K070014 · North East Monitoring, Inc. · MWJ · May 25, 2007 · Cardiovascular
Device Facts
Record ID
K070014
Device Name
AUTO DETECT FOR TELAHEART DR200/E-A
Applicant
North East Monitoring, Inc.
Product Code
MWJ · Cardiovascular
Decision Date
May 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Indications for Use
The Telaheart Digital Recorder can be used in Holter mode and Event Recorder mode. Holter Mode: Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation. Event Recorder: The Telaheart event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted via telephone for evaluation. The device includes the Auto Detect feature which can be used as a looping event recorder to capture events automatically or when activated by the patient. When one or more events are captured the patient may transmit his/her recordings transtelephonically.
Device Story
Ambulatory ECG recorder (Holter/Event monitor) captures cardiac signals via patient-applied electrodes; stores data in internal memory. Features 'Auto Detect' loop memory function for automatic or patient-activated event capture. Device provides audible/visual alerts for data transmission. Data transmitted transtelephonically to physician for analysis via compatible Holter/Event analysis software. Used in clinical or home settings; operated by patient. Output assists physicians in diagnosing arrhythmias and evaluating pacemaker function or drug efficacy. Not intended for real-time telemetry of life-threatening arrhythmias.
Clinical Evidence
Bench testing only. Device passed electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2-47) testing. No clinical data provided.
Technological Characteristics
Ambulatory ECG recorder; 8.6cm x 6.0cm x 2cm; 99.3g with battery. Powered by 1.5V AA alkaline, NiMH, or Lithium battery. Recording bandwidth 0.05-70 Hz; sampling rates 180/360/720 samples/sec. Pacemaker sensitivity 2mV. Connectivity via telephone transmission. Standards: IEC 60601-1, IEC 60601-2-47 (equivalent to AAMI/ANSI EC38:1998), EN 55022 Class B.
Indications for Use
Indicated for patients requiring Holter monitoring for arrhythmia detection, pharmacological treatment efficacy, or pacemaker evaluation, and for patients requiring event recording for diagnostic evaluation of transient symptoms including dizziness, palpitations, syncope, and chest pain.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with two mountains in the background. In front of the mountains, there is a line graph that resembles an EKG. The mountains and the line graph are black.
# NorthEast Monitor
2 Clock Tower Place, Suite 555 Maynard, MA 01754 U.S.A. www.nemon.com
phone: 978-461-3992 fax: 978-461-5991 toll free: 866-346-5837
advancing Holter technology
# 510(k) Sur
Date Prepared [21 CFR 807.92(a)(1)] December 29, 2006
#### Submitter's Information [21 CFR 807.92(a)(1)]
## MAY 2 5 2007
Contact / Regulatory Consultant: Joseph M. Azary C/o NorthEast Monitoring Inc. 543 Long Hill Avenue Shelton, CT. 06484
Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor NorthEast Monitoring Inc.
Sponsor / Manufacturer NorthEast Monitoring Inc. Two Clock Tower Place, Suite 555 Maynard, MA 01754
FDA establishment registration# 1224919.
#### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade names are:
- Auto Detect .
- Telaheart DR200/HE-a �
- DR200/HE-a .
Device Common, Usual, or Classification Names:
Ambulatory ECG Recorder, Ambulatory Electrocardiograph (without analysis), Telephone electrocardiograph transmitter and receiver
{1}------------------------------------------------
510(k) Summary
Pg. 2 of 5
Classification:
Class II, 21 CFR 870.2800, MWJ Class II, 21 CFR 870.2920, DXH
#### Predicate Device [21 CFR 807.92(a)(3)]
- � NorthEast Monitoring Telaheart - K061293
- Heartrak Smart AT K033451 .
#### Auto Detect feature compared to Telaheart
The subject device (DR200/HE-a and Telaheart DR200/E-a) is identical to the predicate device (Telaheart) with one difference. The devices have the same indications for use, identical physical and electrical characteristics, and identical materials as the predicate device. The main difference between the subject device and predicate device are as follows:
- The subject device includes the Auto Detect feature. The Auto Detect feature was included in the . software code of the Telaheart, but was not enabled with the Telaheart units distributed.
#### DR200/HE compared to Heartrak Smart AT
The subject device has the same indications for use as the predicate devices have automatic detection features, are loop memory recorders, and can transmit data via telephone. The differences are as follows:
- . The Heartrak Smart AT uses 2 AA batteries
- . The devices have minor differences dimensionally
- The Heartrak Smart AT is only an event recorder and cannot be used as a Holter Monitor. .
#### Description of the Device [21 CFR 807.92(a)(4)]
NorthEast Monitoring will offer versions with the Auto Detect feature.
- Telaheart DR200/E-a Event Recorder with Auto Detect feature (using a black background). o
- DR200/HE-a Holter and Event Recorder with Auto Detect feature (using white background). o
The Auto Detect feature can be used as a looping event recorder to capture events automatically or when activated by the patient. When one or more events are captured the patient may transmit his/her recordings transtelephonically. The device stores the events in memory. There is a display counter for the events and an audible beep to remind the user to transmit the data to the physician.
