← Product Code [KRD](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRD) · K112103

# INTERLOCK-35 FIBERED IDC OCCLUSION SYSTEM (K112103)

_Boston Scientific Corp · KRD · Aug 12, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRD/K112103

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [KRD](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRD.md)
- **Decision Date:** Aug 12, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Interlock-35 Fibered IDC Occlusion System is indicated to obstruct or reduce rate of blood flow in the peripheral vasculature. This device is not intended for neurovascular use.

## Device Story

The Interlock-35 Fibered IDC Occlusion System is a vascular embolization device consisting of a platinum-tungsten alloy coil with synthetic fibers for increased thrombogenicity, mechanically attached to a delivery wire and housed in an introducer sheath. Used in the peripheral vasculature, the device is delivered under fluoroscopic guidance through a 5F diagnostic catheter. The interlocking delivery wire mechanism allows for controlled advancement, retraction, and withdrawal of the coil prior to final deployment. The device is operated by a physician to achieve vessel occlusion. This submission updates the Directions for Use to include MRI safety information, confirming the device is MR Conditional at 1.5T and 3.0T.

## Clinical Evidence

Bench testing only. Non-clinical MRI safety testing was conducted according to ASTM F2182-02a, ASTM F2052-06e1, ASTM F2213-06, and ASTM F2119-07 to evaluate heating and migration at 1.5T and 3.0T. Results demonstrate the device is MR Conditional.

## Technological Characteristics

Platinum-tungsten alloy coil with synthetic fibers; delivery wire assembly; introducer sheath. Designed for 5F diagnostic catheters. MR Conditional at 1.5T and 3.0T per ASTM F2182-02a, F2052-06e1, F2213-06, and F2119-07.

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- Interlock-35 Fibered IDC Occlusion System ([K110295](/device/K110295.md))

## Reference Devices

- Imager™ II Diagnostic Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 长川之103

## 510(k) Summary
per 21 CFR §807.92 (c)

AUG 1 2 2011

per 21 CFR §807.92 (c)

:

| Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and Information    | Harlan Jones<br>Regulatory Affairs Specialist II<br>Tel: 763-255-0027<br>Fax: 763-494-2222<br>E-mail: Harlan.Jones@bsci.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Date Prepared                      | July 14, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name                         | Interlock™-35 Fibered IDC™ Occlusion System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Common Name                        | Vascular embolization device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                       | KRD, Vascular embolization devices<br>(21 CFR 870.3300)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device                   | Interlock-35 Fibered<br>IDC Occlusion<br>System (Vascular<br>embolization device) K110295 SE: 3 Mar 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Reason for<br>Submission           | The reason for this premarket notification is to seek<br>clearance for a revision to the Directions for Use (DFU) for<br>Boston Scientific Corporation's (BSC) Interlock™-35<br>Fibered IDC™ Occlusion System (Vascular emolization<br>device). The device was previously cleared by FDA under<br>K110295 as Interlock™-35 Fibered IDC™ Occlusion System<br>(Vascular emolization device).<br><br>Accordingly, BSC is submitting this Premarket Notification<br>to incorporate results of non-clinical MRI testing in the DFU.<br>The proposed DFU revisions are described in Section 4.2.1,<br>Proposed MRI Labeling. The addition of MRI information is<br>not prompted by field events or complaints; the additional<br>information allows for safer and more effective use of the<br>device. The revision has no effect on the Indications for<br>Use, the Contraindications, or the device configuration. This<br>Premarket Notification represents no other device or<br>labeling changes |

:

{1}------------------------------------------------

| Device<br>Description               | The Interlock-35 Fibered IDC Occlusion System includes a<br>coil manufactured from a platinum-tungsten alloy that is<br>mechanically attached to a coil delivery wire. This assembly<br>is contained within an introducer sheath. The platinum coil<br>contains synthetic fibers for greater thrombogenicity. The<br>Interlock-35 Fibered IDC Occlusion System is designed to<br>be delivered under fluoroscopy through a 5F (0.035 in [0.89<br>mm] or 0.038 in [0.97 mm] inner lumen) Diagnostic Catheter<br>(e.g. Imager™ II Diagnostic Catheter). The interlocking<br>delivery wire design allows the coil to be advanced and<br>retracted before final placement in the vessel, thus aiding in<br>more controlled delivery including the ability to withdraw the<br>coil prior to deployment. |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use              | The Interlock-35 Fibered IDC Occlusion System is indicated<br>to obstruct or reduce rate of blood flow in the peripheral<br>vasculature. This device is not intended for neurovascular<br>use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Non-Clinical<br>Performance<br>Data | Non-clinical testing demonstrates that the Interlock-35<br>Fibered IDC Occlusion System will not present additional<br>risk to a patient during a MRI procedure in comparison to<br>risks imposed by gravitational, and other normal daily<br>activities, or a temperature rise that is experienced during a<br>feverish condition. The results of testing conducted under<br>methods described by ASTM F2182-02a, ASTM F2052-<br>06e1, ASTM F2213-06 and ASTM F2119-07 have<br>demonstrated that the Coils are MR Conditional and can be<br>scanned safely under a static magnetic field of 1.5 Tesla or<br>3.0 Tesla.                                                                                                                                                                           |
|                                     | Testing at field strengths other than 1.5 Tesla or 3.0 Tesla<br>has not been performed to evaluate coil migration or<br>heating.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Conclusion                          | The modifications do not affect the intended use or alter the<br>fundamental scientific technology of the predicate Boston<br>Scientific Interlock-35 Fibered IDC Occlusion System<br>(K110295).<br><br>Based on the Indications for Use, unaltered technological<br>characteristics, and submitted non-clinical performance data<br>supporting this modification, the Boston Scientific Interlock-<br>35 Fibered IDC Occlusion System is shown to be<br>appropriate for its intended use and demonstrates that the<br>device is as safe, as effective, and performs as well as the<br>predicate device.                                                                                                                                                                                          |

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

:

. .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The text is in uppercase letters and is evenly spaced around the perimeter of the seal.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Boston Scientific Corporation c/o Mr. Harlan Jones Regulatory Affairs Specialist II One Scimed Place Maple Grove. MN 55311

AUG 12 2011

Re: K112103

Trade/Device Name: Interlock™-35 Fibered IDC™ Occlusion System Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: July 21. 2011 Received: July 22, 2011

Dear Mr. Jones :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Harlan Jones

device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Center for Devices and Radiological Health Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Interlock-35 Fibered IDC Occlusion System

## Indications for Use

510(k) Number (if known): K112 (0 3

Device Name: Interlock-35™ Fibered IDC™ Occlusion System

Indications for Use:

The Interlock-35 Fibered IDC Occlusion System is indicated to obstruct or reduce rate of blood flow in the peripheral vasculature. This device is not intended for neurovascular use.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ·

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bivision Sign-Off) Division of Cardiovascular Devices

K112103 510(k) Number_

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRD/K112103](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/KRD/K112103)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
