K101052 · Infinium Medical · MWI · Aug 11, 2010 · Cardiovascular
Device Facts
Record ID
K101052
Device Name
OMNI III PATIENT MONITOR
Applicant
Infinium Medical
Product Code
MWI · Cardiovascular
Decision Date
Aug 11, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The purpose and function of the OMNI III patient monitor is to monitor basic physiological parameters including, ECG, heart rate, NIBP (systolic, diastolic, and mean arterial pressures), SpO2, respiration, and temperature for adult, neonate and pediatric patients. It may be used as a bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities.
Device Story
Omni III Patient Monitor is a compact, portable, bedside monitoring system; collects physiological data including ECG, heart rate, NIBP, SpO2, respiration, and temperature. Device integrates parameter measuring modules, display, and recorder; features built-in battery for transport. Operated by healthcare professionals in hospitals, clinics, and emergency rooms. System processes and displays up to six parameters across eight traces. Output provides real-time physiological data to clinicians for patient assessment and decision-making; portability facilitates continuous monitoring during patient transport.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to the predicate device.
Technological Characteristics
Compact, portable patient monitor; integrates parameter modules, display, and recorder. Powered by built-in battery for transport. Monitors ECG, heart rate, NIBP, SpO2, respiration, and temperature. System displays up to six parameters via eight traces.
Indications for Use
Indicated for adult, pediatric, and neonate patients requiring monitoring of ECG, heart rate, NIBP, SpO2, respiration, and temperature in hospital, clinic, or emergency room settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
K112329 — OMNI PATIENT MONITOR · Infinium Medical · Sep 8, 2011
K103737 — OMNI II PATIENT MONITOR · Infinium Medical · Jun 2, 2011
K103268 — OMNI EXPRESS · Infinium Medical · Nov 23, 2010
K132229 — OMNI PATENT MONITORS · Infinium Medical · Feb 21, 2014
K153079 — Axeo 3 Patient Monitor · Axeo Medical Technologies, LLC · Mar 17, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
1/2 KIDIOS2
## 510(k) SUMMARY (as required by 807.92(c))
AUG 1 1 2010
Regulatory Correspondent: AJW Technology Consultants Inc 962 Allegro Lane Apollo Beach, FL 33572 John O'Brien jobrien@ajwtech.com 813-645-2855 813-677-4787
Submitter of 510(k):
Infinium Medical 12151 62nd Street North #5 Largo, FL 33773 Suleyman Bilgutay sales@infiniummedical.com
| Date of Summary: | February 24, 2010 |
|-------------------------|-------------------------------------------------------------------------|
| Trade/Proprietary Name: | Omni III Patient Monitor. |
| Classification Name: | Monitor, physiological, patient (without arrhythmia detection or alarms |
| Product Code: | MWI |
## Intended Use:
The purpose and function of the OMNI III patient monitor is to monitor basic physiological parameters including, ECG, heart rate, NIBP (systolic, diastolic, and mean arterial pressures), SpO2, respiration, and temperature for adult, neonate and pediatric patients. It may be used as a bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities.
## Device Description:
The OMNI III monitor is a comprehensive monitoring system with eight traces compiling, processing, analyzing and displaying data from up to six different patient parameters. It integrates parameter measuring modules, display and recorder in one device, featuring in compactness, lightweight and portability. Built-in battery facilitates transportation of patient.
{1}------------------------------------------------
Predicate Device:
K021154 - Goldway UT4000F Patient Monitor
Substantial Equivalence:
The proposed device is substantial equivalent to the Goldway UT4000F, which has been cleared under K021154. The proposed device has the same intended use and similar technological characteristics as compared to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is facing to the right. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Infinum Medical, Inc. c/o John O'Brien AJW Technology Consultants, Inc. 962 Allegro Ln. Apollo Beach, Florida 33572
AUG 1 1 2010
Re: K101052
Trade/Device Name: Omni III Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm Regulatory Class: Class II (two) Product Code: MWI Dated: July 6, 2010 Received: July 8, 2010
Dear Mr. O'Brien:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. John O'Brien
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wand
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
10:1052 1/1
Indications for Use
AUG 1 1 2010
510(k) Number (if known): K101052
Device Name: OMNI III Patient Monitor
The purpose and function of the OMNI III patient monitor is to monitor basic physiological parameters including, ECG, heart rate, NIBP (systolic, diastolic, and mean arterial pressures), SpO2, respiration, and temperature for adult, neonate and pediatric patients. It may be used as a bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
i
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K101052 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
3c3 of 62
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.