K032582 · Ge Medical Systems Information Technologies · MSX · Sep 17, 2003 · Cardiovascular
Device Facts
Record ID
K032582
Device Name
UNITY NETWORK PATIENT DATA SERVER
Applicant
Ge Medical Systems Information Technologies
Product Code
MSX · Cardiovascular
Decision Date
Sep 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Unity Network® Patient Data Server (PDS) is intended for use under the direct supervision of a licensed healthcare practitioner. The Unity Network® Patient Data Server is intended to provide centralized intermediate term storage of patient centric events and physiological data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. Patient events and physiological data stored at the Unity Network® Patient Data Server can be accessed via any (authorized) device that implements the PDS data exchange protocol. The Unity Network Patient Data Server is NOT intended to be the sole source for patient data and is to be used in conjunction with the data at the bedside monitor and central monitoring station.
Device Story
Unity Network® Patient Data Server (PDS) provides centralized intermediate-term storage for physiological data and patient events (arrhythmia, ST history) collected from acquisition devices on the Unity Network. System operates on standard server-class hardware; interfaces with Unity Network and hospital Intranet. Clinicians access stored data via authorized devices implementing proprietary data exchange protocol. Device is not a patient monitor; clinicians must reference primary bedside monitors for patient care decisions. Benefits include centralized access to historical patient data for clinical review. Operated by hospital staff in clinical environments.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness established through risk analysis, requirements reviews, design reviews, module verification, system integration testing, final validation, and performance/safety/environmental testing.
Technological Characteristics
Standard server-class hardware; proprietary data exchange protocol. Connectivity via Unity Network (proprietary) and hospital Intranet. Stores minimum 500 alarm histories (400 arrhythmia, 100 ST) and 72 hours of physiological data per patient. Software-based data management system.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in hospital or clinical settings requiring centralized intermediate-term storage of physiological data and patient events. Must be used under direct supervision of a licensed healthcare practitioner. Not for use as a sole source of patient data; must be used with bedside monitors and central monitoring stations.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
GE Marquette Prism Information Server Applications (MPIS) (K001268)
Related Devices
K032346 — UNITY NETWORK IS PATIENT VIEWER · Ge Medical Systems Information Technologies · Aug 5, 2003
K052160 — ACUITY CENTRAL STATION, MODEL 020XXXXX (NOTE: XXXXX = VARIOUS CONFIG.) · Welch Allyn, Inc. · Dec 16, 2005
K993907 — VIRIDIA INFORMATION CENTER SOFTWARE FOR M3154A OPT C22 · Agilent Technologies, Inc. · Dec 6, 1999
K163584 — M3290B Philips IntelliVue Information Center iX · Philips Medical Systems · Jul 7, 2017
K033030 — CALYSTO SERIES IV PATIENT CARE MONITORS AND CENTRAL STATION AND ECG MANAGEMENT SYSTEM · Witt Biomedical Corp · Feb 24, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
K032582 p1/2
SEP 1 7 2003
Unity Network® Patient Data Server 510k Premarket Notification Submission
# Section 2 Summary and Certification
| 510(k) Summary of Safety and Effectiveness | |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | August 18, 2003 |
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA |
| Contact Person: | Karen Lunde<br>Sr. Regulatory Affairs Specialist<br>GE Medical Systems Information Technologies<br>Phone: (414) 362-3329<br>Fax: (414) 918-8114 |
| Device: | |
| Trade Name: | Unity Network® Patient Data Server |
| Common/Usual Name: | Computer, Information Network Server |
| Classification Names: | 21 CFR 870.2300 System, Network and Communication, Physiological |
| Predicate Device: | K001268 GE Marquette Prism Information Server Applications (MPIS) |
