MODEL 6500 UNIPOLAR TEMORARY MYOCARDIAL PACING LEAD
Applicant
Medtronic Vascular
Product Code
LDF · Cardiovascular
Decision Date
Aug 14, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3680
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Model 6500 Unipolar Temporary Myocardial Pacing Lead is designed for temporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 days or less. The device is supplied sterile and intended for single use only.
Device Story
Model 6500 Unipolar Temporary Myocardial Pacing Lead consists of insulated multi-filament conductor wire crimped to an electrode; proximal end features atraumatic myocardial curved needle for implantation; distal end features electrode for myocardial tissue contact. Device used for temporary atrial and ventricular pacing and sensing. Implanted by physician in clinical setting. Atraumatic chest needle allows lead removal; silicone rubber disc remains in body for atrial applications. Single-use device.
Clinical Evidence
Bench testing only. System compatibility testing performed, including environmental conditioning, visual verification, dimensional testing, electrical testing, and mechanical testing. No clinical data provided.
Technological Characteristics
Unipolar temporary myocardial pacing lead. Materials include silicone rubber. Sterilization via Ethylene Oxide (ETO).
Indications for Use
Indicated for temporary atrial and ventricular pacing and sensing in patients requiring temporary cardiac support for a duration of 7 days or less. Single use only.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Related Devices
K012459 — MODEL 6494 UNIPOLAR TEMPORARY MYOCARDIAL PACING WIRE · Medtronic Vascular · Aug 14, 2001
K012454 — MODEL 6491 UNIPOLAR PEDIATRIC TEMPORARY PACING LEAD · Medtronic Vascular · Aug 14, 2001
K012460 — MODEL 6495 BIPOLAR TEMPORARY MYOCARDIAL PACING LEAD · Medtronic Vascular · Aug 14, 2001
K012458 — MODEL 6492 UNIPOLAR TEMPORARY ATRIAL PACING LEAD · Medtronic Vascular · Aug 14, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
MEDTRONIC CONFIDENTIAL AUG 1 4 2001
Special 510(k) Premarke Model 6500 Unipolar Temporary Myocardial Pacing Lead loder 6500 Unipolar Teliffordi y Myobatantial Equivalence
KOL2452
# ATTACHMENT D 510(K) SUMMARY OF SUBSTANTIAL EQUIVALENCE
## Submitter
Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432
Contact: Tina Benoit, Associate Product Regulation Manager Telephone: (763) 514-4112 Fax: (763) 514-6424 E-Mail: tina.benoit@medtronic.com
Date Prepared: July 31, 2001
### Name of Device
Trade Name: Temporary Pacemaker Electrode, 74 LDF Common Name: Temporary Pacing Lead Classification: Class II
{1}------------------------------------------------
## Predicate Devices
The predicate device for the Model 6500 Unipolar Temporary Myocardial Pacing
Parties of the Model 6500 Unicolar Temporary The predicate device for the Model 6500 Unipslar Temporary Myocardial Pacing Lead.
## Device Description
The Model 6500 Unipolar Temporary Myocardial Pacing Lead consists of an The Model 0500 Unipolar Temperary They conductor which are crimped together. electrode and an insulated multi-inament concession of the lead is attached to the A blue monomanent proximally ocheardumatic myocardial curved needle. A
t electrode and terminates distally in an atraumatic myocardial in museum electrode and terminates distanty in an active the lead is implanted in myocardial tissue. An atraumatic chest needle at the proximal end of the conductor wire tissue. An alraumatic chest neceite at the probably of oremove the pacing lead, permits exiting the pacing icad through the enains in the body, except the silicone rubber disc in case of atrial application.
## Packaging
The sterile packaging for the Model 6500 Unipolar Temporary Myocardial Pacing I he sterife packaging for the Model Coop Configuration. The inner pouch (or package liner) Lead consists of a dodoic pouch connigation. Tyvek - polyester/polyethylene laminate. The pouches are heat-sealed.
## Intended Use
The is designed for temporary atrial and ventricular pacing and sensing for a The is designed for temporary attrails release. The device is supplied sterile and intended for single use only.
## Technological Characteristics
The technology used with the Model 6500 Unipolar Temporary Myocardial The technology asoc with the Moderial characteristics as the predicate device.
## Summary of Studies
Medtronic, Inc. performed system compatibility testing to support that the Model 6500 Unipolar Temporary Myocardial Pacing Lead is equivalent to the predicate device. Device testing included:
- Environmental Conditioning
{2}------------------------------------------------
Special 510 Model 6500 Unipolar Temporary Myocardia loder 6500 Umpolar Temporary ikyo ostantial Equivalence
- Visual Verification .
- Dimensional Testing .
- Electrical Testing .
- Mechanical Testing .
All system compatibility tests performed have demonstrated that the modified Model 6500 heartwire meets the specified requirements.
# Sterilization Validation
The Model 6500 Unipolar Temporary Myocardial Pacing Lead is sterilized using a The Model 0300 Oxide (ETO) sterilization process. Processes appropriate for sterilizing the devices were validated.
## Conclusion
Through data and information presented, numerous similarities support a I hrough data and micrination proseence, and show the device modifications do not affect the intended use of the device or alter the fundamental scientific technology of the device. Market clearance of the Model 6500 Unipolar Temporary Myocardial Pacing Lead is supported through this Special 510(k) Premarket Notification.
31
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 4 2001
Ms. Tina L. Benoit Medtronic, Inc. Cardiac Rhythm Management 7000 Central Avenue NE Minneapolis, MN 55432-3576
Re: K012452
Trade Name: Model 6500 Unipolar Temporary Myocardial Pacing Lead Regulation Number: 21 CFR 870.3680 Regulatory Class: Class II (two) Product Code: 74 LDF Dated: July 31, 2001 Received: August 1, 2001
Dear Ms. Benoit:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this
{4}------------------------------------------------
#### Page 2 - Ms. Tina L. Benoit
response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Dore K. Miller
James E. Dillard II Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known):
KO12452 NA
Device Name:
Medtronic® Model 6500 Unipolar Temporary-Myocardial Pacing Lead
Indications For Use:
The Model 6500 Unipolar Temporary Myocardial Pacing I he Model 0500 Omporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 days or less. schsing for a concemplated in and intended for single use only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 1 Over-The-Counter Use__ OR Prescription Use (Per 21 CFR 801.109)
(Optional Format 1-2-96)
Deak Teth
**Division of Cardiovascular & Respiratory Devices**
510(k) Number K012452
4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.