← Product Code [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI) · K113183

# PMS8210A (IRIS) MULTI-PARAMETER PATIENT MONITOR, MODEL CODE 500 (K113183)

_Shanghai 3f Electronics Co, Ltd. · MWI · Feb 9, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K113183

## Device Facts

- **Applicant:** Shanghai 3f Electronics Co, Ltd.
- **Product Code:** [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI.md)
- **Decision Date:** Feb 9, 2012
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

PMS8210A Model Code 500 is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Model Code 500 has certain features and functions. The patient parameters that can be monitored by PMS8210A Model Code 500 are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2), Temperature (TEMP) and End-tidal CO2 (EtCO2) . Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs. The PMS8210A Model Code 500 is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment. The PMS8210A Model Code 500 is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.

## Device Story

Battery or line-powered multi-parameter patient monitor; acquires physiological signals (ECG, RESP, NIBP, SpO2, TEMP, EtCO2) via sensors; converts signals to digital data; processes data for alarm conditions; displays real-time physiological data. Used in hospital clinical areas (ICU, CCU, OR, emergency department) by qualified medical/nursing staff. Provides audible/visual alarms for parameter excursions to notify clinicians of patient status changes; aids clinical decision-making by providing continuous physiological monitoring. Benefits include real-time patient status awareness.

## Clinical Evidence

Bench testing only. Device underwent third-party safety and performance testing in accordance with IEC standards, including environmental, vibration, shock, and water ingress testing. Software of moderate concern evaluated.

## Technological Characteristics

Battery or line-powered; multi-parameter monitoring; ECG (3/5-lead), NIBP, SpO2, TEMP, EtCO2. Standards: IEC 60601-1-2, ISO 81060-2:2009 (NIBP), ISO 9919:2005 (SpO2), ASTM E1112:2000/E1104-03 (TEMP). High voltage isolation (4000V), anti-defibrillation protection.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- PMS8210A (Iris) Patient Monitor/Multi-Parameter Patient Monitor ([K100394](/device/K100394.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. On the left side, the text reads 'K113183'. To the right of this, there is a fraction written as '1/4'. The handwriting is somewhat stylized, with variations in stroke thickness and letter formation.

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

ـ ﺃ- .

Date: October 14, 2011

1. Company and Correspondent making the submission:

Name - Shanghai 3F Electronics Co., Ltd. Address – 77325 Joyce Way Echo, Oregon 97826 Telephone - 931-625-4938 Fax - 541-376-5063 Contact - Charles Mack Email - charliemack@irc-us.com

# 2. Device :

Trade/proprietary name: PMS8210A (IRIS) Multi-Parameter Patient Monitor Model Code 500

Common Name : Multi-parameter Patient Monitor

Classification of the device: Class II

Panels: Cardiovascular, General Hospital, Anesthesiology 21CFR870.2300, Monitor, Physiological, MWI Product code:

Establishment Registration Number : 3008383116

{1}------------------------------------------------

K113183 2/14

Predicate Devices:

| Predicate Model                                                           | Manufacturer                            | K Number | Submitted Device                                                                         |
|---------------------------------------------------------------------------|-----------------------------------------|----------|------------------------------------------------------------------------------------------|
| PMS8210A (Iris)<br>Patient Monitor/<br>Multi-Parameter<br>Patient Monitor | Shanghai 3F<br>Electronics Co.,<br>Ltd. | K100394  | PMS8210A Model Code<br>500 (Iris) Patient<br>Monitor/ Multi-Parameter<br>Patient Monitor |

# 3. Description :

- 3.1 General
PMS8210A Model Code 500 Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), Non-Invasive blood pressure (NIBP), Saturation of pulse oxygen (SPO2), Temperature (TEMP), End-tidal (etCO₂). The signals are converted into digital data and processed, examines the data for alarm conditions and displays the data. The monitor also provides operating control for the user. The submitted device is the same as the predicate with two differences. The new model has added an etCO2 function and the Temp module has been changed. The new Temp module is a previously FDA cleared device.

