← Product Code [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD) · K190270

# FORE-SIGHT ELITE Tissue Oximeter Module (K190270)

_Cas Medical Systems, Inc. · MUD · Oct 21, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K190270

## Device Facts

- **Applicant:** Cas Medical Systems, Inc.
- **Product Code:** [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD.md)
- **Decision Date:** Oct 21, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The noninvasive FORE-SIGHT ELITE Tissue Oximeter "Module" is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The Module is also intended for use as an adjunct monitor of relative changes of oxygenated hemoglobin, deoxygenated hemoglobin, and their summation, total hemoglobin, of blood under the sensors. The Module is intended to allow for the display of StO2 and regional hemoglobin on a third party display/patient monitor and is indicated for use as follows: When used with large sensors, the Module is indicated for use on adults and transitional adolescents ≥40 kg. When used with medium sensors, the Module is indicated for use on pediatric subjects ≥3 kg. When used with small sensors, the Module is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.

## Device Story

Device is a tissue oximeter module for continuous, noninvasive monitoring of regional hemoglobin oxygen saturation (StO2) and relative changes in oxygenated/deoxygenated/total hemoglobin. Inputs: light absorption signals (660-900 nm) from sensors placed on tissue. Operation: multi-distance diffuse reflectance spectroscopy; module controls LED currents, processes light signals, and calculates hemoglobin concentrations. Output: StO2 and hemoglobin concentration data displayed on third-party host monitors via USB/serial interfaces. Used in clinical settings (e.g., cardiac surgery) by clinicians to assess tissue oxygenation; provides real-time data to assist in identifying ischemic states and guiding clinical interventions. Benefits: enables continuous, noninvasive monitoring of regional perfusion, potentially improving patient management during procedures.

## Clinical Evidence

Clinical validation included healthy adult volunteers and pediatric cardiac surgical patients. StO2 precision validated against co-oximetry (arterial/venous blood samples). Relative hemoglobin changes validated against co-oximetry and frequency domain oximetry on liquid optical phantoms. Studies confirmed high precision and equivalence to predicate/reference devices across specified weight ranges.

## Technological Characteristics

Multi-distance diffuse reflectance spectroscopy; LED light source (660-900 nm); host-powered module; USB/serial connectivity; multi-channel capability. Software-based signal processing for StO2 and hemoglobin concentration calculations. No specific material standards or sterilization methods provided.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- FORE-SIGHT ELITE Tissue Oximeter Module ([K180003](/device/K180003.md))

## Reference Devices

- Hamamatsu Photonics Niro-200NX ([K143219](/device/K143219.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 21, 2019

CAS Medical Systems, Inc. Ron Jeffrey Director, Regulatory Affairs 44 East Industrial Road Branford, Connecticut 06405

Re: K190270

Trade/Device Name: FORE-SIGHT ELITE Tissue Oximeter Module Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: February 6, 2019 Received: February 8, 2019

Dear Ron Jeffrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K190270

Device Name

FORE-SIGHT ELITE Tissue Oximeter Module

#### Indications for Use (Describe)

The noninvasive FORE-SIGHT ELITE Tissue Oximeter "Module" is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The Module is also intended for use as an adjunct monitor of relative changes of oxygenated hemoglobin, deoxygenated hemoglobin, and their summation, total hemoglobin, of blood under the sensors. The Module is intended to allow for the display of StO2 and regional hemoglobin on a third party display/patient monitor and is indicated for use as follows:

When used with large sensors, the Module is indicated for use on adults and transitional adolescents ≥40 kg. When used with medium sensors, the Module is indicated for use on pediatric subjects ≥3 kg. When used with small sensors, the Module is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects < 5kg.

| Type of Use (Select one or both, as applicable) | .                                                                                   | A MISSA MARKA A MALA A MAJA JA JA PE SE A A A PERSON SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE S | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|-------------------------------------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                 | 17 Personal de 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 |                                                                                                                                                                                |                                                                                                                                                                                |

