MULTI-PARAMETER MOBILE CAREGUIDE 3100 OXIMETER

K130079 · Reflectance Medical, Inc. · MUD · Jul 19, 2013 · Cardiovascular

Device Facts

Record IDK130079
Device NameMULTI-PARAMETER MOBILE CAREGUIDE 3100 OXIMETER
ApplicantReflectance Medical, Inc.
Product CodeMUD · Cardiovascular
Decision DateJul 19, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

The Multi-Parameter Mobile CareGuide™ 3100 Oximeter is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party device, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB or CAN connection. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter has not been demonstrated in disease states.

Device Story

The Multi-Parameter Mobile CareGuide 3100 Oximeter is a non-invasive monitor using Near Infrared (NIR) spectroscopy to measure muscle oxygen saturation (SmO2) and muscle pH (pHm). The device consists of a reusable sensor and a disposable sleeve. The sensor illuminates the skin with LEDs (700-900 nm) and uses a spectroscopic detector to collect reflected spectra. Internal microprocessors control optical components and perform spectral analysis to calculate SmO2 and pHm. Results are transmitted via USB or CAN connection to a third-party display or patient monitor. Used in hospital settings by trained healthcare professionals, the device provides numeric trends of SmO2 and pHm to assist in patient monitoring. It serves as an adjunct tool and is not intended for standalone diagnostic or therapeutic decision-making.

Clinical Evidence

No clinical data provided. Accuracy was established via a bridging study comparing device values against direct blood measurements using a laboratory analyzer. The prospective clinical value of the data in disease states has not been demonstrated.

Technological Characteristics

NIR spectroscopy platform; 700-900 nm LED light source; reusable sensor with disposable sheath; USB/CAN connectivity; microprocessor-based spectral analysis; powered by internal battery.

