The AutoCAT®2 Intra-Aortic Balloon Pump (IABP) and AC3™ Series Intra-Aortic Balloon Pump (IABP) are clinically indicated for use for the following conditions: 1. Acute Coronary Syndrome 2. Cardiac and Non-Cardiac Surgery 3. Complications of Heart Failure
Device Story
IABP system provides counter-pulsation therapy to adult patients with impaired left ventricular function. System consists of pump control/display module with touchscreen/keypad and pneumatic drive module. Device connects to intra-aortic balloon (IAB) catheter inserted into femoral artery and positioned in descending thoracic aorta. Principle of operation: helium gas inflation of IAB during diastole increases blood pressure/perfusion; deflation during early systole reduces aortic pressure, decreasing cardiac work and oxygen demand. Used in clinical settings by healthcare providers to provide hemodynamic support. Output displayed on screen for monitoring and therapy management. Modification involves updated motor connector on MForce motor driver to IABP stepper motor for increased robustness.
Clinical Evidence
Bench testing only. Verification testing included system-level, manufacturing process, and performance/durability tests. Results met all specified acceptance criteria and demonstrated performance equivalent to predicate devices.
Technological Characteristics
Intra-aortic balloon and control system; pneumatic drive module; helium gas inflation/deflation; compatible with 30, 40, and 50cc IAB catheters; MForce motor driver with stepper motor; touchscreen/keypad interface; Class II device.
Indications for Use
Indicated for adult patients with impaired left ventricular function experiencing acute coronary syndrome, undergoing cardiac or non-cardiac surgery, or suffering complications of heart failure.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 27, 2020
Arrow International Inc. Shajunath Nirupama Senior Reguilatory Affairs Specialist 16 Elizabeth Drive Chelmsford, Massachusetts 01824
Re: K201112
Trade/Device Name: AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABP Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon And Control System Regulatory Class: Class II Product Code: DSP Dated: April 24, 2020 Received: April 27, 2020
Dear Shajunath Nirupama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrb/ofdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
K201112
Device Name
AutoCAT@2 Intra-Aortic Balloon Pump (IABP) and AC3™ Series Intra-Aortic Balloon Pump (IABP)
Indications for Use (Describe)
The AutoCAT®2 Intra-Aortic Balloon Pump (IABP) and AC3™ Series Intra-Aortic Balloon Pump (IABP) arc clinically indicated for use for the following conditions:
1. Acute Coronary Syndrome
2. Cardiac and Non-Cardiac Surgery
3. Complications of Heart Failure
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 2
[As required by 21 CFR 807.92]
Date Prepared: April 24, 2020
510(k) Number: K201112
#### Submitter's Name / Contact Person
Manufacturer Arrow International, Inc. Subsidiary of Teleflex, Inc. 16 Elizabeth Drive Chelmsford, MA 01824 Establishment Registration: 3010532612
## Contact Person Shajunath Nirupama Sr. Regulatory Affairs Specialist Shajunath.Nirupama@teleflex.com usmin-vsireg@teleflex.com Phone: 978-250-5100 ext. 725187
### General Information
| | General Predicate Device Information | |
|-------------------------|-------------------------------------------------------------------------------|------------------------------|
| Trade Name | AutoCAT2® Intra-Aortic Balloon Pump<br>AC3 ™ Series Intra-Aortic Balloon Pump | |
| Common Name | Intra-Aortic Balloon Pump | |
| 510(k) Number | AutoCAT2 ® IABP<br>K060309 | AC3 ™ Series IABP<br>K192238 |
| Reference 510(k) Number | K002256 | K162820 |
| 510(k) Holder | Arrow International, Inc. | |
| Classification | II | |
| Regulatory Description | 21 CFR 870.3535, Intra-aortic balloon and control system | |
| Product Code | DSP | |
| Panel | Cardiovascular | |
#### Device Description
## The AutoCAT Series Intra-Aortic Balloon Pump (IABP) System
The AutoCAT Series IABP System, including the AutoCAT®2 and AC3™ Series IABP's, provide counter-pulsation therapy to adult patients with impaired left ventricular (LV) function. They provide hemodynamic support of blood pressure and reduced cardiac work through volume displacement principles. The IABP is attached to an intra-aortic balloon
{4}------------------------------------------------
(IAB) catheter which is inserted into the femoral artery and positioned in the descending thoracic aorta.
The IABP delivers helium (He) into the IAB catheter during diastole to displace blood above and below the IAB. increasing blood pressure and perfusion to organs close to the IAB catheter. The IABP deflates or removes He from the IAB catheter just prior to or in the early phase of systole, reducing the pressure in the aorta and therefore the LV must generate to open the aortic valve and eject its contents into the circulatory system. This results in a decrease in work and oxygen demand.
The AutoCAT Series IABP Systems consists of two main components:
- The pump control/display module which incorporates a touch screen and keypad for . system operation, and
- . The pneumatic drive module which is incorporated into the body of the device
The AutoCAT Series IABP are designed to be used with 30, 40 and 50cc IAB catheters with the appropriate connectors. The primary difference between the AutoCAT®2 IABP and AC3TM series IABP is the user interface, where the display is a touchscreen.
# Indication for Use
The AutoCAT®2 Intra-Aortic Balloon Pump (IABP) and AC3™ Series Intra-Aortic Balloon Pump (IABP) are clinically indicated for use for the following conditions:
- 1. Acute Coronary Syndrome
- 2. Cardiac and Non-Cardiac Surgery
- 3. Complications of Heart Failure
## Technological Characteristics Comparison
The AutoCAT Series IABP system, including AutoCAT®2 IABP and AC3TM Series IABP, are similar in design and identical in Indications for Use to the predicate IABP system. The motor connector on the MForce motor driver to IABP stepper motor, used in both the AutoCAT®2 and AC3™ Series, are being modified to increase the robustness of the IAB pump system.
The differences between the subject AutoCAT®2 and AC3™ systems and the predicate IABP systems were evaluated to provide evidence that the modification is substantially equivalent to the predicate systems.
The proposed modification to the IABP systems were verified through design verification testing as outlined on the following page.
- . System Level Verification Testing
- Manufacturing Process Verification Test ●
- Performance / Durability Verification Testing .
{5}------------------------------------------------
The results of the verification tests met the specified acceptance criteria and performed similar to the predicate device. The testing demonstrates that the subject devices are substantially equivalent to the predicate devices.
## Substantial Equivalence Conclusion
The subject AutoCAT®2 and AC3™ series Intra-Aortic Balloon Pump (IABP) are substantially equivalent to the specified predicate devices (cleared under K060309 and K192238, respectively) based on a comparison of the device functionality, technological characteristics, and indications for use.
The device modifications and results of design verification tests do not raise new or different questions of safety or effectiveness; therefore, the subject devices are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.