K143304 · Medical Informatics Corp · MWI · May 26, 2015 · Cardiovascular
Device Facts
Record ID
K143304
Device Name
Sickbay Clinical Platform
Applicant
Medical Informatics Corp
Product Code
MWI · Cardiovascular
Decision Date
May 26, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Sickbay Clinical Platform is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Electronic Medical Record (EMR) and Clinical Information Systems. The Sickbay Clinical Platform is a remote monitoring platform that displays physiologic data, waveforms and alarms routed through the Sickbay Clinical Platform for supported devices. The Sickbay Clinical Platform is intended to be used by healthcare professionals in a hospital setting for the following purposes: - To remotely consult, regarding a patient's status . - To remotely review other standard or critical near real-time patient data, . waveforms and alarms in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner. WARNING: The Sickbay Clinical Platform is intended to supplement and not to re-place any part of the hospital's device monitoring. Do not rely on the Sickbay Clinical Platform product as the sole source of alarms.
Device Story
Sickbay Clinical Platform is a software-based remote monitoring system for hospital use. It ingests physiologic data, waveforms, and alarms from various bedside patient monitors via wired or wireless hospital networks. The platform routes this data to a central server, stores it in a proprietary database, and displays it on web-enabled devices for clinicians. It includes three applications: Patient Monitoring App, Patient Alarm Data Monitor, and Alarm Analytics Dashboard. Clinicians use the platform to remotely consult on patient status and review near real-time data to support clinical decisions. The device is intended to supplement, not replace, primary hospital monitoring systems. It provides a centralized view of patient data, potentially improving the timeliness of clinical care.
Clinical Evidence
No clinical or animal testing was submitted. Substantial equivalence was established through non-clinical bench testing, including usability (IEC 62366), software life-cycle (IEC 62304, AAMI TIR 45), human factors (AAMI/ANSI HE75), health informatics (IEEE 11073-10406), risk management (ISO 14971), and electrical safety/alarm standards (IEC 60601-1-8, IEC 60601-2-27).
Technological Characteristics
Software-based platform utilizing general-purpose computers, wired/WIFI networks, and disk-based data storage. Operates as a networked system within hospital infrastructure. Complies with IEC 62304, IEC 60601-1-8, IEC 60601-2-27, and IEEE 11073-10406. Architecture supports data routing, storage, and display via web-enabled devices.
Indications for Use
Indicated for ICU and ER patients with physiological sensors attached, for use by healthcare professionals in hospital settings to remotely monitor, consult, and review near real-time physiologic data, waveforms, and alarms to aid clinical decision-making.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{1}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2015
Medical Informatics Corp % David Makanani Consultant OMEDtech, LLC 1725 Signal Ridge Drive, Suite 150 Edmond, Oklahoma 73013
Re: K143304
Trade/Device Name: Sickbay™ Clinical Platform Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: April 13, 2015 Received: April 14, 2015
Dear David Makanani,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
# Indication for Use Statement
510(k) Number: K143304
Device Name: Sickbay™ Clinical Platform
Indications for Use:
The Sickbay Clinical Platform is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Electronic Medical Record (EMR) and Clinical Information Systems.
The Sickbay Clinical Platform is a remote monitoring platform that displays physiologic data, waveforms and alarms routed through the Sickbay Clinical Platform for supported devices. The Sickbay Clinical Platform is intended to be used by healthcare professionals in a hospital setting for the following purposes:
- To remotely consult, regarding a patient's status .
- To remotely review other standard or critical near real-time patient data, . waveforms and alarms in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner.
WARNING: The Sickbay Clinical Platform is intended to supplement and not to re-place any part of the hospital's device monitoring. Do not rely on the Sickbay Clinical Platform product as the sole source of alarms.
Note: Sickbay Clinical Platform product includes 3 Apps: Patient Monitoring App; Patient Alarm Data Monitor & Alarm Analytics Dashboard App.
Prescription Use: Yes _ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use: NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
510(k)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medical Informatics. The logo consists of a blue square with a stylized "m" inside, followed by the words "Medical" in blue and "Informatics" in gray. The "m" in the square is made up of several smaller squares, some of which are filled in with a lighter shade of blue.
