← Product Code [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI) · K153335

# AlertWatch: OR (K153335)

_Alertwatch, LLC · MWI · Mar 15, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K153335

## Device Facts

- **Applicant:** Alertwatch, LLC
- **Product Code:** [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI.md)
- **Decision Date:** Mar 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management systems and patient medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision. AlertWatch:OR is also intended for use by supervising anesthesiologists outside of operating rooms. Once alerted, the supervising anesthesiologist must contact the clinician inside the operating room or must return to the operating room before making a clinical decision. Once either clinician is alerted, they must refer to the primary monitor or device before making a clinical decision.

## Device Story

AlertWatch:OR is a secondary monitoring system for operating rooms; it aggregates data from networked physiologic monitors, anesthesia information management systems (AIMS), and patient medical records. The system synthesizes this data to identify potential clinical problems and generates alerts when values exceed preset thresholds. It is used by anesthesiology staff (residents, CRNAs, attending anesthesiologists) within the OR and by supervising anesthesiologists outside the OR via mobile devices (iPhone/iPad). The device does not replace primary monitors; clinicians must refer to primary devices before making clinical decisions. The system benefits patients by providing synthesized clinical advisories to help clinicians identify potential complications. The device is installed on hospital networks; it supports paging integration for alert transmission.

## Clinical Evidence

No clinical data. Evidence includes a human factors/usability study for the iPhone interface, verification of default limits/thresholds via literature review and expert committee consensus, wireless coexistence testing, and software verification testing.

## Technological Characteristics

Secondary monitoring software; connects to hospital network for data from AIMS and physiologic monitors. Platforms: PC, iPad, iPhone. Connectivity: Networked. Software: Version 2.50. Compliance: IEC 60601-1-8. Threshold-based alerting system.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- AlertWatch: OR ([K130401](/device/K130401.md))

## Reference Devices

- Amcom CommTech Messenger ([K112047](/device/K112047.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. The profiles are stylized and connected by flowing lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 15, 2016

Alertwatch LLC % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington Street Suite 100 Alexandria, Virginia 22314

Re: K153335

Trade/Device Name: Alertwatch: OR Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: February 16, 2016 Received: February 18, 2016

Dear Donna-Bea Tillman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153335

Device Name AlertWatch: OR

#### Indications for Use (Describe)

AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management systems and patient medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision.

AlertWatch:OR is also intended for use by supervising anesthesiologists outside of operating rooms. Once alerted, the supervising anesthesiologist must contact the clinician inside the operating room or must return to the operating room before making a clinical decision. Once either clinician is alerted, they must refer to the primary monitor or device before making a clinical decision.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY - K153335

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the AlertWatch:OR is provided below.

| Device Common Name:        | Physiological Patient Monitor without arrhythmia<br>detection or alarms                                                                               |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name:         | AlertWatch:OR                                                                                                                                         |
| Applicant:                 | AlertWatch, LLC<br>1600 Huron Pkwy, Bldg. 520, Ste. 2326<br>Ann Arbor, MI 48109                                                                       |
| Contact:                   | Justin Adams, CEO<br>AlertWatch, LLC<br>Phone: 734-998-8344<br>Email: Justin.adams@alertwatch.com                                                     |
| Prepared by:               | Donna-Bea Tillman<br>Senior Consultant<br>Biologics Consulting Group, Inc.<br>Phone: 410-531-6542<br>Fax: 720-293-0014<br>Email: dtillman@bcg-usa.com |
| Date Prepared:             | November 18, 2015                                                                                                                                     |
| Classification Regulation: | 21 CFR 870.2300 - Cardiac monitor (including<br>cardiotachometer and rate alarm)                                                                      |
| Panel:                     | Cardiovascular                                                                                                                                        |
| Product Code:              | MWI                                                                                                                                                   |
| Predicate Device:          | K130401, AlertWatch: OR, AlertWatch, LLC                                                                                                              |

### Indication for Use:

AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management systems and patient medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision.

AlertWatch:OR is also intended for use by supervising anesthesiologists outside of operating rooms. Once alerted, the supervising anesthesiologist must contact the clinician inside the

{4}------------------------------------------------

operating room or must return to the operating room before making a clinical decision. Once either clinician is alerted, they must refer to the primary monitor or device before making a clinical decision.

## Device Description:

AlertWatch:OR is a display and secondary alert system used by the anesthesiology staff residents. CRNA's, and attending anesthesiologists - to monitor patients in operating rooms. The purpose of the program is to synthesize a wide range of patient data and inform clinicians of potential problems that might lead to immediate or long-term complications. Once alerted, the clinician is instructed to refer to the primary monitoring device before making a clinical decision. AlertWatch:OR should only be connected to AIMS systems and physiologic monitors that have been validated for use with AlertWatch:OR. AlertWatch, LLC performs the validation for each installation site.

The purpose of this 510(k) is for marketing clearance of AlertWatch:OR 2.50 which includes minor modifications to some display views, user features, indicators and alerts as well as compatibility with the iPad and the iPhone.