Indications for Use

Indicated for use as an adjunct, non-invasive monitor of hemoglobin oxygen saturation (SmO2) and pH (pHm) of microvascular blood in skeletal muscle tissue in hospital settings by trained healthcare professionals. Suitable for use on pigmented skin. Not for use as sole basis for diagnosis or therapy.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130079 Page 1 of 5 Reflectance Medical, Inc. 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter # SECTION 5 ## 510(k) SUMMARY # JUL 1 9 2013 # SUMMARY OF SAFETY AND EFFECTIVENESS FOR Multi-Parameter Mobile CareGuide™ 3100 Oximeter ## Submitter Information | Name: | Reflectance Medical, Inc. (RMI) | |----------------------|-----------------------------------------------------------------------------------------------------------------------| | Address: | 116 Flanders Road, Suite 1000 | | | Westborough, MA 01581 USA | | Telephone Number: | 508.366.4700 | | Registration Number: | NA (RMI will apply for registration number following 510(k) clearance, prior to commencement of commercial shipment.) | | Contact Person: | Dr. Babs Soller | | Telephone Number: | 508.366.4700, Ext 223 | | Fax Number: | 508.366.4770 | | Email: | Babs.Soller@reflectancemedical.com | | Date Prepared: | June 12, 2013 | Device Name | Trade name of New Device: | Multi-Parameter CareGuide™ Oximeter | |------------------------------------|-------------------------------------------| | Model Number: | 3100 | | 510(k) Holder/Submitter: | Reflectance Medical, Inc. (RMI) | | 510(k) Number: | N/A | | Proposed Additional product codes: | CBZ, 21 C.F.R. § 868.1170, Anesthesiology | | Classification Panel: | Anesthesiology | Predicate Devices | Predicate Device #1: | Multi Parameter Catheter | |----------------------|--------------------------| |----------------------|--------------------------| | Predicate Device #1 : | Multi Parameter Catheter | |--------------------------|------------------------------------------------------------------------------------------------------------------| | Trade name of Device: | Paratrend™ Multi parameter Sensor and Satellite Monitor System<br>with Paratrend 7 Plus multi-parameter catheter | | Model #: | 7 | | 510(k) holder/Submitter: | Diametrics Medical Limited | . 1 {1}------------------------------------------------ 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter | 510(k) Number: | K970906 | |---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product codes: | CBZ, 21 CFR 868.1170, Anesthesiology | | Predicate Device #2:<br>Trade name of device:<br>Model #:<br>510(k) holder/Submitter:<br>510(k) Number:<br>Product codes: | Terumo Khuri<br>Terumo Khuri™ Myocardial pH Monitoring System<br>N/A<br>Terumo Cardiovascular Systems Corporation<br>K020967<br>CBZ, 868.1170, Anesthesiology | | Predicate Device #3:<br>Trade name of Device:<br>510(k) holder/Submitter:<br>510(k) Numbers:<br>Product Codes: | Masimo multi-parameter oximeter<br>Rainbow SET™ Radical 7 Co-oximeter<br>Masimo Corporation<br>K080238, K061204<br>DQA, 21 CFR 870.2700, Anesthesiology<br>JKS, 21 CFR 862.3220, Clinical Toxicology<br>DPZ, 21 CFR 870.2710, Cardiovascular (K080238) | | Predicate Device #4:<br>Trade Name of Device:<br>Model #:<br>510(k) Holder/Submitter:<br>510(k) Number:<br>Product code: | CareGuide<br>Mobile CareGuide Oximeter<br>1100, 2100<br>Reflectance Medical Inc.<br>K113656, CareGuide 1100<br>K122645, CareGuide 2100<br>MUD, 21 CFR 870.2700, Cardiovascular | ## Device Description The Multi-Parameter Mobile CareGuide 3100 Oximeter sensor uses Near Infrared (NIR) Spectroscopy to calculate muscle oxygen saturation (SmO2) and muscle pH (pHm). > -. {2}------------------------------------------------ 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter | Characteristics | Reflectance Medical Multi-Parameter Mobile CareGuide 3100<br>Oximeter | |------------------------|-----------------------------------------------------------------------| | Principle of Operation | NIR spectroscopy | | Components | Monitor with reusable sensor and disposable pad | | Light Source | LEDs | | Parameters Measured | Tissue oxygen saturation (SmO2) and muscle pH (pHm) | The Multi-Parameter Mobile CareGuide 3100 Oximeter is a multiple parameter oximeter. The sensor contains algorithms that calculate SmO2 and pHm from collected spectra and communicates the current SmO2 and pHm results to a 30 party display or patient monitor through a proprietary protocol. The Multi-Parameter Mobile CareGuide 3100 Oximeter reusable sensor contains the optical and electronic elements necessary to collect spectra from skin, fat and muscle. The sensor has a 3m long cord with either a USB connection or CAN connection to the 310 party display/patient monitor. The sensor is identical to the predicate (K122645) Mobile CareGuide 2100 Oximeter hardware containing six major components: (1) light sources to illuminate the skin; (2) a spectroscopic detector to analyze the reflected spectra back from the subject; (3) a microprocessor to control the optical components; (4) a microprocessor to perform the spectral analysis and generate the calculated SmO2 and pHm; (5) one of two different communications components to transmit in CAN or USB format; (6) a battery to power all components. The Multi-Parameter Mobile CareGuide 3100 Oximeter uses the same disposable element as the Mobile CareGuide 2100 Oximeter, a disposable sleeve that isolates the sensor optical elements from the patient's skin. #### Indications for Use The Multi-Parameter Mobile CareGuide™ 3100 Oximeter is intended for use as an adjunct, noninvasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party device, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB or CAN connection. The Multi-Parameter