| Device Description: | The Unity Network® Patient Data Server provides a caregiver the ability<br>to view patient centric events and physiological data that was collected<br>from acquisition devices on the Unity Network prior to the current<br>monitoring session. The Unity Network® Patient Data Server stores at<br>least 500 alarm histories (400 arrhythmia events and 100 ST history<br>events) and at least 72 hours of physiological data per unique patient<br>identifier. Information from the Unity Network® Patient Data Server can<br>be accessed by a GEMS-IT device that implements the Unity Network®<br>Patient Data Server's proprietary data exchange protocol.<br><br>The Unity Network® Patient Data Server is a system of hardware and<br>software that operates on standard commercially available server class<br>hardware and interfaces with the Unity Network®. The Unity Network®<br>Patient Data Server system obtains and stores data from devices that<br>are connected and admitted to a monitoring device on the Unity<br>Network® and relays that data to connected clients for viewing. Only<br>clients that are connected to Unity Network® and have access to a<br>proprietary application protocol interface can access the data stored in<br>the Unity Network® Patient Data Server.<br><br>The Unity Network® Patient Data Server system provides data to<br>central station monitoring devices, and is NOT a patient monitoring<br>device. The clinician is instructed to always reference the primary<br>bedside monitor before making any patient care decisions. In the event<br>that data is not available via the Unity Network® Patient Data Server,<br>the clinician is instructed to obtain the data from the primary bedside |
{1}------------------------------------------------
K052582 p2/2
Unity Network® Patient Data Server 510k Premarket Notification Submission
- Intended Use: The Unity Network® Patient Data Server (PDS) is intended for use under the direct supervision of a licensed healthcare practitioner. The Unity Network® Patient Data Server is intended to provide centralized intermediate term storage of patient centric events and physiological data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. Patient events and physiological data stored at the Unity Network® Patient Data Server can be accessed via any (authorized) device that implements the PDS data exchange protocol. The Unity Network Patient Data Server is NOT intended to be the sole source for patient data and is to be used in conjunction with the data at the bedside monitor and central monitoring station.
- Technology: The Unity Network® Patient Data Server system consists of standard server class hardware purchased from an OEM and loaded with application software developed by GE Medical Systems Information Technologies.
The server is connected to two networks: Unity Network® and the hospital's Intranet. The Unity Network® is a currently marketed proprietary network connecting patient monitors. Applications communicate over the Unity Network® using a proprietary protocol in order to obtain or provide data to devices on the network. The hospital's Intranet refers to the existing network within the hospital from which a user can gain access to the product for configuration and servicing.
#### Test Summary:
The Unity Network® Patient Data Server complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the Unity Network® Patient Data Server:
- Risk Analysis ●
- Requirements Reviews
- . Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- . Final acceptance testing (Validation)
- Performance testing .
- Safety testing .
- Environmental testing ●
#### Conclusion:
The results of these measurements demonstrated that the Unity Network® Patient Data Server is as safe, as effective, and perform as well as the predicate device.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows a black and white seal. The seal is circular and contains an image of an eagle with three lines representing its wings. The text around the seal reads "HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 7 2003
GE Medical Systems Information Technologies c/o Ms. Karen Lunde Senior Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee, WI 53223
Re: K032582
Trade Name: Unity Network® Patient Data Server Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm). Regulatory Class: Class II (two) Product Code: MSX Dated: August 18, 2003 Received: August 21, 2003
Dear Ms. Lunde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
### Page 2 - Ms. Karen Lunde
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Prem D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
Unknown; 510(k) filed on August 11, 2003 510(k) Number (if known):
Unity Network® Patient Data Server Device Name
Indications for Use:
The Unity Network® Patient Data Server (PDS) is intended for use under the direct supervision of a licensed healthcare practitioner. The Unity Network® Patient Data Server is intended to provide centralized intermediate term storage of patient centric events and physiological data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. Patient events and physiological data stored at the Unity Network® Patient Data Server can be accessed via any (authorized) device that implements the PDS data exchange protocol. The Unity Network Patient Data Server is NOT intended to be the sole source for patient data and is to be used in conjunction with the data at the bedside monitor and central monitoring station.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Valeria
510(k) |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.