← Product Code [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI) · K183116

# Unity Network ID (K183116)

_GE Healthcare · MWI · Dec 6, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K183116

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI.md)
- **Decision Date:** Dec 6, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to.

## Device Story

Unity Network ID acts as a data gateway/bridge; acquires digital patient data from independent bedside devices via eight RS-232 serial ports; converts device-specific output into Unity Network protocols; transmits data over monitoring network to central stations, clinical information systems, or GE patient monitors. Used in clinical settings; operated by healthcare staff. Device does not perform monitoring or control connected equipment; serves solely as a communication interface to facilitate data integration into hospital information systems. Benefits include centralized data access and improved clinical information management.

## Clinical Evidence

No clinical data. Bench testing only. System verification and validation performed, including risk analysis, requirements reviews, design reviews, module verification, integration testing, performance testing, safety testing, and simulated use testing.

## Technological Characteristics

Data gateway; converts RS-232 serial protocol from bedside devices to Unity Network protocol. Hardware unchanged from predicate; includes eight serial ports. Connectivity: Networked. Software-based protocol conversion.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- Unity Network ID V8 ([K170199](/device/K170199.md))

## Reference Devices

- Edwards HemoSphere ([K163381](/device/K163381.md))
- GE Aisys CS2 ([K170872](/device/K170872.md))
- GE Giraffe Incubator ([K101788](/device/K101788.md))
- GE Giraffe Omnibed Incubator ([K101788](/device/K101788.md))
- GE Giraffe Warmer ([K070377](/device/K070377.md))
- GE Panda Warmer ([K070377](/device/K070377.md))
- Maquet Flow-i ([K160665](/device/K160665.md))
- Maquet Servo-air
- Maquet Servo-n ([K151814](/device/K151814.md))
- Maquet Servo-s ([K123149](/device/K123149.md))
- Maquet Servo-U ([K151814](/device/K151814.md))
- Nellcor PM1000N ([K141518](/device/K141518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 6, 2018

GE Healthcare Brandon O'Shea Senior Regulatory Affairs Leader 8200 West Tower Ave. Milwaukee, Wisconsin 53223

Re: K183116

Trade/Device Name: Unity Network ID Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: November 8, 2018 Received: November 9, 2018

Dear Brandon O'Shea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Stephen C. Browning -S5

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K183116

Device Name Unity Network ID

#### Indications for Use (Describe)

The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (information systems) it is connected to.

| Type of Use (Select one or both, as applicable)                                      |                                                            |
|--------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K183116

| Date:                                        | 8 November 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                   | GE Healthcare                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                              | 8200 W. Tower Ave.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                              | Milwaukee, WI 53223                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Primary Contact<br>Person:                   | Brandon O'Shea                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              | Regulatory Affairs Leader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                              | GE Medical Systems Information Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                              | Email: brandon.oshea@ge.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                              | Ph: (414) 323-3147                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Secondary Contact<br>Person:                 | Monica Morrison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                              | Senior Regulatory Affairs Director                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                              | GE Medical Systems Information Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                              | Ph: (608) 515-3077                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Trade Name:                           | Unity Network ID                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common/Usual Name:                           | Physiological Patient Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Names:                        | 21 CFR 870.2300 Monitor, Physiological, Patient<br>(without arrhythmia detection or alarms)                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Product Code:                                | MWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Predicate Device(s):                         | Unity Network ID V8 (K170199)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description:                          | The Unity Network ID system communicates patient data<br>from sources other than GE Medical Systems Information<br>Technologies, Inc. equipment to a clinical information<br>system, central station, and/or GE Medical Systems<br>Information Technologies Inc. patient monitors.                                                                                                                                                                                                                                                      |
|                                              | The Unity Network ID acquires digital data from eight serial<br>ports, converts the data to Unity Network protocols, and<br>transmits the data over the monitoring network to a Unity<br>Network device such as a patient monitor, clinical information<br>system or central station.                                                                                                                                                                                                                                                   |
| Intended Use:                                | The Unity Network ID is indicated for use in data collection<br>and clinical information management through networks with<br>independent bedside devices. The Unity Network ID is not<br>intended for monitoring purposes, nor is the Unity Network<br>ID intended to control any of the clinical devices<br>(independent bedside devices / information systems) it is<br>connected to.                                                                                                                                                 |
| Technology:                                  | The device converts the output from independent bedside<br>device's RS-232 protocol into the Unity Network protocol.                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                              | The Unity Network ID V9 and its applications were tested to,<br>and comply with, applicable voluntary standards. The Unity<br>Network ID V9 was tested to assure that the device meets its<br>design specifications. Testing included all new or modified<br>features.                                                                                                                                                                                                                                                                  |
|                                              | The following quality assurance measures were applied to<br>the development and testing of the of the system:<br>Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation)<br>Summary of Clinical Tests:<br>The subject of this premarket submission, Unity Network<br>ID V9, did not require clinical studies to support substantial<br>equivalence. |

### 510(k) Summary

In accordance with 21 CFR 807.92 the following summary information is provided:

{4}------------------------------------------------

{5}------------------------------------------------

# K183116

| Comparison: | Hardware:<br>1) No change to the Unity Network ID hardware<br>2) Update encoding of cables to detect newly supported devices (no change to physical cable)                                                                                                                                                                                                                                                                                                                                           |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|             | Software:<br>Add interface support for the following third-party devices:<br>a) Edwards HemoSphere (K163381)<br>b) GE Aisys CS2 (K170872)<br>c) GE Giraffe Incubator (K101788)<br>d) GE Giraffe Omnibed Incubator (K101788)<br>e) GE Giraffe Warmer (K070377)<br>f) GE Panda Warmer (K070377)<br>g) Maquet Flow-i (K160665)<br>h) Maquet Servo-air (Not for sale in US)<br>i) Maquet Servo-n (K151814)<br>j) Maquet Servo-s (K123149)<br>k) Maquet Servo-U (K151814)<br>l) Nellcor PM1000N (K141518) |
| Conclusion: | GE Healthcare considers the Unity Network ID V9 to be as safe, as effective, and its performance is substantially equivalent to the predicate device(s).                                                                                                                                                                                                                                                                                                                                             |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K183116](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K183116)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
