The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
Device Story
Biotres is a 3-channel ambulatory ECG recorder for home use; captures continuous ECG data via chest-worn electrodes; stores data on internal memory for up to 30 days. Device features power button and patient event marking button; LED status indicator provides feedback on device state. Data transmission occurs via Bluetooth to Biotres Gateway App (iOS/Android) and then to Biotres Configured Secure Server. Clinicians access recordings via Bioflux Software II for manual annotation and review. Device does not perform automated/semi-automated analysis; diagnosis relies on physician interpretation. Benefits include long-term cardiac monitoring for transient symptom correlation.
Clinical Evidence
Bench testing only. Device passed IEC 60601-2-47:2012 (particular requirements for basic safety and essential performance of ambulatory electrocardiographic systems). No clinical data presented.
Technological Characteristics
3-channel ambulatory ECG recorder; rechargeable Li-Ion battery; Bluetooth connectivity; 3-lead electrode interface. Complies with IEC 60601-1 (8.5.5.2) and IEC 60601-2-47:2012. System includes recorder, AC wall charger, mobile gateway app, and secure server. Software is non-analytical/manual review only.
Indications for Use
Indicated for adult patients 18+ years, asymptomatic or experiencing transient symptoms (palpitations, dyspnea, dizziness, lightheadedness, pre-syncope, syncope, fatigue, chest pain, anxiety) requiring continuous cardiac recording up to 30 days. Contraindicated for patients requiring inpatient monitoring, those requiring hospitalization, infants <10kg, patients with implanted pacemakers, or defibrillator use.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 28, 2022
Biotricity Spencer Ladow VP Engineering 203 Redwood Shores Pkwv Suite 600 Redwood City, California 94065
Re: K222017
Trade/Device Name: Biotres Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: MWJ Dated: July 6, 2022 Received: July 8, 2022
Dear Spencer Ladow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name Biotres
## K222107
Indications for Use (Describe)
The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Biotricity. The logo consists of a teal-colored geometric flower-like design on the left, followed by the word "biotricity" in a teal sans-serif font. The flower design has eight petals, each made up of smaller teal shapes. The word "biotricity" is written in lowercase letters and is positioned to the right of the flower design.
# 510(k) SUMMARY
Date Prepared: July 6, 2022
# SUBMITTER
l.
Biotricity 203 Redwood Shores Pkwy, Suite 600 Redwood City, CA 94065 Contact Person: Spencer LaDow sladow(@biotricity.com Phone: 585-414-7407
#### DEVICE II.
Name of Device: Classification Regulation: Common Name: Device Panel: Regulatory Class: Product Code:
Biotres 21 CFR§870.2800 Medical Magnetic Tape Recorder Cardiovascular Class II MWJ Electrocardiograph, Ambulatory (without Analysis)
#### PREDICATE DEVICE III.
The Biotres (Biotricity) was cleared under K211709.
Name of Device: Classification Regulation: Common Name: Device Panel: Regulatory Class: Product Code:
Biotres 21 CFR§870.2800 Medical Magnetic Tape Recorder Cardiovascular Class II MWJ Electrocardiograph, Ambulatory (without Analysis)
Biotricity K22207
Page 1 of 7
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Biotres system consists of 1) the Biotres Recorder, 2) a dedicated Biotres AC wall charger, 3) Biotres Gateway application for use on mobile devices using Android or iOS, and 4) Biotres Configured Secure Server.
#### Biotres Recorder
The Biotres Recorder is an ECG recording only device and does not include ECG analysis, alarms and is not a real time ECG monitor. The Biotres recorder is a small, lightweight, ambulatory, reusable 3 channel ECG recorder that is placed on the center of the chest (Reference Below).
Biotres Recorder – Front View
Image /page/4/Figure/5 description: The image shows a diagram of a device with several labeled components. The components include a 'Status Indicator Light' at the top center, a 'Marking Button' to the right, and a 'Power Button' on the left side. Additionally, there is a 'Charging Port' on the right and a 'Snap for ECG electrode' at the bottom.
