Spiral Laminar Flow™ Vascular Arteriovenous Graft (AV0645)

K252277 · Vascular Flow Technologies Limited · DSY · Oct 20, 2025 · Cardiovascular

Device Facts

Record IDK252277
Device NameSpiral Laminar Flow™ Vascular Arteriovenous Graft (AV0645)
ApplicantVascular Flow Technologies Limited
Product CodeDSY · Cardiovascular
Decision DateOct 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Spiral Laminar Flow Vascular Arteriovenous Graft is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis.

Device Story

The Spiral Laminar Flow (SLF) Vascular Arteriovenous Graft is a straight, subcutaneous conduit used for hemodialysis access. It consists of an expanded polytetrafluoroethylene (ePTFE) tube featuring a helical 'spiral flow inducer'—a deformation in the graft wall designed to promote laminar flow. A nonremovable external polyurethane (ChronoFlex AL-80A) indicator ring marks the start of the inducer segment and identifies a no-cannulation zone. A shaped hood is provided to assist surgeons in achieving optimal attachment angles and alignment during distal anastomosis. The device is implanted by surgeons in a hospital operating theatre. By modifying blood flow patterns, the graft aims to improve vascular access performance. The device is supplied sterile for single use.

Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by bench testing, including mechanical performance (ISO 7198), biocompatibility (ISO 10993-1), and sterilization validation (ISO 11135).

Technological Characteristics

Materials: ePTFE (graft body), ChronoFlex AL-80A polyurethane (inducer indicator ring). Dimensions: 6 mm diameter, 450 mm length. Principle: Helical wall deformation for spiral flow induction. Sterilization: Ethylene Oxide (ISO 11135). Compliance: ISO 7198 (vascular prostheses), ISO 10993-1 (biocompatibility).

Indications for Use

Indicated for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis in patients requiring such access. Implantation is restricted to trained and qualified physicians and/or surgeons in a hospital operating theatre environment.