# 510(K) SUMMARY
| Date | April 10th 2015 |
|-------------------|------------------------------------------------------------------------------------------------------|
| SUBMITTER | Medical Informatics Corp |
| | 6500 Main St. |
| | BioScience Research Collaborative Suite 1020G |
| | Houston TX 77030 |
| CONTACT PERSON | Ralph Krog JD, Chief Compliance Officer |
| | Ralph.Krog@MedicalInformaticsCorp.com |
| | 713-370-3904 |
| DEVICE NAME | Sickbay™ Clinical Platform |
| Classification | Device |
| Trade Name | Sickbay Clinical Platform |
| Common Name | Sickbay |
| Classification | Class II |
| Product Code | MWI, OUG |
| Review Panel | Part 870, |
| | Subpart B Cardiovascular Diagnostic Devices, |
| | Sec. 870.1425 Programmable diagnostic computer. |
| PREDICATE DEVICE: | COVIDIEN VITAL SYNC INFORMATICS MANAGER & VIRTUAL<br>PATIENT MONITORING PLATFORM - 510(k)'s: K132604 |
{5}------------------------------------------------
"The Vital Sync™ Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Electronic Medical Record (EMR) and Clinical Information System (CIS).
DEVICE DESCRIPTION: Medical Informatics Corp (MIC) has developed a software platform and analytics engine, Sickbay™, which gathers physiological data streams coming from patient monitoring devices and other clinical data sources. Medical Informatics Corp's software platform, Sickbay, captures patient data and enables near real-time display and analytics for clinicians to use in clinical decision support.
ENVIRONMENT OF USE: Sickbay Clinical Platform is intended to be used in a hospital environment.
### TECHNOLOGICAL CHARACTERISTIC AND SUBSTANTIAL EQUIVALENCE:
The following table provides more detailed information regarding the basis for the determination of substantial equivalence:
| Parameter | Vital Sync | Sickbay Platform |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Vital Sync Informatics<br>Manager is software that is<br>intended to route and store<br>medical device data and<br>device diagnostic information<br>from supported devices to<br>the Electronic Medical Record<br>(EMR) and Clinical<br>Information System. | The Sickbay Clinical Platform<br>is software that is intended to<br>route and store medical<br>device data and device<br>diagnostic information from<br>supported devices to the<br>Electronic Medical Record<br>(EMR) and Clinical<br>Information Systems. |
| | The Vital Sync Virtual Patient<br>Monitoring Platform is a<br>remote monitoring platform<br>that displays physiologic<br>data, waveforms and alarms<br>routed through the Vital Sync<br>Informatics Manager for<br>supported devices. The Vital<br>Sync Virtual Patient | The Sickbay Clinical Platform<br>is a remote monitoring<br>platform that displays<br>physiologic data, waveforms<br>and alarms routed through<br>the Sickbay Clinical Platform<br>for supported devices. The |
| Parameter | Vital Sync | Sickbay Platform |
| | Monitoring Platform is<br>intended to be used by<br>healthcare professionals for<br>the following purposes:<br>- To remotely consult,<br>regarding a patient's status<br>- To remotely review other<br>standard or critical near real-<br>time patient data, waveforms<br>and alarms in order to utilize<br>this information to aid in<br>clinical decisions and deliver<br>patient care in a timely<br>manner.<br>WARNING: The Vital Sync™<br>Informatics Manager &<br>Virtual Patient Monitoring<br>Platform are intended to<br>supplement and not to<br>replace any part of the<br>hospital's device monitoring.<br>Do not rely on the Vital Sync™<br>Informatics Manager &<br>Virtual Patient Monitoring<br>Platform as the sole source of<br>alarms. | Sickbay Clinical Platform is<br>intended to be used by<br>healthcare professionals in a<br>hospital setting for the<br>following purposes:<br>- To remotely consult,<br>regarding a patient's status<br>- To remotely review other<br>standard or critical near real-<br>time patient data, waveforms<br>and alarms in order to utilize<br>this information to aid in<br>clinical decisions and deliver<br>patient care in a timely<br>manner.<br>WARNING: The Sickbay<br>Clinical Platform is intended<br>to supplement and not to<br>replace any part of the<br>hospital's device monitoring.<br>Do not rely on the Sickbay<br>Clinical Platform as the sole<br>source of alarms.<br>Note: Sickbay Clinical<br>Platform product includes 3<br>Apps: Patient Monitoring<br>App; Patient Alarm Data<br>Monitor & Alarm Analytics<br>Dashboard App. |
