K954610 · Intermedics, Inc. · DTB · Dec 16, 1996 · Cardiovascular
Device Facts
Record ID
K954610
Device Name
MODEL 430-07 ENDOCARDIAL PACING LEAD
Applicant
Intermedics, Inc.
Product Code
DTB · Cardiovascular
Decision Date
Dec 16, 1996
Decision
PT
Submission Type
Traditional
Regulation
21 CFR 870.3680
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The Intermedics leads which are the subject of this 510(k) Notification are endocardial pacing leads designed for use with implantable pulse generators for long-term cardiac pacing. This indication is identical to the indication for use of the legally marketed predicate devices, the Intermedics Models 430-07 and 432-03 Endocardial Pacing Leads. The indications for ventricular pacing include, but are not limited to: sick sinus syndrome, sinus bradycardia, complete heart block, symptomatic second-degree heart block, and certain conditions of asymptomatic second-degree block. In the presence of normal atrioventricular (A-V) conduction, the indications for atrial pacing include, but are not limited to: sinus arrest, sick sinus syndrome, sinus bradycardia and conditions requiring increased cardiac efficiency, enhanced cardiac output or the overdrive of certain cardiac arrhythmias. In the absence of normal A-V conduction, an atrial lead may be used with a ventricular lead in a dual-chamber pacing system to restore A-V synchrony.
Device Story
Endocardial pacing leads (Models 430-07, 431-07, 432-03, 433-03) designed for long-term cardiac pacing; implanted transvenously via cephalic, subclavian, or jugular vein. Tip electrode (cathode) positioned at endocardial surface; fixation via four silicone rubber tines. Lead body carries electrical impulses between heart and implanted pulse generator for sensing and cardiac depolarization. Electrodes feature iridium oxide-coated (IROX) titanium tip. Device enables low stimulation thresholds (mean threshold ≤ 0.38 ms at 1.5V). Used by physicians in clinical settings to manage bradycardia and heart block; provides patient benefit through reliable, low-energy cardiac pacing.
Clinical Evidence
Clinical study of Models 430-07 and 432-03 over 12 months. Primary endpoint: stimulation threshold at 1.5V. Results: mean pacing threshold 0.32 ms (ventricle) and 0.35 ms (atrium) at 2 weeks, decreasing to 0.26 ms in both chambers at 6 months. 97.5% upper confidence bound for capturing pulse width established at ≤ 0.38 ms at 1.5V. Animal study (12 dogs, 24 leads) compared IROX, IROX-PEG, and Biopore electrodes, demonstrating superior low-threshold performance for IROX-based electrodes.
Technological Characteristics
Endocardial pacing leads; bipolar or unipolar configurations. Tip electrode: titanium coated with iridium oxide (IROX). Fixation: four silicone rubber tines. Lead body: insulated conductor coils. Designed for transvenous implantation. No software or energy source integral to the lead itself (passive component).
Indications for Use
Indicated for patients requiring long-term cardiac pacing due to sick sinus syndrome, sinus bradycardia, complete heart block, symptomatic second-degree heart block, asymptomatic second-degree block, sinus arrest, or conditions requiring increased cardiac efficiency/output or overdrive of cardiac arrhythmias.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Predicate Devices
Intermedics Models 430-07 Endocardial Pacing Lead
Intermedics Models 432-03 Endocardial Pacing Lead
Related Devices
K955122 — INTERMEDICS MODEL 436-07 BIPOLAR & MODEL 437-07 UNIPOLAR IMPLANTABLE ENDOCARDIAL PACING LEAD · Intermedics, Inc. · Nov 4, 1996
K962174 — BIPLR IM EN PAC LEADS (430-07,432-03,436-02 & 07, 438-05 & 07) · Intermedics, Inc. · Aug 21, 1996
K955550 — CARDIFIX EZ BIPOLAR/UNIPOLAR ENDOCARDIAL PACING LEAD · Intermedics, Inc. · Jun 19, 1996
K993139 — YP 45/15-BP; YP 53/15-BP; YP 60/15-BP · Biotronik, Inc. · Dec 17, 1999
Submission Summary (Full Text)
{0}
DEC-06-'96 16:19 ID: JEVICES FOR FUTURE TEL NO:713P6646070 #453 P02
K954610
Intermedics Inc.