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K190270

## 510(k) Summary

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

| Submitter:           | CAS Medical Systems, Inc.                                                                                                   |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address:             | 44 East Industrial Rd. Branford CT. 06405 USA                                                                               |
| Contact:             | Ron Jeffrey – Director, Regulatory Affairs<br>Phone - (203) 488-6056<br>Fax – (203) 488-9438<br>Email – rjeffrey@casmed.com |
| Prepared:            | February 6, 2019                                                                                                            |
| Trade Name:          | FORE-SIGHT ELITE® Tissue Oximeter Module                                                                                    |
| Common Name:         | FORE-SIGHT ELITE Module                                                                                                     |
| Classification Name: | Oximeter, Tissue Saturation (870.2700) (MUD)                                                                                |

{4}------------------------------------------------

#### EQUIVALENCE (Predicate Device)

The FORE-SIGHT ELITE® Tissue Oximeter Module is equivalent to the following device:

FORE-SIGHT ELITE Tissue Oximeter Module (K180003);

#### REFERENCE DEVICE

- Hamamatsu Photonics Niro-200NX (K143219); ●
#### DESCRIPTION

The FORE-SIGHT ELITE Tissue Oximeter Module measures hemoglobin under the Sensor, allowing the clinician to continuously and accurately determine absolute levels of blood oxygenation saturation in the tissue (StO2). The Module additionally measures relative changes in oxygenated and deoxygenated hemoglobin concentrations for assessment of oxygenation changes as well as their sum, total hemoglobin concentration, under the sensor.

The Oximeter Module consists of a signal acquisition and processing unit, power isolation box, host power and communication cables, and sensor cables to connect with FDA cleared FORE-SIGHT ELITE small, medium and large sensors. The Sensors use multiple wavelengths in the range of 660 to 900 nm to precisely measure light absorption in tissue. Sensors are sized to provide targeted penetration depths appropriate for the tissue and patient populations of interest. The Module controls the measurement sequence, generating the sensor LED currents and processing the detected light signals after amplification. The FORE-SIGHT algorithm determines the StO2 values for the tissue under the sensor from the light absorption values and measured patient characteristics. The Module provides simultaneous measurements on up to two Sensors.

The Module is a host-powered device. The host configures the Oximeter and provides applicable audible, on-screen, and dedicated visual alarm indicators. The host monitor also provides numeric and/or realtime graphical display for tissue blood oxygenation saturation and changes in hemoglobin concentrations. Measurement data is delivered to the host through various interfaces including USB and serial connections.

#### FORE-SIGHT Oximeter Monitor Intended Use

The noninvasive FORE-SIGHT ELITE Tissue Oximeter Module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The Module is also intended for use as an adjunct monitor of relative changes of oxygenated hemoglobin, deoxygenated hemoglobin, and their summation, total hemoglobin, of blood under the sensors. The Module is intended to allow for the display of StO2 and regional hemoglobin on a third party display/patient monitor and is indicated for use as follows: When used with large sensors, the Module is indicated for use on adults and transitional adolescents ≥40 kg. When used with medium sensors, the Module is indicated for use on pediatric subjects ≥3 kg. When used with small sensors, the Module is indicated for cerebral use on pediatric subjects < 8 kg and noncerebral use on pediatric subjects <5kg.

{5}------------------------------------------------

#### FORE-SIGHT Module Technology Compared to Predicate Devices

The proposed FORE-SIGHT ELITE Tissue Oximeter Module is identical to the primary predicate device K180003 with the singular addition of software enhancement for the processing and display for relative changes in hemoglobin concentrations in addition to %StO2. Relative changes in hemoglobin concentrations are derived and displayed much the same as the NIRO-200NX reference device (K143219). The Module allows integration into CASMED and third-party monitoring systems.

