K963738 · Ela Medical, Inc. · DTB · May 23, 1997 · Cardiovascular
Device Facts
Record ID
K963738
Device Name
STELA MODEL BJ44 AND BJ45 PACING LEADS
Applicant
Ela Medical, Inc.
Product Code
DTB · Cardiovascular
Decision Date
May 23, 1997
Decision
PT
Submission Type
Traditional
Regulation
21 CFR 870.3680
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
ELA Medical endocardial leads are designed to be used with implantable cardiac pacemakers. ELA Medical Stela™ Model BJ44 and BJ45 bipolar pacing leads are intended for atrial implantation only.
Device Story
Stela Model BJ44 and BJ45 are silicone rubber, tined, J-shaped, bipolar transvenous leads; provide permanent electrical pathway between pacemaker and atrium. Used by physicians for cardiac pacing and sensing. Device functions as passive lead; no active processing or software. Clinical benefit includes reliable atrial pacing/sensing. Postmarket surveillance and device tracking required.
Clinical Evidence
Limited clinical study performed; demonstrated appropriate pacing and sensing characteristics with no observed complications. Bench testing included stylet insertion/removal, electrical resistance, IS-1 connector conformity, leak resistance, tensile strength, insulation integrity, and conductor coil fatigue.
Technological Characteristics
Silicone rubber, tined, J-shaped, bipolar transvenous leads. Features IS-1 connector. Materials consistent with previously cleared lead products. No software or electronic components.
Indications for Use
Indicated for patients requiring cardiac pacing and sensing via atrial implantation using an implantable cardiac pacemaker.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Predicate Devices
ELA Medical Focus® Model T44F bipolar tined silicone rubber pacing lead (K904255)
Focus® Model J83F tined silicone rubber pacing lead (K896742)
Related Devices
K972574 — STELA MODEL UJ45 PACING LEADS · Ela Medical, Inc. · Jan 8, 1998
K993448 — STELA UT46 AND BT45/46 STRAIGHT TINED PACING LEADS, STELA UJ45 AND BJ44/45 J-SHAPED TINED PACING LEADS · Ela Medical, Inc. · Apr 10, 2000
K964604 — BIPOLAR IMPLANTABLE ENDOCARDIAL PACING LEADS, WITH SILICONE RUBBER INSULATION AND PASSIVE FIXATION · Biotronik, Inc. · Mar 10, 1997
{0}
K963738
510(k) Notification for Stela leads, Models BJ44/BJ45
Page 11 / 73
# 2. Summary & Certification
MAY 23 1997
## 2.1 Summary of safety and effectiveness information
### 2.1.1 General Information
Device Generic Name: Atrial tined pacing lead.
Device Trade Name: Stela™ Model BJ44 and BJ45 Pacing Leads
Applicant's Name and Address: ELA Medical, Inc., 15245 Minnetonka Boulevard, Minnetonka, MN 55345, Tel. (612) 935-2033
Date of Summary Preparation: September 16th 1996
Contact Person: Cathy G. Goble
510(k) Number: K963698
Date of Judgment of Substantial Equivalence Sent to Applicant:
Predicate Devices: ELA Medical Focus® Model T44F bipolar tined silicone rubber pacing lead (Document Control Number K904255) and Focus® Model J83F tined silicone rubber pacing lead (Document Control Number K896742).
### 2.1.2 Description of Conditions for Which the Devices are Indicated
The Stela™ Model BJ44 and BJ45 leads are indicated for cardiac pacing and sensing, which is the same as other transvenous tined leads.
### 2.1.3 Device Description
Stela™ BJ44 and BJ45 models are silicone rubber, tined, J-shaped, bipolar transvenous leads that provide a permanent electrical pathway between a pacemaker and the atrium. They are similar in design and construction to other such leads in commercial distribution.
### 2.1.4 Alternatives
The alternatives for Stela™ Model BJ44 and BJ45 leads are other commercially available transvenous pacing leads.
ELA Medical Incorporated
Confidential and proprietary
{1}
510(k) Notification for Stela leads, Models BJ44/BJ45
Page 12 / 73
## 2.1.5 Marketing History
Stela™ Model BJ44 and BJ45 leads are not in commercial distribution in the U.S. They were recently introduced into commercial distribution outside the U.S. No unanticipated adverse device effects have been reported for these leads.
## 2.1.6 Potential Adverse Effects
The potential adverse effects of Stela™ Model BJ44 and BJ45 leads are the same as those for tined, J-shape, transvenous leads in commercial distribution. Lead-related complications are described in the ELA Medical generic lead manual.
