← Product Code [PSQ](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ) · K170951

# CardioCel 3D (K170951)

_Admedus Regen Pty, Ltd. · PSQ · Apr 28, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K170951

## Device Facts

- **Applicant:** Admedus Regen Pty, Ltd.
- **Product Code:** [PSQ](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ.md)
- **Decision Date:** Apr 28, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.3470
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

CardioCel 3D is indicated for use as a patch in pericardial closure and the repair of cardiac and vascular defects including intracardiac defects; septal defects, valve and annulus repair; great vessel reconstruction, peripheral vascular reconstruction and suture line buttressing.

## Device Story

CardioCel 3D is a sterile, bovine pericardial patch used in cardiac and vascular surgical procedures. It is manufactured from glutaraldehyde-crosslinked bovine pericardium using ADAPT TEP technology. The device is an acellular collagen sheet, 4 x 6 cm, featuring a 60° pre-formed curve. It is intended for use by surgeons in clinical settings to repair defects or reinforce suture lines. The curved configuration is designed to facilitate surgical application compared to flat patches. The device functions as a structural patch; it is implanted during surgery to provide tissue support and closure. Benefits include biocompatibility and structural integrity for cardiac and vascular repair.

## Clinical Evidence

Bench testing only. Verification and validation included dimensional verification, crosslink stability, tensile testing, and burst pressure testing. Risk assessment covered biocompatibility, sterility, packaging, and shelf life. Results demonstrated that design outputs meet design inputs.

## Technological Characteristics

Bovine pericardium tissue; glutaraldehyde-crosslinked; ADAPT TEP technology; acellular collagen; 4 x 6 cm dimensions with 60° curve; sterile; moist, light yellow to beige sheet.

## Regulatory Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

## Predicate Devices

- CardioCel ([K130872](/device/K130872.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2017

Admedus Regen Pty Ltd % Mary Donlin Vice President Regulatory Affairs and Compliance 860 Blue Gentian Road, Ste 295 Eagan, Minnesota 55121

Re: K170951

Trade/Device Name: CardioCel 3D Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac Patch Or Pledget Made Of Polypropylene, Polyethylene Terephthalate, Or Polytetrafluoroethylene Regulatory Class: Class II Product Code: PSQ Dated: March 30, 2017 Received: March 31, 2017

Dear Ms. Donlin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Wilhelmsen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K170951

Device Name CardioCel 3D

Indications for Use (Describe)

CardioCel 3D is indicated for use as a patch in pericardial closure and the repair of cardiac and vascular defects including intracardiac defects; septal defects, valve and annulus repair; great vessel reconstruction, peripheral vascular reconstruction and suture line buttressing.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 100%;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span>                                                                                    ☑                                                                                   </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></div> |
| <div> <span> <span style="font-size: 100%;"> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span> <span>                                                                                    ☐                                                                                   </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></span></div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# I. Applicant Information

| Date Prepared                  | March 30, 2017                                                     |
|--------------------------------|--------------------------------------------------------------------|
| Submitter                      | Admedus Regen Pty Ltd                                              |
| Address                        | 26 Harris Road<br>Malaga, Western Australia 6090<br>Australia      |
| Establishment Registration     | 3010805634                                                         |
| Contact Person                 | Mary E. Donlin<br>Vice President Regulatory Affairs and Compliance |
| Telephone Number<br>Fax Number | (651) 493-0606 Ext 1008<br>(651) 528-8042                          |

## II. Device Information

| Trade Name          | CardioCel® 3D                                                                                                                                                                                                                                                                                                  |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name         | Cardiovascular Patch                                                                                                                                                                                                                                                                                           |
| Classification Name | Intracardiac Patch or Pledget, Biologically Derived                                                                                                                                                                                                                                                            |
| Classification      | Class II, 21 CFR § 870.3470                                                                                                                                                                                                                                                                                    |
| Product Code        | PSQ                                                                                                                                                                                                                                                                                                            |
| Predicate Device    | CardioCel (K130872)                                                                                                                                                                                                                                                                                            |
| Device Description  | The CardioCel 3D device is a bovine pericardial patch prepared<br>from glutaraldehyde-crosslinked bovine pericardium using the<br>ADAPT® TEP technology. It is a sterile, light yellow to beige<br>colored, moist, pre-cut sheet of acellular collagen.  It is offered in a 4<br>x 6 cm size with a 60° curve. |
| Indications for Use | CardioCel 3D is indicated for use as a patch in pericardial closure<br>and the repair of cardiac and vascular defects including intracardiac<br>defects; septal defects, valve and annulus repair; great vessel<br>reconstruction, peripheral vascular reconstruction and suture line<br>buttressing.          |

{4}------------------------------------------------

| Comparison to<br>Predicate Device            | The predicate device, CardioCel (K130872), was cleared by the<br>FDA for use as a patch in pericardial closure and the repair of<br>cardiac and vascular defects including intracardiac defects; septal<br>defects, valve and annulus repair; great vessel reconstruction,<br>peripheral vascular reconstruction and suture line buttressing. The<br>CardioCel 3D device is identical to the predicate device with respect<br>to intended use, raw materials, tissue processing, sterilization,<br>packaging and principles of operation. Whereas CardioCel is offered<br>in a flat sheet configuration, the manufacturing change implemented<br>during crosslinking enables CardioCel 3D to hold a curved shape. |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Verification and<br>Validation | Verification and validation testing was conducted on CardioCel 3D<br>including dimensional verification, crosslink stability, tensile testing,<br>and burst pressure testing. A risk assessment included evaluation of<br>biocompatibility, sterility, packaging, and shelf life. The results<br>demonstrated that the design outputs of the modified device meet the<br>design inputs in conformance with established design controls.                                                                                                                                                                                                                                                                           |
| Conclusion                                   | The accumulated technical information, intended use, and laboratory<br>verification tests demonstrate that CardioCel 3D is substantially<br>equivalent to the currently marketed predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K170951](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/PSQ/K170951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
