WAVE Clinical Platform

K171056 · Excel Medical Electronics, LLC · MWI · Jan 4, 2018 · Cardiovascular

Device Facts

Record IDK171056
Device NameWAVE Clinical Platform
ApplicantExcel Medical Electronics, LLC
Product CodeMWI · Cardiovascular
Decision DateJan 4, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The WAVE Clinical Platform is software intended to route, store, and display data, alarms, results and diagnostic information from medical devices, Electronic Medical Records (EMR), and Clinical Information Systems (CIS). The WAVE Clinical Platform is a remote monitoring platform that displays physiologic data, waveforms, alarms, results and diagnostic information routed the WAVE Clinical Platform from supported devices and systems. The WAVE Clinical Platform is intended for use in hospital type environments. The WAVE Clinical Platform is intended to be used by healthcare professionals for the following purposes: · To remotely consult regarding patients' statuses: · To remotely review other standard or critical near real-time patient data, waveforms, and results in order to utilize this information to aid in clinical decisions.

Device Story

WAVE Clinical Platform is a software-based remote monitoring system; gathers physiologic data, waveforms, alarms, and diagnostic information from medical devices, EMRs, and CIS; routes and stores data for near real-time and retrospective display. Used in hospital environments by healthcare professionals; operates on standard hospital network hardware and web-enabled display devices. Provides clinicians with remote access to patient data to aid in clinical decision-making and patient care delivery. Does not control connected medical devices; serves as a supplementary monitoring and data management tool.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including system-level tests, user interface verification, device compatibility verification, and performance/safety testing.

Technological Characteristics

Software-based platform; client-server architecture; runs on general-purpose server hardware; utilizes standard hospital wired/wireless networks; web-enabled display; disk-based data storage.