## 2.1.7 Summary of Studies
The following in-vitro functional testing was performed on the Stela™ Model BJ44 and/or BJ45 leads: stylet insertion/removal, stylet bottoming, electrical resistance, IS-1 connector conformity, leak resistance, distal tip electrode/conductor coil tensile strength, insulation integrity, conductor coil fatigue, and tip flexibility. Biocompatibility testing was not performed, due to the successful history with the same materials in other lead products. Sterilization testing and mechanical / environmental packaging validation were performed, because the lead package design and sterilization method changed. All test results demonstrated that the established pass / fail criterion was met in all cases.
A limited clinical study demonstrated that the Model BJ45 lead exhibits appropriate pacing and sensing characteristics. No complications were observed in this study.
## 2.1.8 Conclusion
The information presented in this submission provides reasonable assurance that the Stela™ Model BJ44 and/or BJ45 leads will perform in a safe and effective manner.
ELA Medical Incorporated
Confidential and proprietary
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 23 1997
Ms. Catherine G. Goble
Regulatory Affairs Manager
ELA Medical, Inc.
2950 Xenium Lane North
Plymouth, Minnesota 55441
Re: K963738
Stela Models BJ44 and BJ45 Pacing Leads
Regulatory Class: III (three)
Product Code: DTB
Dated: February 21, 1997
Received: February 25, 1997
Dear Ms. Goble:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (act). The general controls provisions of the act include requirements for registration, listing of devices, good manufacturing practices, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulation.
{3}
Page 2 - Ms. Catherine G. Goble
Under Section 522(a) of the act, manufacturers of certain types of devices identified by the Act or designated by FDA are required to conduct postmarket surveillance studies. FDA has identified under Section 522(a)(1)(A) the device cleared for marketing by this letter as requiring postmarket surveillance.
Within thirty (30) days of first introduction or delivery for introduction of this device into interstate commerce you are required to submit to FDA certification of the date of introduction into interstate commerce, a detailed protocol which describes the postmarket surveillance study, and a detailed profile of the study's principal investigator that clearly establishes the qualifications and experience of the individual to conduct the proposed study. For your information, general guidance on preparing a protocol for a postmarket surveillance study is attached.
Submit five (5) copies to:
Center for Devices and Radiological Health
Postmarket Surveillance Studies Document Center
Room 3083 (HFZ-544)
1350 Piccard Drive
Rockville, Maryland 20850
Within sixty (60) days of receipt of your protocol, FDA will either approve or disapprove it and notify you of the Agency's action in writing. You should not begin your postmarket surveillance study of this device until the protocol has been approved. Data generated under an unapproved protocol may not satisfy your obligation under section 522. Please note that you must continue to collect and report data needed to maintain compliance with Medical Device Reporting regulations (21 CFR 803).
Failure to certify accurately the date of initial introduction of your device into interstate commerce, to submit timely an acceptable protocol, or to undertake and complete and FDA approved postmarket surveillance study consistent with the protocol will be considered violations of section 522. In accordance with the Medical Device Amendments of 1992, failure of a manufacturer to meet its obligations under section 522 is a prohibited act under section 301(q)(1)(C) of the Act (21 U.S.C. 331 (q)(1)(C)). Further, under section 502(t)(3) of the act (21 U.S.C. 352(t)(3)), a device is misbranded if there is a failure or refusal to comply with any requirement under section 522 of the act. Violations of sections 301 or 502 may lead to regulatory actions including seizure of your product, injunction, prosecution, or civil money penalties.
{4}
Page 3 - Ms. Catherine G. Goble
If you have questions specifically concerning postmarket surveillance study requirements, contact the Postmarket Surveillance Studies Branch at (301) 594-0639.
In addition, on August 16, 1993, the Final Rule for Device Tracking was published in the *Federal Register*, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirement of the regulation are found in 21 CFR 821 as described in the August 16, 1993 *Federal Register* beginning on page 43447.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Wolff Saperlin, MO
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosures
{5}
510(k) Notification for Stela leads, Models BJ44/BJ45 Page 10 / 73
## 1.10 Indications for Use Statement
510 (k) Number: K963738
Device Name: Stela™ Model BJ44 and BJ45 pacing leads.
### Indication for Use:
ELA Medical endocardial leads are designed to be used with implantable cardiac pacemakers.
ELA Medical Stela™ Model BJ44 and BJ45 bipolar pacing leads are intended for atrial implantation only.
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K963735
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☐ OR Over-The-Counter Use ☐
(Per 21 CFR 801.109)
ELA Medical Incorporated Confidential and proprietary
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.