MEDICK MHM200 PERSONAL HEART MONITOR

K063339 · Medick Healthcare , Ltd. · MLO · Jun 29, 2007 · Cardiovascular

Device Facts

Record IDK063339
Device NameMEDICK MHM200 PERSONAL HEART MONITOR
ApplicantMedick Healthcare , Ltd.
Product CodeMLO · Cardiovascular
Decision DateJun 29, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2

Indications for Use

The MHM200 Monitor is intended for in-context monitoring of adult patients where symptoms occur infrequently or where resting ECG monitoring is unlikely to capture symptomatic or asymptomatic abnormal rhythm events. Its use includes assessing symptoms (including palpitations, dizzy spells, paroxysmal light-headedness, and pounding of the heart) that may relate to disturbances of the heart. Where it is clinically appropriate, the MHM200 Monitor can assist in ongoing monitoring of the frequency of abnormal rhythm events in patients receiving antiarrhythmic medication or rehabilitation therapy.

Device Story

Medick MHM200 is a portable, single-channel ambulatory ECG monitor; records heart rhythm on beat-by-beat basis for up to 8 hours. Device features LCD for setup and display of cardiac information; powered by two AAA batteries. Input via patient cables; data transferred to computer via USB connection for storage and printing. Patient cable must be disconnected before USB connection. Used in clinical or home settings; operated by patients or clinicians. Provides cardiac rhythm data to assist clinicians in diagnosing heart disturbances and monitoring therapy efficacy. Benefits include capturing infrequent or asymptomatic cardiac events not detected by resting ECG.

Clinical Evidence

Bench testing only. Device performance validated against AAMI/ANSI EC38:1998 (Ambulatory Electrocardiographs) and AAMI/ANSI EC57-98 (Testing and Reporting Performance Results of Cardiac Rhythm and ST-Segment measurement Algorithms). Safety and EMC testing performed per IEC/EN 60601-1 and IEC/EN 60601-1-2.

Technological Characteristics

Single-channel ambulatory ECG monitor. Powered by 2x AAA alkaline batteries. Features LCD display. Connectivity via USB for data transfer. Proprietary patient/USB cables with common input connector. Complies with IEC/EN 60601-1 (safety), IEC/EN 60601-1-2 (EMC), 21 CFR 898 (cables), AAMI/ANSI EC12:2000 (electrodes), AAMI/ANSI EC38:1998 (ECG), and AAMI/ANSI EC57-98 (algorithms).

