← Product Code [DSY](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY) · K202703

# Vascutek Gelsoft Plus ERS Vascular Graft (K202703)

_Vascutek, Ltd. · DSY · Apr 6, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K202703

## Device Facts

- **Applicant:** Vascutek, Ltd.
- **Product Code:** [DSY](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY.md)
- **Decision Date:** Apr 6, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the Gelsoft Plus ERS vascular graft is for the extra-anatomical vascular repair.

## Device Story

Gelsoft Plus ERS is an externally reinforced, gelatin-sealed, knitted polyester vascular graft. Device consists of a tubular knitted polyester base fabric, crimped for flexibility, with an external polypropylene support to provide kink resistance and a smooth flow surface for extra-anatomical applications. The polypropylene support is peelable at the ends to facilitate anastomosis. The graft is impregnated with bovine-derived gelatin to seal interstices, eliminating the need for pre-clotting. Used by vascular surgeons in clinical settings for bypass procedures. The device serves as a conduit for blood flow in patients with arterial occlusive or aneurysmal disease, restoring circulation and potentially preventing complications associated with arterial blockages or aneurysms.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical testing including physical bench testing, biocompatibility, chemical characterization, and an animal performance study.

## Technological Characteristics

Knitted polyester textile graft; externally reinforced with polypropylene support; impregnated with bovine-derived gelatin sealant. Tubular form with crimping. Peelable polypropylene support at ends. Sterilized device.

## Regulatory Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

## Predicate Devices

- Vascutek Gelsoft Plus ERS ([K034010](/device/K034010.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services USA logo, and on the right is the FDA logo. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".

April 6, 2021

Vascutek Ltd Chirag Merchant Senior Regulatory Affairs Associate Newmains Avenue, Inchinnan Renfrewshire PA4 9RR Scotland. UK

Re: K202703

Trade/Device Name: Gelsoft Plus ERS Vascular Grafts Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: Class II Product Code: DSY Dated: February 12, 2021 Received: February 25, 2021

Dear Chirag Merchant:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Rohini Retarekar -S

for Carmen Gacchina Johnson, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K202703

Device Name Vascutek Gelsoft Plus ERS Vascular Graft

Indications for Use (Describe)

Indicated for extra anatomical vascular repair, primarily for axillo-femoral bypass and femoro- popliteal reconstruction.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Terumo Aortic logo. The word "TERUMO" is in green, with a red swoosh above it. Below that, the word "Aortic" is in gray. The logo is simple and modern.

Vascutek Ltd Newmains Avenue, Inchinnan Renfrewshire PA4 9RR Scotland, UK

> Tel. +44 (0) 141 812 5555 terumoaortic.com

## 510 (k) Summary

This 510(k) Summary is being submitted in accordance with 21 CFR 807.92.

| Submitter:            | Vascutek Ltd<br>Newmains Avenue<br>Inchinnan<br>Renfrewshire<br>PA4 9RR<br>Scotland, UK                                  |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------|
|                       | Tel: +44 (0) 141 812 555                                                                                                 |
| Contact:              | Chirag Merchant<br>Senior Regulatory Affairs Associate<br>Tel: +44 (0) 141 812 555<br>Email: c.merchant@terumoaortic.com |
| Date of Preparation:  | April 2, 2021                                                                                                            |
| Trade Name:           | Vascutek Gelsoft Plus ERS Vascular Graft                                                                                 |
| Common or Usual Name: | Vascular Graft Prosthesis                                                                                                |
| Classification Name:  | Prosthesis, vascular graft, of 6mm and<br>greater diameter                                                               |
| Product Code:         | DSY                                                                                                                      |
| Regulation Number:    | 21 CFR 870. 3450                                                                                                         |
| Device Class:         | II                                                                                                                       |

Image /page/3/Picture/7 description: This image is a logo for The Queen's Awards for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design attached to it. Above the "e" is a crown, symbolizing the Queen. The text "THE QUEEN'S AWARDS FOR ENTERPRISE: INNOVATION 2016" is written below the "e".

