SimuPlus Flexible Annuloplasty Ring, SimuPlus Flexible Annuloplasty Band, SimuForm Semi-Rigid Annuloplasty Ring

K201449 · Medtronic, Inc. · KRH · Jul 1, 2020 · Cardiovascular

Device Facts

Record IDK201449
Device NameSimuPlus Flexible Annuloplasty Ring, SimuPlus Flexible Annuloplasty Band, SimuForm Semi-Rigid Annuloplasty Ring
ApplicantMedtronic, Inc.
Product CodeKRH · Cardiovascular
Decision DateJul 1, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SimuPlus flexible annuloplasty ring and band are in patients undergoing surgery for diseased or damaged mitral valves. The band is indicated for the same use for tricuspid valves. The SimulPlus flexible annuloplasty ring and band provide support for the mitral annulus or tricuspid annulus (band only) and restrict expansion of the annulus. The SimuForm semi-rigid annuloplasty rings are for use in patients undergoing surgery for diseased or damaged mitral valves. The SimuForm semi-rigid annuloplasty ring provides support for the mitral annulus and restricts expansion of the annulus.

Device Story

Implantable annuloplasty rings and bands for mitral or tricuspid valve repair; used by surgeons in OR to stabilize annulus and restrict expansion. SimuPlus (flexible) features braided polyester fabric; SimuForm (semi-rigid) features MP35N wire stiffener and coil spring within silicone sheath and polyester cover. Both include radiopaque silicone markers and integrated chordal reference line on holder to assist in artificial chord placement. Devices are positioned and implanted via specialized handle. Output is physical support of valve annulus; clinical benefit is restoration of valve function and prevention of further annular dilation.

Clinical Evidence

No clinical data provided. Performance testing was not required as modifications were limited to labeling and sizer accessory alignment.

Technological Characteristics

Flexible rings/bands: braided polyester fabric. Semi-rigid rings: MP35N wire stiffener, MP35N coil spring, silicone sheath, braided polyester cover. Radiographic visualization via barium sulfate/tungsten-impregnated silicone markers. Mechanical support principle. Sterilized.

Indications for Use

Indicated for patients undergoing surgery for diseased or damaged mitral valves (all models) or tricuspid valves (SimuPlus band only) to provide annular support and restrict expansion.

