← Product Code [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD) · K123700

# FORE-SIGHT ELITE ABSOLUTE TISSUE OXIMETER (K123700)

_Cas Medical Systems, Inc. · MUD · Apr 11, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K123700

## Device Facts

- **Applicant:** Cas Medical Systems, Inc.
- **Product Code:** [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD.md)
- **Decision Date:** Apr 11, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The noninvasive FORE-SIGHT Elite Absolute Tissue Oximeter is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states and is indicated for use as follows: When used with large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and transitional adolescents ≥ 40 kg.

## Device Story

Noninvasive tissue oximeter; monitors absolute regional hemoglobin oxygen saturation (StO2) in brain tissue. Inputs: multi-wavelength light (660-900 nm) via sensors. Operation: multi-distance diffuse reflectance spectroscopy; monitor unit regulates light output, processes detected signals via preamplifier, and applies proprietary algorithm to calculate StO2. Output: numeric and real-time graphical display of oxygenation levels; supports up to four sensors. Used in clinical settings; operated by healthcare professionals. Provides continuous monitoring to assist clinicians in identifying ischemic states; enables timely clinical intervention to prevent tissue damage.

## Clinical Evidence

Clinical validation performed for Cerebral Tissue Oxygen Saturation (StO2%) in adult and transitional adolescent subjects. Testing included hardware, systems verification, and validation studies to demonstrate substantial equivalence.

## Technological Characteristics

Multi-distance diffuse reflectance spectroscopy; 660-900 nm light wavelengths. Mains-powered monitor with battery backup. Touchscreen interface; VGA video output; USB/RS-232 data export. Supports up to 4-channel simultaneous monitoring. Software-based algorithm for StO2 calculation.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- FORE-SIGHT Absolute Tissue Oximeter Monitor ([K112820](/device/K112820.md))
- Nonin Model 7600 Regional Oximeter Equanox ([K102715](/device/K102715.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

CASMED Nov. 28, 2012

#### FORE-SIGHT ELITE ABSOLUTE TISSUE OXIMETER

K123700

Page 18 of 240

# APR 1 1 2013

# 2 (2)

Image /page/0/Picture/6 description: The image shows the word "CASMED" in large, bold, black letters. The letters are slightly slanted to the right, giving the word a dynamic appearance. A small registered trademark symbol is located to the upper right of the letter "D".

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

and the same of the same

1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -

| Submitter:           | CAS Medical Systems, Inc.                                                                                                   |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address:             | 44 East Industrial Rd. Branford CT. 06405 USA                                                                               |
| Contact:             | Ron Jeffrey – Director, Regulatory Affairs<br>Phone – (203) 488-6056<br>Fax – (203) 488-9438<br>Email – rjeffrey@casmed.com |
| Prepared:            | November 28, 2012                                                                                                           |
| Trade Name:          | FORE-SIGHT Elite™ Absolute Tissue Oximeter                                                                                  |
| Common Name:         | FORE-SIGHT Oximeter                                                                                                         |
| Classification Name: | Oximeter, Tissue Saturation (870.2700) (MUD)                                                                                |

Page 1 of 3

.

{1}------------------------------------------------

## EQUIVALENCE (Predicate Device)

The FORE-SIGHT Elite™ Absolute Tissue Oximeter is equivalent to the following devices:

- * FORE-SIGHT Absolute Tissue Oximeter Monitor (K112820)
- * Nonin Model 7600 Regional Oximeter Equanox (K102715)

#### DESCRIPTION

The FORE-SIGHT Elite Absolute Tissue Oximeter measures hemoglobin under the Sensor, allowing the clinician to continuously and accurately determine absolute levels of blood oxygenation saturation in the tissue (StO2).

