← Product Code [DTC](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTC) · K011729

# SIGMA PACE EXTERNAL PACEMAKER ANALYZER MODEL #1000 (K011729)

_Dni Nevada, Inc. · DTC · Aug 30, 2001 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTC/K011729

## Device Facts

- **Applicant:** Dni Nevada, Inc.
- **Product Code:** [DTC](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTC.md)
- **Decision Date:** Aug 30, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3630
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The SigmaPace 1000 External Pacemaker Analyzer is intended for use as regular preventative maintenance and quality assurance checks on external cardiac pacemakers in both the pre-hospital and hospital based environments. The SigmaPace 1000 is limited to preventative maintenance checks for transcutaneous and transvenous types of temporary external cardiac pacemakers. The SigmaPace 1000 will perform a full range of specific testing procedures and protocols developed by the external cardiac pacemaker manufacturers. The SigmaPace 1000 will also provide ECG simulation capabilities for applications in clinical training and product demonstration. The SigmaPace 1000 is intended to be used by biomedical technicians, OEM manufacturers, and third party service facilities to perform preventative maintenance and quality assurance on external cardiac pacemakers, to verify the operation of devices returned from repair, and to assist repair facilities in the diagnosis of intermittent performance problems. It is also used by training facilities to train medical personnel. The SigmaPace 1000 is not used on patients and does not perform any diagnostic, therapeutic, or monitoring functions. Additionally, these devices are not used to test any programmable, implantable pacemakers or any related indwelling cardiovascular catheters or lead wires.

## Device Story

SigmaPace 1000 is a dual-channel external pacemaker analyzer; tests transcutaneous and transvenous temporary pacemakers. Inputs: pacemaker leads connected to device ports; user interface via front panel buttons or RS232 serial port. Device performs pulse current, rate, width, energy, demand/async mode, amplitude sensitivity, noise immunity, refractory period, DC leakage, current drain, and output stability tests. Includes manufacturer-specific algorithms for testing specific pacemaker models; provides ECG simulation for training. Used in clinical/hospital settings by biomedical technicians and service personnel. Output: test results displayed on front panel; used by technicians to verify performance, isolate intermittent problems, and perform preventative maintenance. Benefits: ensures reliability of external pacemakers before clinical use.

## Clinical Evidence

No clinical data. Device is test equipment not used on patients; does not perform diagnostic, therapeutic, or monitoring functions. Verification and validation performed via bench testing of features including menu structure, test loads, battery circuitry, transcutaneous/transvenous channel tests, microprocessor clock, and algorithms.

## Technological Characteristics

Dual-channel analyzer; internal lithium battery; RS232 connectivity. Meets IEC 1010 safety standards. Features include 5-lead and high-level ECG output, selectable test loads, and manufacturer-specific algorithms for pacemaker testing. Dimensions/form factor not specified.

## Regulatory Identification

A pacemaker generator function analyzer is a device that is connected to a pacemaker pulse generator to test any or all of the generator's parameters, including pulse duration, pulse amplitude, pulse rate, and sensing threshold.

## Predicate Devices

- Netech, Model EXPMT 100 External Pacemaker Tester ([K661582](/device/K661582.md))
- Bio-Tek Inc, Model PMA-1 Pacemaker Analyzer ([K903966](/device/K903966.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 3 0 2001

K011729 p.1/3

## DNI NEVADA INC. 510(k) SigmaPace 1000 External Pacemaker Analyzer

# 1 - 510(k) SUMMARY

#### Contact Person: 2-

Kristine G. Boggs, Quality Assurance Manager 775-883-3400 x265 (phone) 775-883-9541 (fax) boggsk@dninevada.com Establishment Registration Number: 2921581 Establishment will be both manufacturing and marketing SigmaPace 1000 Preparation Date: May 29, 2001

Classification Name: 3-

> Common Name: Proprietary Name:

Analyzer, Pacemaker Generator Function, Product, DTC, Class II External Pacemaker Analyzer SigmaPace 1000

The device is substantially equivalent to the 4- Substantially Equivalent To: following two legally marketed predicate devices:

- Netech, Model EXPMT 100 External Pacemaker Tester, -#K661582
- Bio-Tek Inc, Model PMA-1 Pacemaker Analyzer, #K903966 ー

SigmaPace 1000 and the predicate External Pacemaker Analyzers have the identical intended use and there are no new technological features or issues that would raise concern of safety or effectiveness.

5 & 6 - Intended Use and Description of Device: This device is intended for use as an external pacemaker analyzer. It is intended to be used by biomedical technicians, third party service organizations, and manufacturers to perform regular preventative maintenance checks on external pacemakers, to verify operations of external pacemakers returned from repair, to assist repair technicians in isolating intermittent performance problems, and as a tool for training medical personnel in the use of external cardiac pacemakers. This device is not to be used as an external cardiac pacemaker calibrator or to make any adjustments based on SigmaPace 1000 test results. This device is not used on patients and does not perform any monitoring, diagnosing, or therapeutic functions.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text "K011729 p.2/3" in a handwritten style. The text appears to be a document identifier or page number. The characters are written in black ink on a white background, and the handwriting is legible.

The SigmaPace 1000 is a device used to verify the performance of various parameters I he ciginal acc 1000 is a don'ts. This is done by connecting the external leads of the of external cardido pacematers: "Ital inputs on the SigmaPace 1000 and selecting the passemater to eas or by using the press buttons on the front panel display or via approprate test childr by same and port (RS232). The SigmaPace 1000 can perform a variety of Transcutaneous and Transvenous (Atrial and Ventricular) Pacer tests.