Regulatory Classification

Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. July 1, 2020 Medtronic, Inc. Lisa Corbin Sr. Regulatory Affairs Specialist 1851 E. Deere Ave. Santa Ana, California 92705 Re: K201449 Trade/Device Name: SimuPlus Flexible Annuloplasty Ring, SimuPlus Flexible Annuloplasty Band, SimuForm Semi-Rigid Annuloplasty Ring Regulation Number: 21 CFR 870.3800 Regulation Name: Annuloplasty ring Regulatory Class: Class II Product Code: KRH Dated: May 29, 2020 Received: June 1, 2020 Dear Lisa Corbin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jaime Raben Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201449 Device Name SimuPlus™ Flexible Annuloplasty Ring SimuPlus™ Flexible Annuloplasty Band SimuForm™ Semi-Rigid Annuloplasty Ring ## Indications for Use (Describe) The SimuPlus flexible annuloplasty ring and band are in patients undergoing surgery for diseased or damaged mitral valves. The band is indicated for the same use for tricuspid valves. The SimulPlus flexible annuloplasty ring and band provide support for the mitral annulus or tricuspid annulus (band only) and restrict expansion of the annulus. The SimuForm semi-rigid annuloplasty rings are for use in patients undergoing surgery for diseased or damaged mitral valves. The SimuForm semi-rigid annuloplasty ring provides support for the mitral annulus and restricts expansion of the annulus. | Type of Use ( <i>Select one or both, as applicable</i> ) | | |----------------------------------------------------------|--| |----------------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 1.0 510(k) Summary ## 1.1 Applicant Information | Date Prepared: | May 29, 2020 | |--------------------|-----------------------------------------------------------------------------------------------------------------| | Applicant: | Medtronic, Inc.<br>Medtronic Heart Valves Division<br>1851 E. Deere Ave.<br>Santa Ana, CA 92705 | | Contact Person: | Lisa Corbin<br>Sr. Regulatory Affairs Specialist<br>Phone: (949) 399-1670<br>Email: lisa.corbin@medtronic.com | | Alternate Contact: | Zach Larsen<br>Sr. Regulatory Affairs Supervisor<br>Phone: (949) 382-1034<br>Email: zach.s.larsen@medtronic.com | {4}------------------------------------------------ #### 1.2 Device Name and Classification | Trade Name | Predicate Device | Common<br>Name | Classification<br>Name | Regulation<br>Number | Product<br>Classification | Product<br>Code | Classification<br>Panel | Special<br>Controls | |----------------------------------------------------------------|------------------------------------------------------------------------|----------------------|------------------------|----------------------|---------------------------|-----------------|---------------------------------|---------------------| | SimuPlus™<br>Flexible<br>Annuloplasty Ring,<br>Model 7700FR | Simulus™ Flexible<br>Annuloplasty Ring,<br>Model 700FF<br>(K052565) | | | | | | | | | SimuPlus™<br>Flexible<br>Annuloplasty Band,<br>Model 7700FB | Simulus™ Flexible<br>Annuloplasty Band,<br>Model 700FC<br>(K052899) | Annuloplasty<br>Ring | Ring,<br>Annuloplasty | 21 CFR<br>870.3800 | Class II | KRH | Cardiovascular<br>Devices Panel | Yes | | SimuForm™ Semi-<br>Rigid Annuloplasty<br>Ring, Model<br>7800RR | Simulus™ Semi-<br>Rigid Annuloplasty<br>Ring, Model 800SR<br>(K072655) | | | | | | | | # Table 1-1: Device Name and Classification for Rebranded Simulus Devices {5}------------------------------------------------ #### 1.3 Design and Manufacturing Facility Addresses | Design Site: | Medtronic Heart Valves Division<br>1851 E. Deere Ave.<br>Santa Ana, CA 92705<br>Establishment Registration Number: 2025587 | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturing Site: | Medtronic Mexico S. de R.L. de CV<br>Av. Paseo Cucapah 10510 El Lago<br>Tijuana, Baja California Mexico C.P. 22210<br>Establishment Registration Number: 9617601 | #### Device Description 1.4 #### SimuPlus Flexible Annuloplasty Ring and Band 1.4.1 The SimuPlus flexible annuloplasty ring (Model 7700FR) and band (Model 7700FB) consist of a flexible braided polyester fabric with a 3 mm cross-section. The ring or band may be implanted in the mitral annulus to stabilize the annulus following valve repair. The band may be implanted in the tricuspid annulus to stabilize the annulus following valve repair. The ring and band are marked at three points by colored sutures. Two markers correspond to the trigones (mitral valve) or septal commissures (tricuspid valve), and a central marker identifies the centerline