K172959 · Perahealth, Inc. · MWI · May 1, 2018 · Cardiovascular
Device Facts
Record ID
K172959
Device Name
PeraServer and PeraTrend
Applicant
Perahealth, Inc.
Product Code
MWI · Cardiovascular
Decision Date
May 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Pediatric
AI Performance
Output
Acceptance
Observed
Dev DS
Test DS
Rothman Index (RI) Score
Performance equivalent to predicate device (BedsidePEWS)
Performance equivalent to predicate device for both pediatric and adult populations
Not specified
Rothman et. al (2013): Mechanics and validation of the adult RI score; Finlay et. al (2014): Comparison between RI scores and MEWS; Rothman et. al (2017): Mechanics and validation of the pediatric RI score.
Indications for Use
The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values. PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings. PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support. PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g., in the emergency department, on the wards, in intensive care units). The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit.
Device Story
PeraServer/PeraTrend system; software-only platform; imports EHR data (vital signs, nursing assessments, lab values). Computes Rothman Index (RI) score; represents aggregate statistical mortality risk; provides configurable warnings. PeraTrend provides web-based UI for trending/displaying scores; used by healthcare providers in hospital settings (ED, wards, ICU). Adjunct to clinical decision support; does not replace vital signs monitoring. Output assists clinicians in assessing patient physiologic condition; aids clinical decision-making; potentially benefits patients through early identification of status changes.
Clinical Evidence
Three retrospective studies provided: Rothman et al. (2013) (adult RI validation), Finlay et al. (2014) (RI vs. MEWS comparison), and Rothman et al. (2017) (pediatric RI validation). Studies demonstrate performance equivalence to predicate for both pediatric and adult populations.
Technological Characteristics
Software-only system; installed on user-provided hardware. Interfaces with hospital EHR systems. Web-based user interface (PeraTrend). HIPAA compliant. No arrhythmia detection performed; heart rhythm data imported from EHR.
Indications for Use
Indicated for neonatal, pediatric, and adult patients in hospital settings (ED, wards, ICU) to provide a patient status index (Rothman Index) as an adjunct to clinical decision support and vital signs monitoring. Not for use in the Neonatal Intensive Care Unit.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
K241479 — Etiometry Platform (DAV 5.4 RAE 9.2) · Etiometry, Inc. · Feb 12, 2025
K124038 — BEDSIDE PAEDIATRIC EARLY WARNING SYSTEM (BEDSIDEPEWS) · Bedside Clinical Systems, Inc. · Jun 21, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1. 2018
PeraHealth, Inc. % Donna-Bea Tillman Senior Consultant Biologics Consulting 1555 Kingh Street, Suite 300 Alexandria, Virginia 22314
Re: K172959
Trade/Device Name: PeraServer™ and PeraTrend™ System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: March 29, 2018 Received: March 29, 2018
Dear Donna-Bea Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Wilhelmsen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172959
Device Name PeraServer and PeraTrend System
# Indications for Use (Describe)
The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values.
PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings.
PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support.
PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g., in the emergency department, on the wards, in intensive care units).
The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for the PeraServer and PeraTrend System is provided below:
| Device Common Name: | Adjunct to Multiparameter Patient Monitor |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | PeraServerTM and PeraTrendTM System |
| Applicant: | PeraHealth, Inc.<br>6302 Fairview Road, Suite 310<br>Charlotte, NC 28210 |
| Contact: | Joseph Beals, Ph.D.<br>Vice President, Informatics<br>jbeals@perahealth.com |
| Prepared by: | Donna-Bea Tillman, Ph.D.<br>Senior Consultant<br>Biologics Consulting Group, Inc.<br>dtillman@biologicsconsulting.com |
| Date Prepared: | April 26, 2018 |
| Classification Regulation: | 21 CFR 870.2300 (Class II) |
| Regulation Description: | Cardiac Monitor (including Cardiotachometer and Rate Alarm) |
| Panel: | Cardiovascular |
| Product Code: | MWI |
| Product Code Description: | Physiological Patient Monitor (without arrhythmia<br>detection or alarms) |
| Predicate Device: | BedsidePEWS (K124038) |
#### 1. INDICATIONS FOR USE:
The indications for use for the PeraServer and PeraTrend System are the following:
"The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values.
{4}------------------------------------------------
PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings.
PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support.
PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g. in the emergency department, on the wards, in intensive care units).
