ACCUTRANS DISPOSABLE PRESSURE MONITORING SYSTEM

K070710 · Biosensors International-USA · DRS · Apr 19, 2007 · Cardiovascular

Device Facts

Record IDK070710
Device NameACCUTRANS DISPOSABLE PRESSURE MONITORING SYSTEM
ApplicantBiosensors International-USA
Product CodeDRS · Cardiovascular
Decision DateApr 19, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2850
Device ClassClass 2
AttributesPediatric

Indications for Use

Accutrans Disposable Pressure Transducer Kit provides the fluid pathway components that are intended to convert the hemodynamic waveform from the patient's catheter, through the disposable pressure transducer with integrated pressure sensor, into electrical signals which can be displayed using separate monitoring equipment. The whole kit is fully disposable and consists of pressure monitoring lines and the Accutrans Disposable Pressure Transducer. Accutrans Disposable Pressure Transducer: The disposable pressure transducer is provided with or without an integral 3cc/hr or 30cc/hr flush devices. Kits with 3cc/hr flush devices are designed exclusively for use with a infusion pump for neonatal and infant applications. The disposable pressure transducer, with the integrated pressure sensor, is intended for direct coupling to sterile catheters with fluid (saline) flows in contact with the patient bloodstream. Accutrans Disposable Pressure Transducer Kits are available for modular assembly of an IV pole-mount transducer organizer system, or for portable on-patient mounting. AccuTrans Pressure Monitoring Kits and AccuTrans Disposable Transducer are indirectly in contact with the blood path and hence, are supplied sterile and is intended for single use only. Pole-mounts, backplates and transducer adapter cables are accessories that are to be used externally outside the human body, and hence supplied non-sterile but hygienically clean. Between surgical cases of patient, suitable cleaning instructions should be followed to minimize cross contamination between patient and health worker.

Device Story

Disposable pressure monitoring system; converts hemodynamic waveforms from patient catheters into electrical signals for external monitoring equipment. Consists of pressure tubing, integrated pressure sensor, and optional flush device (3cc/hr or 30cc/hr). Pole-mounted or patient-mounted; used in clinical settings. Healthcare providers connect device to sterile catheters; fluid (saline) path contacts bloodstream. Output displayed on separate monitors; assists in hemodynamic assessment. Benefits include real-time pressure monitoring and reduced cross-contamination risk via single-use design.

Clinical Evidence

Bench testing only. Functional performance evaluated per IEC 60601-1-34 and ANSI/AAMI BP22-1994. Biocompatibility testing performed per ISO 10993-1 and USP 27 <88>.

Technological Characteristics

Disposable pressure transducer with integrated sensor; fluid-coupled to catheters. Materials biocompatible per ISO 10993-1. Standards: IEC 60601-1-34, ANSI/AAMI BP22-1994. Configurations: pole-mount or patient-mount; optional 3cc/hr or 30cc/hr flush devices. Sterile, single-use.

Indications for Use

Indicated for hemodynamic waveform monitoring in patients via catheter connection. Includes neonatal and infant populations when using 3cc/hr flush device configurations with infusion pumps.

