The Accutorr 7/VS-900/VS-900c Vital Signs Monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. . The monitor also provides a Modified Early Warning Score (MEWS) for clinical assessment in adult patients.
Device Story
Accutorr 7/VS-900/VS-900c is a vital signs monitor for professional healthcare facilities. Inputs: SpO2 (light absorption), NIBP (oscillometric cuff pressure), CO2 (infrared light absorption), and temperature (thermal resistance or infrared temporal artery scanning). Device processes signals to display numeric values and waveforms; supports data transmission to central monitoring systems or eGateway. Features include touchscreen interface, rechargeable battery, and WiFi connectivity. Provides Modified Early Warning Score (MEWS) for adult clinical assessment. Used by physicians or medical staff to track patient status; output aids clinical decision-making by providing real-time physiological data and early warning scores, potentially improving patient monitoring and timely intervention.
Clinical Evidence
Bench testing only. Performance evaluated via functional and system-level testing, EMC/electrical safety testing (IEC 60601-1, IEC 60601-1-2), and software verification/validation. Compliance with consensus standards (IEC 80601-2-30, IEC 60601-2-49) demonstrated equivalence to the predicate device.
Technological Characteristics
Main unit with touchscreen; rechargeable Li-ion battery (11.1 VDC, 4500 mAh). Sensing: SpO2 (optical), NIBP (oscillometric), CO2 (IR absorption), TEMP (thermal resistance or IR temporal scanner). Connectivity: WiFi (IEEE 802.11 a/b/g/n). Socket material: PBT + Glass Fibre (DR48). Standards: AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 80601-2-30, IEC 60601-2-49.
Indications for Use
Indicated for monitoring SpO2, PR, NIBP, TEMP, and CO2 in adult, pediatric, and neonatal patients in professional healthcare facilities. Provides MEWS for clinical assessment in adult patients.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
March 27, 2019
Shenzhen Mindray Bio-medical Electronics Co., LTD. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department
Hi-tech Industrial Park, Nanshan Shenzhen, 518057 CN
Re: K182821
Trade/Device Name: Accutorr 7/VS-900/VS-900c Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: February 13, 2019 Received: February 19, 2019
Dear Yanhong Bai:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Shawn W. Forrest -S 2019.03.27 14:23:19 -04'00' Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182821
#### Device Name
Accutorr 7/VS-900/VS-900c Vital Signs Monitor
#### Indications for Use (Describe)
The Accutorr 7/VS-900/VS-900c Vital Signs Monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
. The monitor also provides a Modified Early Warning Score (MEWS) for clinical assessment in adult patients.
Type of Use (Select one or both, as applicable)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Accutorr 7/VS-900/VS-900c Vital Signs Monitor is provided below.
| Device Common Name: | monitor, physiological, patient (without arrhythmia<br>detection or alarms) |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | Accutorr 7/VS-900/VS-900c Vital Signs Monitor |
| Submitter: | Shenzhen Mindray Bio-medical Electronics Co., LTD.<br>Mindray Building, Keji 12th Road South,<br>Hi-tech Industrial Park, Nanshan,<br>Shenzhen, 518057, P. R. China<br>Tel: +86 755 8188 8998<br>Fax: +86 755 2658 2680 |
| Contact: | Ms. Yanhong Bai<br>Manager Regulatory Affairs<br>Shenzhen Mindray Bio-medical Electronics Co., LTD.<br>Mindray Building, Keji 12th Road South,<br>High-tech Industrial Park, Nanshan, Shenzhen<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | September 30, 2018 |
| Classification Regulation: | 21 CFR 870.2300, Class II, Cardiac monitor (including<br>cardiotachometer and rate alarm) |
| Panel: | Cardiovascular |
{5}------------------------------------------------
| Primary | | | | |
|-----------------|----------------------|---------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Product<br>Code | Regulation<br>Number | Panel | Regulation description | Device Common Name |
| MWI | 21 CFR<br>870.2300 | Cardiovascular | Cardiac monitor<br>(including<br>cardiotachometer and rate<br>alarm) | monitor, physiological,<br>patient (without<br>arrhythmia detection or<br>alarms) |
| Secondary | | | | |
| Product<br>Code | Regulation<br>Number | Panel | Regulation description | Device Common Name |
| CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas<br>analyzer | analyzer, gas,<br>carbon-dioxide,<br>gaseous-phase |
| DQA | 21 CFR<br>870.2700 | Anesthesiology | Oximeter | oximeter |
| DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood<br>pressure measurement<br>system | system, measurement,<br>blood-pressure,<br>non-invasive |
| FLL | 21 CFR<br>880.2910 | General<br>Hospital | Clinical electronic<br>thermometer | thermometer, electronic,<br>clinical |
#### Classification Regulation, Classification Name and Product Codes:
#### Primary Predicate Device:
K170712 - Accutorr 7/VS-900 Vital Signs Monitor; SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
#### Secondary Predicate Device:
K011291 - TEMPORALSCANNER THERMOMETER, SENSORTOUCH; EXERGEN CORP.