The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit."
### 2. DEVICE DESCRIPTION:
The subject of this submission is the PeraServer and PeraTrend System, consisting of:
- PeraServer a backend hardware/software platform that imports electronic health record ● (EHR) data, which are used to compute and store Rothman Index (RI) scores and related measures
- PeraTrend – the frontend web-based user interfaces for displaying RI scores and related measures, as well as simple manipulations of the display (i.e., graphical representations over time, comparison of Rothman Index scores and input measures)
PeraServer and PeraTrend are software-only devices that are installed on user-provided hardware.
## SOFTWARE: 3.
Documentation has been provided in the submission commensurate with a Moderate Level of Concern.
## SUBSTANTIAL EQUIVALENCE: 4.
### 4.1. Cited Predicate Device:
The cited predicate device is the Bedside Clinical Systems Bedside Paediatric Early Warning System (BedsidePEWS), cleared under K124038.
## Comparison of Intended Use: 4.2.
The PeraServer and PeraTrend System and the predicate BedsidePEWS have the same intended use, namely to provide healthcare providers with a patient status index that reflects the underlying patient condition. Both devices acquire their inputs from the hospital Electronic Medical Record System and are intended to be an adjunct to vital signs monitoring. Both devices are intended to be used by healthcare professionals in the clinical care environment. The intended patient population of BedsidePEWS is the pediatric population, while the PeraServer
{5}------------------------------------------------
and PeraTrend System is intended for both pediatric and adult populations. This difference in indications does not raise different questions of safety and effectiveness.
### Comparison of Technological Characteristics: 4.3.
| | Subject Device | Predicate Device |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | PeraHealth, Inc. | Bedside Clinical Systems, Inc. |
| Device Name | PeraServerTM and PeraTrendTM System | Bedside Paediatric Early Warning<br>System (BedsidePEWSTM) |
| Classification<br>Regulation | 21 CFR 870.2300, Class II | 21 CFR 870.2300, Class II |
| Product Code | MWI: Physiological Patient Monitor<br>(without arrhythmia detection or<br>alarms) | MWI: Physiological Patient Monitor<br>(without arrhythmia detection or<br>alarms) |
| Device Description | PeraServer collects specified<br>demographic and clinical data from the<br>EHR and ancillary systems. PeraServer<br>computes an Early Warning Score<br>called the Rothman Index (RI),<br>determines if RI scores and/or clinical<br>criteria meet configurable EWS<br>criteria; specified input data, and RI<br>score and warning states are stored for<br>analysis, remote viewing, and<br>interfacing.<br>PeraTrend provides a web-based user-<br>interface for trending and review of RI<br>scores and underlying data and<br>warning status.<br>PeraServer/PeraTrend also interface to<br>other medical automation systems such<br>as the Hospital Electronic Health<br>Record (EHR). | BedsidePEWS is a web-based clinical<br>decision-support tool that can also be<br>run as a standalone application for use<br>on computers running Microsoft<br>Software Operating features with a<br>local network connection, which access<br>medical data. It is also interfaces to<br>other medical automation systems such<br>as Hospital Information System (HIS),<br>Electronic Medical Record (EMR), and<br>Practice Management Software. |
| | Subject Device | Predicate Device |
| Indications for Use | The Rothman Index uses commonly<br>recorded vital sign, nursing<br>assessment, and lab data to compute a<br>patient status index. The Rothman<br>Index is a single measure of a patient's<br>physiologic condition based on the<br>aggregate statistical mortality risk<br>associated with the values of the<br>patient's vital signs, nursing<br>assessments, and selected lab values.<br>PeraServer is indicated for use<br>wherever there is interest in generating<br>Rothman Index (RI) scores and/or<br>associated configurable warnings.<br>PeraTrend is indicated for use by<br>healthcare providers whenever there is<br>need for displaying and/or trending RI<br>scores and displaying associated<br>configurable warning states as an<br>adjunct to clinical decision support.<br>PeraServer/PeraTrend is intended for<br>the care of patients throughout the<br>hospital setting (e.g. in the emergency<br>department, on the wards, in intensive<br>care units).