FRONT VIEW
Biotres Recorder – Back View
BACK VIEW
Image /page/4/Figure/9 description: The image shows a top-down view of a round, medical device with three electrode connectors extending from its sides. The device has a label with the Biotricity logo, model number "Biotree 1.0", serial number "1230000089", FCC ID "2AKGEBIOTRES1", and FDA cleared status. The label also includes regulatory symbols such as "FC" and "IP22". An arrow points to one of the electrode connectors with the text "Snap connection for electrodes".
Biotricity K222107
Page 2 of 7
{5}------------------------------------------------
The Biotres Recorder has two buttons: 1) power button and 2) patient event marking button. The pash the patient event marking button when the patient is feeling symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety. There is an LED "status indicator light" in the middle of the patient event marking button that provides recorder status indications when powering on, electrodes are disconnected, low battery, charged, encountered internal error, acknowledgment of the patient event marking button, and study not engaged/start/running/paused/completed state.
The Biotres Recorder attaches to the patient's chest through three standard commercially available ECG electrodes (not include with the system) supplied by a clinic or a monitoring center. Biotres includes a Bluetooth modem to communicate with the Biotres Configured Secure Server through the Biotres Gateway Application. The Biotres Recorder integrated memory can store up to 30-days of ECG data. This data can be uploaded by the clinician to the Biotres Configured Secure Server.
## Biotres AC Wall Charger
The dedicated Biotres AC Wall charger is used to charge the Biotres Recorder daily when disconnected from the patient.
## Biotres Gateway Application
The ECG recorded data is transferred via the Biotres Gateway App (Downloaded from the app store for iOS or Android respectively) via a Bluetooth connection and subsequently to the Biotres Configured Secure Server. Clinicians can use the Biotres Gateway Application using their privileged login credentials to connect Biotres Recorders to the Biotres Configured Secure Server so that patient studies may be started and completed. The Biotres Gateway Application can be used by the patient once registered (nonprivileged login credentials) and the patient can use the application to monitor the status indicators of the Biotres Recorder and to enter symptoms if the Event Marking Button is pressed on the recorder. The ECG recorded data from the Biotres Recorder can be transferred, by the clinician, from the Biotres Configured Secure Server when the recorder is connected to the Biotres Gateway App at the end of a patient study.
## Biotres Secure Server
The Biotres Configured Secure Server is an SFTP server that is configured to work with the Biotres System. The server enables the Clinician, through secured login, to command the Biotres Recorder (e.g. initiate ECG data collection) and to access uploaded MIT16 ECG data recordings when the ECG data collection is complete. Biotres does not provide the patient or clinician with diagnostic or interpretive statements based on the ECG data. Diagnosis and interpretation of ECG data is dependent upon the medical advice and guidance provided by a physician or trained healthcare professional. The Biotres is for home use and not for use in the hospital. Clinical judgment and experience are used to check and interpret the data.
## ECG Viewer Software
Biotres ECG recordings can be viewed through the following software tool by qualified medical personnel. Bioflux Software II (K201040) - The ECG viewer software displays ECG records in a browser on a computer and provides tools for trained clinicians to annotate (store, retrieve and report) those ECG recordings. It is utilized by manually opening ECG files or through an API that allows for third parties to integrate their devices and directly push ECG files into the viewer.
{6}------------------------------------------------
#### INDICATIONS FOR USE V.
The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
#### contraindications for USE VI.
The Biotres device is not intended for use under the following conditions:
- Patients with potentially life-threatening arrhythmias who require ● inpatient monitoring.
- . Patients who the attending physician thinks should be hospitalized.
- Infants weighing less than 10 kg. (22 lbs.).
- . Patients with implanted pacemakers.
- . Defibrillator Use.
