Browse hierarchy: [Cardiovascular (CV)](/submissions/CV) → [Subpart D — Cardiovascular Prosthetic Devices](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices) → [21 CFR 870.3600](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/870.3600) → DTE — Pulse-Generator, Pacemaker, External

# DTE · Pulse-Generator, Pacemaker, External

_Cardiovascular · 21 CFR 870.3600 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE

## Overview

- **Product Code:** DTE
- **Device Name:** Pulse-Generator, Pacemaker, External
- **Regulation:** [21 CFR 870.3600](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/870.3600)
- **Device Class:** 2
- **Review Panel:** [Cardiovascular](/submissions/CV)

## Identification

An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source (
*i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
(i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices (
*e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use;
(ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals;
(iii) The labeling must list all pacing modes available in the device;
(iv) Labeling must include a detailed description of any special capabilities (
*e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.

## Recent Cleared Devices (20 of 65)

Showing 20 most recent of 65 cleared devices.

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K241781](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K241781.md) | Solo Pace Control | Solo Pace, Inc. | Jan 10, 2025 | SESE |
| [K232721](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K232721.md) | Lifetech Cardio Temporary Pacemaker | Shenzhen Lifetech Cardio Medical Electronics Co., Ltd. | Jan 7, 2024 | SESE |
| [K201011](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K201011.md) | Medtronic Model 5392 External Pulse Generator (EPG) | Medtronic, Inc. | May 13, 2020 | SESE |
| [K190825](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K190825.md) | Medtronic Model 5392 External Pulse Generator (EPG) | Medtronic, Inc. | Apr 30, 2019 | SESE |
| [K182839](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K182839.md) | Lifetech Cardio Model 8301 Temporary Pacemaker | Shenzhen Lifetech Cardio Medical Electronics Co., Ltd. | Feb 28, 2019 | SESE |
| [K181973](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K181973.md) | Medtronic Model 5392 External Pulse Generator (EPG) | Medtronic, Inc. | Aug 21, 2018 | SESE |
| [K180873](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K180873.md) | Medtronic Model 53401 External Pulse Generator (EPG) | Medtronic, Inc. | Apr 19, 2018 | SESE |
| [K162054](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K162054.md) | Medtronic Temporary External Pacemaker 53401 | Medtronic, Inc. | Oct 18, 2016 | SESE |
| [K162550](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K162550.md) | Medtronic Model 5392 External Pulse Generator (EPG) | Medtronic, Inc. | Oct 12, 2016 | SESE |
| [K150246](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K150246.md) | Medtronic Model 5392 External Pulse Generator (EPG) | Medtronic, Inc. | Feb 18, 2015 | SESE |
| [K132924](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K132924.md) | MEDTRONIC EXTERNAL PULSE GENERATOR (EPG) | Medtronic, Inc. | Oct 31, 2013 | SESE |
| [K123916](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K123916.md) | PSA SERIES | Osypka Medical, Inc. | Apr 9, 2013 | SESK |
| [K042708](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K042708.md) | ERA 3000 PACING SYSTEM ANALYZER | Biotronik, Inc. | Oct 20, 2004 | SESE |
| [K033130](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K033130.md) | PACE 101/PACE 101H, MODEL 3077 SSI TEMPORARY PULSE GENERATOR | Osypka Medical, Inc. | Oct 31, 2003 | SESE |
| [K023080](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K023080.md) | GALIX PACESTAR | Galix Biomedical Instrumentation, Inc. | Jun 11, 2003 | SESE |
| [K022939](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K022939.md) | OSCOR PACE 101/101H AND ACCESSORIES, ST.JUDE MEDICAL MODEL 3077 AND ACCESSORIES,CARDIOTRONIC PACE 101 AND ACCESSORIES | Osypka Medical, Inc. | Oct 3, 2002 | SESE |
| [K020896](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K020896.md) | OSCOR PACE 203/PACE 203H DUAL-CHAMBER TEMPORARY PACEMAKER, ST. JUDE MEDICAL, MODEL 3085 DUAL-CHAMBER EXTERNAL PULSE GENE | Osypka Medical, Inc. | Jun 14, 2002 | SESE |
| [K011980](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K011980.md) | ADAPTER, PACING ANALYZER/UNIVERSAL PACING ADAPTER, MODELS 4820 AND 4825 | Pace Medical | Sep 12, 2001 | SESE |
| [K002204](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K002204.md) | PACEMAKER EXTENSION CABLE MODELS 4265,4265A, 4265V, 5265, 5265A, 5265V | Pace Medical | Oct 18, 2000 | SESE |
| [K001180](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE/K001180.md) | TEMPORARY PACEMAKER EXTENSION CABLE, MODELS 5235, 5235A, 5235V, TEMPORARY PACEMAKER SURGICAL EXTENSION CABLE, MODELS 528 | Pace Medical | Jun 27, 2000 | SESE |

## Top Applicants

- Seamed Corp. — 13 clearances
- Pace Medical — 9 clearances
- Medtronic, Inc. — 8 clearances
- Medtronic Vascular — 5 clearances
- Cynergy — 4 clearances

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTE)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