#### Indications for Use:
The Accutorr 7/VS-900/VS-900c Vital Signs Monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), Temperature (TEMP) and Carbon Dioxide (CO2) on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
- The monitor also provides a Modified Early Warning Score (MEWS) for clinical assessment in adult patients.
{6}------------------------------------------------
#### Device Description:
The Accutorr 7/VS-900/VS-900c Vital Signs Monitors are composed of the main unit and parameter modules. The main unit has the following functions:
- Provides operation interface, such as keys, touchscreen, and display screen. ●
- Communicates with the parameter modules. ●
- Delivers the setup commands of parameters to the parameter modules. ●
- Obtains the parameter measurements. ●
- Displays and stores the parameter numeric measurements and waveforms, and sends the data to Mindray's central monitoring system or Mindray's eGateway.
The parameter measurement modules include Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), Temperature (TEMP) and Carbon Dioxide (CO2). Their measurement principles are as follows:
Pulse Oxygen Saturation (SpO2) monitoring is a non-invasive technique used to determine the amount of oxygenated hemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the emitter side of the probe is partly absorbed when it passes through the monitored tissue. The amount of transmitted light is detector side of the probe. When the pulse active part of the light signal is examined, the amount of light absorbed by the hemoglobin is measured and the pulse oxygen saturation can be calculated. The SpO2 module processes the electrical signal and displays a waveform and numeric values for SpO2 and pulse rate.
The monitor uses the oscillometric method for measuring the non-invasive blood pressure (NIBP). NIBP measurement is based on the principle that pulsatile blood flow through an artery creates oscillations of the arterial wall. The oscillometric device uses a blood pressure cuff to sense these oscillations, which appear as tiny pulsations in cuff pressure. The Oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff deflates from above systolic pressure. The amplitude suddenly increases as the pulse breaks through the occlusion in the artery. As the cuff pressure decreases further, the pulsations increase in amplitude, reach a maximum (which approximates to the mean pressure), and then diminish. The oscillometric method measures the mean pressure and determines the systolic and diastolic pressures.
CO2 monitoring is based on measuring the absorption of infrared (IR) light of specific wavelengths. CO2 has its own absorption characteristic and the amount of light passing to the gas probe depends on the concentration of the measured CO2. When a specific band of IR light passes through respiratory gas samples, some of the IR light will be absorbed by the CO2 molecules. The amount of IR light transmitted after it has been passed through the respiratory gas sample is measured with a photodetector. From the amount of IR light measured, the concentration of CO2 is calculated.
The monitor supports two kinds of TEMP module: Mindray TEMP and Exergen TEMP. Different TEMP modules cannot be used at same time. The customers need to choose one kind of TEMP module before they buy the monitor.
{7}------------------------------------------------
- Mindray TEMP Module has two measuring modes, Monitor mode and Predictive mode. ● In the Monitor mode, the thermal resistance in the probe reflects the subject's temperature by sensing the temperature change of the subject according to the thermal equilibrium principle. In the Predictive mode, the subject's temperature is predicted by the temperature rise curve (which is obtained by collecting multiple temperature measurements) at the time of initial thermal conduction.
- The Exergen Temperature Module is an infrared thermometer designed for accurate and completely non-invasive temperature assessment by scanning the temporal artery (TA). Temperature is measured by gently swiping the Temporal Scanner across the forehead, and includes a momentary touch of the probe to the neck area behind the ear lobe to account for any cooling of the forehead as a result of diaphoresis. The patented arterial heat balance technology (AHB™) automatically measures the temperature of the skin surface over the artery and the ambient temperature. It samples these readings approximately 1000 times a second, ultimately recording the highest temperature measured (peak) during the course of the measurement. The Temporal Scanner emits nothing - it only senses the natural thermal radiation emitted from the skin.
