← Product Code [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI) · K041376

# C1 SERIES PATIENT MONITOR, MODELS 863051, 863052, 863053, 863054; OPTIONAL PRINTER FOR CI, MODEL M3925A (K041376)

_Analogic Corp. · MWI · Jul 20, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K041376

## Device Facts

- **Applicant:** Analogic Corp.
- **Product Code:** [MWI](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI.md)
- **Decision Date:** Jul 20, 2004
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The purpose and function of the C1 Series Patient Monitor is to monitor ECG, heart rate, non-invasive blood pressure (NBP), functional arterial oxygen saturation (SpO2) and temperature for single patient use on adult and pediatric patients in hospital areas and hospital-type facilities, such as clinics.

## Device Story

C1 Series Patient Monitor is a compact, battery-powered device for measuring ECG, SpO2, NIBP, and temperature; used in hospitals and clinics. Inputs include 3-lead ECG, light transmissivity for SpO2, thermistor probes for temperature, and oscillometric pressure cuffs for NIBP. Device processes signals via internal microprocessors and firmware; displays real-time waveforms and vital signs on an LCD; optional thermal printer provides hard-copy trends. Clinicians use output to assess patient status and guide clinical decisions. Design leverages technology from NPB 3900 (A1) and Welch Allyn SureTemp; incorporates SunTech step-bleed algorithm for NIBP. Benefits include rapid temperature measurement and continuous physiological monitoring.

## Clinical Evidence

Bench testing only. No clinical data provided. Validation performed at unit and system levels against requirements specifications. Compliance verified via electromagnetic compatibility (IEC 60601-1-2) and safety/performance standards including IEC 60601-1, IEC 60601-2-27 (ECG), IEC 60601-2-30 (NIBP), IEC 60601-2-49 (multifunction monitors), IEC 60601-1-8 (alarms), ANSI/AAMI EC-13, and ANSI/AAMI SP10.

## Technological Characteristics

Compact, battery-powered (sealed lead-acid) monitor. Measures ECG (3-lead), SpO2 (dual-wavelength), NIBP (oscillometric), and temperature (thermistor). Features LCD display and optional 50-mm thermal strip chart recorder. Software-based processing using embedded firmware. Complies with IEC 60601-1, 60601-1-2, 60601-2-27, 60601-2-30, 60601-2-49, 60601-1-8, ANSI/AAMI EC-13, and ANSI/AAMI SP10.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- NPB 3900 (A1) Patient Monitor ([K980411](/device/K980411.md))

## Reference Devices

- Welch Allyn SureTemp Portable Thermometer ([K943695](/device/K943695.md), [K964643](/device/K964643.md))
- Analogic C3 Patient Monitor ([K030931](/device/K030931.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K041376
p 1/4

510(k) Summary For Analogic Corporation C1 Series Patient Monitor

#### DATE THIS SUMMARY WAS PREPARED:

May 21, 2004

#### SUBMITTER'S NAME AND ADDRESS:

Analogic Corporation 8 Centennial Drive Peabody, MA 01960

## CONTACT PERSON:

Donald J Sherratt, Regulatory Affairs Manager Telephone (978) 977-3000 extension 3049 Facsimile (978) 977-6808

## DEVICE NAME:

Proprietary or Trade Name: C1 Series Patient Monitor Common Name: Multi-Function Patient Monitor Classification Name: Physiological Patient Monitoring System and Accessories

## PREDICATE DEVICE:

The legally marketed device to which equivalence is being claimed is: The NPB 3900 (A1) Patient Monitor that was cleared under Premarket Notification K980411

{1}------------------------------------------------

## DEVICE DESCRIPTION:

The C1 Series Patient Monitor is a compact, lightweight monitor for measuring, processing, storing, and displaying information derived from five physiological measurements:

- Electrocardiogram (ECG). A three lead ECG is acquired and a waveform can be . displayed real-time on the LCD screen or permanently recorded on the optical strip chart recorder. The design of the ECG function is derived directly from the predicate device, the NPB 3900 (A1).
- Pulse Oximetry (SpO2). Functional Oxygen Saturation is calculated from the ratio of . light transmissivity through the capillary bed at two wavelengths. The SpO2 subsystem uses the same printed wiring board layout as well as the same microprocessor and imbedded firmware that is used in the predicate device, the NPB 3900 (A1).
- The temperature is measured using thermistor probes and a "predictive" algorithm to . enable temperature measurements to be completed more quickly. The temperature subsystem uses technology licensed from the manufactures of the Welch Allyn SureTemp® Portable Thermometer, which was cleared for marketing under premarket notification K943695 and K964643.
- . Blood pressure is measured non-invasively (NIBP) by the oscillometric method. The design of the NIBP subsystem is an enhanced version of the NIBP subsystem used in the predicate device, the NPB 3900 (A1) and cleared for use in the C3 Patient Monitor cleared under K030931.
- . An external optional thermal printer records waveforms, digital vital signs, and tabular trends on a 50-mm wide strip chart.
- . The C1 Series Patient Monitor is powered by internal sealed lead-acid batteries. A fully charged battery will power the monitor for 2.5 hours.

## INTENDED USE:

The purpose and function of the C1 Series Patient Monitor is to monitor ECG, heart rate, non-invasive blood pressure (NBP), functional arterial oxygen saturation (SpO2) and temperature for single patient use on adult and pediatric patients in hospital areas and hospital-type facilities, such as clinics.