Indications for Use

Indicated for adult patients experiencing infrequent symptoms or requiring capture of symptomatic/asymptomatic abnormal heart rhythms. Used for assessing palpitations, dizzy spells, light-headedness, and heart pounding; also for monitoring rhythm frequency in patients on anti-arrhythmic medication or rehabilitation therapy.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K063339 # 510(k) Summary of Safety and Effectiveness JUN 2 9 , 007 # Submitter Information: Medick Healthcare Ltd. 4" Floor Exchange House 494 Midsummer Boulevard Central Milton Keynes Mk9 2EA United Kingdom Phone: +44 (0) 1908 255945 FAX: +44 (0) 1908 255700 Contact: Graham Wilson, Chairman and CEO, Medick Healthcare Limited # USA Contact: B. Harlow & Associates, LLC 803 NE Juanita Drive Kirkland, WA 98034 USA Phone: 425 891-7347 FAX: 425 823-4469 # Device Name: Trade Name: Medick MHM200 Personal Heart Monitor Common Name: Event Recorder Classification Name: Electrocardiograph, Ambulatory, with Analysis Algorithm # Predicate Device: The Medick MHM200 Personal Heart Monitor is substantially equivalent to the C.Net 2100 Single Channel ECG Ambulatory Monitor cleared for market by the United States Food and Drug Administration under 510(k) number K010396. # Device Description: The Medick MHM200 is an ECG testing device that monitors heart on a beat by beat basis to detect any irregularities. It does this over a period of up to 8 hours and reports a person's heart rhythm. It is small, light weight, portable, single channel ECG ambulatory monitor equipped with an LCD that is used to set-up the monitor and display cardiac information. The monitor is powered by two (2) standard AAA alkaline batteries. The MHM200 can communicate with a computer over a standard USB connection to store and print a person's cardiac information. For safety reasons, the MHM200 uses proprietary patient and USB cables. The cables share a common input connector. The patient cable must be removed before the monitor can be connected to a computer through the USB cable. # Indications for use: The MHM200 Monitor is intended for in-context monitoring of adult patients where symptoms occur infrequently or where resting ECG monitoring is unlikely to capture symptomatic or asymptomatic abnormal mythm events. Its use includes assessing symptoms (including palpitations, dizzy spells, paroxysmal lightheadedness, and pounding of the heart) that may relate to disturbances of the heart. Where it is clinically appropriate, the MHM200 Monitor can assist in ongoing monitoring of the frequency of abnormal rhythm events in patients receiving anti-arrhythmic medication or rehabilitation therapy. {1}------------------------------------------------ # 06:33339 ### Technology Characteristics: Technology characteristics of the Medick MHM200 are substantially equivalent to the C.Net 2100, Both products were developed by Cardionetics Ltd. The ECG channel and annythmia analysis algorithm are the same. New software has been added to the MHM200 to drive the LCD display and the housing has been changed to accommodate the display. New directions for use were developed to describe the display and ensure that the operator can use the MHM200 safely and for the purpose for which it is intended. ### Summary of Performance Testing: | Item | Title | Number | |------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | 1 | Safety: IEC/EN 60601-1 Medical Electrical Equipment -<br>Part 1 General Requirements for Safety | IEC/EN 60601-1 | | 2 | Safety: IEC/EN 60601-1 Medical Electrical Equipment -<br>Part 1 General Requirements for Safety, 1988:Amendment<br>1, 1991-11, Amendment 2, 1995 | IEC/EN 60601-1<br>Amendment 1 &<br>Amendment 2 | | 3 | EMC: EN 60601-1-2 Medical Electrical Equipment - Part 1-<br>2: General Requirements for Safety; Electrometric<br>Compatibility - Requirements and Tests | IEC/EN 60601-1-2 | | 4 | Patient Cables: 21 CFR 898 Performance Standard for<br>Electrode Lead Wires and Patient Cables | 21 CFR 898 | | 5 | Disposable ECG electrodes | AAMI/ANSI<br>EC 12:2000 | | 6 | ECG: AAMI/ANSI EC38:1998 Ambulatory<br>Electrocardiographs | AAMI/ANSI<br>EC38:1998 | | 7 | Cardiac Rhythm: AAMI/ANSI EC57-98 Testing and<br>Reporting Performance Results of Cardiac Rhythm and ST-<br>Segment measurement Algorithms | AAMI/ANSI<br>EC57-98 | The Medick MHM200 has been tested and conforms to the appropriate sections of the above noted standards. #### Conclusion: The Medick MHM200 Personal Heart Monitor is as safe and effective as the predicate devices when used according to the instructions in the directions for use supplied with the monitor. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 9 2007 Medick Healthcare Ltd. c/o James W. Sandberg, P.E. B. Harlow & Associates 8303 N.E. Juanita Drive LLC Kirkland, WA 98034 Re: K063339 Trade Name: Medick MHM 200 Personal Heart Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: MLO Dated: June 26, 2007 Received: June 27, 2007 Dear Mr. Sandberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. James W. Sandberg, P.E. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Dvina R. Lechner >Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 # INDICATIONS FOR USE # 510(k) Number: K063339 Device Name: Medick MHM200 Personal Heart Monitor ## Indication for Use: The MHM200 Monitor is intended for in-context monitoring of adult patients where symptoms occur infrequently or where resting ECG monitoring is unlikely to capture symptomatic or asymptomatic abnormal rhythm events. Its use includes assessing symptoms (including palpitations, dizzy spells, paroxysmal light-headedness, and pounding of the heart) that may relate to disturbances of the heart. Where it is clinically appropriate, the MHM200 Monitor can assist in ongoing monitoring of the frequency of abnormal rhythm events in patients receiving antiarrhythmic medication or rehabilitation therapy. Prescription Use _ X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) uma R. hachner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_K063339
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...