Medtronic Model 5392 External Pulse Generator (EPG)

K150246 · Medtronic, Inc. · DTE · Feb 18, 2015 · Cardiovascular

Device Facts

Record IDK150246
Device NameMedtronic Model 5392 External Pulse Generator (EPG)
ApplicantMedtronic, Inc.
Product CodeDTE · Cardiovascular
Decision DateFeb 18, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.3600
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient.

Device Story

Battery-powered, dual-chamber external pulse generator (EPG) provides temporary antibradycardia pacing therapy; used with cardiac pacing lead system. Operated by trained personnel in clinical environments; requires continuous patient monitoring. Device delivers demand (synchronous) or asynchronous pacing pulses to heart. Optional disposable protective cover (53922) prevents accidental control activation. Healthcare providers use device to manage cardiac rhythms, support patients during surgery or pacemaker replacement, and treat acute arrhythmias. Benefits include stabilization of heart rate and cardiac output in acute or temporary clinical settings.

Clinical Evidence

Bench testing only. No clinical data presented. Verification performed via regression bench testing to product-level performance requirements to confirm substantial equivalence following modification of the negative battery contact.

Technological Characteristics

Battery-powered external pulse generator; dual-chamber pacing capability. Design and materials are substantially equivalent to predicate (K132924). Features modified negative battery contact. Includes optional disposable protective cover. Operates as a standalone temporary pacing device.

Indications for Use

Indicated for patients requiring temporary single or dual chamber cardiac pacing for therapeutic, prophylactic, or diagnostic purposes, including complete heart block, sinus bradycardia, sick sinus syndrome, bradycardia with congestive heart failure, atrial/ventricular arrhythmias, cardiac arrest, pre-permanent pacemaker implantation support, permanent pacemaker replacement support, cardiac complications during surgery, post-cardiac surgery support, acute myocardial infarction with heart block, and atrial tachyarrhythmias requiring high-rate burst pacing.

Regulatory Classification

Identification

An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment. (2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source ( *i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following: (i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling; (ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use; (iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices ( *e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use; (ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals; (iii) The labeling must list all pacing modes available in the device; (iv) Labeling must include a detailed description of any special capabilities ( *e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 19, 2015 Medtronic, Inc. c/o Becky Severson Sr. Regulatory Associate 8200 Coral Sea Street NE Mounds View, Minnesota 55112 Re: K150246 Trade/Device Name: Medtronic Model 5392 External Pulse Generator (EPG) Regulation Number: 21 CFR 870.3600 Regulation Name: External Pacemaker Pulse Generator Regulatory Class: Class III Product Code: DTE Dated: January 30, 2015 Received: February 2, 2015 Dear Becky Severson, This letter corrects our substantially equivalent letter of February 18, 2015. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hillebrand for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number: K150246 Device Name: Medtronic Model 5392 External Pulse Generator (EPG) Indications for Use: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block ● - Sinus bradycardia ● - Sick sinus syndrome ● - Bradycardia with congestive heart failure ● - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient prior to permanent pacemaker ● Implantation - Support during permanent pacemaker replacement ● - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery ● - Acute myocardial infarction complicated by heart block ● - o Atrial tachyarrhythmias that require high-rate burst pacing for treatment Prescription Use _ X__________AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ #### 510(k) Summary 2 #### [As required by 21 CFR 807.92] | Date Prepared: | January 30, 2015 | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Medtronic, Inc.<br>Medtronic Cardiac Rhythm Heart Failure<br>8200 Coral Sea Street N.E.<br>Mounds View, MN 55112<br>Establishment Registration Number: 2182208 | | Contact Person: | Becky Severson<br>Senior Regulatory Affairs Specialist<br>Medtronic Cardiac Rhythm Heart Failure<br>Phone: 763.526.2365<br>Fax: 651.367.0603<br>Email: becky.t.severson@medtronic.com | | Alternate Contact: | Ryan Calabrese<br>Director Regulatory Affairs<br>Medtronic Cardiac Rhythm Heart Failure<br>Phone: 763.526.3515<br>Fax: 651.367.0603<br>Email: ryan.s.calabrese@medtronic.com | ### General Information | Trade Name: | Medtronic Model 5392 External Pulse Generator (EPG) | |-----------------------|-------------------------------------------------------------| | Common Name: | External pacemaker pulse generator | | Regulation Number: | 21 CFR 870.3600 | | Product Code: | DTE | | Classification: | Class III | | Classification Panel: | Cardiovascular | | Special Controls: | Not applicable | | Predicate Devices: | Medtronic Model 5392 External Pulse Generator (EPG) K13292- | {4}------------------------------------------------ ## Device Description The Model 5392 EPG is a battery-powered, dual chamber, temporary pacemaker designed primarily for temporary antibradycardia pacing therapy. The 53922 EPG cover accessory is an optional disposable protective cover to reduce accidental activation of the controls of the 5392 EPG. ## Indications for Use The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - . Complete heart block - o Sinus bradycardia - Sick sinus syndrome ● - Bradycardia with congestive heart failure ● - Atrial and/or ventricular arrhythmias - Cardiac arrest ● - Support, management, and evaluation of a patient prior to permanent pacemaker Implantation - Support during permanent pacemaker replacement ● - Cardiac complications during invasive or surgical procedures ● - o Support following cardiac surgery - Acute myocardial infarction complicated by heart block ● - Atrial tachyarrhythmias that require high-rate burst pacing for treatment ● ## Technological Characteristics Intended use, design, materials, performance and technological characteristics are substantially equivalent to the predicate devices referenced. ### Technological Characteristics When compared to the predicate device (K132924), the modified Medtronic Model 5392 External Pulse Generator presented in this submission has the same: - Intended use/indications for use ● - Operating principle ● - Design features ● {5}------------------------------------------------ - Device functionality ● - o Biological safety - Packaging materials - Shelf life ● The modified Model 5392 EPG and the predicate device differ in the following: - Modified negative battery contact design ● ### Substantial Equivalence and Summary of Studies: Technological differences between the subject and predicate devices have been evaluated through bench tests to provide evidence of safe and effective use. The modified Medtronic Model 5392 External Pulse Generator is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design and materials have been verified through the following: - Regression bench testing to product level performance requirements ● #### Conclusion: The results of the above verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the modifications made to the Medtronic Model 5392 External Pulse Generator described in this submission result in a device that is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...