Regulatory Classification

Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 20, 2025 Vascular Flow Technologies Limited Angela Paterson Head of QA/RA Prospect Business Centre Gemini Crescent Dundee, DD2 1TY United Kingdom Re: K252277 Trade/Device Name: Spiral Laminar Flow™ Vascular Arteriovenous Graft (AV0645) Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: Class II Product Code: DSY Dated: July 18, 2025 Received: July 22, 2025 Dear Angela Paterson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252277 - Angela Paterson Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252277 - Angela Paterson Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Rohini Retarekar -S for Carmen Johnson, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252277 | | | Device Name Spiral Laminar Flow Vascular Arteriovenous Graft | | | Indications for Use (Describe) | | | The Spiral Laminar Flow Vascular Arteriovenous Graft is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Page 1 of 5 Vascular Flow Technologies Ltd. Traditional 510(k) For Spiral Laminar Flow™ Vascular Arteriovenous Graft 510(k) Summary K252277 This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92. Submitter's Name: Vascular Flow Technologies Limited. Submitter's Address: Vascular Flow Technologies Limited Prospect Business Centre, Gemini Crescent Dundee DD2 1TY, United Kingdom Contact Person: Angela Paterson Telephone: +44 7884274220 Date Prepared: 23rd September 2025 {5} Page 2 of 5 Vascular Flow Technologies Ltd. Traditional 510(k) For Spiral Laminar Flow™ Vascular Arteriovenous Graft Device Classification Information: | Regulation Number | Device | Device Class | Product Code | Classification Panel | | --- | --- | --- | --- | --- | | 870.3450 | Prosthesis, Vascular Graft, of 6mm and greater Diameter | Class 2 | DSY | Cardiovascular | Device Trade Name: Spiral Laminar Flow™ Vascular Arteriovenous Graft Device Common Name: Spiral Laminar Flow™ Vascular Arteriovenous Graft Predicate Device: Spiral Flow Vascular Arteriovenous Grafts cleared under 510(k) number K141124 and K094044 Indications for Use: The SLF™* Vascular Arteriovenous Graft is a vascular prosthesis, which is intended for use as a subcutaneous arteriovenous conduit for vascular access during hemodialysis. ONLY trained and qualified physicians and/or surgeons, under the controlled conditions of a hospital operating theatre environment are indicated for use of this device for implantation. {6} Page 3 of 5 # Vascular Flow Technologies Ltd. Traditional 510(k) For Spiral Laminar Flow™ Vascular Arteriovenous Graft ## Device Description: SLF™ Vascular Arteriovenous Grafts are straight tubes constructed of expanded polytetrafluoroethylene (ePTFE) with the unique spiral flow inducer and nonremovable external inducer indicator ring (palpable ring) formed from ChronoFlex™ AL-80A, a Biodurable Medical Grade polyurethane. The spiral flow inducer is a helical deformation in the wall of the graft. The inducer indicator ring is on the outer surface of the proximal end of the spiral flow inducer, indicating the beginning of the spiral inducer segment and the no cannulation area of the graft. A Shaped Hood is provided to guide the surgeon when creating the distal anastomosis. Although not critical to the functioning of the device, the hood does provide for an effective attachment angle and alignment of the spiral flow inducer with the native vessel All grafts are supplied sterile and will remain so for the stated shelf life unless the packaging is opened or damaged. SLF™ Vascular Arteriovenous grafts are available in one size (Table 1). Table 1: SLF Graft Dimensions | Device | Diameter | Length | | --- | --- | --- | | SLF™ Vascular Arteriovenous Grafts | 6 mm | 450 mm | ## Comparison to Predicate Device The modifications in comparison to the previously cleared devices are as follows - Supplier of base ePTFE Graft Material - Chemical structure of polyurethane overmould material ## Non-Clinical Tests (Performance/Physical Data) A summary of performance testing carried out on the subject device to confirm that it met its specifications and demonstrate substantial equivalence to the predicate device is provided in Table 3. Table 3: Summary of Non-Clinical Performance- Bench Testing | Test Name | Test Description | Individual tests | Results | | --- | --- | --- | --- | | ISO 7198 Testing | Testing of the Spiral Laminar Flow Vascular Arteriovenous Graft demonstrated compliance to ISO 7198:2016 Cardiovascular implants and extracorporeal systems- Vascular prostheses- Tubular vascular grafts and vascular patches. | Useable Length Relaxed internal Diameter Suture Retention Strength Wall Thickness Water Entry Pressure Longitudinal Tensile Strength Strength after repeated puncture Kink Diameter Circumferential Tensile Strength | Passed | {7} Page 4 of 5 # Vascular Flow Technologies Ltd. ## Traditional 510(k) For Spiral Laminar Flow™ Vascular Arteriovenous Graft | Biocompatibility | Testing and analysis of the finished device demonstrated compliance to ISO 10993-1:2018- Biological evaluation of medical devices: Part 1: Evaluation and testing within a risk management process | Endpoints Evaluated Physical and chemical information Cytotoxicity Sensitization Irritation or intracutaneous reactivity Material mediated pyrogenicity Acute systemic toxicity Sub-acute toxicity Sub-chronic toxicity Chronic toxicity Implantation Hemocompatibility Genotoxicity Carcinogenicity | Passed | | --- | --- | --- | --- | | Ethylene Oxide Sterilization Validation | Validation in accordance with ISO 11135:2014 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices | Overkill method Sterility assurance level 1 x 10⁻⁶ confirmed Half cycle Sub lethal cycle Sterility tests Bioburden Endotoxin Bacteriostasis & Fungistatic Inhibition and Enhancement | Passed | | Transit Testing& Environmental Conditioning | Provides objective evidence that the packaging of the Spiral Laminar Flow Vascular Arteriovenous Graft | Transportation Testing Variable Conditioning Package dye penetration Package Bubble leak Aseptic presentation Visual inspection Package tensile testing Package seal width Product package interaction Useable Length Relaxed internal Diameter Suture Retention Water Entry Pressure Longitudinal Tensile Strength Strength after repeated puncture Kink Diameter Circumferential Tensile Strength | Passed | All testing was successfully completed, and the results show that Spiral Laminar Flow Vascular Arteriovenous Graft is safe and performs as expected when used as intended. Spiral Laminar Flow Vascular Arteriovenous Graft is considered substantially equivalent to the predicate device. ## Animal Studies No animal studies were conducted as part of the submission to prove substantial equivalence. {8} Page 5 of 5 Vascular Flow Technologies Ltd. Traditional 510(k) For Spiral Laminar Flow™ Vascular Arteriovenous Graft Clinical Studies No clinical studies were conducted as part of the submission to prove substantial equivalence. Conclusion: The available data supports that the SLF™ Vascular Arteriovenous Graft is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...