← Product Code [DSY](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY) · K061626

# VASORING* VRC VASCULAR RING CONNECTOR (K061626)

_Sunwei Tech Co. , Ltd. · DSY · Oct 5, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K061626

## Device Facts

- **Applicant:** Sunwei Tech Co. , Ltd.
- **Product Code:** [DSY](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY.md)
- **Decision Date:** Oct 5, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The VASORING VRC Vascular Ring Connector is indicated for use as a connector for attaching a vascular graft to native tissue where conditions preclude a conventional end-to-end sutured anastomosis.

## Device Story

VASORING VRC is a titanium alloy vascular ring connector used to attach vascular grafts to native tissue; provides sutureless anastomosis. Device features two outer grooves: narrow groove for suture fixation to vascular prosthesis; wider groove for ligature fixation to blood vessel using Ethicon Nylon Tape. Thin-wall design maintains mechanical strength while maximizing inner diameter. Used in surgical settings where conventional suturing is not feasible. Benefits include time-saving during anastomosis and secure graft-to-vessel connection.

## Clinical Evidence

Bench testing performed to verify functional requirements. Biocompatibility testing conducted per ISO 10993. Two pre-clinical animal studies performed: a 4-hour study to assess immediate impact on thrombogenicity and a 20-week observation study. Results demonstrated satisfactory performance.

## Technological Characteristics

Biocompatible titanium alloy; cylindrical shape; outside diameter 6mm to 30mm; thin-wall design. Utilizes external braided ligature tape (Ethicon Nylon Tape) for fixation. Biocompatibility per ISO 10993.

## Regulatory Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

## Predicate Devices

- Meadox Intra-Aortic Graft ([K881283](/device/K881283.md))
- Titanium GREENFIELD Vena Cava Filter ([K874096](/device/K874096.md))
- St. Jude Symmetry Aortic Clip ([K003446](/device/K003446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KO61626 p. Lof2

SUNWEI TECH Co. Ltd. Pre-market Notification for VASORING* VRC

OCT - 5 2006

VASORING* VRC Vascular Ring Connector of 6mm and Larger Diameter 510(k) Summary of Safety and Effectiveness

## 1. Sumbitter's Name:

SUNWEI TECH Co. Ltd. 3rd Floor, 42, Sec. 2, Jhongcheng Road, Shih Lin District Taipei 111, TAIWAN Contact: Cassy Mao, Director of Operations.

## 2. Name of Device

| Common/Usual Name:   | Vascular Ring Connector                                 |
|----------------------|---------------------------------------------------------|
| Proprietary Name:    | VASORING* VRC Vascular Ring Connector                   |
| Classification Name: | Prostheses, Vascular Graft, of 6mm and Greater Diameter |

#### 3. Predicate Device

| Trade Name                           | 510(k) Number | Decision Date |
|--------------------------------------|---------------|---------------|
| Meadox Intra-Aortic Graft            | K881283       | 05/26/1988    |
| Titanium GREENFIELD Vena Cava Filter | K874096       | 11/04/1988    |
| St. Jude Symmetry Aortic Clip        | K003446       | 05/21/2001    |

#### 4. Device Description

The vascular ring connector is made of biocompatible titanium alloy. Placed in the vascular prosthesis to form a Graft-Connector Unit, it can be used as an intraluminal graft. The braided ligature tape (Ethicon® Nylon Tape) is for tying the ring connector from outside of the blood vessel to provide sutureless anastomosis that is time-saving. There are two grooves on the outer surface of the connector: the narrow groove is for suture fixation between the connector and the vascular prosthesis; and the wider groove is for ligature fixation between the connector and the blood vessel. The titanium alloy allows the ring connector to be of thin-wall design which gives larger inner diameter while maintaining adequate mechanical strength at the anastomotic sites.

{1}------------------------------------------------

K061626
p. 2 of 2

# 5. Indication for Use

The VASORING* VRC Vascular Ring Connector is indicated for use as a connector for attaching a vascular graft to native tissue where conditions preclude a conventional end-to-end sutured anastomosis.

# 6. Technological Characteristics

The VASORING* VRC Vascular Ring Connector is identical to the Meadox Intra-Aortic Graft polyester connector with respect to intended use. The titanium alloy ring is of cylindrical shape with an outside diameter ranging from 6mm to 30mm for use with various sizes of aorta to achieve proper anastomoses.

# 7. Performance Summary

Bench testing was performed to ensure that VASORING* VRC will meet all functional and performance requirements for its intended use. The titanium alloy connector was also tested for material biocompatibility per ISO 10993 guidelines. Two pre-clinical animal studies were performed to investigate the thrombogenicity phenomena; a 4-hours study to assess immediate impact of using the VRC for anastomoses, and the other was a 20-week observation of the test animals. All bench tests, biocompatibility tests, and thrombogenicity studies demonstrated satisfactory performance of the VASORING* VRC connectors.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 5 2006

SUNWEI TECH Co. Ltd c/o Joseph J. Chang, PhD, P.E. Consultant Innomedtech LLC 7128 Staffordshire Street Houston, TX 77030

Re: K061626

Vasoring VRC Vascular Ring Connector Regulation Number: 21 CFR 870.3460 Regulation Name: Vascular Graft Prosthesis Regulatory Class: Class II (Two) Product Code: DSY Dated: September 21, 2006 Received: September 22, 2006

Dear Dr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Joseph J. Chang, PhD, P.E.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

R. lohner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Hcalth

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K061626

Device Name: VASORING VRC Vascular Ring Connector

Indications For Use:

The VASORING VRC Vascular Ring Connector is indicated for use as a connector for attaching a vascular graft to native tissue where conditions preclude a conventional end-to-end sutured anastomosis.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sung R. backing

(Division September Division of Cardiovascular Devices

510(k) Namia Ko 6 162 162 6 162 6 ----------

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K061626](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSY/K061626)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
