DATEX-OHMEDA S/5 WEB VIEWER, DATEX-OHMEDA S/5 POCKET VIEWER AND DATEX-OHMEDA S/5 CELLULAR VIEWER WITH L-WEB04 SOFTWARE
K052975 · GE Healthcare · MSX · Jan 20, 2006 · Cardiovascular
Device Facts
Record ID
K052975
Device Name
DATEX-OHMEDA S/5 WEB VIEWER, DATEX-OHMEDA S/5 POCKET VIEWER AND DATEX-OHMEDA S/5 CELLULAR VIEWER WITH L-WEB04 SOFTWARE
Applicant
GE Healthcare
Product Code
MSX · Cardiovascular
Decision Date
Jan 20, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Indication for use for S/5 Web Viewer: The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only. Indication for use for S/5 Pocket Viewer: The Datex-Ohmeda S/5 Pocket Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Pocket Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Pocket Viewer clients. The S/5 Pocket Viewer client runs on a generic handheld computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Pocket Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Pocket Viewer is not a primary alarm source. The device is for use by qualified medical personnel only. Indication for use for S/5 Cellular Viewer: The Datex-Ohmeda S/5 Cellular Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Cellular Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Cellular Viewer clients. The S/5 Cellular Viewer client runs on a generic cellular phone that is connected to the hospital local area network. The Datex-Ohmeda S/5 Cellular Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Cellular Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 CellularViewer is not a primary alarm source. The device is for use by qualified personnel only.
Device Story
System provides remote access to real-time patient monitoring data (curves, trends, numeric values, visual alarms) from Datex-Ohmeda bedside monitors. Components include S/5 Web Server, hospital LAN, and client applications running on generic PCs, PDAs, or cellular phones. Clients connect via hospital WLAN (802.11b) or mobile networks (GSM, GPRS, HSCSD, CDMA). System acts as decision-support tool for clinicians outside immediate patient care area; not a primary alarm source. Healthcare providers use output for remote consultation and monitoring. Benefits include increased clinician mobility and access to patient status. System supports up to 30 concurrent cellular users and HTTPS tunneling for security.
Clinical Evidence
Bench testing only. Verification and validation of specifications performed against standards including IEC 60950 (product safety), IEC-CISPR 22/24 (EMC), and ISO 14971 (risk analysis). No clinical data presented.
Technological Characteristics
Software-based remote monitoring system. Operates on generic PC, PDA, or cellular phone hardware. Connectivity via hospital LAN, WLAN (802.11b), or mobile networks (GSM, GPRS, HSCSD, CDMA). Supports HTTPS tunneling. Complies with IEC 60950, IEC 60601-1-4, and EMC standards.
Indications for Use
Indicated for remote viewing and processing of real-time patient data (curves, numeric info, trends, visual alarms) from networked bedside monitors. Intended for use by qualified medical personnel for consultation and remote monitoring of hospital patients. Not a primary alarm source.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer (K033078)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K052975" in a large, bold font. Below this, the text "Page 1 of 4" is printed in a smaller font size. The text suggests that this is the first page of a document that is four pages long, and "K052975" may be a document or identification number.
JAN 2 0 2006
## Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Web Viewer, Datex-Ohmeda S/5 Pocket Viewer and Datex-Ohmeda S/5 Cellular Viewer with L-WEB04 software
## GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
#### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
## NAME OF CONTACT:
Mr. Joel Kent
DATE:
-
October 20, 2005
#### DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5 Web Viewer, Datex-Ohmeda S/5 Pocket Viewer and Datex-Ohmeda S/5 Cellular Viewer with L-WEB04 software
#### COMMON NAME:
Remote monitoring device
#### CLASSIFICATION NAME:
The following Class II classifications appear applicable:
| Product Code | Classification Name | CFR Section |
|--------------|-----------------------------------------------------------|-------------|
| MSX | System, network and communication, physiological monitors | 870.2300 |
{1}------------------------------------------------
## NAME OF LEGALI.Y MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The revised Datex-Ohmeda S/5 Web Viewer version and S/5 Pocket Viewer version and the new S/5 Cellular Viewer are substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer versions (K033078) currently in distribution.
#### DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda S/5 Web Viewer is a supplementary monitoring application running on a generic PC that is connected to the hospital LAN, either directly or via the Internet. It is based on the World Wide Web and Java technologies, and it is intended to be used for remote viewing of real-time patient information and trends from patient monitors that are connected to the Datex-Ohmeda \$/5 Network and Central. The Pocket Viewer is a Web Viewer version running on a Pocket PC PDA that is connected to the hospital LAN via wireless access within the hospital, or via a mobile connection outside the hospital. The PDA uses a standard WLAN (802,111b) or mobile connections (GSM, GPRS, HSCSD, CDMA) to gain access to the Hospital LAN and Web Server. The Cellular Viewer is a Web Viewer version running on a generic cellular phone that is connected to the hospital LAN via a mobile connection. The mobile phone uses standard mobile connections (GSM, GPRS, HSCSD) to gain access to the Hospital LAN and Web Server. The Web Viewer, Pocket Viewer and Cellular Viewer are not primary alarm sources but decisionmaking support tools that offer clinicians access to the patient data also outside the patient care area. The network architecture of the S/5 Web/Pocket/Cellular Viewer system consists of the following components:
- Datex-Ohmeda S/5 Network that connects D-O monitors to one or more D O S/5 Centrals ◆
- The Hospital LAN to which the office PCs in the hospital are connected to .
- . The S/5 Web Server that is connected to both of these networks
- S/5 Web Viewer client programs running in desktop and laptop PCs, S/5 Pocket Viewer client . programs running in PDAs and S/5 Cellular Viewers running in generic cellular phones
- . Optional VPN (virtual private network) or dial-up solutions enabling remote connection to patient monitoring data with the S/5 Web Viewer, Pocket Viewer and Cellular Viewer
The hospital is responsible for ensuring a secure and functional interface between the Datex-Ohmeda S/5 Network and the Hospital LAN, by utilizing, for example, a gateway, router, switch or firewall, as shown in the figure above. If the Web Viewer clients are not connected to a hospital Intranet, a regular hub can be used instead. Wireless LAN access points are required to connect the Pocket Viewer to the WLAN. For Cellular phones the proper subscriptions with the telephone operators are needed.
{2}------------------------------------------------
#### INTENDED USE as required by 807.92(a)(5)
Indication for use for S/5 Web Viewer: The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.
Indication for use for S/5 Pocket Viewer: The Datex-Ohmeda S/5 Pocket Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Pocket Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Pocket Viewer clients. The S/5 Pocket Viewer client runs on a generic handheld computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Pocket Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Pocket Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.
Indication for use for S/5 Cellular Viewer: The Datex-Ohmeda S/5 Cellular Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Cellular Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Cellular Viewer clients. The Datex-Ohmeda S/5 Cellular Viewer client runs on a generic cellular phone that is connected to the hospital local area network. The Datex-Ohmeda S/5 Cellular Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Cellular Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 CellularViewer is not a primary alarm source. The device is for use by qualified personnel only.
### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The revised Datex-Ohmeda S/5 Web Viewer version and S/5 Pocket Viewer version and the new S/5 Cellular Viewer are substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer versions (K033078) currently in distribution,
Similarities:
The indications for use for the S/5 Web Viewer is identical to the predicate.
The indications for use for the S/5 Pocket Viewer is identical to the predicate.
The indications for use for the S/5 Cellular Viewer is the same as in predicate S/5 Pocket Viewer except that the term 'generic handheld computer' has been replaced by a term 'generic mobile phone'.
The structure and functionality of the revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer and the new S/5 Cellular Viewer corresponds to the structure and functionality of the Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer (predicate). The basic architecture of the revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer and the new Cellular Viewer is the same as that of Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer (predicate).
{3}------------------------------------------------
The revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer and the new S/5 Cellular Viewer can show real-time curves, numeric information, graphical and numerical trends and visual alarms from bedside monitors just like the predicate.