The proposed FORE-SIGHT ELITE Tissue Oximeter Module and the predicate Module (K180003) use the same Sensors. The FORE-SIGHT ELITE Tissue Oximeter Module uses the same multi-distance diffuse reflectance spectroscopy operating principle. All cited monitors use light to examine a crosssection tissue microvasculature (a mixed bed of arterioles, capillaries and venules). The FORE-SIGHT ELITE Module and predicate devices analyze the light that is returned after having passed through tissues. The spectroscopic analysis determines concentrations of hemoglobin in its oxygenated and deoxygenated states. All cited monitors calculate oxygen saturation which reflects the percentage of oxygenated hemoglobin in the sampled tissue with the proposed Module and reference device additionally computing relative changes in hemoglobin concentrations.

The FORE-SIGHT ELITE Tissue Oximeter Module compares substantially to the predicates. All have multi-channel capability, an LED light source, similar user interface features on the monitor or host device and the benefits of portability.

#### Non-Clinical Performance Testing to Demonstrate Substantial Equivalence

The proposed FORE-SIGHT ELITE Tissue Oximeter Module has successfully undergone extensive performance, safety, electromagnetic, software and environmental testing (K180003) to ensure substantial equivalence to the predicate devices. In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, software, and systems verification and validation studies.

#### Clinical and Phantom Testing to Show Substantial Equivalence

The FORE-SIGHT ELITE Tissue Oximeter Module technology has successfully undergone extensive clinical validation for the indicated use, as demonstrated in the clearance of K180003 and earlier CASMED Premarket Notifications. As documented in the FORE-SIGHT ELITE Clinical Equivalency Report, there is no significant difference between the clinical functionality of this proposed Module and the Module cited in K180003 for oxygen saturation measurements. Clinical validation of changes in hemoglobin concentrations was successfully performed in healthy adult volunteers demonstrating high precision to co-oximetry blood sample references. Longitudinal performance in adult and pediatric cardiac surgical patients and procedures was also evaluated using convenience co-oximetry blood sample references. Additional validation on liquid optical phantoms was performed to demonstrate equivalence of the relative changes in hemoglobin concentrations to the NIRO-200NX reference device cleared under K143219 as well as concurrent frequency domain oximetry measurements.

Refer to Table 1

#### Conclusions Drawn from Clinical and Non-Clinical Testing

Clinical and phantom evaluation, safety / EMC testing, and software validation demonstrate the FORE-SIGHT ELITE Tissue Oximeter Module is substantially equivalent to the predicate devices.