Regulatory Classification

Identification

An extravascular blood pressure transducer is a device used to measure blood pressure by changes in the mechanical or electrical properties of the device. The proximal end of the transducer is connected to a pressure monitor that produces an analog or digital electrical signal related to the electrical or mechanical changes produced in the transducer.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K070710 P1/3 | BIOSENSORS<br>INTERNATIONAL | PREMARKET NOTIFICATION<br>SPECIAL 510K | |-----------------------------|------------------------------------------| | | DISPOSABLE PRESSURE<br>MONITORING SYSTEM | | Page | 1 of 3 | #### SECTION 5 510K SUMMARY [As required by section 807.92(c)] # APR 1 9 2007 #### Prepared on 19 December 2006 #### 1. Applicant Information | Company Address: | Biosensors USA<br>20280 Acacia Street, Suite 300<br>Newport Beach, CA 92660, USA | | |-----------------------|----------------------------------------------------------------------------------|--| | Establishment Number: | 2084493 | | | Contact Person: | Sara Toyloy<br>Phone: (949) 553 8300<br>Facsimile: (949) 553 9129 | | #### 2. Device Identification | Common Name: | Disposable Pressure Monitoring System | |----------------------|-------------------------------------------------| | Trade Name: | Accutrans Disposable Pressure Monitoring System | | Classification Name: | Transducer, Blood-Pressure, Extravascular | #### 3. Predicate Devices - 1. BIOTRANS™ PRESSURE MONITORING KIT (K981747) - 2. UTAH DISPOSABLE PRESSURE TRANSDUCER (K841788) #### 4. Device Description Biosensors's Accutrans Disposable Pressure Transducer System consists of a pressure monitoring kit with pressure tubing connected to the single-use disposable transducer with an integrated pressure sensor. The transducer can come with or without an integral flush device. The complete configuration can be pole-mounted or patient mounted and is provided sterile and non-pyrogenic. #### 5. Intended Use Accutrans Disposable Pressure Transducer Kit is intended to convert the hemodynamic waveform from the patient's catheter, through the disposable pressure transducer with integrated pressure sensor, into electrical signals which can be displayed using separate monitoring equipment. Accutrans Disposable Pressure Transducer is provided with or without an integral 3cchr or 30cchr flush devices. Kits with 30cc/hr flush devices are designed exclusively for use with an infusion pump for neonatal and infant applications, Prepared and Updated by Sylvia ER Dated 19 December 2006 TFAT-FDA-0001-NC {1}------------------------------------------------ | BIOSENSORS<br>INTERNATIONAL | PREMARKET NOTIFICATION<br>SPECIAL 510K | DISPOSABLE PRESSURE<br>MONITORING SYSTEM | |-----------------------------|----------------------------------------|------------------------------------------| | Page | 2 of 3 | | The disposable pressure transducer, with the integrated pressure sensor, is intended for direct coupling to sterile catheters with fluid (saline) flows in contact with the patient bloodstream. ## 6. Summary of Technological Characteristics of Device in relation to Predicate Devices(s) Biosensors plans to introduce a fully disposable pressure monitoring kit. Accutrans models, in addition to the current Biotrans™ models already cleared by 510K (K981747). The materials used in Accutrans transducer are the same as Biotrans™ transducer, however the Accutrans transducer incorporates an in-built sensor. The technology of Accutrans pressure transducer is substantially equivalence to the Utah Deltrans™ pressure transducer, which is also cleared by 510K (K841788). The subject Accutrans models have a similar intended use and design construction with equivalent components as the predicated Biotrans™ and Deltrans™ models. The safety and effectiveness of the subject device models have been assessed and discussed further in Section 7 below. #### 7. Assessment of Performance Data used to justify Substantial Equivalence #### Test Summary, In-vitro Functional performance tests were performed on the Accutrans transducer to evaluate the performance and the reliability of the transducer in accordance to recommended performance standards, IEC 60601-1-34 and ANSI/AAMI BP22-1994. Based on the test results which meet the acceptance criteria, the transducer is concluded to be safe and effective for its intended use. ### Test Summary, Biocompatibility Biocompatibility tests were performed in accordance to ISO10993-1 Part 1: Evaluation & Testing and USP 27 <88> Biological Reactivity Tests. In Vivo. Based on the test results. the Accutrans Pressure transducer is biocompatible and safe for its intended use. ### 8. Conclusion: The Functional performance Tests and Biocompatibility tests have demonstrated that the Biosensors Accutrans Disposable Pressure Monitoring System is substantially equivalent to the predicate devices and is safe and effective for its intended use. {2}------------------------------------------------ | BIOSENSORS<br>INTERNATIONAL | PREMARKET NOTIFICATION<br>SPECIAL 