#### Non-clinical testing:
EMC and Electrical Safety - To establish the substantial equivalence of the subject Accutorr 7/VS-900/VS-900c Vital Signs Monitor to the previously cleared Accutorr 7/VS-900 Vital Signs Monitor (K170712), Mindray conducted Electromagnetic Compatibility and Electrical Safety testing level testing on the subject device. The testing provides an evaluation of the EMC and electrical safety of the subject devices relevant to each of the modifications since clearance of the predicate Accutorr 7/VS-900 Vital Signs Monitor. The Electromagnetic Compatibility and Electrical Safety testing shows that the subject device continues to meet the specifications and the performance of the subject device is equivalent to the predicate device.
- AAMI ANSI ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and 0 A2:2010/(R)2012* Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
Performance Testing - Bench - To establish the substantial equivalence of the subject Accutorr 7/VS-900/VS-900c Vital Signs Monitor to the previously cleared Accutorr 7/VS-900 Vital Signs Monitor (K170712), Mindray conducted functional and system level testing on the subject devices. The testing provides an evaluation of the performance of the subject devices relevant to each of the modifications since clearance of the Accutorr 7/VS-900 Vital Signs Monitor. The functional and system level testing shows that the subject device continues to meet the specifications and the performance of the subject device is equivalent to the predicate device. In addition, Mindray has conducted testing to ensure the subject device meets relevant consensus
{8}------------------------------------------------
#### standards.
In addition, Mindray has conducted testing to ensure the subject device meets following consensus standards.
- IEC 80601-2-30: 2013-07 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-2-49:2011 Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
Software Verification and Validation Testing - Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the subject device (Accutorr 7/VS-900/VS-900c Vital Signs monitors) was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
#### Substantial Equivalence:
Comparison of Indications - Indications of the predicate device (Accutorr 7/VS-900) and the subject device (Accutorr 7/VS-900/VS-900c Vital Signs monitors) are the same except the new model, VS-900c, is added.
Comparison of Technological Characteristics - The table below compares the key technological feature of the subject device (Accutorr 7/VS-900c) to the primary predicate device (Accutorr 7/VS-900) (K170712). The features in gray are the main features that have been modified since their previous clearances.
{9}------------------------------------------------
### Device Comparison Table
| Feature | Primary Predicate Device (K170712) | Subject Devices |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Integrated display and<br>touchscreen | 8.4" 800*600 pixels | Same |
| Power supply | One rechargeable Lithium-ion battery or<br>AC power supply | Same |
| Battery | Chargeable Lithium-Ion, 11.1<br>VDC,<br>4500 mAh | Same |
| Data Recorder | The thermal recorder can be used to print<br>patient<br>information,<br>measurement<br>numerics, and waveforms. | Same |
| Speaker | Provide alarm tones (45 to 85 dB), key<br>tones; support PITCH TONE<br>multi-level tone modulation | Same |
| WiFi | 2.4G/5G dual band WiFi, compatible<br>with IEEE 802.11 a/b/g/n | Same |
| Feature | Primary Predicate Device (K170712) | Subject Devices |
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Temperature (Temp) | Mindray Temp Module :<br>Technique: Thermal resistance<br>Measurement range:<br>Monitor mode:25 to 44 °C (77 to 111.2 °F)<br>Predictive mode: 35 to 43 °C (95 to 109.4 °F)<br>Accuracy (Monitor mode):<br>25 to 32°C (not include 32°C): ± 0.2 °C<br>32 to 44°C (include 32°C): ±0.1 °C(± 0.2 °F)<br>or<br>77 to 89.6 °F (not include 89.6°F): ± 0.4 °F<br>89.6 to 111.2 °F (include 89.6°F): ± 0.2 °F<br>Response Time:<br>Monitor mode: <60 s<br>Predictive mode: <20 s (typical test: < 12s ) | Same |