<br>The Rothman Index score is validated<br>for use with neonatal, pediatric, and<br>adult patients. It is an adjunct-to and is<br>not intended to replace vital signs<br>monitoring and is not intended for use<br>in the Neonatal Intensive Care Unit. | The Bedside Paediatric Early Warning<br>System (BedsidePEWS) is an<br>electronic documentation tool that is<br>designed to be used in conjunction with<br>multi-parameter patient monitoring. It<br>is indicated for use by healthcare<br>professionals with paediatric patients<br>between the ages of term newborn (>37<br>weeks gestational age) and 18 years,<br>who are hospitalized with any medical<br>or surgical condition.<br>BedsidePEWS allows input by the<br>healthcare professional of commonly<br>recorded vital sign data and provides<br>clinician with a patient status index (the<br>"BedsidePEWS score") based on a<br>weighted average of seven vital signs<br>when entered by the clinician, namely<br>Heart Rate, Respiratory Rate, Blood<br>Pressure, Oxygen Saturation, Oxygen<br>Therapy, Respiratory Effort, and<br>Capillary Refill. The "BedsidePEWS<br>score" is a single measure of a patient's<br>condition and indicates the variation in<br>the patient's vital signs with respect to<br>normality. BedsidePEWS is an adjunct-<br>to and is not intended to replace vital<br>signs monitoring.<br>BedsidePEWS is intended for use in<br>wards and emergency rooms in<br>hospitals that provide care for children<br>between the ages of term newborn (>37<br>weeks gestational age) and 18 years. It<br>is not intended for use in the Neonatal<br>Intensive Care Unit. |
| Intended users | Healthcare Professionals | Healthcare Professionals |
| Patient Population | Adult and pediatric patients | Pediatric patients between the ages of<br>term newborn (>37 weeks gestational<br>age) and 18 years |
| Use environment | Hospital | Hospital |
| Parameters | Vital signs (Temp, Sys. BP, Dias. BP,<br>HR, RR) Pulse-ox, heart rhythm*,<br>nursing assessments, Braden score,<br>selected labs (potassium, sodium,<br>creatinine, chloride, WBC, BUN,<br>HGB) | Vital signs (Heart Rate, Respiratory<br>Rate, Blood Pressure, Oxygen<br>Saturation, Oxygen Therapy,<br>Respiratory Effort, and Capillary<br>Refill) |
| | Subject Device | Predicate Device |
| Score | Rothman Index (RI) and Pediatric RI<br>(pRI) | BedsidePEWS score |
| Material | Software only | Software only |
| Hardware Platform | PeraServer/PeraTrend require<br>dedicated application and database<br>servers. | Windows PC or Tablet |
| Data Source | Hospital EHR systems | Hospital EHR systems or manually<br>input |
| Data Display | PeraTrend Web – a browser based<br>display<br>PeraTrend Kiosk – a browser based<br>display configured for dedicated<br>monitors providing an always-up,<br>auto-refreshed display<br>PeraTrend EHR – a browser based<br>display accessed and viewed from<br>within the EHR | Browser based display |
| Ability to trend | Yes | Yes |
| Real-time display | Yes | Yes |
| Privacy | HIPPA compliant | HIPPA compliant |
{6}------------------------------------------------
{7}------------------------------------------------
*Note: Although the PeraServer and PeraTrend System uses heart rhythm as one of its inputs, this information is obtained from the EHR and therefore the PeraServer and PeraTrend System is not performing arrhythmia detection.
### PERFORMANCE TESTING: 5.
To demonstrate the validity of the RI score, three retrospective studies are described in the submission:
- Rothman et. al (2013), to describe the mechanics and validation of the adult RI score ●
- Finlay et. al (2014), to provide a comparison between RI scores and MEWS ●
- Rothman et. al (2017), to describe the mechanics and validation of the pediatric RI score ●
These studies demonstrate that the performance of the PeraServer and PeraTrend System is equivalent to that of the predicate device for both pediatric and adult populations.
### CONCLUSION: 6.
The subject PeraServer and PeraTrend System and the predicate BedsidePEWS have the same intended use, which is to provide healthcare providers with a patient status index that reflects the
{8}------------------------------------------------
underlying patient condition. The differences in technological characteristics do not raise different questions of safety and effectiveness, and the results of performance testing demonstrate that the PeraServer and PeraTrend System performs in accordance with specifications and meets user needs and intended uses. The software testing provided in the submission verifies the functionality of the software, and the studies provided in the submission validate the Rothman Index score. Therefore, the PeraServer and PeraTrend System has been demonstrated to be substantially equivalent to the predicate BedsidePEWS.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.