← Product Code [MSX](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MSX) · K033078

# DATEX OHMEDA S/5 WEB VIEWER AND S/5 POCKET VIEWER WITH L-WEB03 SOFTWARE (K033078)

_Datex-Ohmeda · MSX · Oct 29, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MSX/K033078

## Device Facts

- **Applicant:** Datex-Ohmeda
- **Product Code:** [MSX](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MSX.md)
- **Decision Date:** Oct 29, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The Datex-Ohmeda S/5 Web Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices.

## Device Story

System comprises S/5 Web Server and client applications (Web Viewer for generic PCs, Pocket Viewer for handheld PDAs); inputs real-time patient data, waveforms, and trends from networked bedside monitors via hospital LAN (wired or wireless 802.11b/mobile connections); server coordinates network connections; clients display patient data, trends, and visual alarms; used by clinicians for remote consultation and monitoring; not a primary alarm source; provides decision-making support outside immediate patient care area; benefits include increased clinician access to patient status.

## Clinical Evidence

Bench testing only. Device verified through validation of specifications and compliance with safety standards including EN60950, IEC60601-1-4, and electromagnetic compatibility standards (EN 55022, EN 55024). No clinical data presented.

## Technological Characteristics

Software-based remote monitoring system; utilizes Java and Web technologies; supports generic PC and Pocket PC PDA hardware; connectivity via hospital LAN (wired/wireless 802.11b) or mobile (GSM/GPRS/HSCSD); SSL encryption for data security; operates as client-server architecture; no specific material construction as it is a software application running on generic hardware.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- Datex-Ohmeda S/5 Web Viewer ([K023497](/device/K023497.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 9 2003

Image /page/0/Picture/1 description: The image is a black and white logo. The logo consists of a large circle with a smaller square in the upper right corner. Inside the circle is a white shape that resembles the letter 'D'. The overall design is simple and modern.

## Premarket Notification 510(k) Summary As required by section 807.92 Datex Ohmeda S/5 Web Viewer and S/5 Pocket Viewer with L-WEB03 Software

K033078

Page 1 of 4

## GENERAL COMPANY INFORMATION as required by 807.92(a){1}

## COMPANY NAME/ADDRESS/PHONE/FAX:

Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344

## NAME OF CONTACT:

Mr. Joel Kent

## DATE:

September 27, 2003

## DEVICE NAME as required by 807.92(a)(2)

## TRADE NAME:

Datex Ohmeda S/5 Web Viewer and S/5 Pocket Viewer with L-WEB03 Software

## COMMON NAME:

Remote monitoring device

## CLASSIFICATION NAME:

The following Class II classification appears applicable:

System, network and communication, physiological monitors 870.2300 MSX

## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The revised Datex-Ohmeda S/5 Web Viewer version and S/5 Pocket Viewer are substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Web Viewer version (K023497).

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a series of handwritten numbers, specifically "033078". Below the numbers, the text "Page 2 of 4" is printed in a simple font. The numbers appear to be part of a page numbering system or a reference code within a larger document.

## DEVICE DESCRIPTION as required by 807.92(a)(4)

The intended use for the modified device is the same as the predicate, Datex-Ohmeda S/5 Web Viewer (K023497). The indications for use for the S/5 Web Viewer is the same as in predicate except that the grammar has been improved. The indications for the S/5 Pocket Viewer is the same as in predicate except that the grammar has been improved, and that the term 'generic computer' has been replaced by a term 'generic handheld computer'. There has been no change to the basic technology from the predicate. Changes to the labeling include revised User's Manuals and Brochures.

The Datex-Ohmeda S/5 Web Viewer is a supplementary monitoring application running on a generic PC that is connected to the hospital LAN, either directly or via the Internet. It is based on the World Wide Web and Java technologies, and it is intended to be used for remote viewing of real-time patient information and trends from patient monitors that are connected to the Datex-Ohmeda S/5 Network and Central.

The Pocket Viewer is a Web Viewer version running on a Pocket PC PDA that is connected to the hospital LAN via wireless access within the hospital, or via a mobile connection outside the hospital. The PDA uses a standard WLAN (802.11b) or mobile connections (GSM, GPRS, HSCSD) to gain access to the Hospital LAN and Web Server.

The Web Viewer and Pocket Viewer are not primary alarm sources but decision-making support tools that offer clinicians access to the patient data also outside the patient care area.

