The LifeBed Patient Vigilance System is intended to measure heart rate and respiratory rate in adult patients, in bed, in a general care hospital environment. The system will also monitor bed exit.
Device Story
LifeBed Patient Vigilance System monitors adult patients in general care hospital settings. System comprises bedside unit and Passive Sensor Array (PSA) mattress coverlet. PSA contains sensor array detecting physiological stimuli through clothing/sheets without direct patient contact. Bedside unit processes signals to calculate heart rate, respiratory rate, and bed exit status; displays data via graphical user interface; integrates with hospital nurse call system. If physiological parameters exceed preset limits or bed exit is detected, system triggers nurse call alarm. Clinicians use trended data and waveforms (10-minute to 12-hour views) to inform patient care. Benefits include non-invasive, continuous monitoring without electrodes or leads.
Clinical Evidence
Bench testing only. Electrical, mechanical, and EMC testing performed per IEC 60601-1-2; all tests passed. Biocompatibility testing conducted per ISO 10993 for surface devices in contact with skin <24 hours; all materials passed.
Technological Characteristics
Passive Sensor Array (PSA) mattress coverlet; bedside unit with digital signal processing. Connectivity: interfaces with hospital nurse call system. Biocompatibility: ISO 10993 compliant. Electrical/EMC: IEC 60601-1-2 compliant. Software: GUI-based monitoring system.
Indications for Use
Indicated for adult patients in a general care hospital environment to measure heart rate, respiratory rate, and monitor bed exit status.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
LG1 Intelligent Medical Vigilance System (K052446)
Related Devices
K052446 — LG1 INTELLIGENT MEDICAL VIGILANCE SYSTEM · Hoana Medical, Inc. · Feb 2, 2006
K092037 — LIFEGURNEY NETWORK PATIENT VIGILANCE SYSTEM · Hoana Medical, Inc. · Oct 2, 2009
K180079 — Hill-Rom Vitals Monitoring System powered by EarlySense: Sensor Packs (different #), Hill-Rom Vitals Monitoring System powered by EarlySense: Sensor Activation · Hill-Rom, Inc. · May 7, 2018
K202018 — Hill-Rom Heart and Respiration Rate Monitoring System: Single Sensor, Hill-Rom Heart and Respiration Rate Monitoring System: Sensor Packs, Hill-Rom Heart and Respiration Rate Monitoring System: Sensor Activation · Hill-Rom, Inc. · Mar 4, 2021
K223246 — Hillrom Heart and Respiration Rate Monitoring System powered by EarlySense · Baxter Healthcare Corproration · Nov 18, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
K082366
pg 1 of 3
## SEP 2 6 2008
### Special 510(k) Summarv
| Submitted By: | Hoana Medical Inc.<br>828 Fort Street Mall. Suite 620<br>Honolulu, HI 96813<br>Phone: (808) 523-1225<br>Fax: (808) 523-5480 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact: | Nancy Gertlar<br>Director of QA&RA<br>Voice 808-523-1225<br>FAX 808-523-5480<br>Email ngertlar@Hoana.com |
| Date Summary Prepared: | August 15, 2008 |
| Trade Name: | LifeBed Patient Vigilance System |
| Common/Classification Name: | Monitor, Cardiac<br>74DRT, 870.2300 |
| Substantially Equivalent Devices: | LG1 Intelligent Medical Vigilance<br>System (K052446) |
#### Description of the LifeBed Patient Vigilance System:
The LifeBed Patient Vigilance System consists of two (2) main components: the bedside unit and the Passive Sensor Array™ (PSA™). The bedside unit houses the digital signal processing algorithms that calculate heart and respiratory rates and bed exit status (in-bed sensor) in real time and displays the data as part of the integrated graphical user interface. The bedside unit houses the alarm logic and interfaces to the existing nurse call system found in the hospital. The PSA is comprised of a mattress ticking ("coverlet" that slips over the entire mattress) that houses an array of sensors, which connect to the bedside unit via a cable with an integrated "quick disconnect" safety feature.
The LifeBed Patient Vigilance System's Passive Sensor Array (PSA) technology analyzes basic heart and respiratory physiology without direct patient contact, eliminating the use of electrodes, wires and leads. Hoana's PSA technology integrates proprietary signal processing algorithms with patented data collection devices that produce an electrical signal in response to physiological stimuli. The LifeBed Patient Vigilance System passively extracts patient measurements even through clothing, gowns or sheets.
{1}------------------------------------------------
#### Intended Use
The LifeBed Patient Vigilance System is intended to measure heart rate and respiratory rate in adult patients, in bed, in a general care hospital environment. The system will also monitor bed exit,
#### Principles of Operation
The LifeBed Patient Vigilance System operates in the following manner. The bed on a general care floor is fitted with the sensing array built into a sleep surface coverlet. Each bed is then fitted with a normal sheet and bedding above the coverlet.
The device measures heart rate and respiration and monitors those against a preset limit. If the bedside unit senses the heart rate or respiration rate is outside the desired limits for a period of time, it will signal the nurse with an alarm indicator using the standard nurse call system and the clinician will respond. If the patient has been evaluated as a risk to fall, the bed exit alarm can also be enabled. When the unit senses the patient is attempting to get out of bed, it will use the nurse call system to signal the nurse.
The system will trend and display graphics of heart rate and respiration rate and wave form graphic for respiration in 10 minute, 2, 4, 8, and 12 hour views of the data.
#### Electrical, Mechanical and EMC Testing
Electrical, Mechanical and EMC Testing per IEC 60601-1-2 was performed and the LifeBed Patient Vigilance System passed all tests.
#### Biocompatibility Testing
Biocompatibility was conducted according to the requirements of ISO 10993. Biological Evaluation of Medical Devices, Table 1 - Initial Evaluation Tests for Consideration for a surface device in contact with skin for less than 24 hrs. All test article materials passed.
{2}------------------------------------------------
K082364
pg 3 of 3
#### Conclusion
This Special 510(k) submission covers only the change in the Graphical User Interface (GUI). This change meets the criteria for Special 510(k) under the New 510(k) Paradigm, which states that if the modification does not affect the intended use of the device or alter the fundamental scientific technology of the device, then summary information that results from the design control process can serve as the basis for clearing the application.
There is no change in the indications for use.
There are no changes to labeling that affects the intended use.
There is no change in the fundamental scientific technology of the device.
There are no changes in materials.
Device modifications that should be appropriate for review as a Special 510(k) include "ergonomics of the patient-user interface." This change in GUI is the only change to the original submission covered by this Special 510(k).
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
SEP 2 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hoana Medical, Inc. c/o Ms. Nancy Gertlar Director of Regulatory Affairs and Quality Assurance 828 Fort Street Mall Suite 620 Honolulu, HI 96825
Re: K082366
Trade/Device Name: LifeBed Patient Vigilance System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: DRT Dated: September 6, 2008 Received: September 10, 2008
Dear Ms. Gertlar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or
{4}------------------------------------------------
#### Page 2 - Ms. Nancy Gertlar
any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k)
premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): _K08 2366
## Device Name: LIFEBED PATIENT VIGILANCE SYSTEM
Indications for Use:
The LifeBed Patient Vigilance System is intended to measure heart rate and respiratory rate in adult patients, in bed, in a general care hospital environment. The system will also monitor bed exit.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division/Sign-Off) | |
|------------------------------------|--------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K02360 |
| Page | 1 of 1 |
(Posted November 13, 2003)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.