The Oximeter consists of a monitor unit, preamplifier assembly, and sensor uses multiple wavelengths of light in the range of 660 to 900 nm to precisely measure light absorption in tissue. Sensors are sized to provide targeted penetration depths appropriate for the tissue and patient populations of interest. The monitor unit controls the measurement sequence, regulating sensor light output and processing the detected light signals after amplification by the dual-channel preamplifier assembly. The FORE-SIGHT algorithm determines the StO2 values for the tissue under the sensor from the light absorption values and measured patient characteristics. The monitor unit provides simultaneous measurements on up to four sensors with both numeric and real-time graphical display formats.

The monitor unit is a mains-powered device with a field-replaceable battery backup module. A touchscreen user interface allows configuration of the Oximeter including audible, on-screen, and dedicated visual alarm indicators. The monitor display can be replicated for simultaneous remote viewing through an auxiliary VGA video output. Measurement data can be exported through various interfaces such as USB and RS-232.

#### FORE-SIGHT Oximeter Monitor Intended Use

The noninvasive FORE-SIGHT Elite Absolute Tissue Oximeter is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states and is indicated as follows:

When used with Large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and transitional adolescents ≥ 40 Kg.

### FORE-SIGHT Monitor Technology Compared to Predicate Devices

The FORE-SIGHT Elite Absolute Tissue Oximeter compares substantially to the cited predicate devices in that they use fundamentally the same optical operating principle, called multi-distance diffuse reflectance spectroscopy. All cited monitors use light to examine a cross-section tissue microvasculature (a mixed bed of arterioles, capillaries and venules). The FORE-SIGHT Elite Monitor and predicate devices analyze the light that is returned after having passed through tissues. The spectroscopic analysis determines concentrations of hemoglobin in its oxygenated and deoxygenated states. All cited monitors calculate oxygen saturation which reflects the percentage of oxygenated hemoglobin in the sampled tissue.

{2}------------------------------------------------

The FORE-SIGHT Elite Absolute Tissue Oximeter compares substantially to the FORE-SIGHT predicate device. Differences between the FORE-SIGHT predicate and the Elite include a 4-channel capability, enhanced user interface features and the benefits of portability.

## Non-Clinical Performance Testing to Demonstrate Substantial Equivalence

The FORE-SIGHT Elite Absolute tissue Oximeter has successfully undergone extensive performance, safety, electromagnetic, clinical, software and environmental testing to ensure it has been found to be substantially equivalent to the FORE-SIGHT predicate device. In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, systems verification and validation studies of the FORE-SIGHT monitor and sensors.

#### Clinical Testing to Show Substantial Equivalence

The FORE-SIGHT Elite Absolute Tissue Oximeter has successfully undergone extensive clinical validation for the indicated use. The premarket notification cites a validation report for Cerebral Tissue Oxygen Saturation (StO2%) in Adult and Transitional Adolescent Subjects.

## Conclusions Drawn from Clinical and Non-Clinical Testing

Clinical evaluation, safety testing and software validation demonstrate the FORE-SIGHT Elite Absolute Tissue Oximeter is substantially equivalent to the predicate devices.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2013

CAS Medical Systems c/o Mr. Ron Jeffrey 44 East Industrial Rd Brandford, CT 06405

Re: K123700

Trade/Device Name: FORE-SIGHT Elite Absolute Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter, Tissue Saturation Regulatory Class: Class II Product Code: MUD Dated: February 19, 2013 Received: March 7, 2013

Dear Mr. Jeffrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Ron Jeffrey

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

for

Sincerely yours.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

| 510(k) Number:       | K123700                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:         | FORE-SIGHT Elite™ Absolute Tissue Oximeter.                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use: | The noninvasive FORE-SIGHT Elite Absolute Tissue Oximeter is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states and is indicated for use as follows:<br>When used with large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and transitional adolescents<br>≥ 40 kg. |

Prescription Use 
 *✓*
 AND/OR
 Over-the-Counter Use

Prescription Use J (Part 21 CFR 801 Subpart D)

ANDIOK

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Con

ﺎﻧ

## Concurrence of CDRH, Office of Device Evaluation (ODE)

Clwer

ven P. Faris -S
13:04.11
13:30:11 -04'00'

Page 1 of _ 1 _

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K123700](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K123700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
