K964197 · Impra, Inc. · DSY · Jan 2, 1997 · Cardiovascular
Device Facts
Record ID
K964197
Device Name
IMPRA CARBOFLO EPTFE VASCULAR GRAFT
Applicant
Impra, Inc.
Product Code
DSY · Cardiovascular
Decision Date
Jan 2, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IMPRA Carboflo™ ePTFE Vascular Grafts are intended for use as vascular prostheses.
Device Story
IMPRA Carboflo™ ePTFE Vascular Graft is a vascular prosthesis constructed from expanded polytetrafluoroethylene (ePTFE) with a luminal layer (20-25% of wall thickness) impregnated with USP grade activated charcoal. The carbon is integrated into the graft wall during a singular extrusion process, creating a monolithic structure. The device is intended for surgical implantation to treat lower extremity vascular disease. The carbon-impregnated luminal surface is designed to reduce thrombogenicity by decreasing platelet accumulation, as demonstrated in pre-clinical animal models. The graft is supplied in various configurations, packaged, and sterilized using the same methods as the predicate IMPRA ePTFE Vascular Graft. Surgeons implant the device to restore or bypass blood flow in peripheral vascular applications.
Clinical Evidence
Multi-center, prospective, randomized clinical trial (1990-1994) in France. 160 patients (81 Carboflo, 79 Standard) treated for lower extremity vascular disease. Primary endpoint: patency over 24 months. Results: Cumulative Primary Patency 36.8% (Carboflo) vs 27.7% (Standard); Cumulative Secondary Patency 42.7% (Carboflo) vs 32.3% (Standard). Differences were not statistically significant (p=0.05). Adverse event profiles were comparable between groups, with no new complication types identified for the Carboflo graft.
Technological Characteristics
Materials: ePTFE and USP grade activated charcoal. Construction: Monolithic graft wall with carbon-impregnated luminal layer (20-25% thickness). Manufacturing: Singular extrusion process. Physical properties verified against AAMI Vascular Graft Standard and FDA Guidance Document on Vascular Prostheses. No active energy source or software components.
Indications for Use
Indicated for patients with lower extremity vascular disease requiring vascular graft implantation.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
Predicate Devices
IMPRA ePTFE Vascular Graft
Related Devices
K962639 — IMPRA CARBOFLO EPTFE VASCULAR GRAFT · Impra, Inc. · Oct 3, 1996
{0}
JAN -2 1997
K964197
# Summary of Safety and Effectiveness for IMPRA Carboflo™ ePTFE Vascular Graft
## SUBMITTER
Rajagopal R. Kowligi, Ph.D
Manager, Clinical Affairs
IMPRA, A Division of C.R. Bard, Inc.
1625 West Third Street
Tempe, AZ 85281
## DATE SUMMARY WAS PREPARED
October 18, 1996
## NAME OF THE DEVICE
IMPRA Carboflo™ ePTFE Vascular Graft
## IDENTIFICATION OF PREDICATE DEVICE
IMPRA ePTFE Vascular Graft
## DESCRIPTION OF THE DEVICE
IMPRA Carboflo™ ePTFE Vascular Grafts are made primarily of expanded polytetrafluoroethylene (ePTFE) using the same manufacturing procedures that are used to manufacture IMPRA ePTFE Vascular Grafts, the devices to which substantial equivalence is claimed. The region of the graft wall adjacent to the lumen, approximately 20-25% of the total wall thickness, is uniformly impregnated with Carbon particles along the entire length of the graft. The carbon impregnated region is formed integral to the outer region of the wall by mixing the PTFE resin mixed with carbon particles, with the non-carbon containing PTFE resin during a singular extrusion process, which results in a monolithic graft wall. The carbon used in the device is USP grade activated charcoal. All other components of the Carboflo graft, namely PTFE, lubricant used as a manufacturing aid, blue pigment incorporated in the orientation lines, and the external support PTFE beading are the same as those used in the manufacture of the predicate devices. These grafts are supplied in the same product configurations as the predicate device, and are packaged, labeled, and sterilized in the same manner as the predicate devices.
Extensive bench testing and microscopic analysis has shown that the carbon
000539
{1}
particles are an integral part of the wall and cannot be separated. The amount of carbon incorporated in each Carboflo vascular graft is less than 1% of the total weight of the graft.
## INTENDED USE
IMPRA Carboflo™ ePTFE Vascular Grafts are intended for use as vascular prostheses.
## COMPARISON OF THE DEVICE CHARACTERISTICS TO THE PREDICATE
Physical properties of the Carboflo vascular grafts were compared to the values for Standard IMPRA ePTFE vascular grafts, using methods recommended by the AAMI Vascular Graft Standard or the FDA Guidance Document on Vascular Prostheses. Testing of a variety of product types shows that the addition of carbon particles into the graft wall did not affect the physical properties of the Carboflo grafts. Both the new device and predicate device undergo the same testing and evaluation procedures. The acceptance criteria for both the new device and predicate devices are the same.
