← Product Code [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI) · K170538

# RVS-100 Vital Signs Monitor (K170538)

_Rudolf Riester GmbH · MWI · Oct 17, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K170538

## Device Facts

- **Applicant:** Rudolf Riester GmbH
- **Product Code:** [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI.md)
- **Decision Date:** Oct 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The RVS-100 vital signs monitor is intended to be used for monitoring, displaying, storing and sending alarms regarding multiple physiological patient parameters, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), and Temperature (Temp). The RVS-100 vital signs monitor is intended to be used in controlled, hospital environments. It is not intended for helicopter transport, hospital ambulance or home use.

## Device Story

Portable vital signs monitor; measures SpO2, PR, NIBP, and body temperature; intended for clinical professionals in hospital environments (outpatient, emergency, low-acuity). Device consists of main unit, NIBP cuff, temperature sensor, and SpO2 sensor. Features DC power connector for optional RVS-200 wall station integration. Provides real-time physiological data display and alarm functionality for single-patient monitoring. Supports clinical decision-making by providing continuous or spot-check vital sign data to healthcare providers.

## Clinical Evidence

Bench testing only. No clinical data provided. Device performance verified through compliance with international standards for safety, electromagnetic compatibility, alarm systems, and specific requirements for NIBP, pulse oximetry, and clinical thermometers.

## Technological Characteristics

Portable monitor; AC power supply; includes NIBP cuff, temperature sensor, and SpO2 sensor. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 80601-2-30 (NIBP), IEC 80601-2-61 (SpO2), ISO 80601-2-56 (thermometry).

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- Vital Signs Monitor V6 ([K153135](/device/K153135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.

October 17, 2017

Rudolf Riester GmbH % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China

Re: K170538

Trade/Device Name: RVS-100 Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI Dated: September 5, 2017 Received: September 12, 2017

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Wilhelm

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170538

Device Name RVS-100 Vital Signs Monitor

### Indications for Use (Describe)

The RVS-100 vital signs monitor is intended to be used for monitoring, reviewing, storing and sending alarms regarding multiple physiological patient parameters, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Noninvasive Blood Pressure (NIBP), and Temperature (Temp).

The RVS-100 vital signs monitor is intended to be used in controlled, hospital environments. It is not intended for helicopter transport, hospital ambulance or home use.

| Type of Use (Select one or both, as applicable)                              |                                                                            |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span>    Prescription Use (Part 21 CFR 801 Subpart D)  </span> </div> | <div> <span>   Over-The-Counter Use (21 CFR 801 Subpart C)  </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary of 510(k) information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: _K170538__

- 1. Date of Submission:8/25/2017
- 2. Sponsor Identification

Rudolf Riester GmbH P.O.B. 35, Bruckstraße 31 DE-72417, Jungingen, Germany

Establishment Registration Number: 8010482

Contact Person: Christof Kleiner Position: Technical Director Tel: +49-7477-9270-43 Fax: +49-7477-9270-70 Email: kleiner@riester.de

- 3. Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Betty Xiao (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{4}------------------------------------------------

- 4. Proposed Device Identification
Proposed Device Name:RVS-100 Vital Signs Monitor Proposed Device Common Name: Vital Signs Monitor

Regulatory Information: Classification Name: Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms): Classification: II; Product Code: MWI; Regulation Number: 21 CFR 870.2300; Review Panel: Cardiovascular;

Intended Use Statement:

The RVS-100 vital signs monitor is intended to be used for monitoring, displaying, storing and sending alarms regarding multiple physiological patient parameters, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), and Temperature (Temp).

The RVS-100 vital signs monitor is intended to be used in controlled, hospital environments. It is not intended for helicopter transport, hospital ambulance or home use.

#### 5. Predicate Device Identification

510(k) Number: K153135 Product Name: Vital Signs Monitor V6 Manufacturer: Guangdong Biolight Meditech Co., Ltd

- 6. Device Description
The Riester RVS-100 vital signs monitor can perform automatic blood pressure, pulse oximetry and body temperature measurements for clinical professionals. The RVS-100 is a portable device, consist of main unit and accessories, including NIBP cuff, Temperature sensor, and SpO2 sensor.

RVS-100 vital signs monitor has a DC power connector, RVS-100 can connect with extension device RVS-200 and supply power for RVS-200 through the DC power connector.

