UNIPOLAR & BIOPLAR IMPLANTABLE ENDOCARDIAL PACING LEADS, WITH SILICONE RUBBER INSULATION & PASSIVE FIXATION

K953044 · Biotronik, Inc. · DTB · Sep 27, 1996 · Cardiovascular

Device Facts

Record IDK953044
Device NameUNIPOLAR & BIOPLAR IMPLANTABLE ENDOCARDIAL PACING LEADS, WITH SILICONE RUBBER INSULATION & PASSIVE FIXATION
ApplicantBiotronik, Inc.
Product CodeDTB · Cardiovascular
Decision DateSep 27, 1996
DecisionPT
Submission TypeTraditional
Regulation21 CFR 870.3680
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The BIOTRONIK TIR and TIJ pacing leads are safe and effective transvenous, implantable, endocardial leads used with implantable cardiac pacemakers.

Device Story

TIR/TIJ leads are transvenous, implantable, endocardial cardiac pacing leads. They function as the electrical interface between an implantable pacemaker and the heart muscle. The device features a lenticular electrode tip with a fractal-surfaced iridium coating (applied via Physical Vapor Deposition) to optimize pacing/sensing. Passive fixation is achieved via four flexible silicone rubber tines that engage with cardiac trabeculae. The lead body is insulated with NuSil silicone rubber and utilizes a quadrifilar MP35N conductor. The TIJ model is pre-formed into a 'J' shape for atrial placement. The leads utilize an IS-1 connection system. They are implanted by physicians in a clinical setting. The device delivers electrical impulses from the pacemaker to the heart and transmits cardiac electrical signals back to the pacemaker, facilitating therapy for patients with cardiac rhythm disorders.

Clinical Evidence

Clinical evidence includes European clinical studies of the IRTI/IRTJ models (European versions of TIR/TIJ). The studies evaluated acute and long-term sensing and pacing thresholds, and long-term impedance. Pre-clinical evidence includes biocompatibility testing (cytotoxicity, hemolysis, pyrogenicity) and long-term corrosion studies confirming the safety and durability of the iridium material. Qualification testing validated lead design, including fatigue strength, environmental resistance, and connection integrity.

Technological Characteristics

Materials: NuSil silicone rubber (insulation), MP35N (conductor), Iridium/Titanium (electrode tip), Platinum-Iridium (ring electrode). Fractal surface treatment via Physical Vapor Deposition (PVD). Connection: IS-1 (ISO 5841.3:1992). Configuration: Unipolar and bipolar. Fixation: Passive (four silicone tines).

Indications for Use

Indicated for patients requiring cardiac pacing via transvenous, implantable, endocardial leads connected to an implantable cardiac pacemaker.

Regulatory Classification

Identification

Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.

Related Devices

Submission Summary (Full Text)