Indications for Use

Indicated for in-hospital patients with physiological sensors attached. Used by healthcare professionals to remotely consult on patient status and review near real-time or historical patient data, waveforms, alarms, and results to aid in clinical decision-making. Not intended to replace hospital device monitoring or electronic data management systems; not to be used as the sole source of alarms.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text. January 4, 2018 Excel Medical Electronics, LLC Lance Burton General Manager 801 Maplewood Dr. Suite 25 Jupiter, Florida 33458 Re: K171056 Trade/Device Name: WAVE Clinical Platform Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: November 30, 2017 Received: December 4, 2017 Dear Lance Burton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ ### Page 2 - Lance Burton and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, M& Hillelneme for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171056 Device Name WAVE Clinical Platform #### Indications for Use (Describe) The WAVE Clinical Platform is software intended to route, store, and display data, alarms, results and diagnostic information from medical devices, Electronic Medical Records (EMR), and Clinical Information Systems (CIS). The WAVE Clinical Platform is a remote monitoring platform that displays physiologic data, waveforms, alarms, results and diagnostic information routed the WAVE Clinical Platform from supported devices and systems. The WAVE Clinical Platform is intended for use in hospital type environments. The WAVE Clinical Platform is intended to be used by healthcare professionals for the following purposes: · To remotely consult regarding patients' statuses: · To remotely review other standard or critical near real-time patient data, waveforms, and results in order to utilize this information to aid in clinical decisions. Warning: The WAVE Clinical Platform is intended to supplement and not to replace any part of the hospital's device monitoring or electronic data management systems. Do not rely on the WAVE Clinical Platform product as the sole source of alarms. **Type of Use (Select one or both, as applicable)**> Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY # WAVE Clinical Platform In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a 510(k) Summary for the use of the WAVE Clinical Platform. - April 7, 2017 1. Date: - 2. SUBMITTED BY: Excel Medical Electronics, LLC 801 Maplewood Drive, Suite 25 Jupiter, Florida 33458 - 3. CONTACT PERSON: Lance Burton General Manager Email: lance.burton@excelmedical.com Phone: 866-573-8807 Fax: 561-748-8889 # 4. DEVICE INFORMATION: WAVE Clinical Platform | | Device Information | |-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary<br>Name: | WAVE Clinical Platform | | Common or Usual<br>Name: | Data Management, Visualization and Clinical<br>Decision Support Software | | Classification: | Class II Device | | Product Code: | MWI | | Regulation Number: | CFR 870.2300 | | Regulation Name: | Cardiac Monitor (including Cardiotachometer and<br>Rate Alarm) | | Classification Name: | Cardiac Monitor (including cardiotachometer and<br>rate alarm) Monitor, Physiological, Patient<br>(Without Arrhythmia Detection or Alarms) | {4}------------------------------------------------ | Review Panel: | Part 870, Subpart B Cardiovascular Diagnostic<br>Devices, Sec. 870.1425 Programmable<br>Diagnostic Computer | |---------------|-------------------------------------------------------------------------------------------------------------| |---------------|-------------------------------------------------------------------------------------------------------------| - 5. PREDICATE DEVICE: COVIDIEN VITAL SYNC™ INFORMATICS MANAGER & VIRTUAL PATIENT MONITORING PLATFORM (K140339; MWI, OUG) INTENDED USE: "The Vital Sync™ Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Electronic Medical Record (eMR) and Clinical Information System (Cis). The Vital Sync™ Virtual Patient Monitoring Platform is a remote monitoring platform that displays physiologic data. waveforms and alarms routed through the Vital Svnc™ Informatics Manager for supported devices. The Vital Sync™ Virtual Patient Monitoring Platform is intended to be used by healthcare professionals for the following purposes: - · To remotely consult reqarding a patient's status · To remotely review other standard or critical near real-time patient data, waveforms and alarms in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner. WARNING: The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform are intended to supplement and not to replace any part of the hospital's device monitoring. Do not rely on the Vital Sync™ Informatics Manager &Virtual Patient Monitoring Platform as the sole source of alarms." | | Predicate Device Information | | |-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--| | Device Proprietary<br>Name: | Vital Sync™ Informatics Manager & Virtual Patient<br>Monitoring Platform | | | Classification: | Class II Device | | | 510(k) Number: | K140339 | | | Product Code: | MWI, OUG | | | Regulation Number: | CFR 870.2300 | | | Regulation Name: | Cardiac Monitor (including Cardiotachometer and<br>Rate Alarm) | | | Classification Name: | Cardiac Monitor (including cardiotachometer and<br>rate alarm) Monitor, Physiological, Patient (Without<br>Arrhythmia Detection or Alarms) | | {5}------------------------------------------------ | Review Panel: | Part 870, Subpart B Cardiovascular Diagnostic<br>Devices, Sec. 870.1425 Programmable Diagnostic<br>Computer | |---------------|-------------------------------------------------------------------------------------------------------------| |---------------|-------------------------------------------------------------------------------------------------------------| - 6. DEVICE DESCRIPTION: Excel Medical's WAVE Clinical Platform is a software platform that gathers and stores data from patient monitoring and other medical devices, hospital information systems, and clinical analytics data sources for near real-time and retrospective remote display for clinicians to use in clinical decision support. The WAVE Clinical Platform can also route and forward these data to third party systems. The WAVE Clinical Platform can utilize the hospital's existing network and hardware for installation and display on any device that is web-enabled. The WAVE Clinical Platform does not control any of the medical devices or systems it is connected with. - 7. INTENDED USE / INDICATIONS FOR USE: The WAVE Clinical Platform is software intended to route, store, and display data, alarms, results and diagnostic information from medical devices, Electronic Medical Records (EMR), and Clinical Information Systems (CIS). The WAVE Clinical Platform is a remote monitoring platform that displays physiologic data, waveforms, alarms, results and diagnostic information routed through the WAVE Clinical Platform from supported devices and systems. The WAVE Clinical Platform is intended for use in hospital or hospital type environments. The WAVE Clinical Platform is intended to be used by healthcare professionals for the following purposes: To remotely consult regarding patients' statuses; To remotely review other standard or critical near real-time patient data, waveforms, alarms, and results in order to utilize this information to aid in clinical decisions. Warning: The WAVE Clinical Platform is intended to supplement and not replace any part of the hospital's device monitoring or electronic data management systems. Do not rely on the WAVE Clinical Platform product as the sole source of alarms. {6}------------------------------------------------ # 8. TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE: | | Predicate - Vital Sync<br>Informatics Manager &<br>Virtual Patient<br>Monitoring Platform | WAVE Clinical<br>Platform | Data Storage | Hour to days of data is<br>buffered and then<br>archived to tape for<br>indefinite storage until<br>deleted | All data are archived to<br>and indefinitely stored<br>on disk drives until<br>deleted | |-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The Vital Sync<br>Informatics Manager is<br>software that is<br>intended to route and<br>store medical device<br>data and device<br>diagnostic information<br>from supported devices<br>to the Electronic<br>Medical Record (EMR)<br>and Clinical Information<br>System (CIS).<br><br>The Vital Sync Virtual<br>Patient Monitoring<br>Platform is a remote<br>monitoring platform that<br>displays physiologic<br>data, waveforms and<br>alarms routed through<br>the Vital Sync<br>Informatics Manager<br> | The WAVE Clinical<br>Platform is software<br>intended to route, store,<br>and display data,<br>alarms, results and<br>diagnostic information<br>from medical devices,<br>Electronic Medical<br>Records (EMR), and<br>Clinical Information<br>Systems (CIS).<br><br>The WAVE Clinical<br>Platform is a remote<br>monitoring platform that<br>displays physiologic and<br>clinically relevant data,<br>waveforms, and alarms<br>routed through the<br>WAVE Clinical<br>Platform from<br>supported devices and<br>data systems. The<br>WAVE Clinical<br>Platform is intended to<br>be used by healthcare<br>professionals for the<br>following purposes:<br>• To remotely consult<br>regarding patients'<br>statuses;<br>• To remotely review<br>other standard or critical<br>near real-time and<br>historical patient data,<br>waveforms, alarms, and<br>results in order to utilize | Where Used | Anywhere clinicians<br>remotely consult on<br>patients | Anywhere clinicians<br>remotely consult on<br>patients | | | this information to aid in<br>clinical decisions and<br>deliver patient care in a<br>timely manner.<br><br>WARNING: The Vital<br>Sync Informatics<br>Manager & Virtual<br>Patient Monitoring<br>Platform are intended<br>to supplement and not<br>to replace any part of<br>the hospital's device<br> | this information to aid in<br>clinical decisions and<br>deliver patient care in a<br>timely manner.<br><br>WARNING: The WAVE<br>Clinical Platform is<br>intended to supplement<br>and not to replace any<br>part of the hospital's<br>device monitoring or<br>electronic data<br>management systems.<br>Do not rely on the<br>WAVE Clinical<br>Platform product as the<br>sole source of alarms. | Wireless | Yes, handheld display<br>devices | Yes, handheld display<br>devices | | Device Regulatory<br>Classification | 21 CFR 870.2300;<br>Cardiac monitor<br>(including<br>cardiotachometer and<br> | 21 CFR 870.2300;<br>Cardiac monitor<br>(including<br>cardiotachometer and<br>rate alarm) | Ability to View Near<br>Real-time Data | Yes | Yes | | Product Code | MWI & OUG | MWI | Synopsis of<br>Functionality | Store, route and display<br>patient monitor, EMR<br>and CIS data, alarms,<br>and smart alarms<br>remotely to aid in<br>clinical decisions and<br>deliver patient care in a<br>timely manner | Store, route and display<br>patient monitor, EMR<br>and CIS data, alarms,<br>and smart alarms<br>remotely to aid in clinical<br>decisions and deliver<br>patient care in a timely<br>manner | | Device Class | II | II | Target<br>Population/Demographi<br>CS | ICU and ER patients<br>that have Covidien<br>physiological sensors<br>attached. | In-hospital patients that<br>have physiological<br>sensors