| Functionality | Display patient monitor data<br>remotely | Display patient monitor data<br>remotely and store it for<br>analysis. |
| Device Regulatory<br>Classification | 21 C.F.R.<br>Part 870, Subpart B<br>Cardiovascular Diagnostic | 21 C.F.R.<br>Part 870, Subpart B<br>Cardiovascular Diagnostic |
| Parameter | Vital Sync | Sickbay Platform |
| | Devices,<br>Sec. 870.1425<br>Programmable diagnostic<br>computer. | Devices,<br>Sec. 870.1425<br>Programmable diagnostic<br>computer. |
| Product Code | MWI & OUG | MWI & OUG |
| Device Class | II | II |
| 510(k) number | K132604 | K143304 |
| Technology | Standard computers,<br>network and WIFI technology | Standard computers,<br>network and WIFI technology |
| Data Storage | Hour to days of data is<br>buffered and then archived<br>on tape. | All physiological data is<br>stored on disk drives forever,<br>until deleted. |
| Wireless | Yes: handheld display devices | Yes: handheld display devices |
| Direct Sensor Contact | Covidien sensors supply data<br>but are not part of this<br>device. | Various hospital Monitors<br>and sensors provide data but<br>are not part of our device. |
| Optimum Range | Within the Hospital's secure<br>WIFI network. | Within the Hospital's secure<br>WIFI network. |
| Wireless vs Wired<br>Function | WIFI from sensor to server,<br>wired and WIFI out to<br>handheld displays. | Both WIFI and wired from<br>sensors to server and both<br>WIFI and wired out. |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Medical Informatics. The logo consists of a blue square with a stylized "mi" inside, followed by the words "Medical" in blue and "Informatics" in gray. The logo is simple and modern, and it is likely used to represent a company or organization that specializes in medical informatics.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for Medical Informatics. The logo consists of a stylized letter "m" inside a rounded square shape on the left, followed by the words "Medical" and "Informatics" on the right. The word "Medical" is in a larger, bolder font than "Informatics". The letter "m" and the word "Medical" are in a teal color, while the word "Informatics" is in a gray color.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medical Informatics. The logo consists of a blue square with rounded corners, inside of which are the letters "m" and "i" in white and gray. To the right of the square is the word "Medical" in blue, and below that is the word "Informatics" in gray.
| Parameter | Vital Sync | Sickbay Platform |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software Main<br>functionality | Patient information is<br>transmitted wirelessly from<br>bedside devices to the<br>hospital's server<br>The Vital Sync software,<br>which resides on a hospital's<br>server, takes this information<br>and makes it viewable on any<br>web-enabled device on the<br>hospital network<br>The information is formatted<br>into HL7 protocol and<br>written into the hospital EMR | Patient information is<br>transmitted wired or<br>wirelessly from bedside<br>devices to the hospital's<br>network<br>The Sickbay Clinical Platform,<br>which resides on a hospital's<br>servers, takes this<br>information stores it and<br>makes it viewable on any<br>web-enabled device on the<br>hospital network<br>Information from EMR (Lab,<br>Meds and ADT records in<br>HL7 format) is stored in<br>Sickbay and available for<br>redisplay. |
| Target population/<br>demographics | ICU and ER patients that have<br>Covidien physiological<br>sensors attached. | ICU and ER patients that have<br>any physiological sensors<br>attached. |
### WHY THE DIFFERENCES ARE NOT CRITICAL TO THE INTENDED USE OF THE DEVICE
Covidien has two devices (Vital Sync Informatics Manager & the Vital Sync Virtual Patient Monitoring Platform) performing two functions (route and store medical device data AND re-display physiologic data, waveforms and alarms). MIC has one device (Sickbay Clinical Platform) that performs the same two functions (route and store medical device data AND re-display physiologic data, waveforms and alarms). This difference has no effect on how the devices are used. Indeed both Vital Sync and Sickbay perform the same two general functions.
