HEMASHIELD WOVEN DOUBLE VELOUR VASCULAR GRAFT MODEL M00202175XXXXX; M00202166XXXXX

K122612 · Maquet Cardiovascular, LLC · MAL · Sep 10, 2012 · Cardiovascular

Device Facts

Record IDK122612
Device NameHEMASHIELD WOVEN DOUBLE VELOUR VASCULAR GRAFT MODEL M00202175XXXXX; M00202166XXXXX
ApplicantMaquet Cardiovascular, LLC
Product CodeMAL · Cardiovascular
Decision DateSep 10, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HEMASHIELD Vascular Grafts are indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. The prostheses are also recommended for use in patients requiring systemic heparinization prior to, or during, surgery.

Device Story

HEMASHIELD Woven Vascular Grafts are woven double velour polyester grafts impregnated with highly purified collagen and glycerol (softening agent). Designed to reduce bleeding at implant site; eliminates operative preclotting step. Collagen resorbed by patient over time. Used by surgeons for arterial replacement or repair in patients with aneurysmal or occlusive disease. Device functions as a permanent vascular prosthesis.

Clinical Evidence

Bench testing only. Testing included water permeability, longitudinal tensile strength, factory anastomotic strength, burst strength, usable length, relaxed/pressurized internal diameter, wall thickness, and suture retention strength. Results met all acceptance criteria.

Technological Characteristics

Woven double velour polyester graft; impregnated with purified collagen and glycerol. Dimensions/form factor: vascular graft prosthesis. Sterilization: not specified. No software or electronic components.

Indications for Use

Indicated for patients requiring replacement or repair of arteries affected by aneurysmal or occlusive disease, including those requiring systemic heparinization during surgery.

Regulatory Classification

Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K122612 Page 1 of 2 GETINGE GROUP SEP 1 0 2012 510(k) Summary (as required by 21 CFR 807.92) 510(k) Number: Date Prepared: Device Owner: Contact Personnel: Title: Email: Phone: Trade Name: Device Generic Name: Classification: Predicate Device: Device Description: Indications for Use: Technological Characteristics Unknown August 23, 2012 MAQUET Cardiovascular LLC 45 Barbour Pond Drive Wayne, New Jersey 07470 Marylou Insinga Regulatory Affairs Specialist II marylou.insinga@maquet.com Fax: 973-807-1658 973-709-7442 HEMASHIELD Woven Vascular Grafts Vascular Graft Prosthesis According to Section 513 of the Federal Food, Drug and Cosmetic Act, the device classification is Class II, Performance Standards. HEMASHIELD Woven Vascular Grafts The HEMASHIELD Vascular Grafts are Woven Double Velour polvester grafts impregnated with highly purified collagen. The HEMASHIELD Vascular Grafts are designed to reduce . bleeding at implant and thereby eliminate the operative preclotting step. The collagen is designed to be gradually resorbed by the patient. In addition to collagen, the grafts also contain glycerol as a softening agent. The HEMASHIELD Vascular Grafts are indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. The prostheses are also recommended for use in patients requiring systemic heparinization prior to, or during, surgery. The proposed HEMASHIELD Vascular Grafts and the predicate HEMASHIELD Vascular Grafts are similar with respect to the following: - Identical materials and finished device characteristics � - Identical indication for use 0 - Identical packaging, sterilization and shelf life 0 {1}------------------------------------------------ The proposed HEMASHIELD Vascular Grafts and the predicate HEMASHIELD Vascular Grafts are different with respect to the following: - An alternate method for the collagen coating process was o used in manufacturing the proposed devices. This difference is not considered a technological difference and is substantially equivalent to the predicate HEMASHIELD Vascular Grafts. Bench testing was performed to support a determination of substantial equivalence. The device was qualified through the following tests: - o Water Permeability - Longitudinal Tensile Strength 0 - Factory Anastomotic Strength o - Burst Strength o - Usable Length 0 - Relaxed Internal Diameter 0 - o Pressurized Internal Diameter - o Wall Thickness - o Suture Retention Strength (Suture Pull Out) - 0 Guideline The results of these non-clinical tests meet the specified acceptance criteria and are substantially equivalent to the predicate device. No new safety or performance issues were raised during the testing regimen. Conclusion: Based on the Indication for Use; technological characteristics and performance testing, the HEMASHIELD Vascular Grafts have been shown to be substantially equivalent to the predicate device. ### Safety and Performance: 00 7 {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 10 2012 Maquet Cardiovascular, LLC % Marylou Insinga 45 Barbour Pond Dr. Wayne, NJ 07470 Re: K122612 Trade/Device Name: Hemashield Woven Double Velour Vascular Graft Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: Class II Product Code: MAL Dated: August 23, 2012 Received: August 27, 2012 Dear Marylou Insinga: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Marylou Insinga Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ HEMASHIELD Vascular Grafts 510(k) August 2012 ## Indications for Use 510(k) Number (if known): K122612 Device Name: HEMASHIELD Vascular Grafts Indications For Use: The HEMASHIELD Vascular Grafts are indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive disease. The prosthesis is also recommended for use in patients requiring systemic heparinization prior to, or during, surgery. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K122612
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...