The patient monitor is intended to be used in a hospital clinical area such as intensive care units, cardiac care units, operation room, emergency department, to provide additional information to the medical and nursing staff about the physiological condition of the patient. The PMS8210A Model Code 500 patient monitor is intended to be used only under regular supervision of clinical personnel. The intended location of use is clinics.

{2}------------------------------------------------

- 4. Indication for use :
PMS8210A Model Code 500 is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Model Code 500 has certain features and functions.

The patient parameters that can be monitored by PMS8210A Model Code 500 are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2), Temperature (TEMP) and End-tidal CO2 (EtCO2) . Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.

The PMS8210A Model Code 500 is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.

The PMS8210A Model Code 500 is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.

- 5. Comparison with predicate device: Please see next page for the comparison table.

{3}------------------------------------------------

osive ອ້າງວ່ານອາ9 ot nozinequo 10 ອາໄດ້ຮັ

The state of the first the first the

(ໄຣວເາງ່ວອໄອ)ໄຣວເຮຣ(dq .ອ.i) ຂໍ້ແດງໄຊວເປີຣິເວທີໄວ້ອອຸຣິ ໄຣາອົກອ

| របរុវជន្ម<br>(elqsiC<br>oma<br>Optional printe<br>sən qalınmışdır. Bu yarın və bir və bir və bir mən<br>ipisips le dimension/weig<br>ອານ ໄວ ຣາ ແນຣ<br>they of themging indicato<br>lıqdan sənədlər   Rəmaniya   Rəsmin  Film   Alanmada   Alanmada   Alanmada  Şamar   Alanmada  Şamar   Alanmada  Şamar   Şamar   Şamar   Şamar   Şamar   Şamar   Şamar   Şamar<br>onos rewod lemet<br>əqmədə (qalınmışdır. Bu mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının<br>oom bus əm<br>sonalisme annesthetics<br>oitibuos apsuo<br>oilibuos pultisted<br>oitsiretoere<br>ttem on open time,<br>əsindən  filmləri   fəsiləsinin  cinsinə aid bitki növü. İstinadlar   Respublikası   Respublikası   Respublikası   Almaniya   Alanmar   Alanmar   Alanmar   Alan  Alan  Alan<br>amas<br>കന്നുടെ<br>കന്നും പ്രവരിച്ചു. അവലംബം  പ്രശസ്തമായി വിശ്വാസത്തിനു വിവ്യക്ഷേത്രം  അവലംബം  കേരളത്തിനു  അവലംബം  കേരളത്തിനു  അവലംബം  കേരളത്തിനു  അവലംബം  കേരളത്തിനു<br>ക്ഷേത്രട്ട<br>emas<br>ക്ഷേത്രട്ട<br>AIN<br>emas<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>ക്ഷേത്രം<br>emble<br>ല്പല്ല<br>ക്ക്<br>səy<br>emes<br>ടക)<br>sa)<br>the spinsit Stephronics Co., Ltd<br>pived toeidu<br>otinoM<br>əpoli<br>ביני המונחים המועד המועד Parameraq-iilluM ADLS82MS<br>નદ દિવેદુનાદનન<br>������������������������������������������������������������������������������������������������������������������������������������������������������������������������������<br>ટક્ષે<br>onod & .xo1qqs : dAm0022<br>ടല്<br>suoisuəmii<br>səy<br>Copyright specifically<br>sys = front page<br>px0.2 : tripisV<br>ടല്‍)<br>əd 32 bəmisl<br>EC 60601-1-2<br>ot suiteble<br>strong 0 .xondqs : dAm00A<br>ələr   Surəsinə  Qurğusu  Qaraqların  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qaraqları  Qar<br>ີ 40 ປີ 7°SE) ວິທີ ປີ 2°04 ວ່າ 2°04 ວ່າ 2°04 ອ້າງອີງ<br>າງການໄວວ "Z.E + 口ヨコ ກັບອົບມົນອຣ<br>I  ខេត្តស្វាប់  រាជ្ជម<br>ələrin əvvəl əsasən və qalınmışdır. Bu mənist<br>əsinin  cinsinə aid bitki növü. İstinadlar   Şəhrin Şəhrin  fəsiləsinin  fəsiləsinin  fəsiləsinin  cinsinə aid bitki növü. İstinadlar   Azərbaycan  Qaranların  Qaranların  Q<br>alsusbuos-non)& 262 2017/01/21111111 avitele<br>(%0レモ)つみんのみる~のけ<br>(ZHEF)ZH09/<br>E<br>um) (ロ) 08 ×(H) 08 ×(M) 08 ×<br>ddus rewod lentelni bus<br>10002:<br>מועד מענט שעולם   2000 מעט 44000 מע<br>instablico-non) of consist<br>AVSA,<br>lər<br>ectronics Co., Ltd<br>ุ่นในปี 2017-04-1 วิที่อิส of 2004<br>on (ਮ)015 ലാഗ്ര<br>ਜਾਂ ਦੀ<br>רכם (880×780 | ---------------- |                                                                       | ยให้ กิโลก อยู่ อย่าง (1881) อามรรจา9 bool8 องเรียงประเทศ                                                                                                                                                                                                              |                                                                                                                                                                                                                         |                                                                                                                                                                                            |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        | əsinə Qaraq və qalında və qalında və qalında qalınmışdır. Bu və qalınmış qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və bir və bir və bir və b                                           | də bəyində qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və bir mənist<br>വ (വ)വിട്ട കാലശല                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | apom pue awe                                                                                                                                                                                                                                                           | Autorial Analysis Android Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana Ana<br>າ ອ ໄອ<br>insite 9<br>, olinoM<br>əpov | AOT 288MA<br>និទ្ធិភាព របស់ពេល ពិធីមានទី 10 ទំនាយនាគ១-រដ្ឋានព្រះពុទ្ធសាសនា នៅក្រោយ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រាសាទ និង ប្រ |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | elupowi 48/1                                                                                                                                                                                                                                                           | emes                                                                                                                                                                                                                    | явий нээтийг                                                                                                                                                                               |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | boyları                                                                                                                                                                                                                                                                | small                                                                                                                                                                                                                   | internalliast                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | adisant the programment                                                                                                                                                                                                                                                | ames                                                                                                                                                                                                                    | served the personal patien                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | ອານຣຣອພ ໄດ້ ຄົນ                                                                                                                                                                                                                                                        | ames                                                                                                                                                                                                                    | ലപ്പുഴു റിപ്പ്സ്                                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | עוון: 0 ~ 290mmHp                                                                                                                                                                          |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | ວ່າ ວາ ຣາ ຣັ   – ອາ ກຣຣອາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ ສາ<br>əməlində məsənə qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və bir mənist | amas                                                                                                                                                                                                                    | Hum001 ~ 04 : 100mmm                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | ationalistics 40 - 130mm                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | עול: 20 ~ 200mmHg                                                                                                                                                                          |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | າອາກອານອະນາ                                                                                                                                                                                                                                                            | amas                                                                                                                                                                                                                    |                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | ə qalınmış və qalınmışdır. Bu mənin mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasın              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | Suit: 26 ~ 220mmHg                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | າ                                                                                                                                                                                                                                                                      | ames                                                                                                                                                                                                                    | sublic. 26 ~ 138mms                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | Humori ~ 92 : ateuce                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | ้นนั้น                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | କ୍ଷ<br>്രഖ്യാഭാ                                                                                                                                                                                                                                                        | emba                                                                                                                                                                                                                    | turn of ney, reteen on green than 8 minute                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                         | SOOS: 01 92 IMAANSIAA                                                                                                                                                                      |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                  |                                                                       | ərin  fəsiləsinə  cinsinə aid bitki növü. İstinadlar   Respublikasının  Filmin  Filmin  Filmin  Sənədlər   Respublikasının  Filmin  Sənədlər   Respublikasının  Filmin  Filmi                                                                                          | emas                                                                                                                                                                                                                    | A-090 IN                                                                                                                                                                                   |
|…

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K113183](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K113183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
