PACE 101H
Device Facts
| Record ID | K970497 |
|---|---|
| Device Name | PACE 101H |
| Applicant | Sulzer Oscor, Inc. |
| Product Code | DTE · Cardiovascular |
| Decision Date | Jun 4, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.3600 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The external pacemaker PACE 101 H is designed for temporary stimulation of the heart in case of rhythm disturbances and conduction defects. According to present clinical experience, the instrument is especially suited for stimulation of the heart in the following cases: - treatment of patients before an operation, whereby an implantable pacemaker is being inserted; - treatment of tachyarrhythmia; - treatment of special cases of acute myocardial infarction; - treatment of patients after heart surgery.
Device Story
PACE 101 H is an external pacemaker for temporary cardiac stimulation. Device provides electrical pulses to heart to manage rhythm disturbances and conduction defects. Used in clinical settings (e.g., post-surgery, pre-implant, acute MI) by healthcare professionals. Device delivers stimulation therapy to stabilize heart rate; helps maintain hemodynamic stability in patients with transient cardiac issues. Operation involves setting stimulation parameters to provide temporary pacing support.
Clinical Evidence
No clinical data provided; substantial equivalence based on regulatory review of device design and intended use.
Technological Characteristics
External temporary cardiac pacemaker; provides electrical stimulation for rhythm management. Device is a standalone unit for clinical use. Technical specifications and materials not detailed in provided documentation.
Indications for Use
Indicated for temporary cardiac stimulation in patients with rhythm disturbances or conduction defects, including pre-operative preparation for permanent pacemaker implantation, tachyarrhythmia management, acute myocardial infarction, and post-cardiac surgery recovery.
Regulatory Classification
Identification
An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment. (2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source ( *i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following: (i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling; (ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use; (iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices ( *e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use; (ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals; (iii) The labeling must list all pacing modes available in the device; (iv) Labeling must include a detailed description of any special capabilities ( *e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.
Related Devices
- K023080 — GALIX PACESTAR · Galix Biomedical Instrumentation, Inc. · Jun 11, 2003
- K150246 — Medtronic Model 5392 External Pulse Generator (EPG) · Medtronic, Inc. · Feb 18, 2015
- K162054 — Medtronic Temporary External Pacemaker 53401 · Medtronic, Inc. · Oct 18, 2016
- K033130 — PACE 101/PACE 101H, MODEL 3077 SSI TEMPORARY PULSE GENERATOR · Osypka Medical, Inc. · Oct 31, 2003
- K022939 — OSCOR PACE 101/101H AND ACCESSORIES, ST.JUDE MEDICAL MODEL 3077 AND ACCESSORIES,CARDIOTRONIC PACE 101 AND ACCESSORIES · Osypka Medical, Inc. · Oct 3, 2002