of the ring or band. The individual ring or band size (26 mm to 42 mm increments) refers to the circumference between the green trigone markers on the ring or band. Silicone markers, impregnated with barium sulfate and tungsten, enable radiographic visualization. The ring and band are mounted to a holder to ease positioning and implantation. The SimuPlus annuloplasty ring and band holder incorporates a chordal reference line to assist the surgeon during chordal repair procedures. A laterally stretched suture is present across the holder opening above the typical area of leaflet coaptation. This suture helps approximate the level of the annular plane. If the surgeon decides to include artificial chords in the valve repair, the chordal reference line may be useful when determining the level at which to tie the artificial chord knots. The chordal reference line does not require any additional steps to remove. It is removed with the holder after the retention sutures are cut. ### 1.4.2 SimuForm Semi-Rigid Annuloplasty Ring The SimuForm semi-rigid annuloplasty ring (Model 7800RR) consists of a MP35N wire stiffener in the posterior segment, running from trigone to trigone. The ring stiffener is enclosed within a close-coiled MP35N spring that passes around the circumference of the annuloplasty ring. The spring is covered by a thin silicone sheath. Braided polyester fabric is used to cover and form the body of the ring. The ring has two green markers to indicate the anterior and posterior trigones. A green demarcation suture runs around the upper face of the ring. The individual ring size (24 mm to 40 mm in 2 mm increments) refers to the inner circumference between the green trigone markers on the ring. The internal spring and stiffener provide radiographic visualization around the circumference of the ring. The SimuForm annuloplasty ring holder incorporates a chordal reference line to assist the surgeon during chordal repair procedures. A laterally stretched suture is present across the ring {6}------------------------------------------------ holder opening above the approximate area of leaflet coaptation. This suture helps approximate the level of the annular plane. If the surgeon decides to include artificial chords in the valve repair, the chordal reference line may be useful when determining the level at which to tie the artificial chords knots. The chordal reference line does not require any additional steps to remove. It is removed with the holder after the retention sutures are cut. #### 1.5 Indications for Use Table 1-2: Indications for Use for SimuPlus, SimuForm Rings and Bands | Device Name | Indications for Use | |-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | SimuPlus Flexible<br>Annuloplasty Ring,<br>Model 7700FR | The SimuPlus flexible annuloplasty ring and band are indicated for use in patients<br>undergoing surgery for diseased or damaged mitral valves. The band is indicated for the<br>same use for tricuspid valves. The SimuPlus flexible annuloplasty ring and band<br>provide support for the mitral annulus or tricuspid annulus (band only) and restrict<br>expansion of the annulus. | | SimuPlus Flexible<br>Annuloplasty Band,<br>Model 7700FB | | | SimuForm Semi-Rigid<br>Annuloplasty Ring,<br>Model 7800RR | The SimuForm semi-rigid annuloplasty rings are for use in patients undergoing surgery<br>for diseased or damaged mitral valves. The SimuForm semi-rigid annuloplasty ring<br>provides support for the mitral annulus and restricts expansion of the annulus. | ## 1.6 Comparison to Predicate Devices #### SimuPlus Flexible Annuloplasty Ring and Band 1.6.1 The following characteristics of the SimuPlus flexible annuloplasty ring and band are identical to the Simulus flexible annuloplasty ring and band: implantable ring/band, intended use, principles of operation, sterilization, shelf life, and packaging. The SimuPlus flexible annuloplasty ring and band are substantially equivalent to the Simulus flexible annuloplasty ring and band as summarized in Table 1-3. | Characteristic | SimuPlus Flexible<br>Annuloplasty Ring and Band<br>(Subject) | Simulus Flexible<br>Annuloplasty Ring and Band<br>(Predicate) | Equivalence | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Indications for Use | The SimuPlus flexible<br>annuloplasty ring and band are<br>indicated for use in patients<br>undergoing surgery for diseased or<br>damaged mitral valves. The band<br>is indicated for the same use for<br>tricuspid valves. The SimuPlus<br>flexible annuloplasty ring and<br>band provide support for the<br>mitral annulus or tricuspid annulus<br>(band only) and restrict expansion<br>of the annulus. | The ATS SIMULUS<br>Annuloplasty Rings are for use in<br>those patients undergoing surgery<br>of diseased or damaged mitral or<br>tricuspid valves in whom the<br>surgeon determines that the valve<br>can be preserved by employing<br>the appropriate surgical repair.