A company of SULZERmedica
DEC 16 1996
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
**Submitter's Name:** Intermedics, Inc.
4000 Technology Drive
Angleton, TX 77515 USA
Telephone: (409) 848-4525
**Contact Person:** Alan D. Totah, Director of Domestic Regulatory Affairs
**Date of Summary:** December 6, 1996
**Device Names:**
- Model 430-07 Endocardial Pacing Lead
- Model 431-07 Endocardial Pacing Lead
- Model 432-03 Endocardial Pacing Lead
- Model 433-03 Endocardial Pacing Lead
**Legally Marketed Devices To Which Equivalence Is Claimed:** The Intermedics Models 430-07 and 432-03 Endocardial Pacing Leads.
The Intermedics lead models which are the subject of this 510(k) Notification are legally marketed devices. This 510(k) Notification is submitted to provide data and evidence to support labeling claims concerning the low threshold performance of the iridium oxide-coated (IROX™) titanium tip electrode. In this submission, the Models 430-07 and 432-03 Endocardial Pacing Leads are considered the legally marketed predicate devices since these models were employed in the clinical study which yielded the low threshold data in humans. Since the IROX tip electrode is identical among all lead models in design, materials and construction, the clinical data gathered on the performance of the Model 430-07 and 432-03 Leads are directly applicable to the performance of the other lead models, and the claims concerning low threshold performance are therefore made for all lead models.
**New Performance Claim:** This 510(k) Notification is submitted to provide data and evidence to support the following labeling claim concerning the low threshold performance of the iridium oxide-coated (IROX) titanium tip electrode:
The mean stimulation threshold, both acute and chronic, is equal to or less than 0.38 millisecond at a pulse amplitude of 1.5 volts, with a 97.5 percent confidence limit.
**Device Description:** The Intermedics Model 430-07, 431-07, 432-03, and 433-03 Endocardial Pacing Leads are designed for use with implantable cardiac pulse generators for long-term cardiac pacing. Passed transvenously through the cephalic, subclavian, or jugular
{1}
DEC-06-'96 16:20 ID:NEVICES FOR FUTURE TEL NO:713P6646070 #453 P03
vein (external or internal), the tip electrode (cathode) is positioned at the endocardial surface to permit contact with the heart. The tip fixation mechanism consists of four silicone rubber tines. The lead body, which consists of insulated conductor coils, carries electrical impulses from the heart to the pulse generator, resulting in sensing. The lead body also carries impulses from the pulse generator to the endocardial surface, causing cardiac depolarization (pacing). The proximal end of the lead is connected to the header of the implanted pulse generator.
The electrodes of the Intermedics Model 430-07, 431-07, 432-03, and 433-03 Endocardial Pacing Leads are made of titanium coated with iridium oxide (IROX™). Clinical trials with the Models 430-07 and 432-03 IROX endocardial pacing leads through twelve months demonstrated low threshold performance: 1.5 volts at or below 0.38 millisecond pulse width, with 97.5% confidence
Intended Use: The Intermedics leads which are the subject of this 510(k) Notification are endocardial pacing leads designed for use with implantable pulse generators for long-term cardiac pacing. This indication is identical to the indication for use of the legally marketed predicate devices, the Intermedics Models 430-07 and 432-03 Endocardial Pacing Leads.