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Terumo Aortic. The word "TERUMO" is in large, green, sans-serif font with a red swoosh above the letters "TE". Below the word "TERUMO" is the word "Aortic" in a smaller, gray, sans-serif font. The logo is simple and modern.

#### Identification of the legally marketed device to which equivalence is being claimed:

Vascutek Ltd. are claiming equivalence to the following legally marketed devices:

- . Vascutek Gelsoft Plus ERS (K034010).
### Device Description:

Gelsoft Plus ERS grafts are an externally reinforced (ERS), gelatin-sealed, knitted polyester grafts.

Vascutek's polyester vascular prosthesis family is based on a polyester textile technology. The starting point for all products is polyester yarn. This is fabricated into tubular form by knitting. This is referred to as the base fabric and exists in a variety of designs to meet particular end uses. Further processing is designed to retain the tubular form of the graft e.g. crimping and addition of an external polypropylene support. In Gelsoft Plus ERS, this external polypropylene support is to provide kink resistance and a smooth flow surface for extra-anatomical applications. The polypropylene support may be peeled where it extends to the ends of the prosthesis, in order to facilitate the fashioning of the anastomosis. Additional branches are also attached in a variety of configurations. The last process is impregnation with a bovine derived gelatin sealant. This process fills the gaps between threads of the polyester base graft and eliminates the need for pre-clotting of the graft prior to implant.

#### Intended Use:

The intended use of the Gelsoft Plus ERS vascular graft is for the extra-anatomical vascular repair.

#### Indications for Use:

Vascutek Gelsoft Plus ERS Graft is indicated for extra anatomical vascular repair, primarily for axillo-femoral/bi-femoral bypass and femoro- popliteal reconstruction.

Gelsoft Plus ERS AX -FEM Vascular Graft: Indicated for vascular repair i.e. primarily for axillo-femoral bypass procedures & femoral- popliteal reconstruction in aneurysmal & occlusive disease of the arteries.

Image /page/4/Picture/12 description: The image is a logo for The Queen's Awards for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design extending from it. Above the "e" is a crown, symbolizing the British monarchy. The text "THE QUEEN'S AWARDS FOR ENTERPRISE: INNOVATION 2016" is written below the "e".

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Terumo Aortic logo. The word "TERUMO" is in green, with a red swoosh above it. Below that, the word "Aortic" is in gray. The logo is simple and clean, and the colors are eye-catching.

Gelsoft Plus ERS FEM Vascular Graft: Indicated for vascular repair i.e. primarily for femoro-femoral bypass procedures in aneurysmal & occlusive disease of the arteries.

Gelsoft Plus ERS AX- BIFEM Vascular Graft: Indicated for vascular repair i.e. primarily for axillo-bi-femoral bypass procedures in aneurysmal & occlusive disease of the arteries.

#### Intended Patient Population:

Patients requiring extra anatomical vascular repair whenever knitted polyester grafts are normally indicated, primarily for axillo-femoral/bi-femoral bypass and femoro-popliteal reconstruction in aneurysmal & occlusive disease of the arteries.

They are contraindicated for use in patients with a sensitivity to polyester or materials of bovine origin. They are contraindicated for coronary vascular repair, blood access fistula (e.g. haemodialysis) and pulmonary shunting. They are contraindicated for thoracic use. They are contraindicated for use in the pulmonary positions, use in arteriovenous shunting or cardiovascular patching.

#### Technological Characteristics:

Equivalency is based on identical design, technology, construction and intended use.

The only change is that the gelatin used to seal the grafts will be purchased from a new supplier.

The nonclinical testing performed, including physical bench, biocompatibility, chemical characterisation and an animal performance study, have demonstrated that the device is substantially equivalent to the predicate device.

Image /page/5/Picture/10 description: The image is a black and white logo for The Queen's Award for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design extending from it. Above the "e" is a crown, symbolizing the British monarchy. Below the "e" and crown, the text "THE QUEEN'S AWARD FOR ENTERPRISE: INNOVATION 2016" is arranged in a stacked format.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K202703](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K202703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
