Intended for use in general electrocardiograph procedures where ECG monitoring is deemed necessary and is ordered by a physician.
Device Story
Disposable ECG monitoring electrode; conductive wet gel interface; silver/silver chloride plated ABS eyelet; nickel plated brass stud; adhesive foam substrate; vinyl label; polystyrene release liner. Used in clinical settings for ECG monitoring; applied to patient skin by healthcare personnel to acquire electrical signals for diagnostic/monitoring equipment. Benefits include reliable signal acquisition for cardiac monitoring.
Clinical Evidence
Bench testing only. Device tested per AAMI EC12-1991 standards for electrical performance, labeling, shelf life, packaging, and safety. Biocompatibility testing performed per ISO-10993; results confirmed non-irritating, non-cytotoxic, and non-sensitizing. Real-time shelf life studies substantiated.
Indicated for patients requiring general electrocardiograph (ECG) monitoring as ordered by a physician.
Regulatory Classification
Identification
An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.
Predicate Devices
Medi-Trace® 500 Electrodes
Related Devices
K971444 — 3M RED DOT 2234 & 2266 RADIOLUCENT MONITORING ELECTRODES · 3M Medical Products Group · Jun 11, 1997
K982521 — SKINTACT ECG ELECTRODES, S&W ECG ELECTRODES · Leonhard Lang Co. · Oct 26, 1999
{0}
10/25/1996 14:58 7168477531 GRAPHIC CONTROLS PAGE 04
Graphic Controls Corporation
189 Van Rensselaer Street
PO Box 1274
Buffalo, New York 14240-1274 USA
716 853-7500
Customer Relations: 900 609 1009
Fax: 800 347-2421
K960968
OCT 28 1996

## 510(k) SUMMARY; K960968
## Graphic Controls
Date: October 25, 1996
Manufacturer: Graphic Controls Corporation
189 Van Rensselaer Street
PO Box 1271
Buffalo, NY 14240
Registration Number 1317188
Manufacturing Location: Graphic Controls, Canada, Ltd.
215 Herbert Street
Gananoque, Ontario
Canada, K7G 2Y7
Telephone: (716) 853-7500
Contact Person: Kathleen H. Selover
Regulatory Affairs Specialist
(716) 853-7500, Extension 7630
Fax Number: (716) 847-7531
Device Trade Name: Medi-Trace® 200 and Medi-Trace® 200-30 ECG Electrodes
Common Name: Disposable ECG Monitoring Electrode
Classification Name: Electrocardiograph Electrode
Regulatory Reference: 74 DRX
Predicate Device: Medi-Trace® 500 Electrodes
{1}
10/25/1996 14:58 7168477531 GRAPHIC CONTROLS PAGE 05
## Description:
A conductive wet gel electrode consisting of a conductive adhesive gel, a silver/silver chloride plated ABS eyelet, nickel plated brass stud, an adhesive foam substrate, vinyl label and polystyrene release liner. Electrodes are packaged either as 100 per pouch (Medi-Trace® 200) or 30 per pouch (Medi-Trace® 200-30); pouches are heat sealed.
Unit of sale for each packaging is a case of either 600 or 1000 electrodes.
## Intended Use:
Intended for use in general electrocardiograph procedures where ECG monitoring is deemed necessary and is ordered by a physician.
## Physical/Technical Comparison:
Medi-Trace® 200 is equivalent to the predicate device. Physical and technical characteristics, including materials used in construction, size, intended use and conductive gel type of device and the predicate are comparable.
## Performance Summary:
The device and the predicate were subjected to AAMI electrical performance tests as described in EC12-1991. Test results indicate that this device meet or exceeds the performance specifications as established in EC12-1991 for Disposable ECG Electrodes. In addition, the device was found to meet EC12-1991 AAMI standard for labeling, shelf life, packaging and safety and Graphic Controls' internal specifications for trace performance.
## Biocompatibility Testing:
The device was subjected to biocompatibility testing as recommended in the May 1, 1995 FDA memorandum entitled: *Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"*. The device was found to be non-irritating, non-cytotoxic and non-sensitizing.
## Shelf Life:
Data obtained in real time shelf life studies was reviewed and found to substantiate our claimed shelf life.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.