The physical network components used by the revised S/5 Web Viewer and S/5 Pocket Viewer and the new S/5 Cellular Viewer are the same as in the predicate.
#### Differences:
The following functionality has been added to the revised Datex-Ohmeda S/5 Web Viewer, S/5 Pocket Viewer and the new S/5 Cellular Viewer:
User interface changes:
- A new viewer type Cellular Viewer is available .
- Up to 30 concurrent Cellular Viewer users .
- Entropy parameter numeric values and trends are provided ●
- User specific configurations are possible through User Configuration Pages .
- . User can change her/his own password
- . No more support for S/5 Light Monitor trends
In addition to the functional changes, the following technical improvements have been implemented in the revised S/5 Web Viewer and S/5 Pocket Viewer:
- . The new version supports standard mobile phone technology
- Support for HTTPS tunneling with advanced communication security is available .
- PC hardware: A new version of the PC for the Web Server computer has been . specified because manufacturing of the earlier one was discontinued
#### Summary:
The changes above do not effect safety and effectiveness of the system, and the new Datex-Ohmeda S/5 Web Viewer, S/5 Pocket Viewer and S/5 Cellular Viewer, described in this submission, are substantially equivalent to the predicate device.
## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
Datex-Ohmeda S/5 Web Viewer, Datex-Ohmeda S/5 Pocket Viewer and Datex-Ohmeda S/5 Cellular Viewer with L-WEB04 software has been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications.
- EN60950: 2000 (IEC60950 3rd edition) Product Safety .
- EN 55022: 1998 (IEC-CISPR 22) Radio Frequency Interface �
- EN 55024: 1998 (IEC-CISPR 24) Electromagnetic Immunity .
- EN 61000-3-2:1995 + A1/A2/A14, Harmonic Currents .
- EN 61000-3-3:1995, Voltage Fluctuation and Flicker .
- EMC Directive 89/336/EEC (including amendments) .
- Low Voltage Directive 73/23/EEC (amended by 93/68/EEC) .
- . ISO 14971:2000, Medical devices - Risk analysis
- IEC 60601-1-4Medical electrical equipment. Part 1: General requirements for safety4. . Collateral Standard: Safety requirements for programmable medical systems.
- CAN/CSA-C22.2 No. 60950-00: Safety on Information Technology Equipment. .
- . UL: IEC 60950 (1999) Third Edition,
#### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the revised Datex-Ohmeda S/5 Web Viewer version and S/5 Pocket Viewer version and the new S/5 Cellular Viewer and they are substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer versions (K033078).
{4}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a cascading manner. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 0 2006
GE Healthcare c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492
Re: K052975
Trade Name: Datex-Ohmeda S/5 Web Viewer, Pocket Viewer and Cellular Viewer with L-WEB04 software Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological Monitors Network and Communication System Regulatory Class: Class II (two) Product Code: MSX Dated: December 19, 2005 Received: December 21, 2005
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 -- Mr. Joel C. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
B/Bummar for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
## Device Name: Datex-Ohmeda S/5 Web Viewer, Datex-Ohmeda S/5 Pocket Viewer and Datex-Ohmeda S/5 Cellular Viewer with L-WEB04 software.
#### Indications for use:
Indication for use for \$/5 Web Viewer: The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of a S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.
Indication for use for S/5 Pocket Viewer: The Datex-Ohmeda S/5 Pocket Viewer displays information received from other networked devices. It is comprised of a \$/5 Web Server and S/5 Pocket Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Pocket Viewer clients. The S/5 Pocket Viewer client runs on a generic handheld computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Pocket Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda \$/5 Pocket Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.
Indication for use for S/5 Cellular Viewer: The Datex-Ohmeda S/5 Cellular Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Cellular Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Cellular Viewer clients. The Datex-Ohmeda S/5 Cellular Viewer client runs on a generic cellular phone that is connected to the hospital local area network. The Datex-Ohmeda \$/5 Cellular Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Cellular Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 CellularViewer is not a primary alarm source. The device is for use by qualified personnel only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | |
| 510(k) Number | K052976 |
| | Page __ of __ |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.