← Product Code [MXC](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MXC) · K081365

# SJM CONFIRM MODEL DM2100 IMPLANTABLE CARDIAC MONITOR AND MODEL DM2100A PATIENT ACTIVATOR (K081365)

_St Jude Medical · MXC · Aug 15, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MXC/K081365

## Device Facts

- **Applicant:** St Jude Medical
- **Product Code:** [MXC](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MXC.md)
- **Decision Date:** Aug 15, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The SJM Confirm™ ICM is an implantable patient-activated and automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - patients who experience transient symptoms that may suggest a cardiac arrhythmia

## Device Story

SJM Confirm ICM is a minimally invasive, implantable diagnostic monitor; records subcutaneous ECG via subcutaneous electrodes; features looping memory for EGM storage. System includes implantable monitor (DM2100), external patient activator (DM2100A), and physician programmer (Merlin PCS Model 3650). Device automatically detects arrhythmias (high/low rate, asystole) or is manually triggered by patient. Physician uses programmer to communicate with monitor, retrieve stored EGM data, and adjust detection parameters. Output assists clinicians in diagnosing cardiac arrhythmias; enables long-term monitoring of transient symptoms; facilitates clinical decision-making regarding patient management.

## Clinical Evidence

Bench testing only. Verification and validation activities included mechanical, electrical, and EMC testing. Firmware, QRS detection accuracy, and AF detection algorithm performance were verified. System validation and GLP studies were performed to confirm design conformance to user needs.

## Technological Characteristics

Titanium (Grade 1) housing; Hysol High Purity Epoxy header; Parylene-C coating. Sensing via subcutaneous electrodes. 128 Hz sampling rate. Dimensions: 18 x 56 x 8 mm; Volume: 6.5 cc; Weight: 12 g. Sterilized via 100% Ethylene Oxide (EtO). Connectivity via external programmer.

## Regulatory Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

## Predicate Devices

- Medtronic Reveal Plus ILR ([K994331](/device/K994331.md))

## Reference Devices

- Epic + ICD ([P910023](/device/P910023.md)/S65)
- Atlas + ICD ([P910023](/device/P910023.md)/S69)
- Affinity pacemaker ([P880086](/device/P880086.md)/S61)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081365
Page 1 of 4

## PREMARKET NOTIFICATION [510(K)] SUMMARY

AUG 1 5 2008

| Date Prepared:                                                            | August 11, 2008                                |
|---------------------------------------------------------------------------|------------------------------------------------|
| Submitter:                                                                | St. Jude Medical, CRMD                         |
| Address:                                                                  | 701 E. Evelyn Avenue<br>Sunnyvale, CA 94086    |
| Phone:                                                                    | 408 522 6494                                   |
| Fax:                                                                      | 408 522 6440                                   |
| Contact Person:                                                           | Elisabeth E. Neely                             |
| Trade Name/Proprietary Name:                                              | SJM Confirm Implantable Cardiac Monitor System |
| Common Name:                                                              | Implantable Cardiac Monitor                    |
|                                                                           | Model Numbers: DM2100, DM2100A                 |
| Classification:                                                           | Class II, 21 CFR 870.2800, 21 CFR 870.2920     |
| Legally marketed device<br>to which your firm is<br>claiming equivalence: | Medtronic Reveal Plus ILR K994331              |

## Device Description:

The SJM Confirm Implantable Cardiac Monitor is a minimally invasive, implantable diagnostic monitoring device with subcutaneous electrodes that are used for sensing and a looping memory for storage of electrograms (EGM). The device is comprised of three main components: the implantable cardiac monitor (Model DM2100) and the external patient activator (Model DM2100A). The third component is the programmer the physician uses to communicate to the cardiac monitor and associated programmer software. The programmer is the legally marketed SJM Merlin PCS programmer Model 3650 (with Software Model 3330 Version 6.8.1 (or higher)).

## The indication for use is as follows:

The SJM Confirm™ ICM is an implantable patient-activated and automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:

- . patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- . patients who experience transient symptoms that may suggest a cardiac arrhythmia

## Technological Characteristics of the Device Compared to the Predicate Device:

The SJM Confirm ICM uses similar technology; has similar intended use, functions, materials and method of operation of the following predicate device described in Table 1 below.

Image /page/0/Picture/12 description: The image shows the logo for St. Jude Medical. The logo consists of a 3x3 grid of black squares on the left, with one square rotated. To the right of the grid is the text "ST. JUDE MEDICAL" in a serif font. Below the text is the tagline "MORE CONTROL LESS RISK" in a smaller, sans-serif font.

{1}------------------------------------------------

K081365
Page 2 of 4

| Feature            | SJM Confirm | Medtronic Reveal Plus            |
|--------------------|-------------|----------------------------------|
| Subcutaneous ECG   | Yes         | Yes                              |
| Recording          |             |                                  |
| Pre and Post Event | Yes         | Yes                              |
| Storage            |             |                                  |
| EGM Storage        | 33 minutes  | 21 minutes (without compression) |
| Sampling Rate      | 128 Hz      | 100 Hz                           |
| Dimensions (mm)    | 18 x 56 x 8 | 19 x 61 x 8                      |
| Width              |             |                                  |
| Volume             | 6.5 cc      | 8 cc                             |
| Weight             | 12 g        | 17 g                             |
| Electrode Spacing  | 39 mm       | 38.5 mm                          |
| Shelf life         | 12 months   | 14 months                        |
| Auto Activation    | Yes         | Yes                              |
| Triggers           |             |                                  |
| Manual (Patient)   | Yes         | Yes                              |
| Activation Trigger |             |                                  |
| High Rate Trigger  | Yes         | Yes                              |
| Programmable High  | Yes         | Yes                              |
| Rate Count         |             |                                  |
| Low Rate Trigger   | Yes         | Yes                              |
| Asystole Trigger   | Yes         | Yes                              |

Table 1: SJM Confirm ICM Compared to Predicate Device

Image /page/1/Picture/3 description: The image contains the logo for St. Jude Medical. The logo consists of a grid of nine squares on the left, with one square rotated. To the right of the grid is the text "ST. Jude Medical" in a serif font. Below the text is the tagline "More control. Less risk."