{7}------------------------------------------------
## VII. SUMMARY OF TECHNOLIGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Category | Identical/<br>Different | Biotres | Biotres (predicate) |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K222107 | K211709 |
| Classification<br>Name | Identical | Medical<br>Magnetic Tape Recorder | Medical<br>Magnetic Tape Recorder |
| Product Code | Identical | MWJ | MWJ |
| Intended Use | Identical | The Biotres is indicated for use on<br>adult patients 18 years or older<br>who may be asymptomatic or<br>who suffer from transient<br>symptoms such as palpitations,<br>shortness of breath, dizziness,<br>light headedness, pre-syncope,<br>syncope, fatigue, chest pain<br>and/or anxiety and may require<br>cardiac recording on a continuous<br>basis for up to 30 days. The signal<br>acquired by the Biotres is not<br>intended and should not be used<br>for automated or semi-automated<br>analysis. | The Biotres is indicated for use on<br>adult patients 18 years or older<br>who may be asymptomatic or<br>who suffer from transient<br>symptoms such as palpitations,<br>shortness of breath, dizziness,<br>light headedness, pre-syncope,<br>syncope, fatigue, chest pain<br>and/or anxiety and may require<br>cardiac recording on a continuous<br>basis for up to 30 days. The signal<br>acquired by the Biotres is not<br>intended and should not be used<br>for automated or semi-<br>automated analysis. |
| Defibrillator<br>Protection | Identical | Not Defibrillator Proof, 8.5.5.2 of<br>IEC 60601-1 passed | Not Defibrillator Proof, 8.5.5.2 of<br>IEC 60601-1 passed |
| Wear Time | Identical | Up to 30 Days | Up to 30 Days |
| Recording Format | Identical | Continuous | Continuous |
| Delivered device<br>includes | Identical | -3 lead ECG monitor<br>-internal rechargeable battery<br>-Wall Battery charger | -3 lead ECG monitor<br>-internal rechargeable battery<br>-Wall Battery charger |
| Monitor functional<br>blocks | Identical | Analog ECG front end,<br>MCU,<br>Flash data storage,<br>BLE modem for<br>data transmission,<br>LED indicator, and | Analog ECG front end,<br>MCU,<br>Flash data storage,<br>BLE modem for<br>data transmission,<br>LED indicator, and |
| Category | Identical/<br>Different | Biotres | Biotres (predicate) |
| The App: | Identical | Record button<br>Mobile App, iOS, Android | Record button<br>Mobile App, iOS, Android |
| The server: | Identical | Facilitate data communication<br>with the Biotres device,<br>provide data storage, and<br>present the data for evaluation by<br>a medical professional | Facilitate data communication<br>with the Biotres device,<br>provide data storage, and<br>present the data for evaluation by<br>a medical professional |
| Device form factor | Identical | Small, lightweight body worn<br>ambulatory cardiac monitors. | Small, lightweight body worn<br>ambulatory cardiac monitors. |
| Wireless technology<br>used to transmit data<br>to server | Identical | Yes | Yes |
| Device is battery<br>powered by a<br>rechargeable Li-Ion<br>battery | Identical | Yes | Yes |
| Using an app, can<br>adjust device<br>programming<br>parameters such as<br>pre-post recording<br>times and auto-<br>triggering<br>configuration. | Identical | Yes, app is android or IOS based | Yes, app is android or IOS based |
| Devices have Record<br>button for manual<br>event recordings and<br>a user led to indicate<br>device status and<br>mode of operation. | Identical | Yes | Yes |
| Device has at least 2<br>ECG channels and 3-<br>lead electrodes | Identical | Yes | Yes |
| Functional,<br>Environmental and<br>Electrical<br>characteristics | Identical | Yes | Yes |
| USB connection | Identical | Yes, used for battery charging<br>cannot be connected during ECG<br>recording | Yes, used for battery charging<br>cannot be connected during ECG<br>recording |
| ECG Viewer Software | Different | Bioflux Software II (K201040) | None |
Page 5 of 7
{8}------------------------------------------------
Page 6 of 7
{9}------------------------------------------------
#### PERFORMANCE TESTING VIII.
The following performance and safety tests have been passed successfully:
- IEC 60601-2-47:2012 Medical electrical equipment Part 2-47: . Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
#### IX. CONCLUSION
The analysis of the differences between Biotres and the predicate device does not raise different questions of safety and effectiveness. Based on device performance test results, Biotricity determined that the Biotres performed within its design specifications and is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.