# Performance Data:

- . Human Factors Study: The iPhone version of AlertWatch:OR has a different user interface than the desktop product. To better understand any potential usability issues that could interfere with the intended use of the product. AlertWatch performed a comprehensive human factors study. The iPad version did not require a human factors study, because the layout and input methodology was the same as the predicate version (assuming a touch screen PC in the operating room).
- . Default Limits and Thresholds: AlertWatch, Inc. used a three-phased approach to ensure that the default limits were clinically valid:
	- 1. Review of References. AlertWatch sought out definitive published studies that highlighted appropriate limits for certain patient conditions. This includes limits and thresholds for the green/yellow/red organ schema, as well as the alerts and formulae.
	- 2. Expert Committee. When references were not available, AlertWatch sought out the opinion and confirmation of anesthesia physicians at the University of Michigan Health System. Each of these clinicians reviewed the limits, provided feedback, and reviewed the final results.
	- 3. External Experts. AlertWatch obtained final review of the default limits from an external group of four anesthesiology experts.
- IEC 60601-1-8: A summary of how AlertWatch:OR addresses the primary issues identified in IEC 60601-1-8 is provided in the submission.

{5}------------------------------------------------

# Substantial Equivalence:

Primary Predicate -The cleared AlertWatch:OR (K130401) is provided as a primary predicate device in terms of Indications for Use and technological characteristics.

Reference Device - Per the FDA Guidance, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued July 28, 2104, the cleared Amcom CommTech Messenger (K112047) is provided as a reference device to support the use of paging to include alerts that originate from AlertWatch:OR.

The table below compares the key technological features of the subject device to the primary predicate device (AlertWatch: OR K130401).

|                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                               | AlertWatch:OR (v2.48)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | AlertWatch:OR (v2.50)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 510(k) number                                             | K130401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K153335                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification                                            | MWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | MWI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation                                                | 870.2300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 870.2300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for Use                                       | AlertWatch:OR is intended for use by<br>clinicians for secondary monitoring of<br>patients within operating rooms.<br>AlertWatch:OR combines data from<br>networked physiologic monitors.<br>anesthesia information management<br>systems and patient medical records<br>and displays them in one place.<br>AlertWatch:OR can only be used with<br>both physiological monitors and AIMS<br>versions that have been validated by<br>AlertWatch. Once alerted, you must<br>refer to the primary monitor or device<br>before making a clinical decision. | AlertWatch:OR is intended for use by<br>clinicians for secondary monitoring of<br>patients within operating rooms.<br>AlertWatch:OR combines data from<br>networked physiologic monitors, anesthesia<br>information management systems and<br>patient medical records and displays them<br>in one place. AlertWatch:OR can only be<br>used with both physiological monitors and<br>AIMS versions that have been validated by<br>AlertWatch. Once alerted, you must refer to<br>the primary monitor or device before<br>making a clinical decision.<br>AlertWatch:OR is also intended for use by<br>supervising anesthesiologists outside of<br>operating rooms. Once alerted, the<br>supervising anesthesiologist must contact<br>the clinician inside the operating room or<br>must return to the operating room before<br>making a clinical decision. Once either<br>clinician is alerted, they must refer to the<br>primary monitor or device before making a<br>clinical decision. |
| Intended Use<br>environment                               | OR - referred to primary monitor for<br>clinical decision.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | OR - referred to primary monitor for<br>clinical decision.<br>The mobile AlertWatch application can be<br>installed on a personal iPhone or used via<br>safari on an iPad for use by supervising<br>anesthesiologists outside of operating<br>rooms, who are connected to the hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                           | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | network.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended Users                                            | Anesthesiologist, Resident, CRNA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Data server                                               | AlertWatch:OR accesses physiologic<br>data from the hospital network, and all<br>other data from the AIMS database<br>server.                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Supported AIMS<br>systems                                 | GE Centricity 7.63                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Centricity, version 7.6.3<br>Picis / Optum, version 421<br>iMDsoft / Metavision, version 5.46.44                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Supported physiological<br>monitors                       | GE Solar 9500<br>GE Carespan B850                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Hardware platforms<br>supported                           | PC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PC, iPad, iPhone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Generates clinical<br>advisories                          | AlertWatch:OR Alerts analyzes data<br>from patient monitors and other<br>sources and alerts when values exceed<br>preset limits.                                                                                                                                                                                                                                                                                                                                                                                                                      | AlertWatch:OR Alerts analyzes data from<br>patient monitors and other sources and<br>alerts when values exceed preset limits.<br>There are been some changes to the content<br>of these advisories (see Section 11.9 and<br>11.10)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Re-display of vital sign<br>data from primary<br>monitors | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended to replace<br>primary monitors                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Paging                                                    | Yes. For clinicians to page each other<br>using the hospital-established paging<br>system.                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes. For clinicians to page each other and<br>for AlertWatch:OR to transmit alerts using<br>the hospital-established paging system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Uses color to display<br>clinical information             | Image: Predicate device color display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Image: Proposed device color display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

# Substantial Equivalence Conclusion:

The subject AlertWatch:OR (v2.50) is modification of the previously cleared device (K130401) that adds several new features that do not alter the fundamental intended use of the device as a secondary monitor for patients in the operating room. The indications for use are identical to the predicate device.

{7}------------------------------------------------

The modified alerts and changes to the display of information, and the addition of an iPhone version, do not raise different questions of safety or effectiveness when compared to the primary predicate device. A Usability Study demonstrated that users were able to successfully complete critical tasks in a simulated use environment. Wireless coexistence testing demonstrated that the system could operate correctly in an environment with other wireless devices. Software verification testing demonstrated that the software met all requirements and performed as expected. Therefore, AlertWatch:OR (v2.50) can be found substantially equivalent to the previously cleared AlertWatch:OR (v2.48).

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K153335](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K153335)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