The Vital Sync Informatics Manager routes and stores data from supported devices to the Electronic Medical Record (EMR) and Clinical Information System. The Sickbay Clinical Platform routes and stores data from supported devices and from the Electronic Medical Record (EMR) and Clinical Information Systems. Covidien's Vital Sync is gathering data from Covidien bedside sensors and storing it in the Hospital's EMR. MIC's Sickbay gathers data from all sensors on the bedside from the hospital network (before they get to the EMR)
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medical Informatics. The logo consists of a blue square with a stylized "m" inside, followed by the words "Medical" in blue and "Informatics" in gray. The font is sans-serif and modern.
and stores data in its own Sickbay data base and Sickbay data repository or file set. This difference has no effect on how the devices are used; both devices gather and store medical data from other medical device sensors or monitors and re-display such data on supported devices. The set of medical sensors may differ, but the function and use of the two devices (Vital Sync and Sickbay) are the same.
# WHY THE DIFFERENCES DO NOT AFFECT THE SAFETY AND EFFECTIVENESS OF THE DEVICE WHEN USED AS LABELED.
Covidien has two products performing two functions and MIC has one product performing the same two functions. This difference in structure or packaging has no effect on safety and effectiveness. The hazards are the same.
The difference in data storage (an EMR database versus Sickbay database) do not affect the safety and effectiveness of the product, because the reliability of data storage and retrieval technology is the same for both products. Sickbay's database does not replace or preclude use of an EMR.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The technological characteristics of the two devices are the same. Both rely on general purpose computers, wired and wireless computer networks, and data storage technology. This technology is several generations old, well established, and its risks and mitigations are well understood. The only technological difference between Vital Sync and Sickbay is one of volume not technology. Sickbay uses more data storage because the data can be analyzed offline as well as redisplayed.
### PERFORMANCE TESTING - (NON-CLINICAL) BENCH
Sickbay Clinical Platform has been determined through engineering bench testing to support substantial equivalence with this device and the predicates.
| Performance Test | Applicable<br>Standard | Results |
|---------------------|--------------------------|----------|
| Usability | IEC 62366 | Complies |
| Software Life-Cycle | IEC 62304<br>AAMI TIR 45 | Complies |
| AAMI/ANSI HE75 | Human Factors | Complies |
| Health Informatics | IEEE 11073-10406 | Complies |
### TESTS CONDUCTED:
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Medical Informatics. The logo consists of a blue square with a stylized "m" inside, followed by the words "Medical" in blue and "Informatics" in gray. The "m" in the square is made up of three vertical bars, with the middle bar being a lighter shade of blue.
| Performance Test | Applicable<br>Standard | Results |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------|
| Risk Management | ISO 14971 | Complies |
| Medical Electrical Equipment – Collateral<br>Standard: Gen. Requirements, Tests and<br>Guidance for Alarm systems in Medical<br>Electrical Equipment and Medical Electrical<br>system | IEC 60601-1-8 | Complies |
| Medical Electrical Equipment - Particular<br>Requirements for the Basic Safety and Essential<br>Performance of Electrocardiographic<br>Monitoring Equipment | IEC 60601-2-27 | Complies |
| | | |
MIC referred to the following documents:
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm08954 3.htm
General Principles of Software Validation; Final Guidance for Industry and FDA Staff http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm085281. htm
Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm07377 8.htm
Draft Guidance for Industry and Food and Drug Administration Staff - Applying Human Factors and Usability Engineering to Optimize Medical Device Design http://www.fda.gov/RegulatoryInformation/Guidances/ucm259748.htm
Content of Premarket Submissions for Management of Cybersecurity in Medical Devices - Guidance for Industry and Food and Drug Administration Staff http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidanceDocuments/UCM35619 0.pdf
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Medical Informatics. The logo consists of a blue square with a stylized "m" and "i" inside, followed by the words "Medical" in blue and "INFORMATICS" in gray. The logo is simple and modern, and it is likely used to represent a company or organization that provides medical informatics services.
Guidance for Industry - Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm07781 2.htm
Guidance for Industry and FDA Staff - Addition of URLs to Electronic Product Labeling http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm22855 3.htm
This testing showed the Sickbay to meet applicable ISO, IEC and FDA safety and performance standards. And guidances.
### PERFORMANCE TESTING - ANIMAL/CLINICAL
There has been no animal/clinical testing submitted with this Notification.
### CONCLUSION:
Based on the results of non-clinical testing, the Sickbay performs safely, as intended, and the comparative discussion of intended use, principle of operation, and technological characteristics, it is determined by MIC that the Sickbay is substantially equivalent to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.