## NON-CLINICAL TESTING
Carbon containing surfaces in medical devices, e.g. Heart valves, have been shown to impart anti-thrombogenic properties¹. Short-term pre-clinical studies have been conducted with IMPRA Carboflo™ ePTFE Vascular Grafts to determine the performance of carbon containing blood contact surface. Animal studies comparing the Carboflo grafts with Standard IMPRA ePTFE grafts have shown that the inside surfaces of the Carboflo grafts have a significantly higher Thrombus Free Surface Area after 3 months². Patencies of both Carboflo and Standard grafts were similar. Short-term animal studies in dogs and rabbits have shown that the Carboflo inside surfaces have reduced platelet accumulation when compared to Standard ePTFE vascular grafts, suggesting that the addition of carbon particles reduces the thrombogenicity of the surface³.
Additional bench testing performed to determine the response of the Carboflo ePTFE graft compared to the Standard ePTFE in repeated puncture test and the
¹ References on file
² Data on file
³ References on file
000540
{2}
particulate flow test showed them to be equivalent.
# CLINICAL INFORMATION
A multi-center, prospective, randomized clinical trial comparing IMPRA Carboflo™ ePTFE Vascular Grafts to commercially available ePTFE Vascular Grafts, was performed in France between 1990-1994. The purpose of the investigation was to compare the patencies of both IMPRA Carboflo ePTFE Vascular Grafts and Standard ePTFE Vascular Grafts. A total of 81 patients received Carboflo grafts and 79 received Standard ePTFE grafts. All grafts were implanted to treat lower extremity vascular disease. Of the 160 grafts, only 5 grafts (3 Carboflo, 2 Standard) were considered to be Above-Knee, and the remaining grafts were all Below-Knee.
73% of the distal anastomoses were direct graft to vessel, with the following types comprising the other anastomoses: venous patch (16%), distal arterio-venous fistula (5.8%), or interpositional vein cuff graft (5.2%). All implanted grafts were followed for at least 24 months, or until failure, lost to follow-up, or death of patient. Adverse events were recorded and documented on Case Report Forms (CRF).
# Analysis
PRIMARY PATENCY was defined as the time between implant date and date of follow-up when the graft was patent, prior to the first intervention to correct complications. Interventions included immediate re-operations to correct any complication post implantation. SECONDARY PATENCY was defined as the time between implant date and date on which the graft is no longer patent or useful, after a series of interventions. Conventional life-tables were then constructed to determine the Cumulative Primary and Secondary Patencies. At the end of the 2 yr follow-up, patencies were as follows:
Cumulative Primary Patency for Carboflo grafts was 36.8% compared to 27.7% for Standard grafts at the end of 24 months.
Cumulative Secondary Patency for Carboflo grafts was 42.7% compared to 32.3% for Standard grafts.
These results are not statistically significantly different at p = 0.05.
000541
{3}
DEVICE SAFETY was demonstrated by determining adverse events for both Carboflo and Standard grafts. Complications were counted as they were encountered and added up at the end of the study period for each patient. The results are tabulated below:
COMPLICATIONS: CARBOFLO VS. STANDARD
| Adverse event | Device Type
Standard | Carboflo |
| --- | --- | --- |
| No. At Start | 79 | 81 |
| No. LFU @2yrs * | 7 | 17 |
| No. Patent @2yrs | 17 | 16 |
| No. Immed. Redos | 6 | 12 |
| No. Failed @2yrs | 52 | 43 |
| Thrombosis | 34 | 26 |
| False aneurysm | 1 | 1 |
| Infection | 5 | 4 |
| Other adverse events | 2 | 3 |
| Amputation | 26 | 28 |
| Deaths | 16 | 13 |
* LFU = Lost to follow-up
It is important to note that no new type of complications were identified with Carboflo grafts.
RISK FACTORS in patients with Carboflo and Standard ePTFE Vascular Grafts were similar (p > 0.05), except for Smoking. There were a higher number of patients who were smokers, in the Carboflo group (p<0.05). The risk factors for the patients in this study were then compared to the risk factors for patients in USA who had treatment for peripheral vascular disease (information was summarized from published literature). Risk factors were comparable for both groups (p>0.05), except for Smoking. There were a higher number of patients who were smokers, in the US (p<0.05).
CONCLUSION
IMPRA ePTFE Carboflo Vascular Grafts are substantially equivalent to the currently marketed IMPRA ePTFE Vascular Grafts.
000542
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.