← Product Code [DSP](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSP) · K192238

# AC3 Series Intra-Aortic Balloon Pump (IABP) (K192238)

_Arrow International, Teleflex · DSP · Nov 12, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSP/K192238

## Device Facts

- **Applicant:** Arrow International, Teleflex
- **Product Code:** [DSP](/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSP.md)
- **Decision Date:** Nov 12, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.3535
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The AC3™ Series IABP system is intended to provide counterpulsation therapy to adult patients with impaired Left ventricular function. The users consist of trained hospital personnel in Intensive care areas, Cardiac Cath Labs, Cardiac Operating Rooms and emergency departments. This device is for in-hospital use and for transport between departments in a specific hospital or between facilities as required by the patient condition and treatment required.

## Device Story

AC3 Series Intra-Aortic Balloon Pump (IABP) provides hemodynamic support via counter-pulsation therapy. Device inputs include ECG and arterial pressure (AP) signals; FiberOptix® technology allows AP measurement via fiber-optic catheters. System uses pneumatic drive module to deliver/remove Helium (HE) into intra-aortic balloon catheter positioned in descending thoracic aorta. Inflation during diastole increases blood pressure/organ perfusion; deflation prior to systole reduces LV workload/oxygen demand. Operated by trained hospital personnel in ICUs, Cath Labs, ORs, and emergency departments. Features 'Autopilot' mode for automatic trigger/timing/source selection and 'Operator' mode for manual control. Output displayed on 13.3" touchscreen; provides real-time hemodynamic monitoring and therapy management. Benefits include reduced cardiac work and improved perfusion for patients with impaired LV function.

## Clinical Evidence

Bench testing only. Software verification and system validation testing conducted per IEC 62304 and FDA software validation guidance. Results met all acceptance criteria and demonstrated performance similar to predicate devices.

## Technological Characteristics

System consists of pump control/display module (13.3" touchscreen) and pneumatic drive module. Uses Helium gas for balloon inflation/deflation. Supports 30, 35, 40, and 50cc intra-aortic balloon catheters. Includes FiberOptix® connection for AP measurement. Software V3.11; non-sterile device. Connectivity includes touchscreen and keypad interface.

## Regulatory Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- AC3™ IAB Pump ([K162820](/device/K162820.md))

## Reference Devices

- AutoCAT2 IABP system

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "HUMAN SERVICES-USA" around the edge. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

November 12, 2019

Arrow International, Teleflex Deb Fleetham Manager, Regulatory Affairs 16 Elizabeth Drive Chelmsford, Massachusetts 01824

Re: K192238

Trade/Device Name: AC3 Series Intra-Aortic Balloon Pump (IABP) Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon and Control System Regulatory Class: Class II Product Code: DSP Dated: October 11, 2019 Received: October 15, 2019

Dear Deb Fleetham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K192238

Device Name

AC3 Series Intra-Aortic Balloon Pump (IABP)

Indications for Use (Describe)

The AC3 Intra-Aortic Balloon Pump is clinically indicated for use for the following conditions:

- · Acute Coronary Syndrome
- · Cardiac and Non-Cardiac Surgery
- · Complications of Heart Failure

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D)    |  |
| Over-The-Counter Use (21 CFR 801 Subpart C)     |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 2 510(k) Summary

[As required by 21 CFR 807.92]

Date Prepared: November 7, 2019

510(k) Number: K192238

## Submitter's Name / Contact Person

| Manufacturer                            |
|-----------------------------------------|
| Arrow International, Inc.               |
| A Subsidiary of Teleflex                |
| 16 Elizabeth Drive                      |
| Chelmsford, MA 01824 USA                |
| Establishment Registration # 3010532612 |

Contact Person Deb Fleetham Manager, Regulatory Affairs Cardiac-Interventional BU Tel: 612-403-3806 Fax: 763-656-4253

## General Information

| Trade Name          | AC3™ SERIES IAB Pump                                |
|---------------------|-----------------------------------------------------|
| Common / Usual Name | IAB Pump                                            |
| Product Code        | DSP                                                 |
| Classification Name | 21 CFR 870.3535 – Intra-Aortic Balloon Pump, System |
| Predicate Device    | K162820, AC3™ IAB Pump – cleared March 31, 2017     |

## Device Description

The AC3™ Series IABP System provides counter-pulsation therapy to adult patients with impaired left ventricular (LV) function. It provides hemodynamic support of blood pressure and reduced cardiac work through volume displacement principles. The IABP is attached to an intra-aortic balloon catheter which is inserted into the femoral artery and positioned in the descending thoracic aorta.

The IABP delivers Helium (HE) into the IAB catheter during diastole to displace blood above and below the IAB, increasing blood pressure and perfusion to organs close to the IAB catheter. The IABP deflates or removes HE from the IAB catheter just prior to or in the early phase of systole, reducing the pressure in the aorta and therefore the pressure the LV must generate to open the aortic valve and eject its contents into the circulatory system. This results in a decrease in work and oxygen demand.

The AC3™ Series IABP System consists of two main components:

- . The pump control/display module which incorporates a touch screen and keypad for system operation, and
- The pneumatic drive module which is incorporated into the body of the device ●

The AC3™ Series IABP is designed to be used with 30, 35, 40 and 50cc Intra-aortic balloons with the appropriate connectors.