{0} K953044 ^{}[] ## B. Safety And Effectiveness Summary The BIOTRONIK TIR and TIJ pacing leads are safe and effective transvenous, implantable, endocardial leads used with implantable cardiac pacemakers. The leads are manufactured in unipolar and bipolar configurations. The lead body insulation of all TIR/TIJ endocardial leads is NuSil silicone rubber tubing, with the conductor of quadrifilar MP35N. The TIR/TIJ endocardial leads provide long-term safe and effective pacing due to the surface structure of the lenticular electrode tip. The tip has a Physical Vapor Deposition (PVD) of iridium over titanium, creating a fractal-surfaced ball-like microstructure. Passive fixation in the heart's trabeculae is provided by four flexible silicone rubber tines. The ring electrode (anode) of the bipolar lead is made of platinum-iridium, also with a fractal iridium surface treatment. The IS-1 connection system of TIR/TIJ leads complies with the International Standard ISO 5841.3:1992: Low Profile Connectors. The TIJ lead is pre-formed into a “J” shape for optimal positioning in the atrium. Pre-clinical testing was performed on the leads to examine biocompatibility and to qualify the manufacturing process of these leads. During biocompatibility testing, cytotoxicity, hemolysis, pyrogenicity (LAL) and other testing were performed, as well as long term studies. Corrosion studies were completed to address both long-term toxicity and durability of the material. The testing conducted for biocompatibility as well as extensive clinical experience confirms that iridium is safe for use as an implantable biocompatible material. Literature has been published illustrating this conclusion. The long term corrosion testing results substantiate that it is a non-toxic and durable material for use in implantable devices. Qualification testing results validate the safety and effectiveness of the lead design and materials used. The TIR and TIJ leads are tested for crimp and weld strength of connections, fatigue strength, DC resistance, environmental resistance, adherence to IS-1 standards, stylet performance, packaging and transportation durability, load tip testing, and sterilization validation. All tests BIOTRONIK, September 23, 1996 {1} FROM : BIOTRONIK INC PHONE NO. : 5036976235 Sep. 23 1996 11:27AM P5 # TIR/TIJ Cardiac Pacing Leads ## Response to K953044 results were within specifications. An overview of pre-clinical testing is found in section III.A. of this premarket notification with additional information provided in the Appendices. The European version of the TIR/TIJ endocardial leads, models IRTI and IRTJ, were evaluated in European clinical studies to examine the performance characteristics of the unipolar and bipolar leads. The protocol used examined acute and long-term sensing and pacing thresholds, and long-term impedance's. Details of this evaluation can be found in section III.C. of this notification. (APPENDIX 3) The results of studies, the in-vitro and qualification testing performed on the TIR/TIJ leads show that the risk to the patient in using these leads is the same as that of any implantable endocardial lead. [2 sentences stricken from this paragraph] Potential complications resulting from the use of endocardial leads include, but are not limited to: thrombosis, embolism, body rejection phenomena, cardiac tamponade, muscle/nerve stimulation, valve damage, fibrillation, and infection. Lead perforation through the myocardium has been rarely observed. The table below summarizes some of the potential symptoms indicating a complication and possible corrective actions: ### Table 6 #### Lead Complications | Symptom | Potential Complication | Potential Corrective Action | | --- | --- | --- | | Loss of pacing or sensing | Electrode displacement Lead fracture Setscrew penetration Improper lead to pacemaker connection | Reposition lead Replace lead Replace lead Reconnect lead to pacemaker | | Increase or decrease in threshold | Fibrotic tissue formation | Adjust pulse generator output; Reposition lead | The fractal iridium surface treatment was introduced in Europe on TIR/TIJ on January 20, 1994. The lead body design and the size of the tip and the ring (where applicable) is the same for all leads, now representing over 100,000 leads sold. BIOTRONIK, September 23, 1996 Page 3 {2} FROM : BIOTRONIK INC PHONE NO. : 5036976235 Sep. 23 1996 11:27AM P6 TIR/TIJ Cardiac Pacing Leads Response to K953044 Table 7 Number of Devices Sold, Worldwide | Type | 1990 | 1990 | 1991 | 1992 | 1993 | 1994* | SUM | | --- | --- | --- | --- | --- | --- | --- | --- | | TIR-UP | 430 | 2370 | 4790 | 7840 | 14790 | 18860 | 49080 | | IRTI-UP | | | | 50 | 650 | 1090 | 1790 | | TIR-BP | 70 | 700 | 2130 | 4480 | 9880 | 19850 | 37090 | | IRTI-BP | | | | 60 | 900 | 2160 | 3120 | | TIJ-UP | | | 210 | 570 | 990 | 2780 | 4550 | | IRTJ-UP | | | | 30 | 210 | 130 | 370 | | TIJ-BP | | | 170 | 700 | 1140 | 2560 | 4570 | | IRTJ-BP | | | | 20 | 510 | 900 | 1430 | | | | | | | | TOTAL: | 102000 | A bibliography of relevant articles on cardiac pacing leads is provided as Appendix 23, with selected copies provided as Appendix 25. BIOTRONIK, September 23, 1996 Page 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...