← Product Code [PSQ](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ) · K181038

# CorMatrix Cor Patch (3 cm x 5 cm)(single pack), CorMatrix Cor Patch (4 cm x 7 cm)(single pack), CorMatrix Cor Patch (7 cm x 10 cm)(single pack) (K181038)

_Cormatrix Cardiovascular, Inc. · PSQ · Oct 10, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K181038

## Device Facts

- **Applicant:** Cormatrix Cardiovascular, Inc.
- **Product Code:** [PSQ](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ.md)
- **Decision Date:** Oct 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3470
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

CorMatrix(R) Cor Patch is intended for epicardial tissue support and repair.

## Device Story

CorMatrix Cor Patch is a 4-ply, lyophilized, sterilized sheet of porcine small intestinal submucosa extracellular matrix (SIS-ECM). It is used by surgeons for epicardial tissue support and repair. The device acts as a scaffold for tissue reinforcement. It is provided in various sizes (up to 7 cm x 10 cm) and is applied during cardiac procedures. The device provides mechanical support to damaged or repaired cardiac structures, facilitating tissue repair. It is a prescription-only device intended for use in a clinical/surgical setting.

## Clinical Evidence

No clinical data. Bench testing only, including tensile strength, suture retention, and burst strength testing, demonstrated that the SIS-ECM material exceeds biomechanical requirements for its intended use.

## Technological Characteristics

Material: Porcine small intestinal submucosa extracellular matrix (SIS-ECM). Construction: 4-ply, lyophilized, sterilized sheet. Principle: Biological scaffold for tissue support. Connectivity: None (standalone). Sterilization: Not specified.

## Regulatory Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

## Predicate Devices

- CorMatrix(R) Tyke(R) ([K152127](/device/K152127.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 10, 2018

CorMatrix Cardiovascular, Inc. % Wendv Perreault Owner/Principal New South Regulatory, LLC 223 Coventry Road Decatur, Georgia 30030

Re: K181038

Trade/Device Name: CorMatrix Cor Patch (3 cm x 5 cm)(single pack), CorMatrix Cor Patch (4 cm x 7 cm)(single pack), CorMatrix Cor Patch (7 cm x 10 cm)(single pack) Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac Patch Or Pledget Made Of Polypropylene, Polyethylene Terephthalate, Or Polytetrafluoroethylene Regulatory Class: Class II Product Code: PSO Dated: April 18, 2018 Received: April 19, 2018

Dear Wendy Perreault:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Rachel E. Neubrander -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181038

Device Name CorMatrix(R) Cor Patch

| Indications for Use (Describe) | CorMatrix(R) Cor Patch is intended for epicardial tissue support and repair. |
|--------------------------------|------------------------------------------------------------------------------|
|--------------------------------|------------------------------------------------------------------------------|

| Type of Use (Select one or both, as applicable) | <div> <span> <input type="checkbox"/> </span> <span> <input type="checkbox"/> </span> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary: K181038

| Company Information   |                                                  |
|-----------------------|--------------------------------------------------|
| Company Name:         | CorMatrix Cardiovascular, Inc.                   |
| Contact Name:         | Wendy Perreault                                  |
| Contact Title:        | Regulatory Consultant                            |
| Address:              | 1100 Old Ellis Rd.                               |
|                       | Roswell, GA 30076                                |
| Phone (office):       | 678-566-2628                                     |
| Phone (mobile):       | 404-542-5854                                     |
| Fax:                  | 678-736-5082                                     |
|                       |                                                  |
| Date Prepared:        | April 18, 2018                                   |
|                       |                                                  |
| Product Information   |                                                  |
| Trade Name:           | CorMatrix® Cor Patch                             |
| Common Name:          | Patch, Pledget and Intracardiac                  |
| Classification Name:  | Patch, Pledget and Intracardiac, 21 CFR 870.3470 |
|                       | Product Code PSQ, Class II                       |
| Classification Panel: | Cardiovascular                                   |
|                       |                                                  |

#### Predicate Devices

The CorMatrix® Cor Patch is substantially equivalent to the CorMatrix® Tyke®, K152127.

#### Device Description

CorMatrix Cor Patch is intended for epicardial tissue support and repair.

CorMatrix Cor Patch is derived from the same multi-laminate SIS-ECM material as the CorMatrix Tyke (2-ply, non-pressed). The CorMatrix® Cor Patch will be supplied as a 4-ply, lyophilized, sterilized sheet of SIS-ECM. With the exception of the differing layer counts and a change to the manufacturing process for the Cor Patch which does not result in a change to the fundamental technology of the device, the device design and construction are identical to the FDA-cleared CorMatrix® Tyke™ (K152127).

#### Indications for Use

The CorMatrix Cor Patch is intended for epicardial tissue support and repair.

#### Substantial Equivalence

The device is constructed from the same SIS-ECM material used for construction of the CorMatrix Tyke (K152127). The device differs from the CorMatrix Tyke only in the number of ECM layers from which

{4}------------------------------------------------

each device is constructed, and additional processing (wash) steps for the Cor Patch device which serve to further remove impurities from the SIS-ECM.

The CorMatrix Cor Patch is a four-layer ECM material, and the CorMatrix Tyke is a two-layer ECM material. The CorMatrix Tyke is intended for use in neonates and infants, and is therefore provided in a maximum size of 4 cm x 7 cm, whereas the CorMatrix Cor Patch is provided in sizes up to 7 cm x 10 cm. Each layer of the Cor Patch device is identical to the ECM layers in the predicate device, with the exception of the additional processing step performed for the Cor Patch product.

The Indications for Use for the predicate CorMatrix Tyke include use as an epicardial covering for damaged or repaired cardiac structures; the Cor Patch is intended for epicardial tissue support and repair.

## Non-clinical Testing

The CorMatrix Cor Patch is intended for epicardial tissue support and repair. The performance testing that supports the Cor Patch includes tensile strength, suture retention, and burst strength testing. The testing demonstrates that the sterile SIS-ECM of the Cor Patch possesses adequate material properties for use in the indicated applications.

### Conclusion

Performance testing of the SIS-ECM material of the Cor Patch demonstrated that it exceeds the biomechanical requirements for its intended use.

The CorMatrix Cor Patch is substantially equivalent to the CorMatrix Tyke; both are intended for epicardial support and repair and are made from porcine SIS-ECM. The Cor Patch undergoes an additional processing step which serves to further remove impurities from the base material, but does not result in a change to the fundamental technology of the devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K181038](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K181038)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