The indications for ventricular pacing include, but are not limited to: sick sinus syndrome, sinus bradycardia, complete heart block, symptomatic second-degree heart block, and certain conditions of asymptomatic second-degree block. In the presence of normal atrioventricular (A-V) conduction, the indications for atrial pacing include, but are not limited to: sinus arrest, sick sinus syndrome, sinus bradycardia and conditions requiring increased cardiac efficiency, enhanced cardiac output or the overdrive of certain cardiac arrhythmias. In the absence of normal A-V conduction, an atrial lead may be used with a ventricular lead in a dual-chamber pacing system to restore A-V synchrony.
Descriptive Summary Of Technological Characteristics And Those Of Predicate Devices: Following is a summary of the technological characteristics of the Models 430-07 and 432-03 Endocardial Pacing Leads (predicate devices) and those of the other lead models which are the subject of this 510(k) Notification:
- The Model 430-07 is a bipolar lead, designed for use in the ventricle.
- The Model 431-07 is a unipolar lead, designed for use in the ventricle.
- The Model 432-03 is a bipolar lead, designed for use in the atrium.
- The Model 433-03 is a unipolar lead, designed for use in the atrium.
All these leads have the same tip electrode, identical in design (slotted blunt tip), materials (titanium coated with iridium oxide), and construction. The differences among the lead models do not pertain to new indications for use nor to new technological characteristics. These differences pertain only to the polarity of the lead and to the heart chamber in which the lead is implanted. The differences among the lead models do not affect the safety and effectiveness of the devices, nor do they affect the performance claim which is the subject of this 510(k) Notification.
{2}
DEC-06-'96 16:21 ID:DEVICES FOR FUTURE TE NO:713P6 45070 #453 P04
Performance Data: Animal and human clinical performance data, acute and chronic, have been gathered on Intermedics Endocardial Pacing Leads with iridium oxide-coated titanium tip electrodes.
## Animal Study
In order to evaluate the electrical performance of new electrode technologies, Intermedics conducted a long-term animal study of endocardial leads. Twenty-four leads were implanted in twelve dogs. All leads were unipolar, 5 French lead body, polyurethane insulated tined leads. Eight of each of the following leads were used in the study:
- Biopore™ Leads ("Biopore"), with a carbon-coated porous titanium electrode;
- IROX™ Leads ("IROX"), with an iridium oxide-coated titanium electrode; and
- IROX™ Leads ("IROX-PEG"), with an iridium oxide-coated titanium electrode layered with a thin coating of polyethylene glycol (PEG).
Stimulation thresholds were measured at regular intervals for 24 weeks. The IROX-PEG stimulation thresholds rose only 0.23 volts from implant to the chronic period (after week 6), and were significantly lower than those of Biopore and IROX. Low acute values were observed for all leads, but the IROX-PEG electrode demonstrated lower subchronic, peak and chronic values, with the mean pacing threshold less than 0.6 volts at 0.6 millisecond pulse width at 24 weeks. The data from this study establish the low threshold characteristics of the IROX-PEG electrode.
## Clinical Study
Clinical performance data have been gathered on the Models 430-07 and 432-03 Endocardial Pacing Leads to support the performance claim of low acute and chronic stimulation threshold. Pacing threshold data were reported at implant, two weeks, four weeks, six weeks, three months, six months and twelve months. The data showed a mean pacing threshold of 0.32 millisecond in the ventricle and 0.35 millisecond in the atrium at 1.5 volts at the two-week follow-up point; this threshold gradually decreased to a mean value of 0.26 millisecond in both chambers at the six-month follow-up visit. Statistical analysis demonstrated that the dependence of capturing pulse width on pulse amplitude remains invariant with respect to time elapsed from implant. An upper confidence bound of 97.5% on the capturing pulse width was computed to be at or below 0.38 millisecond at a pulse amplitude of 1.5 volts.
The results of the clinical study establish the low threshold performance of the Models 430-07 and 432-03 Endocardial Pacing Leads on both an acute and a chronic basis, and support the low threshold labeling claim proposed by Intermedics.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.