510K | |------------------------------------------|----------------------------------------| | DISPOSABLE PRESSURE<br>MONITORING SYSTEM | Page 3 of 3 | #### 510K Substantial Equivalence Decision-Making Tree Image /page/2/Figure/3 description: This image is a flowchart titled "510(k) \"Substantial Equivalence\" Decision-Making Process (Overview)". The flowchart starts with the question "New Device is Compared to Marketed Device". The flowchart then asks a series of questions, including "Does New Device Have Same Intended Use?" and "Does New Device Have Technological Characteristics That Raise New Types of Safety or Effectiveness Questions?". The flowchart ends with the determination of either "Not Substantially Equivalent" or "Substantially Equivalent". Decision to substantial Equivalence is based on 510(k) Memorandum #K86-3. 510(k). Questions in the decision path are evaluated as follow: Yes. The intended use of the new range of models of the Accutrans transducer is the same as the Blotrans transducer currently listed, Q2. Does New Device have Technological Characteristics that raise new types of Safety or effectiveness questions? No. The new models were designed to have a built-in sensor unlike the current models but the additional sensors housing material and gel were evaluated for biocompalibility and functional performance and demonstrated no new safety and effectiveness issue arises due to technological difference. Q3. Does descriptive or Performance information demonstrate equivalence? Yes. The product description and the intended use on the instruction for Use enclosed in Appendix A between the subject models and predicated models demonstrate the substantial equivalence. The physical and performance characteristics evaluated in Table 1 in Section 5 - Description of the device also showed substantial equivalence between the subject device and predicated device. Conclusion: As according to the decision path, the Accutrans Disposable Pressure Monitoring kit can be submitted using substantial Equivalence. Prepared and Updated by Sylvia ER Dated 19 December 2006 TFAT-FDA-0001-NC Q1. Does New Device have same intended use? {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 9 2007 Biosensors International c/o Sara Toyloy 20280 Acacia St. Suite 300 Newport Beach, CA 92660 Re: K070710 Trade/Device Name: Accutrans Disposable Pressure Monitoring System Regulation Number: 21 CFR 870.2850 Regulation Name: Extravascular blood pressure transducer Regulatory Class: Class II Product Code: DRS Dated: March 9, 2007 Received: March 14, 2007 Dear Ms. Toyloy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ # Page 2 - Ms. Toyloy or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, B. Zimmerman Jr Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K0707/0 | BIOSENSORS<br>INTERNATIONAL | PREI | |-----------------------------|------| |-----------------------------|------| MARKET NOTIFICATION SPECIAL 510K DISPOSABLE PRESSURE Page #### SECTION 4 INDICATIONS FOR USE Common Name: Trade Name: #### DISPOSABLE PRESSURE MONITORING SYSTEM ACCUTRANS PRESSURE MONITORING SYSTEM 1. Intended Use / Indications for Use: Accutrans Disposable Pressure Transducer Kit provides the fluid pathway components that are intended to convert the hemodynamic waveform from the patient's catheter, through the disposable pressure transducer with integrated pressure sensor, into electrical signals which can be displayed using separate monitoring equipment. The whole kit is fully disposable and consists of pressure monitoring lines and the Accutrans Disposable Pressure Transducer. Accutrans Disposable Pressure Transducer: The disposable pressure transducer is provided with or without an integral 3cc/hr or 30cc/hr flush devices. Kits with Succhi flush devices are designed exclusively for use with a infusion pump for neonatal and infant applications. The disposable pressure transducer, with the integrated pressure sensor, is intended for direct coupling to sterile catheters with fluid (saline) flows in contact with the patient bloodstream. Accutrans Disposable Pressure Transducer Kits are available for modular assembly of an IV pole-mount transducer organizer system, or for portable on-patient mounting. AccuTrans Pressure Monitoring Kits and AccuTrans Disposable Transducer are indirectly in contact with the blood path and hence, are supplied sterile and is intended for single use only. Pole-mounts, backplates and transducer adapter cables are accessories that are to be used externally outside the human body, and hence supplied non-sterile but hygienically clean. Between surgical cases of patient, suitable cleaning instructions should be followed to minimize cross contamination between patient and health worker. Prescription Use (Per 21CFR801.109) Over-the-Counter Use (Optional Format 1-2-96) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Or Concurrence of CDRH, Office of Device Evaluation (ODE) B. Ammerman (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number Ka20710
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...