The hospital is responsible for ensuring a secure and functional interface between the Datex-Ohmeda S/5 Network and the Hospital LAN, by utilizing, for example, a gateway, router, switch or firewall, as shown in the figure above. If the Web Viewer clients are not connected to a hospital Intranet, a regular hub can be used instead. Wireless LAN access points are required to connect the Pocket Viewer to the WLAN.

#### INTENDED USE as required by 807.92(a)(5)

#### Intended use:

The Datex-Ohmeda S/5 Web Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices.

The Datex-Ohmeda S/5 Pocket Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices.

#### Indication for use for S/5 Web Viewer:

The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of an S/5 Web Server and S/5 Web Viewer clients.

The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network.

The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use.

The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source.

The device is for use by qualified medical personnel only.

#### Indication for use for S/5 Pocket Viewer:

The Datex-Ohmeda S/5 Pocket Viewer displays information received from other networked devices. It is comprised of an S/5 Web Server and S/5 Pocket Viewer clients.

The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Pocket Viewer clients. The S/5 Pocket Viewer client runs on a generic handheld computer that is connected to the hospital local area network.

{2}------------------------------------------------

The Datex-Ohmeda S/5 Pocket Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use.

The Datex-Ohmeda S/5 Pocket Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.

## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The revised Datex-Ohmeda S/5 Web Viewer version and S/5 Pocket Viewer are substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Web Viewer version (K023497) currently in distribution.

#### Similarities:

The indications for use for the S/5 Web Viewer are the same as in the predicate except that the grammar has been improved. The indications for use for the S/5 Pocket Viewer is the same as in predicate except that the grammar has been improved, and that the term 'generic computer' has been replaced by a term 'generic handheld computer'. The structure and functionality of the revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer corresponds to the structure and functionality of the Datex-Ohmeda S/5 Web Viewer (predicate). The basic architecture of the revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer is the same as that of Datex-Ohmeda S/5 Web Viewer (predicate). The revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer can show real-time curves, numeric information, graphical and numerical trends and visual alarms from bedside monitors just like the predicate. The physical network and the PC hardware components used by the revised S/5 Web Viewer and S/5 Pocket Viewer are the same as in the predicate.

## Differences:

The following functionality has been added to the revised Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer:

User interface changes:

- The users can configure additional waveform views with fixed parameter sets in addition to ● the view that is currently displayed on the source monitor.
- Separate parameter configurations for graphical and numerical trends were created, and the ● maximum number of configurable pages was increased. Division between anesthesia and critical care pages was removed.
- Digit field layouts have been improved. .
- Possibility to hide the numeric values of the waveform parameters was added (Pocket Viewer ● client only) in order to give more space to waveforms.
- In Pocket Viewer due to small display size only the highest-level active alarm is indicated ● in the waveform view with a color-coded symbol, and active alarms can be viewed in a separate Alarm View page. In the Web Viewer the alarms are displayed as in the previous version.
- A separate Demographics page was added to display patient demographic data. ●
- To make selecting of a patient easier in the small Pocket Viewer display, a possibility to . display selected patients first in the patient selection menu was added.
- Viewing of EEG waveforms, and EEG and NMT numeric values and trends were added. .
- Possibility to set and change individual user preferences through setup pages that are ● accessible directly from the client application was added.
- In Pocket Viewer, the patient selection is always made through a Patient menu. The menu is . included also in the Web Viewer client, and therefore a possibility to hide the (permanently displayed) monitor selector area has been included in the Web Viewer client.

{3}------------------------------------------------

In addition to the functional changes, the following technical improvements have been implemented in the revised S/5 Web Viewer and S/5 Pocket Viewer:

- The new version supports Microsoft Pocket PC based PDAs in addition to PC based clients. .
- SSL (Secure Sockets Layer) encryption is now utilized in securing the communication ● between the Web Server and the clients.
- The licensing scheme was enhanced from a fixed license of 6 concurrent users to cover . license options for 6, 12, 18, 24 or 30 concurrent Web Viewer clients and for 6, 12, 18, 24 or 30 concurrent Pocket Viewer clients.
- Automatic log-off after pre-selected time of inactivity was implemented .
- Enhanced event logging for administrative purposes. .
- A possibility to define the monitors that individual users can access has been added in the new . version.

Summary:

.