{4}------------------------------------------------

The AC3™ Series IABP System offers two modes of operation:

- Autopilot Mode, where most functions are automatically selected and controlled by the . IABP
- Operator Mode, where an operator can control most settings and selections.

## Indications for Use

The AC3™ Intra-Aortic Balloon Pump is clinically indicated for use for the following conditions:

- Acute Coronary Syndrome
- . Cardiac and Non-Cardiac Surgery
- Complications of Heart Failure

## Verification and Validation Testing

The software changes were developed in accordance with IEC 62304, Medical Device Software – Software Lifecycle Processes and in accordance with FDA Guidance "General Principles of Software Validation: Final Guidance for Industry and FDA Staff." Software Verification testing has been completed to demonstrate that the software requirements have been met. All testing met the expected results. System verification and validation testing was also conducted, and all results met the required specifications.

## Technological Comparison to Predicate

The AC3™ IABP system with Ver. 3.11 software is similar in design and identical in Indications for Use to the predicate, AC3™ IABP system (V3.7). The AC3 IABP system with Ver. 3.11 software was aligned with the AutoCAT2 IABP system (predicate to AC3) for the Helium Loss 2 alarm. The differences between the subject AC3 system and the predicate IABP systems were evaluated to provide evidence that the AC3 IABP System is substantially equivalent to the predicate systems. The IABP systems were verified through software validation and system verification testing. The results of the testing met the acceptance criteria and performed similar to the predicate device (AutoCAT2 and AC3 Ver.3.7). The testing demonstrates that the catheter is substantially equivalent to the predicate device.

| Table: Substantial Equivalence Comparison     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                 |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Characteristic                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device                                |
|                                               | AC3TM IABP System with SW V3.11                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | AC3TM IABP System with<br>SW V3.07<br>(K162820) |
| Device<br>Classification                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                            |
| Regulation                                    | 21 CFR 870.3535                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                            |
| Product Code                                  | DSP – Intra-aortic balloon and control system                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                            |
| Table: Substantial Equivalence Comparison     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                 |
|                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device                                |
| Characteristic                                | AC3™ IABP System with SW V3.11                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | AC3™ IABP System with<br>SW V3.07<br>(K162820)  |
| Indications for<br>Use                        | The AC3™ Intra-Aortic Balloon Pump is clinically<br>indicated for use for the following conditions:<br>• Acute Coronary Syndrome<br>• Cardiac and Non-Cardiac<br>• Surgery Complications of Heart Failure<br>There are no changes to the Indications for Use                                                                                                                                                                                                                                     | Same                                            |
| Intended Use                                  | The AC3™ Series IABP system is intended to provide<br>counterpulsation therapy to adult patients with<br>impaired Left ventricular function. The users consist of<br>trained hospital personnel in Intensive care areas,<br>Cardiac Cath Labs, Cardiac Operating Rooms and<br>emergency departments. This device is for in-hospital<br>use and for transport between departments in a specific<br>hospital or between facilities as required by the patient<br>condition and treatment required. | Same                                            |
| Compatible<br>Disposable<br>Balloon Catheters | 30, 35, 40, and 50cc Intra-aortic balloon catheters                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                            |
| Display                                       | 13.3" High-Resolution touchscreen display including 6<br>hard keys for frequently used or safety functions.<br>Hard keys and text are both backlit (LED) and color<br>coded for visibility in the dark<br>AC3™ IABP Display head has a color-coded corner switch<br>that illuminates when an alarm is active.                                                                                                                                                                                    | Same                                            |
| Operational<br>Modes                          | Two Operational Modes: Autopilot and Operator mode<br>Autopilot mode includes:<br>• Automatic trigger selection<br>• Automatic ECG/Arterial Pressure (AP) source<br>selection<br>• Automatic timing method selection and timing<br>control<br>• Best Signal Analysis<br>• Deflation Timing Management                                                                                                                                                                                            | Same                                            |
| Table: Substantial Equivalence Comparison     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                 |
| Characteristic                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device                                |
|                                               | AC3™ IABP System with SW V3.11                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | AC3™ IABP System with<br>SW V3.07<br>(K162820)  |
|                                               | Operator mode includes:<br>Seven modes for trigger selection ECG Source: 2 Selection and 1 key press to change leads ECG gain: Automatic, Manual and User adjustable AP Source: 3 Selections AP Scaling: Automatic, Manual and User selectable Touchscreen has the timing controls under the Timing Key. A reminder message is shown if the user leaves the timing screen in 1 2 or lower assist                                                                                                 |                                                 |
| FiberOptix® Technology                        | Allows Arterial Pressure (AP) to be measured by FiberOptic method while using Arrow FiberOptix IAB catheters. The AC3™ Series IABP system has an updated position for the FOS connection                                                                                                                                                                                                                                                                                                         | Same                                            |
| Sterility                                     | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                            |

{5}------------------------------------------------

{6}------------------------------------------------

## Substantial Equivalence Conclusion

The AC3™ IABP with V3.11 software is substantially equivalent to the specified predicate device based on comparisons of the device functionality, principle of operation, technological characteristics, and indications for use. The change to the software and results of subsequent software verification and validations tests do not raise new or different questions of safety or effectiveness; therefore, the AC3™ IABP with V3.11 software is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSP/K192238](https://fda.innolitics.com/submissions/CV/subpart-d%E2%80%94cardiovascular-prosthetic-devices/DSP/K192238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