## COMPARISION OF TECHNOLIGICAL CHARACTERISTICS:

The design of the C1 Series Patient Monitor is derived from the design of the NPB 3900 (A1) Patient Monitor. Clinically the two devices are nearly identical.

The C1 Series Patient Monitor uses the same base technology as the predicate and incorporates a step-bleed algorithm for NIBP licensed from SunTech that has been cleared by FDA for use in the Analogic C3 patient monitor (K030931). Software changes were made to allow the use of a color display and update the alarms to bring them in line with current technology and international standards.

The materials of construction, physical size, weight and footprint are the same as predicate.

{2}------------------------------------------------

# NONCLINICAL TESTS USED IN DETERMINATION OF SUBSTANIAL EQUIVALENCE:

The design of the C1 Patient Monitor has been thoroughly validated at the unit and system level and meets all element of its Requirements Specification. The C1 patient monitor will comply with the following recognized standards:

- Electromagnetic Compatibility Tests to determine if it was in compliance with IEC . 60601-1-2: 2001 - Group 1, and Class B emissions limits.
	- CISPR 11 Industrial, scientific and medical (ISM) radio-frequency equipment o Radio disturbance characteristics - Limits and methods of measurement.
	- IEC 61000-3-2 Electromagnetic compatibility (EMC) Part 3-2: Limits o Limits for harmonic current emissions (equipment input current <= 16A per phase)
	- IEC 61000-3-3 Electromagnetic compatibility (EMC) Part 3-3: Limits -O Limitation of voltage changes, voltage fluctuations and flicker in public lowvoltage supply systems, for equipment with rated current <= 16A per phase and not subject to conditional connection
	- IEC 61000-4-2 Electromagnetic compatibility (EMC) Part 4-2: Testing and O measurement techniques - Electrostatic discharge immunity test
	- IEC 61000-4-3 Electromagnetic compatibility (EMC) Part 4-3: Testing and o measurement techniques -- Radiated, radio frequency, electromagnetic field immunity test
	- IEC 61000-4-4 Electromagnetic compatibility (EMC) -- Part 4-4: Testing and o measurement techniques -- Section 4: Electrical fast transient/burst immunity test. Basic EMC Publication
	- IEC 61000-4-5 Electromagnetic compatibility (EMC) Part 4-5: Testing and o measurement techniques - Surge immunity test
	- IEC 61000-4-6 Electromagnetic compatibility (EMC) Part 4-6: Testing and O measurement techniques - Immunity to conducted disturbances, induced by radiofrequency fields
	- IEC 61000-4-8 Electromagnetic compatibility (EMC) Part 4-8: Testing and o measurement techniques – Power frequency magnetic field immunity test
- IEC 61000-4-11 Electromagnetic compatibility (EMC) Part 4-11: Testing and 0 measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K041376 p. 4/4". The text appears to be handwritten. The number "K041376" is on the first line, and "p. 4/4" is on the second line. 

- . Safety and Performance Tests
	- IEC 60601-1:1988 +A1:1991 +A2:1995, Medical Electrical Equipment-O Part 1: General Requirements for Safety
	- ECG Performance Testing According to ANSVAAMI EC-13 Cardiac o Monitors, Heart Rate Meters, and Alarms
	- ANSI/AAMI SP10 Manual, Electronic, or Automated o Sphygmomanometers
	- IEC 60601-2-30 -- Medical electrical equipment -- Part 2-30: particular 0 requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment.
	- IEC 60601-2-27 Medical electrical equipment-- Part 2: Particular O requirements for the safety of electrocardiographic monitoring equipment
	- IEC 60601-2-49 -- -- Medical electrical equipment -- Part 2-49: Particular O requirements for the safety of multifunction patient monitoring equipment
	- IEC 60601-1-8 Medical electrical equipment Part 1-8: General 0 requirements for safety - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

# CONCLUSIONS FROM NONCLINCAL TESTING

The testing of the C1 Patient Monitor demonstrates that the performance is substantially equivalent to the predicate device cited above.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles a stylized eagle or bird with three curved lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 0 2004

Mr. Donald J. Sherratt Regulatory Affairs Manager Analogic Corporation 8 Centennial Drive Centennial Industrial Park Peabody, MA 01960

Re: K041376

Trade Name: C1 Series Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor (without arrhythmia alarms) Regulatory Class: II (two) Product Code: MWI Dated: May 21, 2004 Received: May 24, 2004

Dear Mr. Sherratt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 – Mr. Donald J. Sherratt

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r loase be deviced that i bring that your device complies with other requirements of the Act that I Dr I has Intatutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF R Fart 6075, its ens (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. production of the realling your device as described in your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicated device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you deem's office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Neil R.P. Ogden for

Bram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "ANALOGIC" in bold, black letters. There is a registered trademark symbol to the right of the "G". To the right of the word is a black square.

# Indications for Use Form

K041376 510(k) Number (if known):

Device Name: C1 Series Patient Monitor

Indications For Use:

The C1 Series Patient Monitor series are multi-parameter patient monitors that are used to monitor ECG waveforms, heart rate, noninvasive blood pressure (systolic, dialstolic), functional atterial oxygen saturation and temperature for adult and pediatric patients in all hospital areas and health care facilities.

|                                    | Nil R. Okun |
|------------------------------------|-------------|
|                                    |             |
| (Division Sign-Off)                |             |
| Division of Cardiovascular Devices |             |
| 510(k) Number                      | K041376     |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |    |                              |
|--------------------------------------------------------|----|------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109)               | OR | Over-The-Counter Use _______ |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K041376](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MWI/K041376)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