{6}------------------------------------------------

| Patient<br>Population                                                                                | Sensor | Parameter      | Cerebral<br>Reference                                                                                                                                                       | Non-Cerebral<br>Reference                                                                                                                                                   | Type<br>Measurement                              | Subject<br>Weight<br>Range |
|------------------------------------------------------------------------------------------------------|--------|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|----------------------------|
| Adult                                                                                                | Large  | StO2           | Co-oximetry of jugular bulb and<br>arterial blood samples¹<br>[Range: 48 to 88%]                                                                                            | Co-oximetry of central venous<br>and arterial blood samples¹<br>[Range: 50 to 90%]                                                                                          | Single Point                                     | ≥ 40 kg                    |
|                                                                                                      |        | ΔHHb,<br>ΔO₂Hb | Co-oximetry of jugular bulb and<br>arterial blood samples²<br>[Range: ±5 g/dL, ±19 μM]                                                                                      | Co-oximetry of central venous<br>and arterial blood samples²<br>[Range: ±5 g/dL, ±19 μM]                                                                                    | Differential<br>Changes                          |                            |
| Pediatric -<br>Adolescents,<br>Children,<br>Infants, &<br>Neonates                                   | Medium | StO2           | Co-oximetry of internal jugular<br>vein and arterial blood samples<br>[Range: 48 to 92%]                                                                                    | Co-oximetry of central venous<br>and arterial blood samples<br>[Range: 53 to 88%]                                                                                           | Single Point                                     | ≥ 3 kg                     |
|                                                                                                      |        | ΔHHb,<br>ΔO₂Hb | Co-oximetry of jugular bulb and<br>arterial blood samples³<br>[Range: 0 to 2 g/dL, 7.4 uM]<br>and frequency domain oximeter<br>on liquid optical phantom<br>[Range: ±85 μM] | Co-oximetry of jugular bulb<br>and arterial blood samples³<br>[Range: 0 to 2 g/dL, 7.4 uM]<br>and frequency domain oximeter<br>on liquid optical phantom<br>[Range: ±85 μM] | Differential<br>Changes                          |                            |
| Pediatric -<br>Adolescents,<br>Children,<br>Infants, &<br>Neonates                                   | Small  | StO2           | Co-oximetry of internal jugular<br>vein and arterial blood samples<br>[Range: 50 to 90%]                                                                                    | Co-oximetry of central venous<br>and arterial blood samples<br>[Range: 66 to 96%]                                                                                           | Single Point                                     | 3 to 8 kg                  |
|                                                                                                      |        | ΔHHb,<br>ΔO₂Hb | Frequency domain oximeter on<br>liquid optical phantom<br>[Range: ±85 μM]                                                                                                   | Frequency domain oximeter on<br>liquid optical phantom<br>[Range: ±85 μM]                                                                                                   | Differential<br>Changes                          |                            |
| Pediatric -<br>Neonates<br>(Term,<br>Premature,<br>Low Birth<br>Weight, Very<br>Low Birth<br>Weight) | Small  | StO2           | FORE-SIGHT MC3010⁴<br>[Range: 50 to 90%]                                                                                                                                    | Co-oximetry of umbilical venous<br>and pulse oximetry samples<br>[Range: 66 to 96%]                                                                                         | StO2 data<br>averaged in two-<br>minute windows⁵ | < 5 kg                     |
|                                                                                                      |        | ΔHHb,<br>ΔO₂Hb | Frequency domain oximeter on<br>liquid optical phantom<br>[Range: ±85 μM]                                                                                                   | Frequency domain oximeter on<br>liquid optical phantom<br>[Range: ±85 μM]                                                                                                   | Differential<br>Changes                          |                            |

Table 1: StO2 and Changes in Hemoglobin Concentration Validation Methodology

<sup>1</sup> Validations performed on healthy adult volunteers (Age: 19 to 40 years, BMI: 18 to 30.5, Gender: 12 Male and 13 Fementation: 15 Light, 5 Moderate, 5 Dark)

<sup>2</sup> Validations performed on healthy volunteer subjects (Age: 19 to 40 years, BMI: 18 to 31.6, Gender: 22 Male and 21 Female, Skin Pigmentation: 21 Light, 14 Moderate, and 8 Dark). Convenience sample measurements evaluated on cardiac surgical patients undergoing cardiopulmonary bypass procedures (Age: 47 to 89 years, BMI: 23.8 to 29.6, Gender: 3 Male and 2 Female, Skin Pight, 1 Moderate).

<sup>3</sup> Performance on cardiac surgical patients evaluated using convenients on pediatric patients undergoing interventional catheterization procedures (Age: 0.4 to 9.2 years, BMI: 14.1 to 18.2, Gender: 4 Male and 1 Female, Skin Pigmentation: 2 Light, 1 Moderate, 2 Dark).

Unlike the other FORE-SIGHT ELITE validation studies, this cerebral validation study did not include invasive measurements because of the challenge for medical centers to obtain consent to internal jugular venous catheter in very small subjects.

StO2 data was averaged in two-minute windows for term, premature low birth weight (LBW), and very low birth weight (VLBW) neonates for the following reasons: 1) to reduce the influence of acute changes in body position or touch as the hemodynamics BW and VLBW neonates are not as stable compared to normal birth weight neonates, and 2) to enable measurements for both 3010 and FORE-SIGHT ELITE Sensors or across multiple abdominal locations at nominally the same time for the tes for which only one Sensor can be fitted on the head or specific abdominal location

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K190270](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K190270)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
