TM2000 EASYTRACE PLUS RECEIVING CENTER

K013175 · Card Guard Scientific Survival , Ltd. · DXH · Oct 30, 2001 · Cardiovascular

Device Facts

Record IDK013175
Device NameTM2000 EASYTRACE PLUS RECEIVING CENTER
ApplicantCard Guard Scientific Survival , Ltd.
Product CodeDXH · Cardiovascular
Decision DateOct 30, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2920
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The TM2000 EasyTrace Plus Telemedicine Receiving Center is intended for supporting transtelephonic monitoring of Electrocardiography (ECG) and Spirometric parameters of patients.

Device Story

System receives, stores, displays, measures, updates, prints, and re-transmits patient ECG and spirometric data. Inputs include analog/digital signals via acoustic/modem or direct IR interface from Card Guard transmitters. Operates on MS Windows-based PC; used by clinicians in remote monitoring centers. Provides database management for up to 40,000 transmissions. Enables healthcare providers to view and analyze patient data to support clinical decision-making and remote patient management.

Clinical Evidence

Bench testing only. System compliance verified through design controls, risk analysis (EN1441, IEC 1025, FMEA), and adherence to safety standards (IEC 601-1, IEC 601-1-1, IEC 601-1-4). No clinical data presented.

Technological Characteristics

PC-based software system running on MS Windows. Interfaces: analog/digital (acoustic/modem) and IR. Database capacity: up to 40,000 transmissions (2 GB max). Moderate level of concern. Complies with IEC 601-1, IEC 601-1-1, IEC 601-1-4, and ISO 9002.

Indications for Use

Indicated for patients requiring remote transtelephonic monitoring of ECG and spirometric parameters. Prescription use only.

Regulatory Classification

Identification

A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | | | K013175 p.1/2 | | |------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------|--------------| | TM2000 EasyTrace Plus Transtelephonic Receiving Center<br>510(k) Summary of Safety and Effectiveness | | | | | Submitter: | Card Guard Scientific Survival Ltd.,<br>2 Pekeris St. P.O.B. 527<br>Rehovot 76100, Israel<br>Tel: 972-8-9484600<br>Fax: 972-8-9484605 | | OCT 3 0 2001 | | Contact Person: | Alex Gonorovsky,<br>Deputy Chief Engineer, Regulatory Affairs<br>Tel: 972-8-9484624<br>E-mail: alexanderg@cardguard.com | | | #### Date Prepared: August 12, 2001 # 1. Definition and Intended Use The TM2000 EasyTrace Plus Transtelephonic Receiving Center system comprises a receiver, a PC with associated equipment, and a package of software tools. The System is intended to support a remote monitoring, i.e. receiving, storing, displaying, measuring, updating, printing and re-transmitting of patient ECG and Spirometric parameters and other relevant data. ## 2. Device Class The TM2000 EasyTrace Plus Transtelephonic Receiving Center system is classified as Class II medical device (21 C.F.R. Par. 870.2920 (1992)). ## 3. Applicable Standards No performance standards have been developed under Section 514 of the Federal Food, Drug and Cosmetic Act for telephone ECG and Spirometric transmitter devices. TM2000 EasyTrace Plus meets the requirements of the following standards and guidances: - EN1441: 1997 Medical Devices Risk Analysis ● - IEC 1025: 1990 Fault tree analysis (FTA) ● - IEC/TR 513: 1994 Fundamental aspects of safety standards for medical ● electrical equipment - IEC 801-1, 1984, "General Introduction" ● - IEC 601-1, 1996, "Medical Electrical Equipment, General Requirements for Safety" ● - IEC 601-1-1, 1996, "Safety Requirements for Medical Electrical Systems" ● - . IEC 601-1-4, 1996, "Part 1-4, Programmable Electrical Medical Systems" - IEC 812: 1985 Analysis techniques for system reliability - Procedure for failure mode and effects analysis (FMEA) - IEC 300-3-9: 1995 Dependability management, Part 3: Application guide -. Section 9, Risk analysis of technological systems - . "Reviewer Guidance for Computer Controlled Medical Devices", FDA Aug 29, 1991 - ISO/IEC Guide 51: 1990 Guidelines for the inclusion of safety aspects in standards . - ISO 9002 guidelines ● {1}------------------------------------------------ **CARD GUARD** Scientific Survival LTD. TM2000 EasyTrace Plus Transtelephonic Receiving Center 510(k) Summary of Safety and Effectiveness - EN-46002 - IEEE Standard for Software Quality Assurance Plan . - FDA's Guidance for the Content of Premarket Submissions for Software . Contained in Medical Devices, May 29, 1998 - FDA's "New 510(k) Paradigm, Alternate Approaches to Demonstrating Substantial ● equivalence in Premarket Notifications" Final Guidance, CDRH, March 20, 1998. #### 4. Features - Runs on any MS Windows operating system . - Manual entry of patient and physician detail . - Analog/digital (acoustic/modem) receiving or direct download via IR interface ● - Data processing capabilities . - ECG and Spirometric event recording . - Receiving, viewing and printing medical data from Card Guard's transmitters . - Storage of up to 10 40 thousand transmissions (depending on transmitter type) . - Maximum database file size of 2 GB ● #### User Interface 5. The TM2000 EasyTrace Service Plus GUI enables access to all categories of data through 3 built-in subsystems. #### Substantial Equivalence ે. Card Guard hereby claims that the TM2000 EasyTrace Plus is substantially equivalent to Telemedicine 2000, the Transtelephonic Receiving Center, K992164 The proof of substantial equivalence in all that concerns the intended use, principles of operation, features and technological characteristics is provided in Chapter 7. Substantial Equivalence to Cleared Devices. #### Design Controls and Hazard Analysis 7. The Card Guard's product design procedure, and QA and QC policy, formalize the design and production process and assure that all the respective requirements are met. In the framework of the Design Controls the testing was conducted to verify the system compliance with all its design specifications. The device Level of Concern criteria were evaluated and the system was determined to be a moderate level of concern system. The rigorous design evaluation and the System Safety and Risk analysis expose potential failures or possible system flaws which could directly or indirectly effect the patient. #### Conclusions 8. The system constitutes a safe and reliable means for receiving, storing, displaying, analyzing, updating, printing and re-transmitting of patient ECG and Spirometric parameters and other patient related data. Its operation present no adverse health effect or safety risks to patients when used as intended. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a person with their arms outstretched. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 4 2002 Card Guard Scientific Survival Ltd. c/o Mr. Alex Gonorovsky Deputy Chief Engineer, Regulatory Affairs 2 Pekeris Street P.O. Box 527 Rehovot 76100 ISRAEL Re: K013175 Trade Name: TM2000 Easy Trace Plus Receiving Center Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitters and Receivers Regulatory Class: Class II (two) Product Code: DXH Dated: October 15, 2001 Received: October 17, 2001 Dear Mr. Gonorovsky: This letter corrects our substantially equivalent letter of October 30, 2001, regarding the indications for use statement. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Alex Gonorovsky Please be advised that FDA issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Card Guard, a company that specializes in scientific survival. The logo is in black and white and features the words "CARD GUARD" in large, bold letters. Below the company name is the text "Scientific Survival LTD." The logo also includes a graphic of a heartbeat line. The TM2000 EasyTrace Plus Telemedicine Receiving Center is intended for supporting transtelephonic monitoring of Electrocardiography (ECG) and Spirometric parameters of patients. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Deyice Evaluation (ODE) Division of Cardiovascular & Respiratory Devices 510(k) Number K013175 V Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use OR (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...