<br>The annuloplasty rings provide<br>support for the mitral or tricuspid<br>annulus and restrict expansion of<br>the annulus. | Substantially<br>Equivalent | Table 1-3: Flexible Rinq and Band Substantial Equivalence Comparison {7}------------------------------------------------ | Characteristic | SimuPlus Flexible<br>Annuloplasty Ring and Band<br>(Subject) | Simulus Flexible<br>Annuloplasty Ring and Band<br>(Predicate) | Equivalence | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | | | The ATS SIMULUS FC<br>Annuloplasty Bands are for use in<br>those patients undergoing surgery<br>of diseased or damaged mitral or<br>tricuspid valves in whom the<br>surgeon determines that the valve<br>can be preserved by employing<br>the appropriate surgical repair.<br>The annuloplasty bands provide<br>support for the mitral or tricuspid<br>annulus and restrict expansion of<br>the annulus | | | Labeled Size Range* | 26 mm, 28 mm, 30 mm, 32 mm,<br>34 mm, 36 mm, 38 mm, 40 mm,<br>42 mm | 23 mm, 25 mm, 27 mm, 29 mm,<br>31 mm, 33 mm, 35 mm, 37 mm,<br>39 mm | Substantially<br>Equivalent | | Product Labeling | Update to product name,<br>indications, labeled size range Update handle compatibility<br>to Model 7686 annuloplasty<br>handle Update to 29-language eIFU,<br>6-language paper IFU | Compatible with Model 752<br>annuloplasty handle 16-language eIFU, 16-<br>language paper IFU | Substantially<br>Equivalent | | *Note that there is no change to the construction or dimensions of the implantable ring/band. Modifications to the<br>labeled size range reflect dimensional changes to the sizer accessory to align with standard industry sizing. | | | | # Table 1-3: Flexible Ring and Band Substantial Equivalence Comparison #### 1.6.2 SimuForm Semi-Rigid Annuloplasty Ring The following characteristics of the SimuForm semi-rigid annuloplasty ring are identical to the Simulus semi-rigid ring: implantable ring, intended use, principles of operation, size range, sterilization, shelf life, and packaging. The SimuForm semi-rigid ring is substantially equivalent to the Simulus semi-rigid ring as summarized in Table 1-4. Table 1-4: Semi-rigid Ring Substantial Equivalence Comparison | Characteristic | SimuForm Semi-rigid<br>Annuloplasty Ring (Subject) | Simulus Semi-rigid<br>Annuloplasty Ring<br>(Predicate) | Equivalence | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Indications for Use | The SimuForm semi-rigid<br>annuloplasty rings are for use in<br>patients undergoing surgery for<br>diseased or damaged mitral<br>valves. The SimuForm semi-rigid | The ATS SIMULUS™ Semi-<br>Rigid Mitral Annuloplasty Rings<br>are for use in those patients<br>undergoing surgery of diseased or<br>damaged mitral valves in whom | Substantially<br>Equivalent | {8}------------------------------------------------ | Characteristic | SimuForm Semi-rigid<br>Annuloplasty Ring (Subject) | Simulus Semi-rigid<br>Annuloplasty Ring<br>(Predicate) | Equivalence | |------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | | annuloplasty ring provides support<br>for the mitral annulus and restricts<br>expansion of the annulus. | the surgeon determines that the<br>valve can be preserved by<br>employing the appropriate<br>surgical repair. The annuloplasty<br>rings provide support for the<br>natural annulus and restrict<br>expansion of the annulus. | | | Product Labeling | Update to product name Update handle compatibility<br>to Model 7686 annuloplasty<br>handle Update to include 29-<br>language eIFU, 6-language<br>paper IFU | Compatible with Model 752<br>annuloplasty handle 16-language eIFU, 16-<br>language paper IFU | Substantially<br>Equivalent | ### 1.7 Summary of Performance Data Performance testing was not required to support labeling changes made to align with standard industry sizing. ### 1.8 Conclusion The modifications to the Simulus devices do not impact the intended use or alter the fundamental scientific technology of the devices. There is no change to the implantable rings and bands. The SimuPlus and SimuForm rings and bands are therefore substantially equivalent to the currently marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...