL of $10^{-6}$ organisms. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and see Validation of Sterilization Processes under Sterilization Information. c) Bovine Spongiform Encephalopathy Pericardium used for the raw material for these devices is obtained from cattle from government approved abbatoirs in Transvaal, S. Africa. The cattle of the Republic of South Africa are certified free from B.S.E. (bovine spongiform encephalopathy) See Figure 5 in Addenda, Certificate of B.S.E.-free Status ## 9. Substantial Equivalence a) Comparison with Predicate Device The *Glycar Pericardial Patches* are substantially equivalent to the Peri-Guard™ family of glutaraldehyde-treated bovine pericardium in that they are: i) manufactured from the same material, glutaraldehyde-treated bovine pericardium, ii) have the same intended use, iii) are sterilized to a minimum assurance against contamination of $10^{-6}$ organisms (Sterility Assurance Level, SAL = $10^{-6}$), and iv) have other performance features equal to or better than those of the predicate device. Oct. 1, 1996 iv {4} 510 (K) SUMMARY - GLYCAR PERICARDIAL PATCH GLYCAR See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence and see Supporting Data ## b) Technological Differences ### i) Implanted Material- Bovine Pericardium There is no difference in the implanted material used for the *Glycar Pericardial Patches* and the predicate device, Bio-Vascular Peri-Guard™. Both devices are manufactured from glutaraldehyde-treated bovine pericardium. See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence. ### ii) Sterilization Method There is no difference in the chemical sterilization method used for Glycar Pericardial Patches versus that of the predicate device. *See Supporting Data in Addenda Under Sleeve Configuration* ### iii) Manufacturing Material One difference in the manufacturing process which Glycar Inc. considers proprietary is the use of the manufacturing material, XX. Glycar Inc. has demonstrated no adverse effects on safety or efficacy of the use of this manufacturing material, and we feel it improves safety by virtue of the properties that it confers to the material. See Proprietary Manufacturing Material in Addenda under Confidential Information. Oct. 1, 1996 v {5} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 31 1997 Dirk Frater, M.D. Official Contact Glycar Inc. 4504 Beverly Drive Dallas, Texas 75205 Re: K963368 and K963967 Glycar Pericardial Patch and Glycar Vascular Patch Regulatory Class: II Product Code: DXZ Dated: July 30, 1997 Received: August 4, 1997 Dear Dr. Frater: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please be advised that you may not make any claims about the effectiveness of the anticalcification treatment or durability of the device in any labeling or advertisements for this device until results of human clinical studies are submitted and found to substantiate such claims. You may, however, summarize results of animal studies performed, which should include the number and type of animal(s), location and duration of the implant, controls used, endpoints and results. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: {6} Page 2 - Dirk Frater, M.D. General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, Thomas J. Callahan, Ph.D. Director, Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7} Page 1 of 1 510(k) Number (if known): K963967 and K963368 Device Name: **Glycar Pericardial Patch and Glycar Vascular Patch** Indications For Use: The Glycar Pericardial Patch and Glycar Vascular Patch of the Glycar Tissue Repair Patch line of products are intended for pericardial closure, peripheral vascular reconstruction and repair, and cardiac and great vessel reconstruction and repair. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Respiratory, and Neurological Devices 510(k) Number K963967/K963368 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...