{2}------------------------------------------------

K081365
Page 3 of 4

#### Summary of Studies:

Verification and validation activities necessary to ensure that the SJM Confirm ICM product and system requirements were fulfilled, and to ensure that the product design conforms to the user needs and intended uses were identified and successfully performed. Product verification was successfully performed and is documented in the following reports:

| Appendix 7.1 | QTR2288: Confirm ICM Mechanical Test Report                 |
|--------------|-------------------------------------------------------------|
| Appendix 7.2 | QTR2280: Confirm ICM Electrical Test Report                 |
| Appendix 7.3 | QTR2286: Confirm ICM EMC Test Report                        |
| Appendix 7.4 | QTR2267: Patient Activator Test Report                      |
| Appendix 7.5 | Firmware SVRs                                               |
| Appendix 7.6 | QRS Detection Accuracy and AF Detection<br>Algorithm Report |
| Appendix 7.7 | System Validation Report                                    |
| Appendix 7.8 | GLP Study Report                                            |

#### Biocompatibility:

There is no change to the blood/tissue contact materials of the SJM Confirm ICM device as compared to legally marketed St. Jude Medical pacemakers and ICDs (e.g., Epic + ICD P910023/S65 and Affinity pacemaker P880086/S61). Therefore, no additional biocompatibility testing was considered necessary per ISO 10993-1. See Table 2 below.

Image /page/2/Picture/6 description: The image shows the logo for St. Jude Medical. The logo consists of a square grid of smaller squares on the left, followed by the text "ST. JUDE MEDICAL" in a bold, sans-serif font. Below the text, there is a smaller tagline that reads "MORE CONTROL. LESS RISK."

{3}------------------------------------------------

KC81365
Page 4 of 14

| Component           | Material           | Trade Name of Material                   | Predicate SJM<br>Device    | PMA Number<br>(Approval Date)                           |
|---------------------|--------------------|------------------------------------------|----------------------------|---------------------------------------------------------|
| Can                 | Titanium (grade 1) | n/a                                      | Epic + ICD, Atlas<br>+ ICD | P910023/S65<br>(4/23/03) ,<br>P910023/S69<br>(10/17/03) |
| Header<br>Electrode | Titanium (grade 1) | n/a                                      | Epic + ICD, Atlas<br>+ ICD | P910023/S65<br>(4/23/03) ,<br>P910023/S69<br>(10/17/03) |
| Header              | Epoxy              | Hysol High Purity Epoxy                  | Epic + ICD, Atlas<br>+ ICD | P910023/S65<br>(4/23/03) ,<br>P910023/S69<br>(10/17/03) |
| Parylene<br>Coat    | Parylene           | Parylene-C Conforming<br>Coating polymer | Affinity SR                | P880086/S61<br>(6/17/99)                                |

|  |  |  | Table 2: SJM Confirm ICM Patient Contact Materials |
|--|--|--|----------------------------------------------------|
|--|--|--|----------------------------------------------------|

## Sterilization Validation:

The SJM Confirm ICM is sterilized using the same validated 100% Ethylene Oxide (EtO) sterilization process as legally marketed St. Jude Medical pacemakers. The sterility assurance level (SAL) is 10 % A sterilization assessment report is provided in Appendix 6. The patient activator, Model DM2100A, is not sterile.

#### Conclusion:

St. Jude Medical considers the SJM Confirm ICM system to be substantially equivalent to the legally marketed predicate and referenced device.

The results of the tests and compliance with applicable standards (reference Section 6) provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its indications for use.

Image /page/3/Picture/8 description: The image shows the logo for St. Jude Medical. The logo consists of a 3x3 grid of black squares on the left, followed by the text "St. Jude Medical" in a serif font. Below the company name is the tagline "More Control. Less Risk."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 5 2008

St. Jude Medical c/o Ms. Elizabeth Neely Regulatory Affairs Cardiac Rhythm Management Division 701 East Evelyn Avenue Sunnyvale, CA 94086

Rc: K081365

Trade/Device Name: Confirm Model DM2100 Implantable Cardiac Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC, DSH, DXH Dated: July 9, 2008 Received: July 10, 2008

Dear Ms. Neely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);

{5}------------------------------------------------

Page 2 - Ms. Elizabeth Ncely

labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

ell.mall

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 3 - Ms. Elizabeth Neely

# Indications for Use

510(k) Number (if known):

Device Name:____Confirm Model DM2100 Implantable Cardiac Monitor

Indications For Use:

The SJM Confirm ICM is an implantable patient-activated and automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:

- patients with clinical syndromes or situations at increased risk of cardiac arrhythmias .
- patients who experience transient symptoms that may suggest a cardiac arrhythmia .

Prescription Use
(Part 21 CFR 801 Subpart D)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                |         |
|------------------------------------|---------|
| Division of Cardiovascular Devices |         |
| 510(k) Number                      | K081365 |
| Page 1 of                          | 1       |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MXC/K081365](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MXC/K081365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