Note: RVS-200 Wall Diagnostic Station mainly consists of reservoir for ear specula and probe cover. It is classified as a Class I 510(k) Exempt medical device, and not included in this 510(k) submission

- 7. Non-Clinical Test Conclusion

{5}------------------------------------------------

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- A IEC 60601-1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- A IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- A IEC 80601-2-30:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- A IEC 80601-2-61:2011 Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- > ISO 80601-2-56: 2009 Medical electrical equipment - Part 2-56: Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
- Clinical Test Conclusion 8.

Non-clinical testing is included in this submission, because all modules have been previously cleared by FDA.

- Substantially Equivalent (SE) Conclusion 9.
The following table compares the DEVICE to the predicate device with respect to intended use, energy source and fluency, etc.

| ITEM           | Proposed Device                                                        | Predicate Device                                                              | Remark  |
|----------------|------------------------------------------------------------------------|-------------------------------------------------------------------------------|---------|
|                | RVS-100 Vital Signs Monitor                                            | Vital Signs Monitor V6                                                        |         |
| Product Code   | MWI                                                                    | MWI                                                                           | Same    |
| Regulation No. | 870.2300                                                               | 870.2300                                                                      | Same    |
| Class          | II                                                                     | II                                                                            | Same    |
| Intended use   | The RVS-100 vital signs monitor is intended to be used for monitoring, | The Vital Signs Monitor V6 is intended to be used for monitoring, displaying, | Similar |

Table 3-1 Comparison of Technology Characteristics

{6}------------------------------------------------

|                                                | displaying, reviewing, storing and<br>sending alarms regarding multiple                                                                                                                                                                                                                                              | reviewing, storing and alarming of<br>multiple physiological parameters of                                                                                                                                                                                 |         |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|                                                | physiological patient parameters,                                                                                                                                                                                                                                                                                    | patients, including Pulse Oxygen                                                                                                                                                                                                                           |         |
|                                                | including Pulse Oxygen Saturation<br>(SpO2), Pulse Rate (PR), Non-invasive                                                                                                                                                                                                                                           | Saturation (SpO2), Pulse Rate (PR),<br>Non-invasive Blood Pressure (NIBP),                                                                                                                                                                                 |         |
|                                                | Blood Pressure (NIBP), and Temperature<br>(Temp).                                                                                                                                                                                                                                                                    | Carbon dioxide (CO2) and Temperature<br>(Temp). The Vital Signs Monitor V6 is                                                                                                                                                                              |         |
|                                                | The RVS-100 vital signs monitor is<br>intended to be used in controlled,<br>hospital environments. It is not<br>intended for helicopter transport,<br>hospital ambulance or home use.                                                                                                                                | intended to be use in outpatient<br>departments and emergency treatment<br>rooms of hospitals, community clinics,<br>private clinics and other medical<br>institutions. It is not intended for<br>helicopter transport, hospital ambulance<br>or home use. |         |
| Intended<br>Population                         | The monitors are intended for adult,<br>pediatric and neonatal.                                                                                                                                                                                                                                                      | The monitors are intended for adult,<br>pediatric and neonatal.                                                                                                                                                                                            | Same    |
| Intended<br>Environment                        | The RVS-100 vital signs monitor is<br>intended to be used in outpatient<br>departments, emergency treatment<br>rooms, and low-acuity areas of hospitals,<br>community clinics, private clinics and<br>other medical institutions. It is not<br>intended for helicopter transport, hospital<br>ambulance or home use. | The vital signs monitor is used to be used<br>in hospital environments including<br>out-patient outpatient departments, wards<br>and NICU. It is not intended for helicopter<br>transport, hospital ambulance or home<br>use.                              | Same    |
| Power Supply                                   | The monitors can be supplied by AC<br>power.                                                                                                                                                                                                                                                                         | The monitors can be supplied by AC<br>power.                                                                                                                                                                                                               | Similar |
| Number of Patient<br>be monitored each<br>time | Single patient each time                                                                                                                                                                                                                                                                                             | Single patient each time                                                                                                                                                                                                                                   | Same    |
|                                                | No                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                         | Same    |

The proposed device, RVS-100 Vital Signs Monitor, is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K170538](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K170538)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
