K254114 · Baylis Medical Technologies, Inc. · DSA · Jan 18, 2026 · Cardiovascular
Device Facts
Record ID
K254114
Device Name
Baylis Connector Cable
Applicant
Baylis Medical Technologies, Inc.
Product Code
DSA · Cardiovascular
Decision Date
Jan 18, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture Generators to separately cleared compatible RF devices.
Device Story
Baylis Connector Cable is a sterile, single-use 10-foot cable; connects Baylis RF Puncture generators (e.g., PrecisePath, RFP-100A) to compatible RF delivery devices (e.g., RF guidewires). Device functions as a physical conduit for RF power transfer from generator to puncture device. Used in clinical settings by healthcare professionals. Output is the delivery of RF energy to the compatible device; enables clinical procedures involving RF puncture. Benefits include reliable power transmission for RF-based interventions.
Clinical Evidence
Bench testing only. Verification and validation activities included electrical safety, EMC, generator compatibility, mechanical integrity (connector strength, cable durability), and sterilization/packaging validation. Results demonstrate equivalence to the predicate device.
Technological Characteristics
10-foot single insulated conductor cable; multi-pin positive locking generator connector; spring-loaded wire connection connector. Sterile, EO sterilized. 4-year shelf life. Complies with IEC 60601-1 and IEC 60601-2-2. No software or active electronic components; passive electrical cable.
Indications for Use
Indicated for connecting Baylis RF Puncture generators to separately cleared compatible RF devices for use in clinical procedures requiring RF energy delivery.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
K230571 — RFP-100A Connector Cable (Single Use) · Baylis Medical Technologies, Inc. · May 30, 2023
K023334 — BMC CATHETER CONNECTOR CABLE, MODEL RFP-101 · Baylis Medical Co., Inc. · Dec 31, 2002
K132374 — BAYLIS MEDICAL RF PERFORATION PROBE V2.0 · Baylis Medical Company, Inc. · Aug 30, 2013
K250275 — PrecisePath Radiofrequency Puncture Generator and PrecisePath Footswitch · Baylis Medical Technologies, Inc. · Oct 28, 2025
K232562 — PowerWire Radiofrequency Guidewire Kit · Baylis Medical Technologies, Inc. · Feb 28, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 18, 2026
Baylis Medical Technologies, Inc.
Barb Boyce
Senior Regulatory Affairs Specialist
2645 Matheson Blvd. E, Mississauga, On L4w 5m2
Mississauga, ON L4W 5S4
Canada
Re: K254114
Trade/Device Name: Baylis Connector Cable
Regulation Number: 21 CFR 870.2900
Regulation Name: Patient Transducer And Electrode Cable (Including Connector)
Regulatory Class: Class II
Product Code: DSA
Dated: December 18, 2025
Received: December 19, 2025
Dear Barb Boyce:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254114 - Barb Boyce
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254114 - Barb Boyce
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARCO CANNELLA -S
for
- Aneesh Deoras
- Assistant Director
- Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices
- Office of Cardiovascular Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254114 | ? |
| Please provide the device trade name(s). | | ? |
| Baylis Connector Cable | | |
| Please provide your Indications for Use below. | | ? |
| The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture Generators to separately cleared compatible RF devices. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
# 510(k) Summary
21 CFR 807.92
## I. Submitter Information
Company Name: Baylis Medical Technologies Inc.
Address: 2645 Matheson Blvd. East, Mississauga, Ontario L4W 5S4, Canada
Phone: (905) 948-5800
Contact Person: Barb Boyce
Summary Prepared On: 16-Dec-2025
## II. Device Identification
Device Trade Name: Baylis Connector Cable
Device Common Name: Cable, transducer and electrode, patient, including connector
Classification Name: 21 CFR 870.2900
Device Class: Class II
Product Code: DSA
## III. Identification of Predicate Device
Predicate Device: RFP-100A Connector Cable (Single Use), cleared under 510(K) number K230571
Predicate Device Manufacturer: Baylis Medical Technologies Inc.
Recall Status: Predicate device has not been the subject of a design-related recall.
## IV. Device Description
The Baylis Connector Cable is a sterile, single-use device designed to connect Baylis-approved Radiofrequency (RF) generators (including the PrecisePath RF Puncture Generator and RFP-100A Generator) to separately cleared compatible RF devices (e.g., RF guidewires).
Special 510(k)
Special 510(K) Summary
{5}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
The Baylis Connector Cable is 10 feet long with a multi-pin positive locking Generator Connector at one end and a spring-loaded wire connection connector at the other end which allows connection to the RF delivery device. Between the connector ends is a single insulated conductor cable.
Principle of Operation: Enables RF power transfer from generator to RF puncture device; no changes from predicate device.
# V. Indications for Use
The Baylis Connector Cable is intended to connect Baylis Radiofrequency (RF) Puncture generators to separately cleared compatible RF devices.
# VI. Comparison to Predicate Device
The Baylis Connector Cable and its predicate share the same intended use, principles of operation, and fundamental technology. Differences are minor and do not raise new questions of safety or effectiveness.
| Detail | Equivalence | Comment |
| --- | --- | --- |
| Legal Manufacturer | Identical | Baylis Medical Technologies Inc. |
| Indications for Use | Similar | Change in the name of the cable and updated list of compatible generators |
| Classification | Identical | 21 CFR 870.2900 |
| FDA Product Code | Identical | DSA |
| Sterility | Identical | Sterile, EO |
| Design | Similar | Both the predicate and subject device share the same fundamental design, the subject device differs as follows:
- Minor design change to connector ends |
| Materials | Similar | Both the predicate and subject device share the same fundamental design, the subject device differs in materials as follows:
- Minor change to insulation and connector material |
| Compatible Generators | Minor difference | Update to list of compatible generators to include PrecisePath Radiofrequency Puncture Generator (PRG-1) |
| Packaging | Identical | Tyvek/poly pouch in shelf box |
| Shelf-life | Identical | 4 years |
Special 510(k)
Special 510(K) Summary
{6}
K254114
Baylis Connector Cable
Baylis Medical Technologies Inc.
# VII. Performance Data
Verification and validation activities confirm that the Baylis Connector Cable meets the same requirements as the predicate device. Testing included:
- Electrical Safety, EMC, and Generator Compatibility: IEC 60601-1 and IEC 60601-2-2 applicable requirements
- Mechanical Integrity: Connector strength, cable durability
- Sterilization & Packaging: EO sterilization validation, shelf-life testing (leveraged from predicate device’s 510(K) since these parameters remain unchanged)
Results demonstrate that the subject device performs equivalently to the predicate device and does not introduce new risks.
# VIII. Conclusions
The Baylis Connector Cable is substantially equivalent to the predicate device (RFP-100A Connector Cable (Single Use), K230571). Both devices share the same intended use, principles of operation, and fundamental technology. Verification and validation of design and material changes to the Baylis Connector Cable further support substantial equivalence to the legally marketed predicate device.
Special 510(k)
Special 510(K) Summary
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.