LASERFLO BLOOD PERFUSION SINGLE USE PROBES MODELS SUP-440, SUPR-434 SUP-433
Device Facts
| Record ID | K951832 |
|---|---|
| Device Name | LASERFLO BLOOD PERFUSION SINGLE USE PROBES MODELS SUP-440, SUPR-434 SUP-433 |
| Applicant | Vasamedics, LLC |
| Product Code | DPT · Cardiovascular |
| Decision Date | Mar 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2120 |
| Device Class | Class 2 |
Indications for Use
The Single Use Probes are intended for extravascular monitoring of microcirculation blood flow in buried tissues. An example of this application is the monitoring of buried muscle or esophagus following free muscle transfer or esophageal reconstruction. Another example is monitoring soft tissue microcirculation following reconstructive surgery, such as in oral and facial reconstruction. Yet another example is monitoring cerebral blood flow during and following neurosurgery for head trauma.
Device Story
Fiber optic probes (SUP-440, SUPR-434, SUP-433) designed for use with Vasmedics BPM2 Blood Perfusion Monitor; input: laser light reflected from microcirculatory tissue; output: blood perfusion measurements; used in clinical settings (OR, post-op) by physicians; probes interface with BPM2 via fiber optic cable; provides real-time microvascular flow data to assist clinical assessment of tissue viability; benefits patient by enabling early detection of compromised blood flow in buried tissue grafts or post-neurosurgical sites.
Clinical Evidence
Bench testing only; no clinical data provided. Equivalence established through design comparison and performance characteristics consistent with predicate devices.
Technological Characteristics
Fiber optic probes; stainless steel tubing; soft flexible encapsulation materials; right-angle light emission; connects to BPM2 monitor via fiber optic interface; single-use design.
Indications for Use
Indicated for patients requiring extravascular monitoring of microcirculation blood flow in buried tissues, including post-operative monitoring of free muscle transfers, esophageal reconstruction, oral/facial reconstructive surgery, and cerebral blood flow monitoring during/following neurosurgery for head trauma.
Regulatory Classification
Identification
An extravascular blood flow probe is an extravascular ultrasonic or electromagnetic probe used in conjunction with a blood flowmeter to measure blood flow in a chamber or vessel.
Predicate Devices
- Model P-433 Needle Probe (K875179)
- Model PR-434 Implantable Prism Probe (K875179)
- Model P-440 Soflex Probe (K912446)
Reference Devices
- BPM2 Blood perfusion Monitor (K896515)
Related Devices
- K961368 — TRIMFLO PARENCHYMAL BLOOD FLOW MONITORING PROBE KIT · Vasamedics, LLC · Oct 11, 1996
- K140740 — TRANSONIC TISSUE PERFUSION MONITOR · Transonic Systems, Inc. · Jul 23, 2015
- K022649 — COOK-SWARTZ DOPPLER FLOW PROBE · Cook Vascular, Inc. · Sep 6, 2002
- K182494 — FlowMet-R · Laser Associated Sciences, Inc. · Feb 27, 2019
- K152749 — moorLDI2-IR Infrared laser Doppler imager, moorLDI2-HIR High Resolution Infrared laser Doppler imager, moorLDI2-VR Visible Red laser Doppler imager · Moor Instruments, Ltd. · May 12, 2016