K073230 · Biotronik, Inc. · DTA · Dec 20, 2007 · Cardiovascular
Device Facts
Record ID
K073230
Device Name
MODELS ERA 300 AND 3105 PACING SYSTEM ANALYZERS
Applicant
Biotronik, Inc.
Product Code
DTA · Cardiovascular
Decision Date
Dec 20, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3720
Device Class
Class 2
Indications for Use
The ERA 300 is intended for use during invasive pacemaker procedures in the following activities: . Temporary External Pacing Provides temporary stimulation under DDD, DDI, DOO, VVI, VDD, VOO, AAI, or AOO modalities during implantable pacemaker procedures or physician evaluations. . Lead Threshold Determination Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time. . Pacemaker Function Test Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the following parameters: pulse amplitude and width, sensitivity, refractory period, A/V delay, and rate/interval.
Device Story
ERA 300 is a portable, dual-chamber pacing system analyzer; used during invasive pacemaker implantation or troubleshooting. Device functions as temporary external pulse generator or diagnostic tool for lead/generator evaluation. Inputs include electrical signals from pacing leads or pulse generators. Device processes signals to measure impedance, capture threshold, P/R wave amplitude/slew rate, retrograde conduction time, and pulse generator parameters (amplitude, width, sensitivity, refractory period, AV delay, rate/interval). Operated by physicians in clinical settings. Output displayed for clinician review to guide pacing therapy adjustments and ensure proper device function. Touch-proof configuration prevents hazardous electrical connections. Modification involves sourcing 8V battery from alternative supplier (pbq) meeting original BIOTRONIK specifications.
Clinical Evidence
No clinical data. Bench testing only; verification that the alternative battery meets all existing performance specifications.
Technological Characteristics
Portable, dual-chamber pacing system analyzer. Touch-proof electrical configuration. Powered by 8V battery. Performs electrical signal analysis for pacing lead and pulse generator testing. No changes to core sensing/actuation technology.
Indications for Use
Indicated for patients undergoing invasive pacemaker procedures or physician evaluations requiring temporary external pacing, lead threshold determination, or pulse generator function testing.
Regulatory Classification
Identification
A pacemaker electrode function tester is a device which is connected to an implanted pacemaker lead that supplies an accurately calibrated, variable pacing pulse for measuring the patient's pacing threshold and intracardiac R-wave potential.
{1}------------------------------------------------
November 14, 2007
# ERA 300 Pacing System Analyzer Additional Battery Supplier
Special 510(k) Premarket Notification
#### 1. 510(K) SUMMARY
| Name and Address of Sponsor: | BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035 | DEC 2 0 2007 |
|--------------------------------------|-------------------------------------------------------------|--------------|
| Establishment Registration Number: | 1028232 | |
| Device Name: | | |
| Proprietary Name:<br>Classification: | ERA 300 Dual Chamber Pacing System Analyzer<br>Class II/III | |
Classification Name: External Pacemaker Pulse Generator (21 CFR 870.3600) Pacemaker Electrode Function Tester (21 CFR 870.3630) Pacemaker Generator Function Analyzer (21 CFR 870.3720) Product Code: DTA, DTC, DTE
### General Description:
The ERA 300 is a portable, dual chamber pacing system analyzer designed to test the electrical performance of the pulse generator and the pacing lead system or operate as a temporary external pulse generator at the time of pacemaker implantation and during invasive pacemaker troubleshooting or evaluation procedures. The ERA 300 utilizes a touch-proof configuration to help prevent hazardous connection between patients and electrical power sources.
### Device Modifications:
The 8 Volt battery currently supplied with the ERA 300 Pacing System Analyzer (PSA) is manufactured by Panasonic. However, in the future this supplier will no longer manufacture this battery model. Therefore, an additional supplier for the 8 Volt battery supplied with the ERA 300 PSA was identified, pbq. The pbq manufactured battery meets the same internal BIOTRONIK specifications including purchasing specifications, incoming inspection, and performance specifications.
### Predicate Devices:
BIOTRONIK proposes the following predicate device for the battery modification to the ERA 300 Pacing System Analyzer:
- . BIOTRONIK's ERA 300 Pacing System Analyzer (#K964190, cleared 07-10-97 and #K033613, cleared on 12-08-03)
{2}------------------------------------------------
### Indications for Use:
The ERA 300 is intended for use during invasive pacemaker procedures in the following activities:
#### . Temporary External Pacing
Provides temporary stimulation under DDD, DDI, DOO, VVI, VDD, VOO, AAI, or AOO modalities during implantable pacemaker procedures or physician evaluations.
#### . Lead Threshold Determination
Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time.
#### . Pacemaker Function Test
Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the following parameters: pulse and width, sensitivity, refractory period, AV delay, and rate/interval.
| Name and Address of Manufacturer: | BIOTRONIK GmbH & Co. KG<br>Woermannkehre 1,<br>12359 Berlin, Germany<br>011-49-30-689-05-1210 |
|-------------------------------------|-----------------------------------------------------------------------------------------------|
| Manufacturer's Registration Number: | 9610139 |
Contact Person(s) and Phone Number:
Jon Brumbaugh Vice President, Regulatory Affairs and Compliance Phone (888) 345-0374 Fax (503) 635-9936 jon.brumbaugh@biotronik.com
{3}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 0 2007
Biotronik, Inc. c/o Mr. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, OR 97035
Re: K073230
Trade/Device Name: ERA 300 Pacing System Analyzer Regulation Number: 21 CFR 870.3600 Regulation Name: External Pacemaker Pulse Generator Regulatory Class: Class III (three) Product Code: DTA, DTC, DTE Dated: November 14, 2007 Received: November 15, 2007
Dear Mr. Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Jon Brumbaugh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfsimmmon for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K073230
Device Name: ERA 300 Pacing System Analyzer
### Indications for Use:
The ERA 300 is intended for use during invasive pacemaker procedures in the following activities:
#### . Temporary External Pacing
Provides temporary stimulation under DDD, DDI, DOO, VVI, VDD, VOO, AAI, or AOO modalities during implantable pacemaker procedures or physician evaluations.
#### . Lead Threshold Determination
Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time.
#### . Pacemaker Function Test
Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the following parameters: pulse amplitude and width, sensitivity, refractory period, A/V delay, and rate/interval.
Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhummer
øn Sign-Off Division of Cardiovascular Devices 510/k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.