Mobile CareCuide 3100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter should not be used as the sole basis for diagnosis or {3}------------------------------------------------ 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter has not been demonstrated in disease states. #### Rationale for Substantial Equivalence This device's oximetry feature has been already cleared under classification regulation 21 C.F.R 8 870.2700, Oximeter. The remaining component of the device is pH (21 CFR 868.1170). The Multi-Parameter CareGuide 3100 has the same intended use as one of the identified predicates, the Terumo Khuri Regional pH monitor (K020967). There is no change in how the User would use the information generated by the Multi-Parameter Care Guide 3100 relative to the predicate devices. They are all intended for monitoring of respective parameters. Neither the Multi-Parameter CareGuide 3100 nor any of the predicate devices identified by RMI provide diagnostic output. The Multi-Parameter CareGuide 3100 has the same principle of operation (an optical technological platform that relies on light absorption) as the Paratrend 7 Sensor (pH) (K970906) predicate device. The Multi-Parameter Mobile CareGuide 3100 includes a sensor and monitor and, outputs a numeric trend like these predicate devices. While the optical technology used in the Multi-Parameter CareGuide 3100 is not identical to that of the Paratrend monitor, the accuracy of the CareGuide 3100 is comparable to that predicate device. Much like the testing strategy used by other multi-parameter monitors, accuracy of the CareGuide 3100 was established via a bridging study, comparing CareGuide 3100 values against direct blood measurements using a laboratory analyzer. The Multi-Parameter CareGuide 3100 also has the same technological characteristics as the previously cleared RMI devices, the CareGuide 1100 (K113656) and the Mobile CareGuide 2100 (K122645). - . The principle of operation of the Multi-Parameter Mobile Care Guide 3100 Oximeter is identical to that of the predicate CareGuide devices. They use the exact same NIR spectroscopic platform to measure tissue oxygen saturation and muscle. The same software quantitative algorithm for SmO2 is used in both devices. - The Multi-Parameter Mobile CareGuide 3100 Oximeter is equivalent to the Mobile . CareGuide 2100 predicate in reusable components. Both devices use the exact same sensor hardware: main sensor CPU board, battery, optical board (light sources, spectrometer and microprocessor), CAN/USB interfaces, plastic housing and cables. - The Multi-Parameter Mobile CareGuide 3100 Oximeter is equivalent to the CareGuide � predicates in disposable components. Both devices use the exact same disposable sheath ("Ray") and disposable sensor check device ("Cradle"). {4}------------------------------------------------ 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter - The Multi-Parameter Mobile CareGuide 3100 Oximeter has the identical underlying LED . light source as the CareGuide predicates, with the exact same range of wavelength (700-900 nm). #### Summary of Safety and Effectiveness Data Testing demonstrates that the Multi-Parameter Mobile CareGuide 3100 Oximeter is a safe and effective oximeter meeting all relevant consensuses and FDA recognized standards. The test results in this submission demonstrate that the Multi-Parameter Mobile CareGuide 3100 Oximeter meets the expected performance requirements for an Oximeter, and is therefore equivalent to the predicate relative to safety and mechanical properties. The accuracy and safety of the Multi-Parameter Mobile CareGuide 3100 Oximeter is the same as the predicate device. #### Conclusion The Multi-Parameter Mobile CareGuide 3100 Oximeter is equivalent to predicate device in terms of technology (NIR Spectroscopy) and intended use. The Multi-Parameter Mobile CareGuide 3100 Oximeter does not raise new questions of safety or effectiveness, as compared to the predicate. Therefore, the Multi-Parameter Mobile CareGuide 3100 Oximeter is substantially equivalent to the predicate device. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, which is a symbol of medicine and healthcare. The caduceus features a staff with two snakes coiled around it and a pair of wings at the top. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID-20993-0602 July 19, 2013 Reflectance Medical, Inc. c/o Ms. Nandini Murthy 116 Flaunders Road, Suite 1000 Westborough. MA 01581 Re: K130079 Trade/Device Name: Multi-Parameter Mobile CareGuide 3100 Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD CBZ Dated: June 13, 2013 Received: June 18, 2013 Dear Ms. Nandini Murthy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {6}------------------------------------------------ Page 2 - Ms. Nandini Murthy comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Owen P. Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ 510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter # Indications for Use Form Indications for Use 510(k) Number (if known): K130079 Device Name: Multi-Parameter Mobile CareGuide™ 3100 Oximeter Indications for Use: The Multi-Parameter Mobile CareGuide™ 3100 Oximeter is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on a third party device, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB or CAN connection. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter provides output of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter has not been demonstrated in disease states. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) #### Concurrence of CDRH, Office of Device Evaluation (ODE) Digitally signed by Owen P. Faris -S Date: 2013.07.19 08:19:45-04'00'
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...