attached. | | 510(k) Number | K140339 | TBD | | | | | Other Technology Used | Standard computers,<br>network, and WIFI<br>technology | Standard computers,<br>network, and WIFI<br>technology | | | | | Host server | Uses hospital's existing<br>server hardware or<br>server pre-loaded with<br>Vital Sync software and<br>off the shelf operating<br>system | Uses hospital's existing<br>server hardware or<br>server pre-loaded with<br>the WAVE Clinical<br>Platform software and<br>off-the-shelf operating<br>system | | | | {7}------------------------------------------------ {8}------------------------------------------------ # 9. Substantial Equivalence Discussion on Comparison to Predicate Device: Excel Medical's WAVE Clinical Platform and Covidien's Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform perform the same functions with no difference in how the devices are used. The WAVE Clinical Platform is a single medical device that performs the functions of (a) storing, (b) routing and (c) displaying data from patient monitors, EMR and CIS systems, and smart alarms. Covidien's two devices (Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform) perform these same functions. The WAVE Clinical Platform's data is displayed remotely to aid in clinical decisions and deliver patient care in a timely manner. The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform's data is displayed remotely to aid in clinical decisions and deliver patient care in a timely manner. Excel {9}------------------------------------------------ Medical's WAVE Clinical Platform data are archived to and indefinitely stored on disk drives until deleted. Covidien's Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform Hour to days of data is buffered and then archived to tape for indefinite storage until deleted. The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform gathers ICU and ER patient data from Covidien physiological sensors. Excel Medical's WAVE Clinical Platform gathers data from in-hospital patients that have physiological sensors attached. These differences have no effect on how these devices are used. The patient's in-hospital location and the physiological sensors may be different, but Excel Medical's WAVE Clinical Platform and Covidien's Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform function and intended use are the same. Risk and Hazards are the determined to be the same for both devices. ### 10. Substantial Equivalence Discussion on Safety and Effectiveness: There are three differences between Excel Medical's WAVE Clinical Platform and Covidien's Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform: > (1) The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform are essentially two products and the WAVE Clinical Platform is one product that perform the same functionality. (2) The WAVE Clinical Platform gathers data from inhospital patients that have physiological sensors attached. The Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform gathers ICU and ER patient data from Covidien physiological sensors. This is based on a business decision to include other manufacturer's devices and systems and is no difference in functionality or intended use of the product. (3) The WAVE Clinical Platform has the ability to store a larger volume of data and made available online. Whereas, the Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform requires long term storage of data to be archived offline. These differences have no effect on safety and effectiveness or the intended use for the device. # 11.Summary of Technology Characteristics Technology characteristics of the WAVE Clinical Platform and the Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform are determined to be the same. Both devices are software devices with client server architecture and support web- {10}------------------------------------------------ enabled displays of data. Both software devices acquire and distribute data across general purpose hospital networks, both wired and wireless. Both devices have server software that run on general purposes server computer hardware and use data storage technology mediums widely available in hospitals. Both devices allow users to access and display data from web-enabled devices over wired and wireless general purpose networks. All of these technologies are proven and any risks and their mitigations are widely understood in hospital environments. The platforms have identical indications for use and both display parameters and waveforms from connected medical devices to a remote, webenabled display. ### 12. Non-clinical Performance: Bench Safety, efficacy and substantial equivalence was shown through system level tests, user interface verification, device compatibility verification, performance tests and safety testing. The WAVE Clinical Platform test results on non-clinical testing demonstrate the device is substantially equivalent to the predicate device - the Covidien Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform in terms of safety, features and performance. No new questions of safety and effectiveness have been raised. ### 13. Clinical Performance: Not Applicable. Clinical evidence was not necessary to show substantial equivalence. ### 14. Conclusions from Non-clinical Performance Testing: The WAVE Clinical Platform test results on non-clinical testing demonstrate the WAVE Clinical Platform is substantially equivalent to the predicate device - the Covidien Vital Sync™ Informatics Manager & Virtual Patient Monitoring Platform in terms of safety, features and performance.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...