The data obtained during monitoring is not analyzed at the time of recording. After the recording is complete, the data must later be downloaded to a compatible NorthEast Monitoring Holter or Event
{2}------------------------------------------------
analysis system. The Holter Analysis Software was cleared by FDA under K930564. The TelePro software was included in the Telaheart 510(k) cleared by FDA under K061293.
#### The device is not intended to replace real time telemetry monitoring for patient suspected of having life threatening arrhythmias.
The device has the identical hardware as the Telaheart device (cleared by FDA) and is powered by one 1.5 volt AA alkaline battery (MN1500 or the equivalent), one AA rechargeable NiMH (nickel metal hydride) battery, or one AA Eveready Lithium L91 battery. Batteries should not be re-used for a second patient. The batteries are not included; users are instructed to purchase 2 AA batteries.
The device is compatible with standard silver / silver chloride ECG electrodes. Electrodes are not provided with the subject device. The user is instructed to purchase standard silver / silver chloride ECG electrodes.
| Characteristic | Specification |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions | 8.6cm (length) x 6.0cm (width) x 2cm (depth) |
| Weight | 70.9 grams (2.5 oz) without battery<br>99.3 grams (3.5 oz) with battery |
| Recording Bandwidth | 0.05 to 70 hertz in 180 samples/sec mode |
| Prefilter Sampling Rate | 720 samples/sec |
| Data Stored | In 180 samples/sec mode, data stored at 180 samples/sec (4<br>sample average), in 360 samples/sec mode, data stored at 360<br>samples/sec (2 sample average), in 720 samples/sec mode,<br>data stored at 720 samples/sec. |
| Pacemaker Sensitivity | 2 millivolts |
| Pacemaker Pulse Duration | 150 to 2500 microseconds |
#### Physical and Electrical Specifications:
Electrical Safety and Electromagnetic Compatibility
The hardware and electrical circuitry has not changed. The subject device conforms to the following standards:
- . IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety
- IEC 60601-1-2-47 Medical Electrical Equipment Part 2 47: Particular Requirements for the � Safety including Essential Performance of Ambulatory Electrocardiographic Systems.
#### *IEC 60601-2-47 is the equivalent to AAMI / ANSI EC38:1998.
- EN 55022 Class B ITE emissions ●
{3}------------------------------------------------
#### Patient Contacting Materials
The subject device does not make direct contact with the patient. The subject device is used with commercially available electrodes purchased by the customer.
#### Reprocessing, Cleaning, Disinfection, and Sterilization
The subject device does not make direct contact with the patient and does not require sterilization. The Operator's Manual includes a section on maintenance and care, which includes cleaning and infection control.
#### Software
The software in the subject device is identical to the software included in the Telaheart device K061293. The only difference is the Auto Detect feature was not enabled in the Telaheart device when it was sold to the public. The Auto Detect feature was not utilized in the Telaheart, but existed in the software code.
#### Intended Use [21 CFR 807.92(a)(5)]
The Telaheart Digital Recorder can be used in Holter mode and Event Recorder mode.
#### Holter Mode
Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation.
#### Event Recorder
The Telaheart event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted via telephone for evaluation.
The device includes the Auto Detect feature which can be used as a looping event recorder to capture events automatically or when activated by the patient. When one or more events are captured the patient may transmit his/her recordings transtelephonically.
{4}------------------------------------------------
510(k) Summary
pg. 5 of 5
#### Technological Characteristics [21 CFR 807.92(a)(6)]
NorthEast Monitoring, Inc. believes that the subject device is substantially equivalent to the predicate device.
#### Performance Data [21 CFR 807.92(b)(1)]
The subject device has been subjected to and passed electrical safety and EMC testing requirements.
#### Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minor and conclude that the subject device is as safe and effective as the predicate devices
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NorthEast Monitoring, Inc. C/O Joseph M. Azary Azary Technologies, LLC 543 Long Hill Avenue Shelton, CT 06484
MAY 2 5 2007
Re: K070014
Trade/Device Name: Auto Detect for Telaheart Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory ECG Recorder, Ambulatory Electrocardiograph (without analysis) Regulatory Class: Class II Product Code: MWJ, DXH Dated: May 17, 2007 Received: May 21, 2007
Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Mr. Azary
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B/Zimmerman for
Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): KO70014
Device Name: Auto Detect for Telaheart
Indications For Use:
The Telaheart Digital Recorder can be used in Holter mode and Event Recorder mode.
#### Holter Mode
Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation.
#### Event Recorder
The Telaheart event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once adale is recorded, the data is transmitted via telephone for evaluation.
The device includes the Auto Detect feature which can be used as a looping event recorder to capture events automatically or when activated by the patient. When one or more eyents are captured the natient may transmit his/her recordings transtelephonically.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------|
| (Division Sign-Off) | <div style="text-align:left;"> <span style="font-size: 16px;">Blummuma</span> </div> Division of Cardiovascular Devices |
| | Page 1 of 1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.