Pulse current, rate width , and energy ー

- Qualitative demand and async mode tests -
- Amplitude sensitivity tests -
- Noise immunity tests
- Paced and Sensed Refractory tests -
- Selectable model specific transcutaneous algorithms 1
- Selectable test loads -
- DC leakage current (tranvenous) -
- Current drain test (transvenous)
- Long term test for pacer output stability
- ECG simulation for training activities

7 ~ Summary of Technological Characteristics of Device Compared to Predicate: The SigmaPace 1000 is a dual channel device capable of testing external cardiac pacemakers. The predicate devices have the same intended use and perform most of the same functions as the SigmaPace 1000.

The SigmaPace 1000 is different in that is has dual channel input instead of single channel inputs to allow four pacemaker leads to be tested instead of two. The SigmaPace 1000 also has 5 lead and high level ECG output to allow it to be used with monitors and similar devices for training purposes. DNI has also included a number of manufacturer specific algorithms for improving testing of specific external cardiac pacemaker models. Last, the SigmaPace 1000 has increased test load selections over the predicate devices.

### 8 - Performance Testing: (Verification and Validation)

This device has undergone and continues to go through a variety of verification and validation testing. Each feature is tested to ensure it meets the requirements as defined in the product specification and advertising. To date, testing has been successfully completed for menu structure, test load selections, battery circuitry, transcutaneous test (ventricular channel), tranvenous tests (both atrial and ventricular channels), internal microprocessor clock frequency, algorithms testing, baud rate, and serial port operation. The data report for the first level validation as well as the full validation plan can be found in the Appendix of this submission.

{2}------------------------------------------------

K011729 p-3/3

9- Clinical Testing: Clinical testing was not required since the device is test equipment, and is never in contact with a patient nor do they have any diagnostic, therapeutic, or monitoring patient function. It is not intended to be used for calibrations and should not be used for clinical calls.

10 - Conclusions from testing: The testing conducted to date and that will be conducted prior to release will support all product claims for intended use including accuracy and full feature operation.

### 11 - Other Information of Interest to the FDA:

Potential System Hazards: The primary system function hazards were determined, reviewed, and addressed. The hazard analysis was aimed at identifying were error could occur which could cause believable but incorrect results. The hazard analysis detail can be found in the Appendix of this submission. In summary, the analysis looked at potential errors with keypad failures, display problems, serial port failures, LED indicator failure, failure of measurement circuitry or stimulus circuitry to be out of tolerance or improperly working, battery failure conditions, including shorts and low or no battery, short circuit of DN input node, and defibrillator discharge into Sigma Pace 1000.

User Safety Consideration: The SigmaPace 1000 is designed to meet IEC 1010 safety standards. The device itself uses an internal lithium battery with a battery charger that meets UL/CSA safety requirements for test equipment. IEC 1010 safety test leads are sent with each unit sold for further user protection.

The above information is certified to be truthful and accurate to the best of my knowledge.

1Lt. Lt. Bopp

Quality Assurance Manager

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 3 0 2001

Ms. Kristine G. Boggs Quality Assurance Manager DNI Nevada, Inc. 2000 Arrowhead Drive Carson City, Nevada 89706-0403

Re: K011729

Trade Name: SigmaPace 1000 External Pacemaker Analyzer Regulation Number: 21 CFR 870.3630 Regulatory Class: II (two) Product Code: 74 DTC Dated: May 31, 2001 Received: June 4, 2001

Dear Ms. Boggs:

We have reviewed your Section 510(k) notification of intent to market the device referenced above we have determined the device is substantially equivalent (for the indications for use stated in and we acteenmally marketed predicate devices marketed in interstate commerce prior to the cholosure) to regally manifore of the Medical Device Amendments, or to devices that have been May 20, 1970, the enables with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The r ou may, are on on of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your I spec rary, it the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the

{4}------------------------------------------------

Page 2 – Ms. Kristine G. Boggs

Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noutication: "The FDF intaing of succement for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific arriver diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in This Chaplestions on the promotion and advertising of your device, (201) 594-1015. Traditionally, 2011 proce at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mornfacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsma/dsmamain.html.

Sincerely yours,

K.G. Till

mes E. Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices Radiological Health

Enclosure

{5}------------------------------------------------

# 510(k) Number (if known): _ KO1 | 729

## Device Name: SigmaPace 1000 External Pacemaker Analyzer

Indications for Use: The SigmaPace 1000 External Pacemaker Analyzer is intended for-Indications for USe. "The Digmatenance and quality assurance checks on external cardiac use as regular proventual oth the pre-hospital and hospital based environments.

The SigmaPace 1000 is limited to preventative maintenance checks for transcutancous The Urginal aod 1000 is temporary external cardiac pacemakers. The SigmaPace 1000 and transvenous types of temporary vecific testing procedures and protocols developed will perform a full fungo of anaker manufacturers. The SigmaPace 1000 will also by the external cardias pacemanes ulation capabilities for applications in clinical training and product demonstration.

The SigmaPace 1000 is intended to be used by biomedical technicians, OEM The Sigmal acc 1000 is metrosa vice facilities to perform preventative maintenance and manufacturers, and unra pary a cardiac pacemakers, to verify the operation of devices returned from repair, and to assist repair facilities in the diagnosis of intermittent devices routhed from repair as a sed by training facilities to train medical personnel.

The SigmaPace 1000 is not used on patients and does not perform any diagnostic, the orginial aos 1000 is ng functions. Additionally, these devices are not used to test any programmable, implantable pacemakers or any related indwelling cardiovascular catheters or lead wires.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number R011729

Prescription Use L (Per 21 CFR Part 801.109)

OR

Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTC/K011729](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DTC/K011729)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
