← Product Code [KRH](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRH) · K083623

# CARPENTIER-MCCARTHY-ADAMS IMR ETLOGIX ANNULOPLASTY RING, MODEL 4100 (K083623)

_Edwards Lifesciences, LLC · KRH · Apr 15, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRH/K083623

## Device Facts

- **Applicant:** Edwards Lifesciences, LLC
- **Product Code:** [KRH](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRH.md)
- **Decision Date:** Apr 15, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Carpentier-McCarthy-Adams IMR ETlogix Mitral Annuloplasty Ring is indicated for the correction of mitral valve insufficiency where the lesions are not so severe as to require total valve replacement. The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendinae. The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus. For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures or chordae tendineae, or chordal hypertrophy.

## Device Story

Prosthetic annuloplasty ring (model 4100) for surgical correction of mitral valve insufficiency; implanted by cardiac surgeons during open-heart valvuloplasty. Device provides structural support to the mitral annulus to restore proper valve geometry and function. Implanted via suturing to the fibrous mitral annulus; facilitates remodeling of the valve ring. Benefits include correction of regurgitation and preservation of native valve tissue, avoiding total valve replacement. Device is a passive mechanical implant; no electronic or software components.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Constructed of titanium alloy core; sewing ring margin consists of silicone rubber covered with woven polyester cloth. Passive mechanical implant; dimensions/form factor specific to mitral annulus remodeling. No energy source, connectivity, or software.

## Regulatory Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

## Predicate Devices

- Carpentier-Edwards Classic® ring ([K912554](/device/K912554.md))
- Carpentier-Edwards Physio® annuloplasty ring ([K926138](/device/K926138.md))
- GeoForm® annuloplasty ring ([K032250](/device/K032250.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K083623

Image /page/0/Picture/1 description: The image shows a logo with the word "Edwards" written below it. The logo is a square grid with a stylized letter "E" in the center. The letter "E" is surrounded by circular shapes and lines, creating a complex and abstract design.

# 510(k) Summary

APR ] 5 2009

| Submitter:                     | Edwards Lifesciences LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                | Martin A. Kaufman<br>Director - Regulatory Affairs<br>Edwards Lifesciences<br>One Edwards Way<br>Irvine, CA 92614<br>martin_kaufman@edwards.com<br>(ph) 949.250.2708<br>(fax) 949.250.3630                                                                                                                                                                                                                                                                                                           |
| Date Prepared:                 | December 3, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Trade Name:                    | Carpentier-McCarthy-Adams IMR ETlogix™ Mitral Annuloplasty Ring, or IMR ETlogix™<br>Mitral Annuloplasty Ring                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name:           | Class II, CFR 870.3800 Annuloplasty Ring, KRH                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device(s):           | Carpentier-Edwards Classic® ring (K912554)<br>Carpentier-Edwards Physio® annuloplasty ring (K926138)<br>GeoForm® annuloplasty ring (K032250)                                                                                                                                                                                                                                                                                                                                                         |
| Device Description:            | The Carpentier-McCarthy-Adams IMR ETlogix annuloplasty ring, model 4100 (IMR<br>ETlogix) is constructed of titanium alloy and has a sewing ring margin that consists of a<br>layer of silicone rubber covered with a woven polyester cloth.                                                                                                                                                                                                                                                          |
| Indications for Use:           | The IMR ETlogix Annuloplasty Ring is indicated for the correction of mitral valvular<br>insufficiency where the lesions are not so severe as to require total valve replacement.                                                                                                                                                                                                                                                                                                                     |
|                                | The decision to undertake valvuloplasty can be made only after visual analysis of the<br>lesion present. The most favorable conditions for valvuloplasty using an annuloplasty<br>ring are a combination of a distended natural valve ring associated with supple valve<br>cusps and normal chordae tendineae.                                                                                                                                                                                       |
|                                | The remodeling valvuloplasty technique with a prosthetic ring may be used in all<br>acquired or congenital mitral insufficiencies with dilatation and deformation of the<br>fibrous mitral annulus.                                                                                                                                                                                                                                                                                                  |
|                                | For Type I mitral insufficiencies with no subvalvular lesions and normal valvular<br>movements, this ring technique used alone is sufficient. However, the ring technique<br>must be associated with mitral valvuloplasty repair in Type II insufficiencies with<br>prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III<br>insufficiencies with limitation of valvular movements due to fusion of the commissures<br>of chordae tendineae, or chordal hypertrophy. |
| Comparative Analysis:          | It has been demonstrated that the Carpentier-McCarthy-Adams IMR ETlogix mitral<br>annuloplasty ring is comparable to the predicate devices in design, intended use,<br>materials, and principal of operation.                                                                                                                                                                                                                                                                                        |
| Functional/<br>Safety Testing: | The Carpentier-McCarthy-Adams IMR ETlogix mitral annuloplasty ring has successfully<br>completed design verification testing.                                                                                                                                                                                                                                                                                                                                                                        |
| Conclusion:                    | The Carpentier-McCarthy-Adams IMR ETlogix mitral annuloplasty ring is substantially<br>equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                          |

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a bird-like figure with three curved lines.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 5 2009

Edwards Lifesciences, LLC. Martin A. Kaufman Director of Regulatory Affairs One Edwards Way Irvine CA 92614

Re: K083623

Trade/Device Name: IMR ETlogix annuloplasty ring, model 4100 Regulation Number: 21 CFR 870.3800 Regulatory Class: Class II Product Code: KRH Dated: March 13, 2009 Received: March 16, 2009

Dear Mr. Kaufman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (DMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Mr. Kaufman - Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

### 510(k) Number (if known): K083623

## Device Name: Carpentier-McCarthy-Adams IMR ETlogix™ annuloplasty ring, model 4100, or IMR ETlogix™ annuloplasty ring, model 4100

Indications for Use:

The Carpentier-McCarthy-Adams IMR ETlogix Mitral Annuloplasty Ring is indicated for the correction of mitral valve insufficiency where the lesions are not so severe as to require total valve replacement.

The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendinae.

The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus.

For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures or chordae tendineae, or chordal hypertrophy.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office Of Device Evaluation (ODE)

(Division Sign-Off)
Vision of Cardiovascular Devices
10(k) Number ko 83623

Page of

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRH/K083623](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRH/K083623)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