| Feature | Primary Predicate Device (K170712) | Subject Devices |
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Exergen Temperature<br>(Temp)<br>Module | Not supported. | Integrated an optional OEM temp module (the Exergen Temporal Scanner Thermometer<br>cleared in K011291)<br>Measurement range:<br>$16.0°C$ to $43.0°C$ ( $61°F~110°F$ )<br>Measurement accuracy: $\pm 0.2°C$ |
| Pulse oxygen<br>saturation (SpO2) | Is compatible with the following 3<br>modules to measure oxygen saturation:<br>• Mindray SpO2 Module<br>• Masimo SpO2 Module<br>• Nellcor SpO2 Module<br>Mindray SpO2 Module:<br>Measurement range: 0 to 100%<br>Accuracy:<br>70 to 100%: $\pm$ 2% (without motion in<br>adult/pediatric mode)<br>70 to 100%: $\pm$ 3% (without motion in<br>neonate mode)<br>0% to 69%: Not specified. | Same |
| Feature | Primary Predicate Device (K170712) | Subject Devices |
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| | <b>Masimo SpO2 Module:</b> | |
| | Measurement range: 1 to 100% | |
| | Accuracy:70 to 100%: $\pm$ 2% (without motion in adult/pediatric mode) | |
| | 70 to 100%: $\pm$ 3% (without motion in neonate mode) | |
| | 70 to 100%: $\pm$ 3% (with motion) | |
| | 1% to 69%: Not specified. | |
| | <b>Nellcor SpO2 Module:</b> | |
| | Measurement range: 0 to 100% | |
| | Accuracy: | |
| | 70 to 100%: $\pm$ 2% (adult/pediatric) | |
| | 70 to 100%: $\pm$ 3% (neonate) | |
| | 0% to 69%: Not specified. | |
| | Primary Predicate Device (K170712) | Subject Devices |
| Feature | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Pulse rate (PR) | Pulse rate may be obtained from the<br>SpO2 module or the NIBP module. | |
| | PR from Mindray SpO2 Module<br>Measurement range: 20 to 254 bpm<br>Accuracy: ±3 bpm (without motion)<br>PR from Masimo SpO2 Module<br>Measurement range: 25 to 240 bpm<br>Accuracy:<br>±3 bpm (without motion) | The Measurement range of PR from the NIBP Module has been expanded to: 30 to 300 bpm |
| | ±5 bpm (with motion) | Other specifications are the same |
| | PR from Nellcor SpO2 Module<br>Measurement range: 20 to 300 bpm<br>Accuracy:<br>20 to 250 bpm: ±3 bpm<br>251 to 300 bpm, not specified | |
| | PR from NIBP Module<br>Measurement range: 40 to 240 bpm<br>Accuracy: ±3bpm or ±3%, whichever is<br>greater | |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
| Feature | Primary Predicate Device (K170712) | Subject Devices | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Non-invasive blood<br>pressure (NIBP) | Uses the oscillometric method for<br>measuring non-invasive blood pressure<br>(NIBP). This measurement can be used<br>for adults, pediatrics and neonates.<br><br>Measurement range:<br>Adult Pediatric Neonate Systolic: 40 to 270 40 to 200 40 to 135 Diastolic: 10 to 210 10 to 150 10 to 100 Mean: 20 to 230 20 to 165 20 to 110 Accuracy:<br>Maximum average error: ±5 mmHg<br>Maximum standard deviation: 8mmHg<br>Maximum measurement time:<br>Adult, pediatric: 180 s Neonate: 90 s<br>Static pressure measurement range:<br>0mmHg to 300mmHg<br>Static pressure measurement accuracy:<br>±3 mmHg | | | | | | | | | | | | | | | | | Uses the oscillometric method for measuring non-invasive blood pressure (NIBP). This<br>measurement can be used for adult, pediatric and neonatal patients.<br>Expand the display range of NIBP to comply with IEC 80601-2-30: 2013:<br>Measurement range:<br>Adult Pediatric Neonate Systolic: 25 to 290 25 to 240 25 to 140 Diastolic: 10 to 250 10 to 200 10 to 115 Mean: 15 to 260 15 to 215 15 to 125 All other specifications are the same. | | | | | | | | | | | | | | | | | | |
{15}------------------------------------------------
| Feature | Primary Predicate Device (K170712) | Subject Devices | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------|
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c | |