The changes above do not affect safety and effectiveness of the system, and the new Datex-Ohmeda S/5 Web Viewer and S/5 Pocket Viewer, described in this submission, are substantially equivalent to the predicate device.

## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

Datex Ohmeda S/5 Web Viewer and S/5 Pocket Viewer with L-WEB03 Software complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly tested through validation and verification of specifications. Verification of compliance with the following mandatory and voluntary standards has been made:

- EN60950: 2000 (IEC60950 3rd edition) Product Safety .
- EN 55022: 1998 (IEC-CISPR 22) Radio Frequency Interface .
- EN 55024; 1998 (IEC-CISPR 24) Electromagnetic Immunity .
- EN 61000-3-2:1995 + A1/A2/A14, Harmonic Currents ●
- EN 61000-3-3:1995, Voltage Fluctuation and Flicker .
- EMC Directive 89/336/EEC (including amendments) ●
- Low Voltage Directive 73/23/EEC (amended by 93/68/EEC) ●
- EN 1441, Medical devices Risk analysis .
- IEC 60601-1-4
- Medical electrical equipment. Part 1: General requirements for safety .
- 4. Collateral Standard: Safety requirements for programmable medical systems. .
- CAN/CSA-C22.2 No. 60950-00: Safety on Information Technology Equipment. ●
- . UL: IEC 60950 (1999) Third Edition.
- FDA/ODE Guidance for the Content of Premarket Submission for Software Contained in . Medical Devices, May 29, 1998.
- FDA/ODE Guidance for the Off-The-Shelf Software Use in Medical Devices, September 9, . 1999.

#### CONCLUSION:

The summary above shows that there are no new questions of safety and effectiveness for the Datex Ohmeda S/5 Web Viewer and S/5 Pocket Viewer with L-WEB03 Software as compared to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 9 2003

Datex-Ohmeda c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492

## Re: K033078

Trade Name: Datex Ohmeda \$/5 Web Viewer and S/5 Pocket Viewer with L-WEB03 Software Regulation Number: 21 CFR 870.2300 Regulation Name: System, network and communication, physiological monitors Regulatory Class: Class II (2) Product Code: MSX Dated: September 27, 2003 Received: September 29, 2003

Dear Mr. Kent:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Joel C. Kent

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mefglaintep
Bram D. Zuckerman, M.D.

Image /page/5/Picture/5 description: The image shows a handwritten text that appears to be the word "for". The letters are written in a cursive style, with the "f" having a loop at the top and extending downwards, and the "or" connected to the "f" in a fluid motion. The writing is simple and clear, suggesting it could be a quick note or a part of a larger text.

Bram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# 510(k) Number (if known): _ KO33078

#### Device Name: Datex-Ohmeda S/5 Web Viewer and Datex-Ohmeda S/5 Pocket Viewer with L-WEB03 software

#### Indication for use for S/5 Web Viewer:

The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Web Viewer clients.

The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network.

The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use.

The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source.

The device is for use by qualified medical personnel only.

#### Indication for use for S/5 Pocket Viewer:

The Datex-Ohmeda S/5 Pocket Viewer displays information received from other networked devices. It is comprised of a S/5 Web Server and S/5 Pocket Viewer clients.

The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the devices in the Datex-Olimeda Network and S/5 Pocket Viewer clients. The S/5 Pocket Viewer client runs on a generic handheld computer that is connected to the hospital local area network.

The Datex-Olimeda S/5 Pocket Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Pocket Viewer can be used for patients in the hospital and it is meant for consultation and remote monitoring use.

The Datex-Ohmeda S/5 Pocket Viewer is not a primary alarm source.

The device is for use by qualified medical personnel only.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|                      | Concurrence of CDRH, Office of Device Evaluation (ODE)          |
|----------------------|-----------------------------------------------------------------|
|                      | <div style="text-decoration: overline;">Mo Markan for BDZ</div> |
|                      | (Division Sign-Off)                                             |
|                      | Division of Cardiovascular Devices                              |
| Prescription Use     | Over-The-Counter Use                                            |
| (Per 21 CFR 801.109) | OR                                                              |
|                      | <div style="font-size: 3em;">✓</div>                            |
| 510(k) Number        | <div style="text-decoration: underline;">10033078</div>         |
|                      | (Optional Format 1-2-96)                                        |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MSX/K033078](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MSX/K033078)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
