K091077 · Cordis Corp., A Johnson & Johnson Co. · DTK · Feb 4, 2010 · Cardiovascular
Device Facts
Record ID
K091077
Device Name
OPTEASE VENA CAVA FILTER AND RETRIEVAL CATHETER
Applicant
Cordis Corp., A Johnson & Johnson Co.
Product Code
DTK · Cardiovascular
Decision Date
Feb 4, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3375
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OPTEASE Vena Cava Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava in the following situations: Pulmonary thromboembolism when anticoagulants are contraindicated, Failure of anticoagulant therapy in thromboembolic disease, Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced, Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. The OPTEASE Filter may be retrieved according to the instructions supplied in the Section labeled: Optional Procedure for Filter Retrieval. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to the vena cava. The Cordis OPTEASE Retrieval Catheter is has been designed for retrieval of an implanted OPTEASE Filter in patients who no longer require a filter. Retrieval of the filter can be performed only by femoral approach.
Device Story
System comprises flexible, self-expanding vena cava filter, 6F sheathed introduction kit, and retrieval catheter. Filter deployed in infra-renal inferior vena cava to trap large emboli while maintaining caval patency. Angiographic vessel dilator provides visualization and linear measurement of vasculature using radiopaque contrast media. Retrieval catheter allows removal of filter via femoral approach when no longer needed. Used in clinical settings by physicians. Device benefits patients by preventing recurrent pulmonary embolism in high-risk populations where anticoagulation is not viable.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and intended use similarities to predicate devices.
Technological Characteristics
Self-expanding vena cava filter; 6F sheathed introduction kit; includes angiographic vessel dilator, directional filter storage tube, catheter sheath introducer, and obturator. Mechanical device; no energy source or software. Sterilization and biological evaluation performed per ISO-10993 and FDA guidance.
Indications for Use
Indicated for patients requiring prevention of recurrent pulmonary embolism via percutaneous inferior vena cava placement when anticoagulants are contraindicated, anticoagulant therapy has failed, or in emergency cases of massive pulmonary embolism. Retrieval indicated for patients no longer requiring the filter via femoral approach.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
Predicate Devices
Cordis OPTEASE® Vena Cava Filter and OPTEASE® Retrieval Catheter (K034050)
G2 EXPRESS Filter System, Bard Peripheral Vascular, Inc. (K082305)
Submission Summary (Full Text)
{1}------------------------------------------------
K091077
FEB - 4 2010
# 510(k) Summary
| Applicant | Cordis Corporation, a Johnson & Johnson Company<br>430 Rt. 22 East<br>Bridgewater, New Jersey 08807<br>Telephone: 908-541-4888<br>Fax: 908-541-4559 | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Contact | Joan Martin, Manager, Regulatory Affairs | |
| Date | April 13, 2009 | |
| Subject Device | Trade Name: | OPTEASE® Vena Cava Filter and OPTEASE® Retrieval Catheter |
| | Common or Usual Name: | Cardiovascular Intravascular Filter (per 21 CFR 870.3375) |
| | Classification: | Class II |
| | Classification Panel | Cardiovascular |
| Name of Predicate Devices | The device is substantially equivalent to:<br>Cordis OPTEASE® Vena Cava Filter and OPTEASE® Retrieval Catheter (ref. 510(k) K034050) G2 EXPRESS Filter System, Bard Peripheral Vascular, Inc. (ref. 510(k) K082305) | |
| Performance Standards | As per 21 CFR 870.3375, the following special controls were established for cardiovascular intravascular filters:<br>Use of International Standards Organization's ISO-10993 'Biological Evaluation of Medical Devices Part I: Evaluation and Testing, FDA's Updated 510(k) Sterility Review Guidance (K90-1); Final Guidance for Industry and FDA, August 30, 2002 and FDA's Guidance for Cardiovascular Intravascular Filter 510(k) Submissions, dated November 26, 1999 | |
{2}------------------------------------------------
| Indications for<br>Use for Filter | The OPTEASE Vena Cava Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava in the following situations: <ul><li>Pulmonary thromboembolism when anticoagulants are contraindicated,</li><li>Failure of anticoagulant therapy in thromboembolic disease,</li><li>Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced,</li><li>Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.</li></ul> The OPTEASE Filter may be retrieved according to the instructions supplied in the Section labeled: Optional Procedure for Filter Retrieval. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to the vena cava. |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use for<br>Retrieval<br>Catheter | The Cordis OPTEASE Retrieval Catheter is has been designed for retrieval of an implanted OPTEASE Filter in patients who no longer require a filter. Retrieval of the filter can be performed only by femoral approach. |
| Device<br>Description | The subject device is a system that consists of a flexible, self-expanding vena cava filter to be deployed in the infra-renal inferior vena cava via a 6F sheathed introduction kit. The filter is designed to trap large, life threatening emboli and therefore prevent recurrent pulmonary embolism, while maintaining caval patency. The OptEase Permanent Vena Cava Filter is packaged with a filter introduction kit that includes the Angiographic Vessel Dilator, a directional filter storage tube, catheter sheath introducer and obturator for safe and accurate deployment of the filter. |
| Summary of<br>Substantial<br>Equivalence | The subject OPTEASE Vena Cava Filter is substantially equivalent to the predicate devices, the OPTEASE Vena Cava Filter and the G2 EXPRESS Filter System in terms of intended use, application and user application. The change to the predicate devices, OPTEASE Vena Cava Filter and OPTEASE Retrieval Catheter, only affects the labeling. |
・
. . . . .
"
・
.
l
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human services. The symbol consists of a series of curved lines that form a wing-like shape, suggesting care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
FEB - 4 2010
Cordis Corporation c/o Ms. Joan Martin 430 Rt. 22 East Bridgewater, NJ 08807
Re: K091077
OPTEASE Vena Cava Filter and OPTEASE Retrieval Catheter Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: January 4, 2010 Received: January 5, 2010
Dear Ms. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for use in the prevention of recurrent pulmonary embolism to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against resporancing and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Joan Martin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Dina R. Lehner
E
L
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Cordis Corporation, a Johnson & Johnson Company OPTEASE® Vena Cava Filter and OPTEASE® Retrieval Catheter
SPECIAL 510(k) Premarket Notification
#### Indications for Use
510(k) Number (if known): K091077
## Device Name: OPTEASE® Vena Cava Filter and OPTEASE Retrieval Catheter
Indication for Use:
The OPTEASE Filter is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated, .
- Failure of anticoagulant therapy for thromboembolic disease, �
- Emergency treatment following massive pulmonary embolism where anticipated benefits of . conventional therapy are reduced.
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is . contraindicated
The OPTEASE Filter may be retrieved according to the instructions supplied in the Section labeled: Optional Procedure for Filter Retrieval.
The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopague contrast media to the vena cava.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Vihunen
(Division Sign-Off) (Division Sign of Cardiovascular Devices
510(k) Number_10 9 10 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.