| Carbon dioxide (CO2) | CO2 parameter measuring.<br>CO2 measurement range: 0-20%<br>Resolution: 1mmHg<br>CO2 Accuracy:<br>After Warm-up (FULL Accuracy):<br>0%≤CO2<1%: $\pm$ 0.1%<br>1%≤CO2<5%: $\pm$ 0.2%<br>5%≤CO2<7%: $\pm$ 0.3%<br>7%≤CO2<12%: $\pm$ 0.4%<br>12%≤CO2≤13%: $\pm$ 0.5%<br>13%<CO2≤20%: $\pm$ (0.43%+8%rel)<br>20%<CO2≤30%, unspecified<br><br>During Warm-up(ISO Accuracy):<br>Add $\pm$ 0.3%ABS to the FULL accuracy<br>Accuracy drift: Meet the requirement for<br>measurement accuracy within 6 hours.<br>Sampling flow rate: 50ml/min | Same | |
| | Add a new model named VS-900c | Not supported | Supported |
{16}------------------------------------------------
| | Primary Predicate Device (K170712) | Subject Devices | | | |
|--------------------------------------------------------|------------------------------------|------------------------------|--------------|-------------------------------------------------------|--------------------------------|
| Feature | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c | | | |
| Support<br>the<br>new<br>Masimo RD SpO2<br>accessories | Not supported | Mindray<br>Part<br>No. | Masimo Model | Description | Corresponding<br>510(k) Number |
| | | 040-003376-00 | 4050 | RD SET DCI, Adult Reusable<br>Sensor | K051212 |
| | | 040-003377-00 | 4051 | RD SET DCI, Pediatric Reusable Sensor | K051212 |
| | | 040-003378-00 | 4080 | RD SET MD 14-05, PC 5 ft | K042536 |
| | | 040-003379-00 | 4081 | RD SET MD 14-12, Patient<br>Cable 12 ft | K042536 |
| | | 040-003380-00 | 4053 | RD Set TC-I SpO2 Reusable<br>Tip-Clip Ear Sensor, 3ft | K051212 |
| | | 040-003381-00 | 4089 | RD to LNC Adapter Cable | K042536 |
| | | 040-003382-00 | 4000 | RD SET Adhesive Sensor | K042346 |
| | | 040-003383-00 | 4001 | RD SET PDT Adhesive Sensor | K042346 |
| | | 040-003384-00 | 4002 | RD Set Infant Adhesive Sensor | K042346 |
| | | 040-003385-00 | 4003 | RD Set Neo Adhesive Sensor | K042346 |
| | | 040-003386-00 | 4004 | RD Set NeoPt Adhesive Sensor | K042346 |
| | | 040-003387-00 | 4005 | RD Set NeoPt-500<br>Non-adhesive sensor | K042346 |
| | | 040-003426-00 | 4092 | LNCS to RD Adapter Cable | K042346 |
| 040-003310-00 | 583A | 8pin Masimo Cable ( RD SET ) | / | | |
| Support timer tool | Not supported | Supported | | | |
{17}------------------------------------------------
| | Primary Predicate Device (K170712) | Subject Devices |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Feature | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Support the Visit<br>Number (VID) of the<br>patient demographics | Not supported | Supported |
| Support the Facility<br>of the location<br>information of patient<br>monitor | Not supported | Supported |
| Support DNS for<br>ADT and EMR server<br>address | Not supported | Supported |
| Support authorization<br>of the user by the<br>LDAP | Not supported | Supported |
| Support password<br>protection for<br>accessing the Alarm<br>Setup Menu | Not supported | Supported |
| Support encrypting of<br>patient private<br>information sent to<br>the network | Not supported | Supported |
| Feature | Primary Predicate Device (K170712) | Subject Devices |
| | Accutorr 7/VS-900 | Accutorr 7/VS-900/VS-900c |
| Support expanded<br>cleaning and<br>disinfecting agents<br>list | Not supported | Supported<br>Please refer to Section 12.3.12 for details. |
| Change the material<br>of the socket on the<br>SpO2/CO2 signal<br>cord | The material of the socket on the<br>SpO2/CO2 signal cord is Polypropylene<br>(PP). | The material of the socket on the SpO2/CO2 signal cord is changed to Polybutyene<br>Terephthatate (PBT) + Glass Fibre (DR48). |
{18}------------------------------------------------
{19}------------------------------------------------
#### Substantial Equivalence Conclusion:
Based on the detailed comparison of specifications for each of the modifications to the primary predicate device (Accutorr 7/VS-900) (K170712) and the secondary predicate device (Exergen Temporal Scanner Thermometer )(K011291), and the performance testing and conformance with applicable standards, the subject device (Accutorr 7/VS-900/VS-900c) can be found substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.